Clinical laboratory records (399,821 Pap tests); College of American Pathologists (CAP) National Percentage Data benchmarks
Retrospective analysis of routine clinical Pap test data to demonstrate equivalency in specimen adequacy and diagnostic categories (The Bethesda System) between the ExCellerator spatula and the predicate device.
Retrospective; Clinical laboratory data; Specimen adequacy; The Bethesda System
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
RWE Data Procurement and Analysis; Retrospective observational analysis of clinical laboratory data; Follow-up/Duration: Data from years 2003, 2006, and 2008; Study Period: 2003-2008
Patients undergoing routine Pap tests at two CLIA-certified clinical laboratories; Sample Size: 399,821 Pap tests; Number of Sites: 2
The ExCellerator Cervical Collection Device is a cervical spatula intended for collecting ectocervical specimens for the ThinPrep Pap test. The device is also intended to be used to aid in a more complete transfer of the specimen material from the endocervical brush (cytobrush) to the PreservCyt solution for ThinPrep testing. The ExCellerator device must be used with an IVD labeled endocervical brush.
Device Story
Single-use, non-sterile polypropylene cervical spatula; features dog-bone end for ectocervical sampling and oblong opening for cytobrush specimen transfer. Used by medical professionals in clinical settings. Ectocervical sampling involves 360-degree rotation against cervix; specimen transferred to PreservCyt solution via vigorous swirling. Endocervical brush transfer involves moving brush bristles through spatula opening while rotating shaft to release material. Device aids in more complete specimen transfer to PreservCyt solution, potentially improving diagnostic yield. Clinical decision-making relies on subsequent laboratory analysis of Pap test slides.
Clinical Evidence
Real World Evidence (RWE) analysis of 399,821 Pap tests compared subject device to predicate. Results showed comparable specimen adequacy and diagnostic categories (TBS system). Use of subject device associated with 54% reduction in unsatisfactory rates. Validation study (n=108) confirmed 94% satisfactory Pap smear rate when using device to transfer cytobrush specimens. Biocompatibility testing (ISO 10993-5, 10993-10) confirmed non-cytotoxic, non-sensitizing, and non-irritant status. Mechanical testing (bend, tension, torsion) confirmed structural integrity.
Technological Characteristics
Material: inert food-grade polypropylene plastic. Design: manual cervical spatula with an integrated oblong hole (~0.250 x 0.130-in) for brush cleaning. Dimensions: ~7.13-in length. Non-sterile, single-use, disposable. No energy source. No software.
Indications for Use
Indicated for use by medical professionals for collecting ectocervical specimens for ThinPrep Pap testing and to assist in transferring specimen material from an IVD-labeled endocervical brush into PreservCyt solution.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
Predicate Devices
Pap Smear Kit consisting of a cytobrush and spatula (K861389)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K182049
B. Purpose for Submission:
New device
C. Manufacturer and Instrument Name:
ExCell Company LLC
ExCellerator Cervical Collection Device
D. Type of Test or Tests Performed:
Not applicable.
E. System Descriptions :
1. Device Description :
The ExCellerator Cervical Collection Device is a non-sterile, single use, disposable cervical sampling device (spatula), comprised of inert food grade polypropylene plastic, intended for sampling the ectocervix during the cervical sample collection procedure. Additionally, it aids in a more complete transfer of the cytological sample collected with an IVD labeled endocervical brush (cytobrush) into the PreservCyt vial for the preparation of ThinPrep Pap test slides.
2. Principles of Operation:
For collection of cytological samples from the ectocervix, apply the dog-bone end of the spatula to the cervix and, while maintaining firm contact, rotate it 360 degrees to collect material from the entire ectocervical surface. Remove the spatula and place directly into the vial of PreservCyt Solution. Transfer the spatula sample into the vial by swirling the dog-bone end of the spatula vigorously ten (10) times in the PreservCyt Solution. Leave the spatula in the vial.
For transferring the collected specimen into the PreservCyt solution, after sampling the endocervix using an IVD labeled endocervical brush, transfer the brush sample into the vial by placing the cytobrush into the PreservCyt solution per instructions specified in the cytobrush label or ThinPrep Pap test label. Then, run the brush down along the "rails" of the ExCellerator spatula, and while keeping them both submersed, move the brush
{1}
bristles in-and-out of the hole in the ExCellerator spatula 10 times while rotating the shaft between the fingertips. Swirl the brush vigorously to release additional material. Remove the brush and discard it. Swirl the ExCellerator spatula in the solution up to five more times, then discard it. Replace and tighten the vial cap so that the torque line on the cap passes the torque line on the vial.
