Omni Hysteroscope, Omni Hysteroscope Kits, Omni Sheaths
K182006 · Hologic, Inc. · HIH · Dec 6, 2018 · Obstetrics/Gynecology
Device Facts
Record ID
K182006
Device Name
Omni Hysteroscope, Omni Hysteroscope Kits, Omni Sheaths
Applicant
Hologic, Inc.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Dec 6, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Omni Hysteroscope is used to provide viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.
Device Story
Omni Hysteroscope is a rigid, multi-use hysteroscope system consisting of a base scope and three interchangeable sheaths (3.7mm diagnostic, 5.5mm operative, 6mm operative) with removable outflow channels. Device provides visualization of cervical canal and uterine cavity during diagnostic and surgical procedures. Operated by physicians in clinical or operating room settings. System connects to external light sources (Metal Halide or Xenon ≤300w) for illumination. Provides direct optical visualization to the clinician to guide surgical instruments or diagnostic assessment. Benefits include improved access and visualization for intrauterine procedures. Device is supplied non-sterile and requires cleaning and sterilization between uses.
Clinical Evidence
Bench testing only. Performance verified through dimensional analysis, objective lens testing (field of view, focal length, depth of field), illumination levels, luminous energy transmission, magnification, and image quality (resolution, distortion). Biocompatibility confirmed per ISO 10993-1:2009. Electrical safety verified per ANSI AAMI ES 60601-1 and IEC 60601-2-18. Usability testing conducted per IEC 62366-1. Cleaning and sterilization validated per ISO 14937 and ISO 17665-1.
Technological Characteristics
Rigid rod lens hysteroscope. Materials: Stainless steel, Sapphire, Optical Glass, PEEK, Glass Fiber, PPSU (black, blue, green colorants), Nickel Silver. Dimensions: 200mm working length; 3.7mm to 6mm sheath diameters. Light source: Metal Halide or Xenon ≤300w. Sterilization: Moist heat (ISO 17665-1) or general sterilization (ISO 14937). Biocompatibility: ANSI AAMI ISO 10993-1:2009. Electrical safety: ANSI AAMI ES 60601-1:2005/(R)2012 & A1:2012, IEC 60601-2-18:2009.
Indications for Use
Indicated for viewing the cervical canal and uterine cavity to perform diagnostic and surgical procedures in patients requiring hysteroscopy.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 6, 2018
Hologic, Inc. Rachelle Fitzgerald Senior Regulatory Affairs Specialist 250 Campus Drive Marlborough, Massachusetts 01752
#### Re: K182006
Trade/Device Name: Omni Hysteroscope - Omni Hysteroscope (60-200), Omni 3.7 mm Diagnostic Sheath (60-201), Omni 5.5 mm Operative Sheath (60-202), Omni 6 mm Operative Sheath (60-203), Operating Room set (60-205-1), Office set (60-250-2) Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: Class II Product Code: HIH Dated: November 8, 2018 Received: November 9, 2018
Dear Rachelle Fitzgerald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sharon M. Andrews -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K182006
#### Device Name
Omni Hysteroscope - Omni hysteroscope (60-200), Omni 3.7 mm Diagnostic Sheath (60-201), Omni 5.5mm Operative Sheath (60-202), Omni 6mm Operative Sheath (60-203), Operating Room set (60-250-1), Office set (60-250-2)
#### Indications for Use (Describe)
The Omni Hysteroscope is used to provide viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(K) SUMMARY
Date: November 8, 2018
### 510(k) Submitter:
Hologic, Inc. 250 Campus Drive Marlborough, MA 01752 Attn: Rachelle D. Fitzgerald P: 508.263.8631 F: 877.793.6434
### Establishment Registration Number: 1222780
Trade Name: Omni Hysteroscope - Omni hysteroscope (60-200), Omni 3.7 mm Diagnostic Sheath (60-201), Omni 5.5mm Operative Sheath (60-202), Omni 6mm Operative Sheath (60-203), Operating Room set (60-250-1), Office set (60-250-2)
Common/Usual Name: Hysteroscope
Regulation Name: Hysteroscope and Accessories
Regulation Number: 21 CFR 884.1690
Product Code: HIH, Hysteroscope and Accessories
Classification: Class II
Panel: Obstetrics/Gynecology
## PREDICATE DEVICES
Tradename: MyoSure XL Rod Lens Hysteroscope Submitter/510(k) Holder: Hologic, Inc. 510(k) #: K122563 Product Code: HIH Regulation: 21CFR 884.1690
Tradename: MyoSure Rod Lens Hysteroscope Submitter/510(k) Holder: Hologic, Inc. 510(k) #: K091465 Product Code: HIH
Regulation: 21CFR 884.1690
The MyoSure and MyoSure XL Rod Lens Hysteroscopes have not been subject to a designrelated recall.
## DEVICE DESCRIPTION
The Omni Hysteroscope is used to provide viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and operative procedure. The Omni Hysteroscope is a
{4}------------------------------------------------
hysteroscope consisting of a (base) rigid hysteroscope with optics and lighting connections. The Omni Hysteroscope includes three compatible sheaths, one diagnostic and two operative, that attach to the base scope; Omni 3.7 mm Diagnostic Sheath, Omni 5.5 mm Operative Sheath, and Omni 6mm Operative Sheath. The Omni Hysteroscope also includes removable outflow channels for fluid drainage.
The Omni Hysteroscope is supplied non-sterile and is designed for multiple use with the device being cleaned and sterilized prior to each use.
