ApexMV is a non-implantable muscle stimulator intended to provide electrical stimulation and/or visual biofeedback (via manometry) for the treatment of stress, urge, or mixed urinary incontinence and/or fecal incontinence in adult women.
Device Story
ApexMV is a non-implantable muscle stimulator for home use by adult women to treat urinary and fecal incontinence. The device consists of a handheld control unit connected via flexible tubing to a customizable inflatable probe. The user inserts the probe vaginally or rectally and manually inflates it for a customized fit. The device delivers electrical stimulation via stainless steel electrodes on the probe to strengthen pelvic floor muscles (high frequency) and calm the detrusor muscle (low frequency). It also provides visual biofeedback via manometry, displaying pressure and contraction/relaxation cues on an OLED screen. The user controls stimulation levels and session start/stop via push-button controls. The device automatically stops after an approximately 11-minute session. By strengthening pelvic floor muscles and providing biofeedback, the device helps users improve muscle control, potentially reducing incontinence symptoms. It is distributed as an OTC kit including the control unit, probe, electrode gel, and instructional materials.
Clinical Evidence
No clinical trials were performed. Evidence includes a clinical literature evaluation of the predicate devices (ApexM and InToneMV) supporting the safety and efficacy of electrical stimulation for urinary and fecal incontinence. A Human Factors Usability Study was conducted to validate OTC use, confirming that users could self-identify conditions, identify contraindications, and safely operate the device using provided labeling and instructions.
Indicated for adult women with stress, urge, or mixed urinary incontinence and/or fecal incontinence. Contraindications include pregnancy, attempting pregnancy, cardiac demand pacemaker or implanted electrical device, active urinary tract infection, vaginal infections, localized lesions, extra-urethral or overflow incontinence, severe urine retention, poor pelvic sensation, cognitive disabilities (e.g., Alzheimer's, dementia), active pelvic cancer, intestinal clamp, or less than 6 weeks post-pelvic surgery/childbirth.
Regulatory Classification
Identification
A nonimplanted electrical continence device is a device that consists of a pair of electrodes on a plug or a pessary that are connected by an electrical cable to a battery-powered pulse source. The plug or pessary is inserted into the rectum or into the vagina and used to stimulate the muscles of the pelvic floor to maintain urinary or fecal continence. When necessary, the plug or pessary may be removed by the user. This device excludes an AC-powered nonimplanted electrical continence device and the powered vaginal muscle stimulator for therapeutic use (§ 884.5940).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 15, 2019
InControl Medical, LLC Jennifer Koch Director of Quality & Regulatory Affairs 3225 Gateway Road Suite 250 Brookfield. WI 53045
K182022 Re: Trade/Device Name: ApexMV Regulation Number: 21 CFR§ 876.5320 Regulation Name: Nonimplanted Electrical Continence Device Regulatory Class: II Product Code: KPI Dated: February 5, 2019 Received: February 7, 2019
Dear Jennifer Koch:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Glenn B. Bell -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182022
Device Name ApexMV
Indications for Use (Describe)
ApexMV is a non-implantable muscle stimulator intended to provide electrical stimulation and/or visual biofeedback (via manometry) for the treatment of stress, urge, or mixed urinary incontinence in adult women.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for InControl Medical, followed by their address. The address is 3225 Gateway Road, Ste. 250 in Brookfield, WI 53045. The logo is a light blue circle with the word "InControl" in dark blue letters. The word "MEDICAL" is in smaller, dark blue letters to the right of "InControl".
Image /page/3/Picture/3 description: The image contains the logo for ApexMV. The logo features the text "apexMV" in a sans-serif font, with the "apex" portion in a lighter gray and the "MV" portion in a darker gray. Above the "apex" portion of the text is a green, stylized flower-like graphic with six petals.
# 7. 510(k) Summary
## Date Prepared
Friday, March 15th, 2019
## Submitter Information
Corey Olson Regulatory Affairs Manager InControl Medical, LLC 3225 Gateway Road, Suite 250, Brookfield, WI 53045 USA Phone: (262) 373.0422 Fax: (262) 373.0463 Email: colson@incontrolmedical.com
## Device Information
Table 7.1: Device Information
| Type of 510(k): | Traditional 510(k) |
|--------------------------------|-------------------------------------------|
| Common Name: | Pelvic Floor Muscle Stimulator |
| Trade Name (proprietary name): | ApexMV |
| Classification name: | Nonimplanted Electrical Continence Device |
| Classification Regulation: | 21 CFR 876.5320 |
| Class: | Class II |
| Product Code: | KPI |
## Legally Marketed Device for Substantial Equivalence
Table 7.2: Predicate Device Information
| 510(k) | Name | Product Code | Owner |
|---------|----------|--------------|-----------------------------------------------------------------------------------|
| K134020 | InToneMV | KPI | InControl Medical, LLC<br>3225 Gateway Road, Ste. 250<br>Brookfield, WI 53045 USA |
| K150183 | ApexM | KPI | InControl Medical, LLC<br>3225 Gateway Road, Ste. 250<br>Brookfield, WI 53045 USA |
{4}------------------------------------------------
# Traditional 510(k) Subm
Image /page/4/Picture/2 description: The image shows the logo and address of InControl Medical. The logo features the text "InControl MEDICAL" in a serif font, with the word "MEDICAL" in a smaller font size. The address is "3225 Gateway Road, Ste. 250 Brookfield, WI 53045".