3. Modes of Operation:
Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ________ or No ________ X ________
Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ________ or No ________ X ________
4. Specimen Identification:
Not applicable.
5. Specimen Sampling and Handling:
Collected specimens are placed in PreservCyt solution and transported to the laboratory.
6. Calibration:
Not applicable.
7. Quality Control:
Not applicable.
8. Software:
FDA has reviewed applicant's Hazard Analysis and Software Development processes for this line of product types:
Yes ________ or No ________ X ________
Not applicable since the device is a cytological sampling device.
F. Regulatory Information:
1. Regulation section:
{2}
21 CFR 884.4530 Obstetric-gynecologic specialized manual instruments.
2. Classification:
Class II
3. Product code:
HHT – Spatula, cervical, cytological
4. Panel:
Obstetrics/Gynecology
G. Intended Use:
The ExCellerator Cervical Collection Device is a cervical spatula intended for collecting ectocervical specimens for the ThinPrep Pap test. The device is also intended to be used to aid in a more complete transfer of the specimen material from the endocervical brush (cytobrush) to the PreservCyt solution for ThinPrep Pap testing.
The ExCellerator device must be used with an IVD labeled endocervical brush.
1. Indications for Use:
See Intended Use above.
2. Special Conditions for Use Statement(s):
For prescription use only.
It is not intended for use in pregnant women.
“Caution: For use by medical professionals only.”
H. Substantial Equivalence Information:
1. Predicate Device Name(s) and 510(k) numbers:
Pap Smear Kit consisting of a cytobrush and spatula (K861389).
2. Comparison with Predicate Device:
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Material | Plastic | Same |
| Dimensions | Overall Length: 7.12” Paddle End Width: .5” Head End Width: .6875” | Same
Same
Same |
| Supplied | Non-Sterile-Disposable | Same |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Technological Characteristics | Additional oblong opening 0.250x0.130” near paddle end of the spatula | No additional oblong opening |
| Intended Use | The ExCellerator Cervical Collection Device is a cervical spatula intended for collecting ectocervical specimens for the ThinPrep Pap test. The device is also intended to be used to aid in a more complete transfer of the specimen material from the endocervical brush (cytobrush) to the PreservCyt solution for ThinPrep Pap testing. The ExCellerator device must be used with an IVD labeled endocervical brush. | Predicate device (spatula) is intended for ectocervical sampling |
I. Special Control/Guidance Document Referenced (if applicable):
ISO 10993-5: Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity (2009).
ISO 10993-10: Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization (2010).
ISO 10993-12: Biological evaluation of medical devices – Part 12: Sample preparation and reference material (2012).
J. Performance Characteristics:
1. Analytical Performance:
{4}
a. Accuracy:
Not applicable.
b. Precision/Reproducibility:
Not applicable.
c. Linearity:
Not applicable.
d. Carryover:
Not applicable.
e. Interfering Substances:
Not applicable.
2. Other Supportive Instrument Performance Data Not Covered Above:
The following clinical and non-clinical studies were performed to assess the clinical performance of the subject device:
a) Real World Evidence (RWE) Data Procurement and Analysis
RWE data was collected to demonstrate the equivalency of specimen adequacy and diagnoses of Pap test per The Bethesda System (TBS) for reporting cervical or vaginal cytologic diagnoses when using ExCellerator spatula compared to the predicate device (Pap-Perfect Plastic Spatula, K861389).
RWE data was collected from 399,821 Pap tests performed in two CLIA certified clinical laboratories for years 2003, 2006 and 2008. The two participating laboratories had annual volume between 88,800 and 90,400, and 34,200 and 57,200 Pap tests, respectively. In 2003, only the predicate device (Pap Perfect Spatula) was used for the Pap tests, however, in 2006, 50% of Pap test specimens were collected with the predicate device and 50% were collected using the ExCellerator spatula. In 2008, all Pap test specimens were collected with the ExCellerator spatula. The ExCellerator spatula was used according to the Instructions for Use. The dataset (as shown in Tables 1 and 2) includes data on specimen adequacy and on TBS diagnostic categories for Pap test specimens collected with both the predicate and the subject device. ThinPrep Pap test was used on all specimens. In 2003, all specimens were reviewed by manual microscopy and in 2006 and 2008 all specimens were reviewed using the ThinPrep Imaging System.
Specimen adequacy and diagnoses were also compared with benchmark College of American
5
{5}
Pathologists (CAP) National Percentage Data published on 27 September 2007 and 17 June 2010.
