Renovis Tesera C/Tesera SC Anterior Cervical Fusion (ACF) System
Applicant
Renovis Surgical Technologies
Product Code
OVD · Orthopedic
Decision Date
Dec 7, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Renovis Tesera SC Stand-alone ACF System is indicated for intervertebral body fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Renovis Tesera SC Stand-alone ACF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six weeks of non-operative treatment prior to implantation. The Renovis Tesera SC Stand-alone ACF System is a stand-alone system when used with the cover plate and screws provided, and requires no additional supplemental fixation. When used as a stand-alone system, the use of three (3) screws and the cover plate assembly. When used without the cover plate and three screws the Renovis Tesera SC Stand-alone ACF System is a non-stand-alone system and requires additional supplemental fixation cleared by the FDA for use in the cervical spine to augment stability. The Renovis Tesera C ACF System is indicated for intervertebral body fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Renovis Tesera C ACF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six weeks of non-operative treatment prior to implantation. The Renovis Tesera C ACF System is a non-stand-alone system and requires additional supplemental fixation cleared by the FDA for use in the cervical spine to augment stability.
Device Story
System comprises cervical interbody fusion cages, bone screws, and anterior cover plate assembly. Cages feature proprietary Tesera Trabecular Technology porous titanium structure on superior/inferior surfaces to facilitate fusion. Used in cervical spine (C2-T1) for degenerative disc disease. Tesera SC functions as stand-alone system when used with cover plate and three screws; otherwise, or when using Tesera C, requires supplemental fixation. Implants are additively manufactured from Ti-6Al-4V. Surgeon implants device during fusion procedure to restore stability and promote bone growth. Benefits include structural support and potential for improved fusion via porous surface architecture.
Clinical Evidence
Bench testing only. Mechanical performance validated per ASTM F2077 (static/dynamic compression, shear, torsion) and ASTM F2267 (subsidence). Sterilization and packaging validation completed.
Technological Characteristics
Additively manufactured Ti-6Al-4V (ASTM F2924-14) interbody cages with porous trabecular structure. Instrumentation uses stainless steel (ASTM A564/A64M-13e1) or titanium alloy (ASTM F136-13). Terminally gamma sterilized. Stand-alone or non-stand-alone configurations.
Indications for Use
Indicated for intervertebral body fusion in skeletally mature patients with cervical degenerative disc disease at one level (C2-T1). Requires autogenous bone graft and at least six weeks of prior non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System (K153250)
Renovis S180 Lateral Lumbar Interbody Fusion System (K181655)
Submission Summary (Full Text)
{0}------------------------------------------------
December 7, 2018
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Renovis Surgical Technologies % Sharyn Orton Senior Consultant MEDIcept, Inc. 200 Homer Ave Ashland, Massachusetts 01721
Re: K182007
Trade/Device Name: Renovis Tesera C/Tesera SC Anterior Cervical Fusion (ACF) System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE, ODP Dated: November 5, 2018 Received: November 7, 2018
Dear Ms. Orton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone
(1-800-638-2041 or 301-796-7100).
Sincerely,
# Melissa Hall -S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182007
### Device Name
Renovis Tesera C/Tesera SC Anterior Cervical Fusion (ACF) System
### Indications for Use (Describe)
The Renovis Tesera SC Stand-alone ACF System is indicated for intervertebral body fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Renovis Tesera SC Stand-aCF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six weeks of non-operative treatment prior to implantation.
The Renovis Tesera SC Stand-alone ACF System is a stand-alone system when used with the cover plate and screws provided, and requires no additional supplemental fixation. When used as a stand-alone system, the use of three (3) screws and the cover plate assembly. When used without the cover plate and three screws the Renovis Tesera SC Stand-alone ACF System is a non-stand-alone system and requires additional supplemental fixation cleared by the FDA for use in the cervical spine to augment stability.
The Renovis Tesera C ACF System is indicated for intervertebral body fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Renovis Tesera C ACF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six weeks of non-operative treatment prior to implantation.