#### Device Summary
The ApexMV device is a non-implantable muscle stimulator intended to provide electrical stimulation and/or visual biofeedback (via manometry) for the treatment of stress, urge, or mixed urinary incontinence and/or fecal incontinence in adult women for Over-the-Counter use. The ApexMV device is to be distributed as a kit that contains the following items:
- · 1 ApexMV device
- 1 4 pack of AA batteries
- · 1 2 oz. tube of InControl Medical Electrode Gel**
- 1 ApexMV IFU/ User Manual
- · 1 ApexMV Quick Reference Card
- 1 Screw driver
- 1 Travel Bag
- 1 Animation Video
**The InControl Medical Electrode Gel is privately labeled for InControl Medical. It is a Pre-Amendment Class II Medical Device under D039725.
The kit contents are substantially equivalent to the ApexM (K150183) and InToneMV (K134020) that are distributed on the market by InControl Medical. The inclusion of the screw driver in the ApexMV kit is for the screw that holds the battery cover panel in place on the back side of the control unit.
The ApexMV device consists of a hand held control unit with bio-feedback and inflation pump attached to a customizable inflatable probe via flexible tubing. The customizable inflatable probe is inserted vaginally or rectally and manually inflated by the end user to ensure a customized fit. Electrical stimulation is delivered via stainless steel electrodes on the inflatable probe to induce a contraction of the pelvic floor muscles. Muscle stimulation is used to train and strengthen the pelvic floor muscles in a controlled manner. Muscle stimulation is also used to improve the ability of the muscles to hold a contraction for an extended period of time for the treatment of stress, urge, or mixed urinary and/or fecal incontinence in adult women. During a session, high and low frequency stimulation is delivered by the device. The higher frequency stimulation is delivered to specific muscles to encourage their contraction, strengthening the muscles and helping the end user recognize which muscles to activate during self-directed contractions. Lower frequency stimulation calms the detrusor muscle, decreasing symptoms of urgency. The level of electrical stimulation is easily controlled by the end user using manual, push-button controls.
## Indications for Use
ApexMV is a non-implantable muscle stimulator intended to provide electrical stimulation and/or visual biofeedback (via manometry) for the treatment of stress, urge, or mixed urinary incontinence and/or fecal incontinence in adult women.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for InControl Medical, with the words "InControl" in large, bold letters and the word "Medical" in smaller letters below. The logo also includes a circular design with curved lines. Below the logo is the address "3225 Gateway Road, Ste. 250".
Image /page/5/Picture/3 description: The image shows the logo for ApexMV. The logo is in a light gray color, with the letters "apexMV" written in a simple, sans-serif font. Above the "x" in "apex" is a small, stylized flower or starburst design in a teal color. The logo is clean and modern in appearance.
Brookfield, WI 53045
# Equivalence Comparison to the Predicate
Electrical muscle stimulation is the technological principle for ApexMV and the predicate devices, ApexM (K150183) and InToneMV (134020). It is based on the use of electrical muscle stimulation to strengthen the pelvic floor muscles and surrounding structures. The chart below summarizes the shared and unique technological elements between ApexMV and the predicate devices, ApexM (K150183) and InToneMV (K134020). The intended use, technology, engineering, and performance for the ApexMV is substantially equivalent to the predicate devices, ApexM (K150183) and InToneMV (134020).
