K182032 · Aesculap, Inc. · KCT · Dec 21, 2018 · General Hospital
Device Facts
Record ID
K182032
Device Name
SterilContainer S System
Applicant
Aesculap, Inc.
Product Code
KCT · General Hospital
Decision Date
Dec 21, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The SterilContainer S System is a reusable sterilization container system (consisting of perforated bottoms and perforated lids with filter retention plates, and single use polypropylene filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. This container system has been validated with flexible endoscopes and accessories, cables, and camera heads. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use with the STERRAD 100NX DUO Cycle. The SterilContainer S System includes accessories such as silicone mats, baskets, trays, racks, filters, indicator cards, locks, and instrument holders.
Device Story
SterilContainer S System is a reusable rigid container for sterilization and storage of medical devices; constructed of non-anodized aluminum with disposable polypropylene filters. Used in healthcare facilities to enclose instruments (flexible endoscopes, cables, camera heads) during STERRAD 100NX DUO sterilization cycles. System includes accessories: silicone mats, baskets, trays, racks, and tamper-proof locks. Device functions as a protective barrier; allows sterilant penetration while maintaining sterility post-process. Healthcare providers place instruments in containers; process in STERRAD 100NX DUO; store until needed. Ensures sterility maintenance for up to 365 days. Benefits include protection of delicate instruments and sterile storage.
Clinical Evidence
Bench testing only. Validated using biological overkill method and half-cycle validation in STERRAD 100NX DUO cycle. Demonstrated 12-log reduction and SAL of 10^-6. Simulated use testing achieved 6-log reduction of G. stearothermophilus. Sterility maintenance validated for 365 days.
Indicated for use by healthcare providers to enclose medical devices, including flexible endoscopes, accessories, cables, and camera heads, for sterilization and maintenance of sterility until use.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA acronym is in blue, followed by the words "U.S. Food & Drug Administration" in blue.
December 21, 2018
Aesculap, Inc. Paul Amudala Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K182032
Trade/Device Name: SterilContainer S System Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: November 20, 2018 Received: November 23, 2018
Dear Paul Amudala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Elizabeth F. Claverie -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182032
Device Name SterilContainer™ S System
#### Indications for Use:
The SterilContainer S System is a reusable sterilization container system (consisting of perforated bottoms and perforated lids with filter retention plates, and single use polypropylene filters) intended to be used to enclose another medical device that is to be sterilized by a healthcare provider. This container system has been validated with flexible endoscopes and accessories, cables, and camera heads. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use with the STERRAD 100NX DUO Cycle. The SterilContainer S System includes accessories such as silicone mats, baskets, trays, racks, filters, indicator cards, locks, and instrument holders.
Validation testing for event related sterility maintenance has been conducted for up to 365 days.
| Load Configuration | Container # 1 | Container # 2 |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | • 135mm Full Size Container Bottom<br>• Full Size Container Lid<br>• Full Size Basket<br>• Full Size Mat<br>• Flexible Endoscope (1.2 mm x 785 mm)<br>• Guide Cable (1x) | • 135mm Full Size Container Bottom<br>• Full Size Container Lid<br>• Full Size Basket<br>• Full Size Mat<br>• Flexible Endoscope (1.2 mm x 785 mm)<br>• Guide Cable (2x) |
| 2 | • 187mm Full Size Container Bottom<br>• Full Size Container Lid<br>• Full Size Basket<br>• Full Size Mat<br>• Flexible Endoscope (1.2 mm x 785 mm)<br>• Guide Cable (2x) | • N/A – Top shelf is removed to allow the container to fit in the chamber |
| 3 | • 90mm Full Size Container Bottom<br>• Full Size Container Lid<br>• Full Size Basket<br>• Full Size Mat<br>• Flexible Endoscope (1.2 mm x 785 mm)<br>• Guide Cable (2x) | • 135mm Half Size Container Bottom<br>• Half Size Container Lid<br>• Half Size Basket<br>• Half Size Mat<br>• Camera Head for Endoscope |
#### Table 1: Validated STERRAD 100NX DUO Cycle Load Configurations
{3}------------------------------------------------
| Lid | Bottom | Description | Container Load Weight * |
|-------|--------|------------------------|---------------------------------------------------------------|
| JM489 | JM440 | Full Size 90mm (4 ¼") | |
| | JM441 | Full Size 120mm (5 ½") | |
| | JM442 | Full Size 135mm (6") | |
| | JM444 | Full Size 187mm (8") | |
| JM789 | JM740 | ¾ Size 90mm(4 ¼") | 13.2 lbs total weight. (1 or 2<br>shelves based on container) |
| | JM741 | ¾ Size 120mm (5 ½") | |
| JM389 | JM340 | ½ Size 90mm (4 ¼") | |
| | JM341 | ½ Size 120mm (5 ½") | |
| | JM342 | ½ Size 135mm (6") | |
| | JM344 | ½ Size 187mm (8") | |
#### Table 2: STERRAD 100NX DUO Compatible SterilContainer S Container Systems
*Loads containing a flexible endoscope or bronchoscope should follow Table 1 recommendations
#### Table 3: STERRAD 100NX DUO Cycle Compatible Accessories
| Accessories | STERRAD 100NX DUO<br>Cycle |
