The GMC Upper Arm Automatic Blood Pressure Monitor, Model X3, is a non-invasive blood pressure measurement device that is used for measuring systolic and diastolic blood pressure and pulse rate using the oscillometric method for adults at home. The device has an irregular heartbeat (IHB) indicator. The device detects the appearance of an irregular heartbeat during measurement, and displays an IHB symbol on the LCD with the reading once the irregular heartbeat is detected.
Device Story
Device measures systolic/diastolic blood pressure and pulse rate via oscillometric method; uses electronic pressure sensor to convert cuff pressure variations into electrical signals; signals analyzed to determine BP and pulse. Features include irregular heartbeat (IHB) detection (triggered by >25% deviation from average pulse), body movement detection, and hypertension classification indicator. Intended for adult home use; operated by patient. Output displayed on LCD; includes BP, pulse, memory results, date/time, and status icons (IHB, movement, battery). Data used by patients for personal health monitoring.
Clinical Evidence
Clinical validation performed per ISO 81060-2. Study included 85 patients (39 male, 46 female). Accuracy of the device was compared against the standard auscultatory method using alternating measurements in the same arm. Results confirmed clinical accuracy.
Technological Characteristics
Oscillometric measurement; semi-conductor pressure sensor; pump-driven inflation; automatic air release valve. Powered by batteries or AC adapter. LCD display. Complies with ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 80601-2-30, ISO 81060-2, and ISO 10993-1. Materials are biocompatible per ISO 10993-1.
Indications for Use
Indicated for adult patients for non-invasive measurement of systolic and diastolic blood pressure and pulse rate at home. Includes irregular heartbeat (IHB) detection.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
April 4, 2019
GMC Inc. Yolanda Lin Director of Quality Management Div. No. 686, Su Chu Rd. Chuzhou, 239000 Cn
Re: K181931
Trade/Device Name: GMC Upper Arm Automatic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: January 11, 2019 Received: January 11, 2019
Dear Yolanda Lin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{3}------------------------------------------------
Section 5- 510(k) Summary
Image /page/3/Picture/3 description: The image shows the logo for GMC 尚钧医疗. The logo consists of the letters "GMC" in a dark gray sans-serif font, followed by a red circle with a white plus sign inside. To the right of the circle are the Chinese characters 尚钧医疗, also in a dark gray color. The characters are written in a traditional Chinese calligraphy style.
GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China
X3
## 510(k) Summary
| 510(k) owner's information: | GMC Inc.<br>No. 686, Su Chu Rd., Chuzhou,<br>Anhui, China<br>Tel: +86-550-2172888<br>Fax: +86-550-2172889 |
|-----------------------------|-----------------------------------------------------------------------------------------------------------|
| Official Contact: | Yolanda Lin<br>Tel: +86-550-2172888<br>Fax: +86-550-2172889<br>E-mail: linyolanda23@gmail.com |
| Proprietary or Trade Name: | GMC Upper Arm Automatic Blood<br>Pressure Monitor, Model X3 |
| Common/Usual Name: | Non-Invasive Blood Pressure Monitor |
| Classification Name: | System, Measurement, Blood-Pressure,<br>Non-Invasive |
| Regulation | CFR 870.1130 |
| Product Code | DXN |
| Device Class: | II |
| Predicate Device: | Omron HEM-7311 (K133379) |
| Date Summary Prepared: | 07/16/2018 |
{4}------------------------------------------------
Section 5- 510(k) Summary
Image /page/4/Picture/3 description: The image shows the logo for GMC 尚钧医疗. The logo consists of the letters "GMC" in a dark gray sans-serif font, followed by a red circle with a white plus sign in the center. To the right of the circle are four Chinese characters in a dark gray, traditional font.
