K181955 · Medical Cables S.L. · NGX · Feb 12, 2019 · Physical Medicine
Device Facts
Record ID
K181955
Device Name
WiEMSpro
Applicant
Medical Cables S.L.
Product Code
NGX · Physical Medicine
Decision Date
Feb 12, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
WIEMSPRO is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The WIEMSPRO is not intended to be used in conjunction with therapy or treatment of medical conditions of any kind. None of WIEMSPRO training programs is designed for injured or ailing muscles and its use on such muscles is contraindicated. WIEMSPRO is a machine with electronic muscle stimulation based on EMS technology. Regarding is specifically designed as an addition to other sports and for training muscles. It must be used for only healthy muscles and clients, not for rehabilitation purposes. WIEMSPRO electrical impulses allow the triggering of action potentials on motoneurones of motor nerves (excitations). These excitations of motoneurones are transmitted to the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration, duration, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.
Device Story
WIEMSPRO is an electro-medical device for electronic muscle stimulation (EMS) used for fitness and muscle training. It delivers electrical impulses to trigger action potentials on motor nerves, causing mechanical muscle fiber responses. The device is designed for use on healthy muscles only and is not intended for rehabilitation. It is used in conjunction with a mobile application. The system is worn over cotton undergarments, avoiding direct patient contact. The device is battery-operated (3.7V LiPo) and features 10 channels. It is not intended for use while the user is in motion or lifting weights. The healthcare provider or user selects stimulation parameters (frequency, pulse duration, rest duration) to impose specific types of muscle work, aiming to improve muscle performance.
Clinical Evidence
Bench testing only. No clinical studies were submitted. Performance was verified through electrical safety, electromagnetic compatibility, and FCC radio frequency testing.
Technological Characteristics
Powered muscle stimulator; battery-operated (3.7V LiPo); 10 output channels; symmetrical biphasic waveform; frequency 1-100 Hz; pulse duration 100-400 µsec; max output 125mA/62.5V; max current density 1.92 mA/cm². Complies with IEC 60601-1, IEC 60601-2-10, IEC 60601-1-6, IEC 62304, ISO 14971, and ANSI/AAMI ES60601-1.
Indications for Use
Indicated for healthy adults to stimulate muscles to improve or facilitate muscle performance. Contraindicated for use on injured or ailing muscles, or for rehabilitation, therapy, or treatment of any medical condition.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 12, 2019
Medical Cables, S.L. José Fuertes Peña Manager Duque de la Victoria 6, 1º Malaga, Spain 29015
Re: K181955
Trade/Device Name: WIEMSPRO Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: December 18, 2018 Received: January 16, 2019
Dear José Fuertes Peña:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181955
Device Name WIEMSPRO
#### Indications for Use (Describe)
WIEMSPRO is intended to stimulate healthy muscles in order to improve or facilitate muscle performance. The WIEMSPRO is not intended to be used in conjunction with therapy or treatment of medical conditions of any kind. None of WIEMSPRO training programs is designed for injured or ailing muscles and its use on such muscles is contraindicated.
WIEMSPRO is a machine with electronic muscle stimulation based on EMS technology. Regarding is specifically designed as an addition to other sports and for training muscles. It must be used for only healthy muscles and clients, not for rehabilitation purposes.
WIEMSPRO electrical impulses allow the triggering of action potentials on motoneurones of motor nerves (excitations). These excitations of motoneurones are transmitted to the motor endplate where they generate mechanical muscle fiber responses that correspond to muscle work. Depending on the parameters of the electrical impulses (pulse frequency, duration, duration, duration of rest, total session duration), different types of muscle work can be imposed on the stimulated muscles.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) Premarket medicalcables WIEMSpro Notification DATE OF SUBMISSION: 2018-03-02 SUBMITTER NAME: Medical Cables, S.L. SUBMITTER ADDRESS: Duque de la Victoria 6, 1º 29015, Malaga SPAIN CONTACT: José Fuertes Peña TELEPHONE: +34 95 221 89 91 E-mail: ifuertes@medicalcables.eu DEVICE TRADE NAME: WIEMSpro COMMON NAME: Powered muscle stimulator. Stimulator, Muscle, Powered, For Muscle Conditioning CLASSIFICATION NAME: (21 CFR 890.5850) E-Fit EF-1280 (K133225) PREDICATE DEVICE(S): Compex Wireless USA (K170903)
#### DEVICE DESCRIPTION:
The device described in this submission is an electro-medical device intended for stimulating healthy muscles in order to improve or facilitate muscle performance. It is designed for personal training performances.