|--------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Stainless Steel baskets, basket lids, and dividers | Yes |
| Instrument Organization System (Silicone and Stainless Steel racks,<br>brackets, instrument holders, and clamps) | Yes |
| Silicone mats | Yes |
| Stainless Steel racks, trays, instrument holders, clamps, brackets,<br>tamper-proof locks, indicator cards and platforms | Yes |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) SUMMARY (as required by 21 CFR 807.92)
SterilContainer" S System for use in the STERRAD 100NX DUO Cycle
Dec 13, 2018
| COMPANY: | Aesculap®, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
|----------------------|--------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Paul Amudala<br>Regulatory Affairs Specialist<br>610-984-9303 (phone)<br>610-791-6882 (fax) |
| TRADE NAME: | SterilContainer S System |
| COMMON NAME: | Sterilization Container Wrap |
| CLASSIFICATION NAME: | Sterilization Wrap |
| REGULATION NUMBER: | 21 CFR 880.6850 |
| PRODUCT CODE: | KCT |
| DEVICE CLASS: | Class II |
## PREDICATE DEVICE
K093493 - SterilContainer™ S System for use in STERRAD 200, STERRAD NX (STANDARD (Std.) & ADVANCED (Adv.) and STERRAD 100NX (Std. & FLEX)
#### REFERENCE DEVICE
K142970- SterilContainer™ S for use in STERRAD 100NX EXPRESS Cycle via K142970.
## DEVICE DESCRIPTION
The SterilContainer™S System is a container system that will allow for sterilization and storage of medical devices. This container system can be used in the STERRAD 100NX DUO Sterilization System. The SterilContainer" S System rigid containers are made from nonanodized Aluminum and utilize disposable (single use) polypropylene filters. The SterilContainer" S System includes accessories such as mats, baskets, trays, instrument holders, organizers, filters, indicator cards and tamper proof locks.
#### Indications for Use:
The SterilContainer S System is a reusable sterilization container system (consisting of perforated bottoms and perforated lids with filter retention plates, and single use polypropylene filters) intended to be used to enclose another medical device that is to be sterilized by a
{5}------------------------------------------------
healthcare provider. This container system has been validated with flexible endoscopes and accessories, cables, and camera heads. It is intended to allow sterilization of the enclosed device and also maintain sterility of the enclosed device until used. This container system is compatible for use with the STERRAD 100NX DUO Cycle. The SterilContainer S System includes accessories such as silicone mats, baskets, trays, racks, filters, indicator cards, locks, and instrument holders.
Validation testing for event related sterility maintenance has been conducted for up to 365 days.
| Load<br>Configuration | Container # 1 | Container # 2 |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | • 135mm Full Size Container<br>Bottom<br>• Full Size Container Lid<br>• Full Size Basket<br>• Full Size Mat<br>• Flexible Endoscope (1.2 mm x 785 mm)<br>• Guide Cable (1x) | • 135mm Full Size Container<br>Bottom<br>• Full Size Container Lid<br>• Full Size Basket<br>• Full Size Mat<br>• Flexible Endoscope (1.2 mm x 785 mm)<br>• Guide Cable (2x) |
| 2 | • 187mm Full Size Container<br>Bottom<br>• Full Size Container Lid<br>• Full Size Basket<br>• Full Size Mat<br>• Flexible Endoscope (1.2 mm x 785 mm)<br>• Guide Cable (2x) | • N/A – Top shelf is removed to<br>allow the container to fit in the<br>chamber |
| 3 | • 90mm Full Size Container<br>Bottom<br>• Full Size Container Lid<br>• Full Size Basket<br>• Full Size Mat<br>• Flexible Endoscope (1.2 mm x 785 mm)<br>• Guide Cable (2x) | • 135mm Half Size Container<br>Bottom<br>• Half Size Container Lid<br>• Half Size Basket<br>• Half Size Mat<br>• Camera Head for Endoscope |
Table 1: Validated STERRAD 100NX DUO Cycle Load Configurations
#### Table 2: STERRAD 100NX DUO Compatible SterilContainer S Container Systems
| Lid | Bottom | Description | Container Load Weight * |
|-------|--------|------------------------|---------------------------------------------------------------|
| JM489 | JM440 | Full Size 90mm (4 ¼") | 13.2 lbs total weight. (1 or 2<br>shelves based on container) |
| | JM441 | Full Size 120mm (5 ½") | |
| | JM442 | Full Size 135mm (6") | |
| | JM444 | Full Size 187mm (8") | |
| JM789 | JM740 | ¾ Size 90mm(4 ¼") | |
| | JM741 | ¾ Size 120mm (5 ½") | |
| JM389 | JM340 | ½ Size 90mm (4 ¼") | |
| | JM341 | ½ Size 120mm (5 ½") | |
| | JM342 | ½ Size 135mm (6") | |
| | JM344 | ½ Size 187mm (8") | |
*Loads containing a flexible endoscope or bronchoscope should follow Table 1 recommendations
{6}------------------------------------------------
| Accessories | STERRAD 100NX DUO<br>Cycle |
|--------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Stainless Steel baskets, basket lids, and dividers | Yes |
| Instrument Organization System (Silicone and Stainless Steel racks,<br>brackets, instrument holders, and clamps) | Yes |
| Silicone mats | Yes |
| Stainless Steel racks, trays, instrument holders, clamps, brackets,<br>tamper-proof locks, indicator cards and platforms | Yes |
#### Table 3: STERRAD 100NX DUO Compatible Accessories
# TECHNOLIGICAL CHARACTERISTICS COMPARISON TABLE
The SterilContainer S System for proposed use in the STERRAD 100NX DUO is the same container system that was cleared in K093493 & K142970. The materials and design have not changed.
