K182053 · Alesi Surgical, Ltd. · PQM · Sep 7, 2018 · General, Plastic Surgery
Device Facts
Record ID
K182053
Device Name
Ultravision™ Visual Field Clearing System
Applicant
Alesi Surgical, Ltd.
Product Code
PQM · General, Plastic Surgery
Decision Date
Sep 7, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.5050
Device Class
Class 2
Indications for Use
The Ultravision™ Visual Field Clearing System is indicated for the clearance of smoke and other particulate matter that is created during laparoscopic surgery. The Ultravision™ 5mm Trocar component also establishes a path of entry for instruments used in laparoscopic surgery.
Device Story
System removes surgical smoke from laparoscopic visual field via electrostatic precipitation; consists of generator, Ionwand sterile pack, and 5mm trocar. Ionwand introduced into body cavity near smoke-generating electrosurgical device; generator provides energy. Trocar provides instrument entry path. Used in laparoscopic surgery by surgeons. Modified system adds mains power capability via converter accessory alongside existing rechargeable battery. Output is clear visual field for surgeon; benefits include improved visibility during procedures.
Clinical Evidence
No clinical data. Bench testing only, including electromagnetic compatibility and electrical safety compliance per IEC 60601.
Technological Characteristics
Electrostatic precipitation; polymers, elastomers, metals; rechargeable battery or mains power (via converter); 5mm trocar form factor; ethylene oxide sterilization for sterile accessories (SAL 10^-6); software-controlled with alarms.
Indications for Use
Indicated for clearance of surgical smoke and particulate matter during laparoscopic surgery in patients undergoing such procedures.
Regulatory Classification
Identification
A surgical smoke precipitator is a prescription device intended for clearance of the visual field by precipitation of surgical smoke and other aerosolized particulate matter created during laparoscopic surgery.
Special Controls
In combination with the general controls of the FD&C Act, the Surgical smoke precipitator is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Adverse tissue reaction must be mitigated through the following:
(i) Chemical characterization and toxicological risk assessment of the treated surgical smoke.
(ii) Demonstration that the elements of the device that may contact the patient are biocompatible.
(2) Electrical safety and electromagnetic compatibility testing must demonstrate that the device performs as intended.
(3) Software verification, validation, and hazard analysis must be performed.
(4) Performance data must demonstrate the sterility of the patient contacting components of the device.
(5) Performance data must support the shelf life of the sterile components of the device by demonstrating continued functionality, sterility, and package integrity over the identified shelf life.
(6) Animal simulated-use testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Device must be demonstrated to be effectively inserted, positioned, and removed from the site of use.
(ii) Device must be demonstrated to precipitate surgical smoke particulates to clear the visual field for laparoscopic surgeries.
(iii) Device must be demonstrated to be non-damaging to the site of use and animal subject.
(7) Labeling must identify the following:
(i) Detailed instructions for use.
(ii) Electrical safety and electromagnetic compatibility information.
(iii) A shelf life.
Predicate Devices
Ultravision™ Visual Field Clearing System (K170178)
Submission Summary (Full Text)
{0}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 7, 2018 Alesi Surgical Ltd. % Michele Lucev Regulatory Affairs Advisor Lakeshore Medical Device Consulting, LLC 128 Blye Hill Landing, Newbury, New Hampshire 03255
Re: K182053
Trade/Device Name: Ultravision™ Visual Field Clearing System Regulation Number: 21 CFR 878.5050 Regulation Name: Surgical Smoke Precipitator Regulatory Class: Class II Product Code: PQM Dated: August 6. 2018 Received: August 9, 2018
Dear Michele Lucey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Clarence W. Murray lii III -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K182053
Device Name Ultravision™ Visual Field Clearing System
#### Indications for Use (Describe)
The Ultravision™ Visual Field Clearing System is indicated for the clearance of smoke and other that is created during laparoscopic surgery.
The Ultravision™ 5mm Trocar establishes a path of entry for instruments used in laparoscopic surgery and includes the Ionwand for the clearance of smoke and other particulate matter that is created during laparoscopic surgery.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------|-----------------------------------------|
| <span style="text-decoration: underline;"><b>Prescription Use</b></span> (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
| <span style="text-decoration: underline;"><b>Over-The-Counter Use</b></span> (21 CFR 801 Subpart C) | □ |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### K182053 510(K) SUMMARY
## Submitter Information:
| Submitter's Name: | Alesi Surgical Ltd | |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Address: | Cardiff Medicentre<br>Heath Park<br>Cardiff<br>CF14 4UJ | |
| Telephone:<br>Fax: | ++44 (0)2920 682106<br>++44 (0)2920 750239 | |
| Contact Person:<br>Telephone: | Michele Lucey<br>603-748-1374 | |
| Date Prepared: | September 5, 2018 | |
| Device Trade Name: | Ultravision™ Visual Field Clearing System | |
| Classification: | Class II | |
| Product Code(s): | PQM | |
| Regulation Number(s): | 878.5050 | |
| Common Name: | Surgical Smoke Precipitator | |
| Predicate Devices: | Ultravision™ Visual Field Clearing System, K170178 | |
| Intended Use: | The Ultravision™ Visual Field Clearing System is indicated<br>for the clearance of smoke and other particulate matter that is<br>created during laparoscopic surgery.<br><br>The Ultravision™ 5mm Trocar component also establishes a<br>path of entry for instruments used in laparoscopic surgery | |
#### Device Description:
The Ultravision™ Visual Clearing System removes surgical smoke from the visual field during laparoscopic surgery by means of electrostatic precipitation. The System consists of the Ultravision™ Generator, the Ionwand Sterile Pack, and the Ionwand 5mm Trocar. The Ionwand is connected to the energy source and is then introduced into the body cavity near the smoke generating electrosurgical device. The Ultravision™ 5mm Trocar is intended for use only with the Ultravision™ Visual Field Clearing System to introduce the Ionwand while providing a pathway for laparoscopic instruments through one 5 mm trocar incision. The
K182053 Page 1 of 3
is
a
{4}------------------------------------------------
Ultravision™ 5mm Trocar provides an alternative to the currently available Ultravision™ Visual Field Clearing System Ionwand™ Sterile Pack for introduction of the Ionwand. The trocar may be used with or without the Ionwand component of the system. The system is powered using a rechargeable battery or through mains power. Accessories include the rechargeable battery, battery recharging station, mains converter power supply, mains converter, power supply unit, and patient return adaptor.