Table 1: Specimen Adequacy and Diagnoses of Pap Tests* using ExCellerator Spatula
| | | | Unsat | Neg | LSIL | HSIL | ASC-US | ASC-H | AGC | CA |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Collection Device Used** | Year | Total Cases | # | # | # | # | # | # | # | # |
| Pap Perfect Spatula | 2003 | 124238 | 1362 | 112243 | 4154 | 706 | 5295 | 298 | 162 | 18 |
| | | | 1.1% | 90.3% | 3.3% | 0.6% | 4.3% | 0.2% | 0.1% | 0.0% |
| ~50% Pap Perfect Spatula (76, 289)/ ~50% Ex Cellerator (70, 864) | 2006 | 129506 | 949 | 112143 | 5460 | 785 | 9568 | 413 | 172 | 16 |
| | | | 0.7% | 86.6% | 4.8% | 0.8% | 7.2% | 0.3% | 0.1% | 0.0% |
| Ex Cellerator | 2008 | 146077 | 742 | 125608 | 6967 | 1148 | 10527 | 745 | 323 | 17 |
| | | | 0.5% | 86.0% | 4.8% | 0.8% | 7.2% | 0.5% | 0.2% | 0.0% |
# Note:
*Results based on The Bethesda System (TBS) for reporting cervical or vaginal cytologic diagnoses:
ASC-US: Atypical Squamous Cells of Undetermined Significance
ASC-H: Atypical Squamous Cells, Cannot Rule Out High Grade Squamous Intraepithelial Lesion
LSIL: Low Grade Squamous Intraepithelial Lesion
HSIL: High Grade Squamous Intraepithelial Lesion
ASC/SIL: ASCUS to SIL Ratio
AGC: Atypical Glandular Cells
UNSAT: Unsatisfactory
**In year 2006 approximately $50\%$ of pap smears were collected with Pap Perfect Spatula and $50\%$ with the ExCellerator Cervical Collection Device. Breakdown by device type data are not available.
{6}
Table 2: College of American Pathologists Benchmarks on ThinPrep Laboratory Percentile-Reporting Rate in 2007*
| CATEGORY | 5th | 10th | 25th | 50th | 75th | 90th | 95th |
| --- | --- | --- | --- | --- | --- | --- | --- |
| ASC-US (%) | 1.7 | 2.5 | 3.3 | 4.9 | 6.7 | 9.6 | 11.5 |
| ASC-H (%) | 0.0 | 0.1 | 0.2 | 0.3 | 0.6 | 1.0 | 2.0 |
| LSIL (%) | 1.1 | 1.5 | 2.2 | 3.0 | 4.0 | 6.0 | 7.3 |
| HSIL (%) | 0.1 | 0.2 | 0.4 | 0.6 | 0.9 | 1.3 | 2.0 |
| ASC/SIL | 0.7 | 0.8 | 1.1 | 1.5 | 2.0 | 2.5 | 3.4 |
| AGC (%) | 0.0 | 0.0 | 0.1 | 0.2 | 0.3 | 0.9 | 1.4 |
| UNSAT | 0.1 | 0.3 | 0.6 | 1.1 | 1.7 | 2.9 | 3.4 |
Note: *College of American Pathologists. (2007). Cytopathology Checklist: CYP.07600. Rev. 09/27/2007
Table 3: College of American Pathologists Benchmarks on ThinPrep Laboratory Percentile-Reporting Rate in 2010*
| CATEGORY | 5th | 10th | 25th | 50th | 75th | 90th | 95th |
| --- | --- | --- | --- | --- | --- | --- | --- |
| ASC-US (%) | 2.2 | 2.8 | 3.7 | 4.8 | 6.7 | 9.4 | 12.1 |
| ASC-H (%) | 0.0 | 0.1 | 0.2 | 0.3 | 0.4 | 0.7 | 1.0 |
| LSIL (%) | 1.2 | 1.5 | 2.0 | 2.8 | 3.7 | 5.0 | 6.0 |
| HSIL (%) | 0.1 | 0.2 | 0.3 | 0.4 | 0.7 | 1.1 | 1.5 |
| ASC/SIL | 0.7 | 0.9 | 1.2 | 1.6 | 2.2 | 2.8 | 3.2 |
| AGC (%) | 0.0 | 0.0 | 0.1 | 0.1 | 0.3 | 0.5 | 0.7 |
| UNSAT | 0.3 | 0.4 | 0.7 | 1.1 | 1.8 | 2.7 | 3.4 |
Note: *College of American Pathologists (2010). Cytopathology Checklist: CYP.07600. Rev. 06/17/2010
As compared to College of American Pathologists (CAP) benchmarks, RWE data showed comparable performance of the ExCellerator cervical collection device as a cervical spatula to the predicate device. Use of the ExCellerator cervical collection device resulted in a $54\%$ reduction in unsatisfactory rates and increased detection in key diagnostic categories while ASCUS/SIL ratio remains within $20^{\text{th}}$ percentile of the CAP National Data.