The Renovis Tesera C ACF System is a non-stand-alone system and requires additional supplemental fixation cleared by the FDA for use in the cervical spine to augment stability.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Traditional 510(k) Summary as required by 21 CFR 807.92(a) K182007
- A ) Submitted by: Renovis Surgical Technologies Inc. 1901 W. Lugonia Ave, Ste 340 Redlands, CA 92374 Phone: 909-557-2360 Fax: 909-307-8571
- Official Contact: Anthony DeBenedictis Vice President of Quality Assurance
- Consultant: Sharyn Orton, Ph.D. MEDIcept, Inc. 200 Homer Ave Ashland, MA 01721
- Date prepared: December 7, 2018
- B) Device Name: Intervertebral fusion device with integrated fixation, cervical Intervertebral fusion device with bone graft, cervical
- Proprietary Name: Renovis Tesera C/Tesera SC Anterior Cervical Fusion (ACF) System
- Device Class: Class II
- Regulation number: 21 CFR 888.3080
- Regulation name: Intervertebral body fusion device
- Product code: OVE - intervertebral fusion device with integrated fixation, cervical ODP - intervertebral fusion device, cervical
- Classification panel: Orthopedic
- C) Primary predicate: K153250-Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System
- D) Additional Predicate: K181655-Renovis S180 Lateral Lumbar Interbody Fusion System
- E) Device Description:
Tesera (or T3) describes Renovis' proprietary Tesera Trabecular Technology™ Porous Structure for the manufacture of the porous titanium on the superior and inferior surfaces of the implant.
{4}------------------------------------------------
The Renovis Tesera C/Tesera SC Anterior Cervical Fusion (ACF) System is a single, cervical implant System that includes two (2) "subsystems", the:
- Renovis Tesera C Anterior Cervical Fusion (ACF) System ●
- Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System .
The Renovis Tesera C/Tesera SC Anterior Cervical Fusion (ACF) System cages are additively manufactured from Ti-6Al-4V and are available in a variety of sizes (widths, heights, and lordotic angles) to suit the individual pathology and anatomical conditions of the patient.
The Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System includes cages, and existing bone screws (self-drilling and self-tapping options) and anterior cover plate assembly. When used with the cover plate and screws, the Tesera SC ACF System cages require no supplementary fixation systems. The screws protrude through the interbody portion of the device into the adjacent vertebral bodies. The cages have windows in the side for visualization of the bone graft. When used without the cover plate and screws, the Tesera SC ACF cages require supplemental fixation.
The Renovis Tesera C Anterior Cervical Fusion (ACF) System is not stand alone and requires supplemental fixation.
The Renovis Tesera C/Tesera SC ACF System cages comply with ASTM F2924-14 Standard Specification for Additive Manufacturing Titanium-6 Aluminum-4 Vanadium with Powder Bed Fusion.
The system also consists of reusable instrumentation. The instruments comply with ASTM A564/A64M-13e1 Standard Specification for Hot-Rolled and Cold-Finished Age-Hardening Stainless Steel Bars and Shapes or ASTM F136-13 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications.
# F) Indications For Use:
The Renovis Tesera SC Stand-alone ACF System is indicated for intervertebral body fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Renovis Tesera SC Stand-alone ACF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six weeks of non-operative treatment prior to implantation.
The Renovis Tesera SC Stand-alone ACF System is a stand-alone system when used with the cover plate and screws provided, and requires no additional supplemental fixation. When used as a stand-alone system, the cages require the use of three (3) screws and the cover plate assembly. When used without the cover plate and three screws the Renovis Tesera SC Stand-alone ACF System is a non-stand-alone system and requires additional supplemental fixation cleared by the FDA for use in the cervical spine to augment stability.
{5}------------------------------------------------
The Renovis Tesera C ACF System is indicated for intervertebral body fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2-T1. Renovis Tesera C ACF System implants are to be used with autogenous bone graft. Patients should be skeletally mature and have at least six weeks of non-operative treatment prior to implantation.
The Renovis Tesera C ACF System is a non-stand-alone system and requires additional supplemental fixation cleared by the FDA for use in the cervical spine to augment stability.
G) Substantial Equivalence Discussion
The Renovis Tesera C/Tesera SC ACF System has the same Indication for Use, includes the same geometric and lordotic offerings, are provided standalone or non-standalone, are terminally gamma sterilized, and are additively manufactured from biocompatible materials like the predicate devices.