## Comparison to Predicate Device(s):
This Special 510(k) is for the device modification of the device to make it compatible with mains power. This change allows the system to be powered using the rechargeable battery or be used connected to mains electric power using the Ultravision™ Mains Converter, this is regarded as a new system accessory. This change does not affect the intended use of the device or alter the fundamental scientific technology of the device. The comparison chart below shows that there are no other design or material changes to the system, only the addition of an option to use mains power versus battery power.
| Substantial Equivalence Comparison Chart | | | |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Feature/Specification | Ultravision™<br>Visual Clearing<br>System (modified) | Ultravision™<br>Visual Clearing<br>System<br>(predicate) | Comparison |
| Indications | Laparoscopic surgery | Laparoscopic surgery | Same |
| Intended Use | The Ultravision™<br>Visual Field Clearing<br>System is indicated for<br>the clearance of smoke<br>and other particulate<br>matter that is created<br>during laparoscopic<br>surgery.<br>The Ultravision™ 5mm<br>Trocar component also<br>establishes a path of<br>entry for instruments<br>used in laparoscopic<br>surgery | The Ultravision™<br>Visual Field Clearing<br>System is indicated for<br>the clearance of smoke<br>and other particulate<br>matter that is created<br>during laparoscopic<br>surgery.<br>The Ultravision™ 5mm<br>Trocar component also<br>establishes a path of<br>entry for instruments<br>used in laparoscopic<br>surgery | Same |
| Materials of Construction | Polymers, elastomers, metals | Polymers, elastomers, metals | Same, materials<br>are identical |
| Regulatory Clearance/<br>Approval Reference | NA | K170178 | NA |
| Product Code | PQM | PQM | Same |
| Regulation Number | 878.5050 | 878.5050 | Same |
| General Description | Surgical Smoke<br>Precipitator | Surgical Smoke<br>Precipitator | Same |
| Mechanism of Action | Electrostatic<br>precipitation, identical to<br>the predicate | Electrostatic precipitation | Same |
{5}------------------------------------------------
| Substantial Equivalence Comparison Chart | | | |
|-----------------------------------------------|------------------------------------------------------------------------------------------------------|-------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Feature/Specification | Ultravision™ Visual Clearing System (modified) | Ultravision™ Visual Clearing System (predicate) | Comparison |
| Where used | Laparoscopic surgical procedures | Laparoscopic surgical procedures | Same |
| Power source | Battery and mains | Battery | Difference is<br>inclusion of a<br>mains power<br>adapter. This<br>difference does<br>not raise new<br>questions of safety<br>and efficacy |
| Software | Yes, identical to<br>predicate | Yes | Same |
| Alarms | Yes, identical to<br>predicate | Yes | Same |
| Instructions for use | Yes, identical to<br>predicate except for the<br>inclusion of the<br>Ultravision™ Mains<br>Converter | Yes | The addition of<br>the use of the<br>Mains Converter<br>to the IFU does<br>not raise new<br>questions of safety<br>and efficacy |
| Accessories | Yes, identical to<br>predicate except for the<br>inclusion of the<br>Ultravision™ Mains<br>Converter | Yes | The addition of<br>the use of the<br>Mains Converter<br>accessory does not<br>raise new<br>questions of safety<br>and efficacy |
| How Supplied | Nonsterile with nonsterile<br>and sterile accessories | Nonsterile with nonsterile<br>and sterile accessories | Same |
| SAL (sterile accessories) | $10^{-6}$ | $10^{-6}$ | Same |
| Sterilization Method (sterile<br>accessories) | Ethylene Oxide | Ethylene Oxide | Same |
## Nonclinical Performance Data:
In consideration of design control activities including risk analysis the non-clinical performance testing include testing in accordance with electromagnetic compatibility standards and an assessment of electrical safety compliance with IEC 60601.
Conclusion: Based on the performed nonclinical tests, the device is as safe, as effective, and performs as well as the legally marketed predicate device, K170178.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.