| Feature/<br>Function | K134020 InToneMV<br>(Predicate)<br>manufactured by<br>InControl Medical,<br>LLC. | K150183 ApexM<br>(Predicate)<br>manufactured by<br>InControl Medical,<br>LLC. | ApexMV | Comparison | Impact on<br>Safety and<br>Performance |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>An explicit<br>description of<br>all clinical<br>functions<br>performed by<br>the device<br>Indications for<br>Use<br>Explain when<br>the device is to<br>be clinically<br>used and the<br>intended<br>patient<br>population | InToneMV is intended<br>to provide electrical<br>stimulation and/or<br>visual biofeedback<br>(via manometry) for<br>the treatment of<br>male and female<br>urinary and fecal<br>incontinence | ApexM is a non-<br>implanted muscle<br>stimulator designed<br>to treat, stress, urge,<br>and/or mixed urinary<br>incontinence in<br>women. It applies<br>stimulation to the<br>pelvic floor muscles<br>to improve strength<br>and support. | ApexMV is a non-<br>implantable muscle<br>stimulator intended to<br>provide electrical<br>stimulation and/or<br>visual biofeedback (via<br>manometry) for the<br>treatment of stress,<br>urge, or mixed urinary<br>incontinence and/or<br>fecal incontinence in<br>adult women. | Substantially<br>equivalent | None:<br>These devices<br>apply the same<br>electrical<br>stimulation to<br>strengthen the<br>pelvic floor<br>muscles, as<br>supported by<br>literature. |
| Primary<br>Function | Delivery of electrical<br>stimulation<br>Visual biofeedback | Delivery of electrical<br>stimulation | Delivery of electrical<br>stimulation<br>Visual biofeedback | Identical to<br>InToneMV | None |
| Warnings or<br>Precautions | (See product labeling) | (See product<br>labeling) | (See product labeling) | Substantially<br>equivalent | None:<br>ApexMV<br>warnings and<br>precautions are<br>identical to<br>ApexM and<br>InToneMV,<br>which were |
| Feature/<br>Function | K134020 InToneMV<br>(Predicate)<br>manufactured by<br>InControl Medical,<br>LLC. | K150183 ApexM<br>(Predicate)<br>manufactured by<br>InControl Medical,<br>LLC. | ApexMV | Comparison | Impact on<br>Safety and<br>Performance |
| | | | | | defined in<br>accordance<br>with the FDA<br>guidance<br>document and<br>validated<br>through the<br>Human Factors<br>Usability<br>Testing. |
| Contraindications<br>Explain when<br>the device is<br>not to be<br>clinically used | • Patients with a<br>pacemaker or<br>implanted<br>defibrillator require<br>cardiac clearance<br>before electrical<br>stimulation is<br>offered.<br>• Patients with<br>symptoms of active<br>urinary tract<br>infection, vaginal<br>infections, localized<br>lesions, or other<br>undiagnosed<br>symptoms.<br>• Patient has extra-<br>urethra<br>incontinence, (i.e.<br>syrinx, ectopic,<br>urethra).<br>• Patient has<br>overflow<br>incontinence<br>caused by<br>evacuation<br>problems.<br>• Patient has severe<br>urine retention in<br>the upper urethra<br>or other symptoms<br>of urine retention | • Do not use if you<br>are pregnant<br>• Do not use if you<br>are attempting to<br>get pregnant<br>• Do not use if you<br>have a cardiac<br>demand<br>pacemaker or<br>implanted<br>defibrillator<br>• Do not use if you<br>have symptoms of<br>active urinary tract<br>infection, vaginal<br>infections, or<br>localized lesions<br>• Do not use if you<br>have a diagnosis of<br>extra-urethral or<br>overflow<br>incontinence<br>• Do not use if you<br>have severe urine<br>retention<br>• Do not use if you<br>have poor<br>sensation in the<br>pelvic region<br>• Do not use if you<br>have cognitive<br>disabilities, i.e.: | • Do not use if you are<br>pregnant<br>• Do not use if you are<br>attempting to get<br>pregnant<br>• Do not use if you<br>have a cardiac<br>demand pacemaker<br>or implanted<br>electrical device<br>• Do not use if you<br>have symptoms of<br>active urinary tract<br>infection, vaginal<br>infections, or<br>localized lesions<br>• Do not use if you<br>have a diagnosis of<br>extra-urethral or<br>overflow<br>incontinence<br>• Do not use if you<br>have severe urine<br>retention<br>• Do not use if you<br>have poor sensation<br>in the pelvic region<br>• Do not use if you<br>have cognitive<br>disabilities, i.e.;<br>Alzheimer's disease<br>or dementia | Substantially<br>Equivalent | None |