# b) Validation Study
This study assessed the performance of the subject device when used as an aid in transferring samples from the IVD labeled cytobrush into the PreservCyt solution.
Cervical specimens were collected from 108 patients for a routine Pap test using an FDA cleared spatula and endocervical brush. These specimens were processed for Pap testing per standard of care. Following the routine Pap test collection, rather than discarding the cytobrush, the clinician transferred the residual specimen material from the cytobrush into a new vial containing the PreservCyt solution using the ExCellerator spatula and the ExCellerator technique, as described in Principles of Operation. The specimen was then sent to the study lab, processed and assessed for adequacy according to TBS 2014.
{7}
The Study results showed that out of 108 validation samples tested, 102 (94%) yielded satisfactory pap smears according to TBS 2014 and thus met the study acceptance criterion. This study demonstrated that the ExCellerator spatula is effective in aiding in more complete transfer of cytological material from the endocervical brush (cytobrush) into the PreservCyt solution.
c) Biocompatibility Studies
The subject device is a surface device that contacts the skin and mucosal membrane for a limited contact duration (<24 hour). Per the FDA Guidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical device – Part 1: Evaluation and testing within a risk management process", the following tests are conducted:
- Cytotoxicity (ISO 10993-5: 2009)
- Sensitization (ISO 10993-10: 2010)
- Vaginal Irritation or Intracutaneous Reactivity (ISO 10993:10)
Cytotoxicity
The MEM (Minimum Essential Medium) Elution Cytotoxicity Assay was performed to determine the cytotoxic response from the ExCellerator Cervical Collection Device. The study was conducted in compliance with the International Organization for Standardization (ISO) 10993-5: 2009. The test article (ExCellerator) was extracted in cell culture media which was then plated onto the L-929 mouse fibroblast cell monolayer. The test and control article extracts were plated onto the cell monolayer. The monolayers were evaluated for cell death at three time points: 24, 48 and 72 hours (+/- 3 hours).
Validity Criteria: The control results met the validity criteria and the test was considered valid. No reactivity was observed in the Control Article 01 (cell media control) wells and Control Article 02 (negative control) wells as they both received a grade of '0' at 72 hours and were considered non-toxic. Complete or nearly complete destruction of the cell monolayer was observed in Control Article 03 (positive control) wells as they received a grade of '4' at 72 hours and was considered cytotoxic. Therefore, the validation criteria were met.
Final Result: No reactivity was observed; no cell lysis or reduction of cell growth was observed in the triplicate wells. The test article received a grade of '0' at 24, 48 and 72 hours and was considered Non-Cytotoxic.
Study Outcome: Pass (Non-Cytotoxic).
Sensitization
The Guinea Pig Maximization Test was performed to evaluate the sensitization or allergenic potential of the ExCellerator Cervical Collection Device. This test is used as a method for screening contact allergens in guinea pigs. The results are used as predictive measures for detecting potential sensitizers in humans. The study was conducted in
8
{8}
compliance with the International Organization for Standardization (ISO) 10993-10: 2010. A total of 18 animals (12 test and 6 control) per test article (ExCellerator) were used. The study was comprised of two induction phases (intradermal and topical) and one challenge phase. The intradermal induction phase (Induction I) used a series of intradermal injections consisting of Freund's complete adjuvant and the test article and negative control extract and/or selected solvent with Freund's to enhance the state of immunological activity of the animal to the material of interest. Prior to the start of the topical induction phase (Induction II), a pretreatment of 10% sodium lauryl sulfate (SLS) was used as an inflammatory precursor to maximize uptake of test article or negative control applied to the treatment site. Subsequently, within 24 +/- 2 hours, the test article was applied to the intrascapular region of each animal using a patch (filter paper) covering the Induction I injection sites. After the induction phases, the animals were on an observatory period for 14 +/- 1 day. All test and control animal groups were then challenged with the test article (challenge phase). The test article and negative control were administered to the upper right and left flanks of each animal for 24 +/- 2 hours. The challenge skin sites were then observed for evidence of skin sensitization at 24 +/- 2 hours and 48 +/- 2 hours after the removal of the dressings. All the animals on study were euthanized at the end of the survival period.