Differences include manufacturing method, new geometric and lordotic options and packaging. These differences do not raise different issues of safety or effectiveness. Risk analysis conducted under Design Control assessed all differences. Successful manufacturing validation was conducted. The T3 porous shell was evaluated and successfully tested per standards. Mechanical testing of the implant/implant assemblies were tested per standards and found to be acceptable.
The Tesera C/Tesera SC ACF System is substantially equivalent to the predicate devices.
H) Performance Testing - Bench
Mechanical testing of the Renovis Tesera C/Tesera SC ACF System was conducted and found to be acceptable for the following:
- Static Compression, Dynamic Compression, Static Shear Compression, Dynamic Shear . Compression, Static Torsion and Dynamic Torsion Testing per ASTM F2077.
- Expulsion testing ●
- Subsidence testing with and without screws per ASTM F2267
Sterilization validation was completed.
# Conclusion
The Renovis Tesera C/Tesera SC ACF System has the same Indication for Use as the Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System.
Differences between the Renovis Tesera C/Tesera SC ACF System and the predicate devices were assessed under Design Control by risk analysis and FEA raises no new safety or effectiveness questions when compared to the predicate device. Mechanical testing was conducted and found to be acceptable. The Renovis Tesera SC ACF System is substantially equivalent to the Renovis Tesera SC Stand-alone Anterior Cervical Fusion (ACF) System.
{6}------------------------------------------------
- I) Consensus Standards and FDA Guidance
- AAMI/ANSI/ISO 10993-1:2009/(R)2013 Biological Evaluation Of Medical Devices ● Part 1: Evaluation And Testing Within a Risk Management Process
- ASTM F2077-14: Test Methods For Intervertebral Body Fusion Devices ●
- . ASTM F1854-15: Standard Test Method for Stereological Evaluation of Porous Coatings on Medical Implants
- ASTM F1147-05 (R)2017e1 Standard Test Method for Tension Testing of Calcium Phosphate and Metallic Coatings
- . ASTM F1044-05 (R)2017e1 Standard Test Method for Shear Testing of Calcium Phosphate Coatings and Metallic Coatings
- ASTM F1160-14 (R)2017e1 Standard Test Method for Shear and Bending Fatigue . Testing of Calcium Phosphate and Metallic Medical and Composite Calcium Phosphate/Metallic Coatings
- ASTM F1978-17 Standard Test Method for Measuring Abrasion Resistance of Metallic ● Thermal Spray Coatings by Using the Taber Abraser
- ASTM F2267-01(R)2011 Standard Test Method for Measuring Load Induced Subsidence ● of Intervertebral Body Fusion Device Under Static Axial Compression
- AAMI/ANSI ST72:2011/(R)2016 Bacterial Endotoxins - Test Methods, Routine Monitoring, And Alternatives To Batch Testing
- AAMI/ANSI/ISO 11137-2:2013 Sterilization Of Health Care Products Radiation Part ● 2: Establishing The Sterilization Dose
- . AAMI/ANSI/ISO 11737-1:2006 R(2011) Sterilization of health care products microbiological methods - part 1: determination of the population of microorganisms on product
- ISO 11607-2:2006 Packaging for Terminally Sterilized Medical Devices - Part 2: Validation requirements for forming, sealing and assembly processes
- ISTA 2A:2011 Packaged-products weighing 150 lbs. (68kg) or less
- ASTM D4169:2016 Standard Practice for Performance Testing of Shipping Containers ● and Systems
- ISO 17665-1 1st Edition 2006-08-15 Sterilization Of Health Care Products - Moist Heat -Part 1: Requirements For The Development, Validation And Routine Control Of A Sterilization Process For Medical Devices
- ASTM F 983-86 (Reapproved 2013) Standard Practice for Permanent Marking of ● Orthopaedic Implant Components
- ASTM F565-04 (Reapproved 2013) Standard Practice for Care and Handling of ● Orthopedic Implants and Instruments
- . Class II Special Controls Guidance Document: Intervertebral Body Fusion Device, June 2007
- Guidance for Industry and FDA Staff Technical Considerations for Additive . Manufactured Medical Devices, December 2017
- Guidance for Industry and FDA Staff Reporting of Computational Modeling Studies in ● Medical Device Submissions, September 2016
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.