| Feature/<br>Function | K134020 InToneMV<br>(Predicate)<br>manufactured by<br>InControl Medical,<br>LLC. | K150183 ApexM<br>(Predicate)<br>manufactured by<br>InControl Medical,<br>LLC. | ApexMV | Comparison | Impact on<br>Safety and<br>Performance |
| | • Patients with<br>neurological<br>deficiency that does<br>not permit proper<br>sensory perception<br>of stimulation or<br>complete<br>denervation of the<br>pelvic floor.<br>• Patients with<br>cognitive<br>disabilities, i.e.;<br>Alzheimer's disease<br>or dementia.<br>• Patients who are<br>currently pregnant<br>or attempting to get<br>pregnant.<br>• Patients with<br>anatomical pelvic<br>structures that do<br>not permit proper<br>and complete<br>placement of the<br>Insertion Unit.<br>• Patients with active<br>pelvic malignancy.<br>• Patients with an<br>intestinal clamp.<br>• Patients should be 6<br>weeks post-pelvic<br>surgery or vaginal<br>childbirth.<br>• Device should not<br>be used for<br>diagnostic purposes<br>or critical patient<br>monitoring.<br>• Device is not<br>(external)<br>defibrillator-proof. | Alzheimer's disease<br>or dementia<br>• Do not use if you<br>are unable to<br>properly insert the<br>device per<br>instructions<br>• Do not use if you<br>have active pelvic<br>cancer<br>• Do not use if you<br>have an intestinal<br>clamp<br>• You must be 6<br>weeks post-pelvic<br>surgery or vaginal<br>childbirth to use<br>this device<br>• Do not use this<br>device for<br>diagnostic<br>purposes or critical<br>patient monitoring<br>• This device is not<br>(external)<br>defibrillator-proof | • Do not use if you<br>have active pelvic<br>cancer<br>• Do not use if you<br>have an intestinal<br>clamp<br>• You must be 6<br>weeks post-pelvic<br>surgery or vaginal<br>childbirth to use this<br>device | | |
| Feature/<br>Function | K134020 InToneMV<br>(Predicate)<br>manufactured by<br>InControl Medical,<br>LLC. | K150183 ApexM<br>(Predicate)<br>manufactured by<br>InControl Medical,<br>LLC. | ApexMV | Comparison | Impact on<br>Safety and<br>Performance |
| Labeling<br>Summary<br>Clarity to<br>ensure safer or<br>more effective<br>use | User Manual and<br>Quick Reference<br>Patient Guide Card | User Manual and<br>Quick Reference<br>Patient Guide Card | User Manual and<br>Quick Reference<br>Patient Guide Card | Identical to<br>predicates | None:<br>ApexMV User<br>Manual and<br>Quick<br>Reference<br>Patient Guide<br>Card user were<br>based on the<br>labeling and<br>use for ApexM<br>for Over-The-<br>Counter use<br>and was<br>validated<br>through the<br>Human Factors<br>Usability<br>Testing. |
| Environmental<br>Specifications | For indoor use only | For indoor use only | For indoor use only | Identical to<br>predicates | None |
| Power Source | 4/5 AA nickel metal<br>hydride battery | 4-AAA Alkaline<br>battery | 4 -AA Alkaline battery | Substantially<br>equivalent | None |
| Method of line<br>current<br>isolation | N/A (battery) | N/A (battery) | N/A (battery) | Identical to<br>predicates | None |
| Patient<br>leakage<br>current | N/A (battery) | N/A (battery) | N/A (battery) | Identical to<br>predicates | None |
| Number of<br>output modes | 1 | 1 | 1 | Identical to<br>predicates | None |
| Number of<br>output<br>channels | 1 | 1 | 1 | Identical to<br>predicates | None |
| Regulated<br>current or<br>voltage? | Regulated voltage | Regulated voltage | Regulated voltage | Identical to<br>predicates | None |
| Firmware<br>controlled? | Yes | Yes | Yes | Identical to<br>predicates | None |
| Automatic<br>Overload Trip? | No | No | No | Identical to<br>predicates | None |
| Feature/<br>Function | K134020 InToneMV<br>(Predicate)<br>manufactured by<br>InControl Medical,<br>LLC. | K150183 ApexM<br>(Predicate)<br>manufactured by<br>InControl Medical,<br>LLC. | ApexMV | Comparison | Impact on<br>Safety and<br>Performance |
| Automatic No-<br>Load Trip? | No | No | No | Identical to<br>predicates | None |
| Automatic<br>Shut Off? | Yes | Yes | Yes | Identical to<br>predicates | None |
| Indicator<br>Display<br>• On/Off<br>Status<br>• Low<br>Battery | Yes (via display<br>illumination)<br>Yes | Yes (via display<br>illumination)<br>No | Yes<br>Yes (via on screen<br>warning) | Substantially<br>equivalent to<br>InToneMV | None |
| Waveform,<br>shape | Monophasic,<br>alternating polarity,<br>square pulse | Monophasic,<br>alternating polarity,<br>square pulse…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.