Validity Criteria: Magnusson and Kligman grades of 1 or greater in the test group generally indicate sensitization, provided grades of less than 1 are seen in control animals. The outcome of the test is presented as the frequency of positive challenge results in the test and control animals. All assay validity criteria were met.
Final Result: The overall pattern, intensity and duration of the reactions in the test group were compared with the negative control group. The overall frequency of the test scores was similar to the control scores. The test article showed no evidence of delayed dermal contact sensitization in the guinea pig.
Study Outcome: Completed (Non-sensitizer)
## Irritation
Vaginal Mucosal Irritation Test was performed to evaluate the irritation potential of the ExCellerator Cervical Collection Device. This test is used as a method for screening irritants in rabbits. The results are used as predictive measures for detecting potential irritants in humans. The study was conducted in compliance with the International Organization for Standardization (ISO) 10993-10: 2010. A total of 6 animals per extract were used. At a dosage regimen of 1 mL per day for five consecutive days, the test article (ExCellerator) was administered intravaginally. Observations were made prior to initial dosing, at 24 +/- 2 hours after the initial application, immediately prior to each subsequent treatment and prior to euthanasia. The appearance of vaginal opening and perineum was documented for signs of discharge, erythema and edema. At 24 +/- 2 hours after the final dose the animals were euthanized. The entire vagina was dissected free, opened longitudinally and examined for signs of irritation, injury to the epithelial layer of tissue and necrosis. The animals were processed for histological examination performed by the study pathologist. The results of the study were interpreted based upon the frequency of animals displaying irritation responses.
Validity Criteria: All assay validity criteria were met.
{9}
Final Test Score: To obtain the final irritation index, the microscopic evaluation grades for each group were added and the sum was divided by the number of observations to obtain a test group average. The control group average was subtracted from the test group average to obtain the irritation index. The final test article score was 0 for Normal Saline (NS) and 0 for Sesame Seed Oil (SSO), indicating that an irritation response was not observed for either extract.
Study Outcome: Completed (Non-irritant)
Based on the data provided, the conducted biocompatibility tests are acceptable.
## d) Mechanical Tests
Structural Assessments – Summary:
Two groups of ExCellerator devices (A-new, B-aged 7 years), were assessed for Static Bend (Static Cantilever Bend Test), Static Tension and Static Torsion.
Structural Assessments – Criteria:
**Static Bend**
- Clinically Relevant Load (CRL): Ranges from .01 - .05 lbs. or .044 - .222 N
- Acceptance Criteria: At Max CRL of .222 N - Device experiences ≤ 2-3mm displacement (bend)
**Static Tension**
- Clinically Relevant Force (CRF): Ranges from .10 - .50 lbs. or .44 - 2.22 N
- Acceptance Criteria: At Max CRF of 2.22 N - Device experiences ≤ 1mm deformation
**Static Torsion**
- Clinically Relevant Torque (CRT): Ranges from .01 - .05 inch-lbs. or .0011 - .0056 Nm
- Acceptance Criteria: At Max CRT of .0056 Nm - Device experiences ≤ 45 degrees displacement (angular)
Structural Assessments – Results:
**Static Bend**
- Acceptance Criteria: ≤ 2-3mm displacement (bend)
- Group Mean Displacement Results
- Group A (New – Produced 2017): Mean Displacement - 0.147mm
- Group B (Aged – Produced 2009): Mean Displacement - 0.157mm
**Static Tension**
- Acceptance Criteria: ≤ 1mm deformation
- Group Mean Deformation Results
- Group A (New – Produced 2017): Mean Deformation - 0.0147mm
- Group B (Aged – Produced 2009): Mean Deformation - 0.0137mm
**Static Torsion**
10
{10}
- Acceptance Criteria: ≤ 45.00 degrees displacement (angular)
- Group Mean Angular Displacement Results
- Group A (New – Produced 2017): Mean Displacement - 14.44 degrees
- Group B (Aged – Produced 2009): Mean Displacement - 11.19 degrees
Mechanical Testing Interpretation:
Mirrored structural assessment of New (2017) and (Aged 2009) devices – Results:
- No indication of any meaningful variance in performance between the groups
- No individual device from either group approaches failure to acceptance criteria
All tests met the acceptance criteria.
K. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Parts 801 and 809, as applicable.
L. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.