| System | SterilContainer S<br>System (K182032) | SterilContainer S System<br>(K093493) Primary<br>Predicate | SterilContainer S<br>System (K142970)<br>Reference Device | Comparison |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Sterilization<br>process | STERRAD 100NX<br>DUO | STERRAD 200 ,<br>STERRAD NX &<br>Adv. , STERRAD<br>100 NX Std. &<br>FLEX Cycle | STERRAD 100NX<br>EXPRESS | Different |
| Material | Non-<br>anodized<br>aluminum | Non-anodized aluminum | Non-anodized<br>aluminum | Same |
| Container<br>type | Perforated | Perforated | Perforated | Same |
| Filter type | Polypropylene | Polypropylene | Polypropylene | Same |
| Gasket | Silicone | Silicone | Silicone | Same |
| Sizes | Full, 1/2, 3/4 | Full, ½, ¾, ¼, XL Mini | Full,1/2 | Similar |
| Accessories | Stainless Steel<br>baskets, basket lids,<br>dividers, racks, trays,<br>holders clamps;<br>instrument<br>organization systems;<br>silicone mats; tamper<br>proof locks; indicator<br>cards | Stainless Steel baskets,<br>basket lids, dividers,<br>racks, trays, holders<br>clamps; instrument<br>organization systems;<br>silicone mats; tamper<br>proof locks; indicator<br>cards | Stainless Steel<br>baskets, basket lids,<br>dividers, racks,<br>trays, holders<br>clamps; instrument<br>organization<br>systems; silicone<br>mats; tamper proof<br>locks; indicator<br>cards | Same |
| Reuse | Reusable | Reusable | Reusable | Same |
{7}------------------------------------------------
## NON-CLINICAL PERFORMANCE DATA
The Aesculap SterilContainer S System has been validated for the STERRAD 100NX DUO Cycle. These validations were conducted by a qualified testing laboratory. The performance testing demonstrates that the device meets the acceptance criteria for each test.
| Performance Testing | Results |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization Efficacy/<br>Lethality Study | Testing demonstrated a 12 log reduction and a sterility assurance<br>level (SAL) of $10^{-6}$ using the biological (BI) overkill method and<br>half-cycle validation. |
| Maintenance of<br>Sterility | Maintenance Sterility for the proposed devices remains<br>unchanged and continues to provide an effective barrier for<br>maintaining sterility of the contents after sterilization for 365<br>days event related storage under conditions which simulate<br>hospital sterile package handling and storage conditions still<br>applies for the DUO cycle. |
| Simulated Use | Testing demonstrated that a 6-log reduction of G.<br>stearothermophilus suspended in artificial soil medium is<br>achieved for the Aesculap, Inc. SterilContainer S Containers<br>when processed through a full cycle in the STERRAD 100NX<br>DUO Cycle in healthcare facilities. |
| Inactivation Kinetics | No growth was observed from three different injection<br>doses for load configuration 1, 2 and 3 respectively. |
| Aerosol Challenge | The Aerosol challenge remains unchanged from the predicates as<br>the container as well as the gaskets of the container remain the<br>same as cleared in predicates.<br>Predicate device data is still relevant for the proposed device. |
| Material Compatibility | Material Compatibility test data has not changed and remains<br>the same as cleared in predicate. |
| Biocompatibility | The biocompatibility of the container remains unchanged from<br>the predicates as the materials of the container remains the<br>same as cleared in the predicate. |
## CONCLUSION
The non-clinical testing demonstrates that the SterilContainer S System for use in the STERRAD 100NX DUO Cycle is as safe, as effective and performs as well as or better than the legally marketed predicate device, the Aesculap SterilContainer S System for use in the STERRAD 200, STERRAD NX (Std. & Adv.) and STERRAD 100NX (Std. and Flex) Cycles (K093493) .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.