### INDICATIONS FOR USE:
The Omni Hysteroscope is used to provide viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.
The Omni Hysteroscope intended use is identical to the predicates MyoSure XL Hysteroscopes.
### TECHNOLOGICAL CHARACTERISTICS:
Primary specifications for the Omni Hysteroscope including light source, thermal and electrical safety are identical to the predicate MyoSure XL Hysteroscope. The hysteroscope dimensions and materials of construction in the Omni Hysteroscope are similar to the predicate MyoSure and MyoSure XL Rod Lens Hysteroscope.
The mode of operation including method of use and mechanism of action, are identical to the predicate MyoSure XL Rod Lens Hysteroscope and are similar to the secondary predicate MyoSure Rod Lens Hysteroscope.
The primary differences between the proposed Omni Hysteroscope and the predicate MyoSure and MyoSure XL Rod Lens Hysteroscopes are as follows:
- The Omni Hysteroscope does not have an integrated working channel as the predicates do. ● Rather the subject devices have removable sheaths for the working channels of the Omni Hysteroscope.
- The Omni Hysteroscope utilizes simlar materials of construction as the predicate hysteroscopes with the addition of polyphenylsulfone (PPSU) and black, blue, and green colorants.
- . The working length of the Omni Hysteroscope is longer than the predicates MyoSure and MyoSure XL Rod Lens Hysteroscopes.
- The outer diameters of the Omni Hysteroscope 3.7mm diagnostic sheath and 5.5mm operative sheath are smaller than the outer diameter of the predicate MyoSure Rod Lens Hysteroscope.
- The outer diameter of the Omni Hysteroscope 6 mm operative sheath is smaller than the outer diameter predicate MyoSure XL Rod Lens Hysteroscope.
The differences in technological characteristics do not raise different questions of safety and effectiveness.
{5}------------------------------------------------
| Characteristic | Omni Hysteroscope | Primary Predicate<br>Device (K122563) | Secondary Predicate<br>Device (K091465) |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The Omni Hysteroscope is used<br>to provide viewing of the<br>cervical canal and the uterine<br>cavity for the purpose of<br>performing diagnostic and<br>surgical procedures. | The MyoSure Rod Lens Hysteroscope is used to<br>provide viewing of the cervical canal and the uterine<br>cavity for the purpose of performing diagnostic and<br>surgical procedures. | The MyoSure Rod Lens Hysteroscope is used to<br>provide viewing of the cervical canal and the uterine<br>cavity for the purpose of performing diagnostic and<br>surgical procedures. |
| Prescription Only | Yes | Same | Same |
| Design | Rigid Hysteroscope | Same | Same |
| Patient Contacting<br>Materials | Stainless steel, Sapphire,<br>Optical Glass, PEEK, Glass<br>Fiber, PPSU (black, blue, and<br>green colorant), Nickel Silver | Stainless steel,<br>Sapphire, Optical<br>Glass, PEEK, Glass<br>Fiber, PPSU,Nickel<br>Silver | Stainless steel, Sapphire,<br>Optical Glass, PEEK,<br>Glass Fiber, Nickel Silver |
| Working length | 200 mm | 184 mm | 184 mm |
| Outer diameter | 3.7-4.8 mm x 3.7- 7.5 mm | 6.5 x 7.6 mm | 6.0 x 6.4 mm |
| Light Source | Metal Halide or Xenon ≤300w | Same | Same |
| Biocompatibility | According to ANSI AAMI ISO<br>10993-1:2009 | Same | Same |
#### PERFORMANCE TESTING
Design verification testing demonstrates devices comply with design specifications. Design verification testing of the Omni Hysteroscope included:
- Dimensional verification
- . Objective lens testing (Field of view, focal length, depth of field, direction of view)
- o Level of Illumination
- Luminous Energy Transmission Ratio
- Magnification (eyepiece) ●
- o Image Quality (Resolution, Distortion)
Risk management activities in accordance with ISO 14971:2007 demonstrate the risks associated with the use of the Omni Hysteroscope are mitigated to an acceptable level. Analyses of these activities indicate the benefits associated with the use of the Omni Hysteroscope outweigh the residual risks. Material analysis and testing demonstrate the patient contacting materials are biocompatible and comply with the requirements of ANSI AAMI ISO 10993-1:2009. Electrical safety testing was performed and passed ANSI AAMI ES 60601-1:2005/(R)2012 & A1:2012, C1:2009/R)2012 & A2:2010/(R)2012 and IEC 60601-2-18:2009. Usability testing was conducted in accordance with the FDA Guidance, Applying Human Factors and Usability Engineering to Medical Devices, issued on February 3, 2016 and IEC 62366-1:2015 Medical Devices – Part 1: application of usability engineering to medical devices. A cleaning validation of the Omni Hysteroscope was successfully performed and sterilizations of the Omni hysteroscope were conducted in accordance with ISO 14937:2009/(R)2013 Sterilization of health care products - General requirements
{6}------------------------------------------------
for characterization of a sterilizing agent and the development, validation and routine control of a sterilization process for medical devices and ANSI/AAMI/ISO 17665-1:2006/(R)2013 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices. Performance testing demonstrates that the performance of the Omni Hysteroscope is substantially equivalent to the predicate devices.
#### CONCLUSION
Based on the intended use, descriptive information and performance provided in this submission, the Omni Hysteroscope has been shown to be substantially equivalent to the predicate MyoSure and MyoSure XL Rod Lens Hysteroscopes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.