WIEMSero is a device with electronic muscle stimulation based on EMS technology. Regarding its use, the device is specifically designed as an addition to other sports and for training muscles. It must be used for only healthy muscles and clients, not for rehabilitation purposes. The device must be used over user's cotton undergarments and is therefore not patient-contacting
WIEMSPRO system cannot be used while the user is in motion or lifting weights.
#### SUMMARY OF COMPARISON WITH PREDICATE DEVICE:
In the establishment of substantial equivalence, the WIEMSPRO device is compared with the following previously cleared devices:
- E-Fit EF-1280 (K133225) -
- Compex WirelessUSA (K170903) -
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "medicalcables" in white font on a blue background. The word is written in lowercase letters. The font is sans-serif and appears to be bolded.
# 510(k) Premarket Notification
Comparison of the proposed devices with the predicate devices is summarized in the following table:
| COMPARISON OF TECHNOLOGICAL CHARACTERISTICS | | | | |
|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic /<br>Feature | PROPOSED DEVICE | PREDICATE DEVICE | PREDICATE DEVICE | Comparison |
| | WIEMSPRO | E-Fit EF-1280 | Compex Wireless USA | |
| | | GENERAL COMPARISON | | |
| Classification<br>name | Powered muscle stimulator | Powered muscle<br>stimulator | Powered muscle stimulator | Same |
| Product code | NGX | NGX | NGX | Same |
| Regulation<br>number | 21 CFR 890.5850 | 21 CFR 890.5850 | 21 CFR 890.5850 | Same |
| Panel | Physical Medicine | Physical Medicine | Physical Medicine | Same |
| Class | Class II | Class II | Class II | Same |
| 510(K) Number | -- | K133225 | K170903 | N/A |
| | | INTENDED USE | | |
| Indication for<br>use | WIEMSPRO is intended to<br>stimulate healthy muscles in<br>order to improve or facilitate<br>muscle performance. The<br>WIEMSPRO is not intended to<br>be used in conjunction with<br>therapy or treatment of medical<br>diseases or medical conditions<br>of any kind. None of<br>WIEMSPRO training programs is<br>designed for injured or ailing<br>muscles and its use on such<br>muscles is contraindicated.<br><br>WIEMSPRO is a machine with<br>electronic muscle stimulation<br>based on EMS technology.<br>Regarding its use, the device is<br>specifically designed as an<br>addition to other sports and for<br>training muscles. It must be<br>used for only healthy muscles<br>and clients, not for<br>rehabilitation purposes.<br><br>WIEMSPRO electrical impulses<br>allow the triggering of action<br>potentials on motoneurones of<br>motor nerves (excitations).<br>These excitations of<br>motoneurones are transmitted<br>to the muscle fibers via the<br>motor endplate where they<br>generate mechanical muscle<br>fiber responses that<br>correspond to muscle work.<br>Depending on the parameters<br>of the electrical impulses<br>(pulse frequency, duration of<br>contraction, duration of rest,<br>total session duration),<br>different types of muscle work | E-fit EF-1280 is a machine<br>with electronic muscle<br>stimulation based on EMS<br>technology. Regarding its<br>use, the device is<br>specifically designed as an<br>addition to other sports<br>and for training muscles. It<br>must be used for only<br>healthy muscles and<br>clients, not for<br>rehabilitation purposes.<br><br>The E-Fit EF-1280<br>intended to stimulate<br>healthy muscles in order to<br>improve or facilitate<br>muscle performance. The<br>E-Fit EF-1280 is not<br>intended to be used in<br>conjunction with therapy or<br>treatment of medical<br>diseases or medical<br>conditions of any kind.<br>None of the E-Fit EF-1280<br>training programs is<br>designed for injured or<br>ailing muscles and its use<br>on such muscles is<br>contraindicated.<br><br>The E-Fit EF-1280<br>electrical impulses allow<br>the triggering of action<br>potentials on<br>motoneurones of motor<br>nerves (excitations).<br>These excitations of<br>motoneurones are<br>transmitted to the muscle<br>fibers via the motor | The Compex Wireless USA is an<br>Over-The-Counter device<br>intended to stimulate healthy<br>muscles in order to improve or<br>facilitate muscle performance. It<br>is to be used by adults only.<br><br>The Compex Wireless USA is not<br>intended for adjunctive therapy in<br>the treatment of medical<br>diseases and conditions of any<br>kind. None of the Compex<br>Wireless USA stimulation<br>programs are designed for<br>injured or disease afflicted<br>muscles. Its use on such<br>muscles is contraindicated. The<br>work imposed on the muscles by<br>the Compex Wireless USA<br>programs is definitely not suitable<br>for rehabilitation and<br>physiotherapy. The Compex<br>Wireless USA electrical impulses<br>allow the triggering of action<br>potentials on motoneurones of<br>motor nerves (excitations). These<br>excitations of motoneurones are<br>transmitted to the muscle fibers<br>via the motor endplate where<br>they generate mechanical<br>muscle fiber responses that<br>correspond to muscle work.<br>Depending on the parameters of<br>the electrical impulses (pulse<br>frequency, duration of<br>contraction, duration of rest, total<br>session duration), different types<br>of muscle work can be imposed<br>on the stimulated muscles. The<br>Compex Wireless USA may<br>therefore be considered a<br>technique of muscle training. | Same<br>WIEMSPRO<br>does not<br>include<br>TENS |
| | | | | |
| Intended use | can be imposed on the stimulated muscles. | endplate where they<br>generate mechanical<br>muscle fiber responses<br>that correspond to muscle<br>work. Depending on the<br>parameters of the<br>electrical impulses (pulse<br>frequency, duration of<br>contraction, duration of<br>rest, total session<br>duration), different types of<br>muscle work can be<br>imposed on the stimulated<br>muscles.<br>The various types of<br>muscle work that the E-Fit<br>EF-1280 can impose on<br>the stimulated muscles are<br>able to improve or facilitate<br>muscle performance. The<br>E-Fit EF-1280 may<br>therefore be considered a<br>technique of muscle<br>training.<br>Must be used for only<br>healthy muscles and<br>client. It is not intended to<br>be used in conjunction<br>with therapy or treatment<br>of medical diseases or | The Compex Wireless USA<br>TENS is used for:<br>· temporary relief of pain<br>associated with sore and aching<br>muscles due to strain from<br>exercise or normal household<br>and work activities.<br>· the symptomatic relief and<br>management of chronic,<br>intractable pain and relief of pain<br>associated with arthritis.<br>Must be used for only healthy<br>muscles and client. It is not<br>intended to be used in<br>conjunction with therapy or<br>treatment of medical diseases or<br>medical conditions of any kind. | Similar.<br>Difference<br>is the<br>software<br>application<br>which<br>controls the<br>device |
| | Must be used for only healthy<br>muscles and client. It is not<br>intended to be used in<br>conjunction with therapy or<br>treatment of medical diseases<br>or medical conditions of any<br>kind.<br>It is designed to be used<br>together with a WIEMSPRO<br>Mobile Application. | medical conditions of any<br>kind. | | which<br>controls the<br>device |
| | Mobile Application. | TECHNOLOGICAL CHARACTERISTICS AND PERFORMANCE | | |
| Powered Muscle | | | | |
| Stimulator | YES…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.