No. 686, Su Chu Rd., Chuzhou Anhui, China
X3
## Device Description:
The GMC Upper Arm Automatic Blood Pressure Monitor, Model X3, measures systolic and diastolic blood pressure and pulse rate of an adult individual. The method used to define blood pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The device utilizes the oscillometric method whereby the electronic pressure sensor converts variation in cuff pressure to electrical signals. The electrical signals are analyzed and used to define the systolic and diastolic measurements and determine the pulse rate. An irregular heart beat (IHB) indicator is displayed when there is more than a 25% deviation from the averaged pulse rate measurements.
## Intended Use:
The GMC Upper Arm Automatic Blood Pressure Monitor, Model X3, is a non-invasive blood pressure measurement device that is used for measuring systolic and diastolic blood pressure and pulse rate using the oscillometric method for adults at home.
The device has an irregular heartbeat (IHB) indicator. The device detects the appearance of an irregular heartbeat during measurement, and displays an IHB symbol on the LCD with the reading once the irregular heartbeat is detected.
## Substantially Equivalent (SE) Comparison
The GMC Upper Arm Automatic Blood Pressure Monitor, Model X3 is substantially equivalent to the Omron HEM-7311 (K133379).
{5}------------------------------------------------
# GMC0尚钩医疗
GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China
X3
| Model name | Predicate Device | Subject Device | Comparison |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | Omron HEM-7311 (K133379) | GMC X3 | |
| 510(k) submitter/holder | Omron Healthcare, Inc. | GMC Inc. | |
| Intended use and Indications<br>for Use | The device is a digital monitor intended<br>for use in measuring blood pressure and<br>pulse rate in adult patient population.<br>The device detects the appearance of<br>irregular heartbeats during<br>measurement and gives a warning<br>signal with readings. | The device is a non-invasive blood pressure<br>measurement device that is used for<br>measuring systolic and diastolic blood<br>pressure and pulse rate using the<br>oscillometric method for adults at home.<br>The device has an irregular heartbeat (IHB)<br>indicator. The device detects the appearance<br>of an irregular heartbeat during<br>measurement, and displays an IHB symbol<br>on the LCD with the reading once the<br>irregular heartbeat is detected. | Identical |
| Use Environment | Home use | Home use | Identical |
| Patient Population | Adult | Adult | Identical |
{6}------------------------------------------------
## GMC0尚钧医疗
GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China
Specifications / Features
| Model name | Predicate Device<br>Omron HEM-7311 (K133379) | | Subject Device<br>GMC X3 | Comparison |
|-----------------------------|----------------------------------------------|--------------------------------|--------------------------|---------------------------------------------------------------------------------------------|
| Measurement method | Oscillometric methods | | Oscillometric methods | Identical |
| Cuff fits arm circumference | 17-22cm/ 22-32cm/ 22-42cm | | 24-40 cm (9.4"~15.7") | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
| Measurement range | Pressure | 0~299 mmHg | 30~260 mmHg | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
| | Pulse | 40~180 beats/min | 40~199 beats/min | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
| Accuracy | Pressure | $\pm$ 3 mmHg or 2 % of reading | $\pm$ 3 mmHg | Similar<br>There is no difference in the<br>accuracy range based on mmHg<br>scale |
| | Pulse | $\pm$ 5% of reading | $\pm$ 5% of reading | Identical |
{7}------------------------------------------------
#### X3
# GMC0尚钧医疗
GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China
| Model name | Predicate Device<br>Omron HEM-7311 (K133379) | Subject Device<br>GMC X3 | Comparison |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Reusable | Yes | Yes | Identical |
| Display Type | LCD | LCD | Identical |
| Display Parameter | Current cuff pressure, Systolic blood<br>pressure, Diastolic blood pressure,<br>Pulse rate, Error message,<br>measurement results in the memory | Systolic pressure, Diastolic pressure,<br>Pulse rate, Memory sequence<br>number, Date/Time, memory zone<br>icons, pulse mark, weak battery mark,<br>AVG, Hypertension Classification<br>Indicator, IHB, Movement indicator | Similar<br>The different display parameters do not impact safety and effectiveness of the device. |
| Inflation | Pump driven | Pump driven | Identical |
| Deflation | Automatic Air Release Valve | Automatic Air Release Valve | Identical |
| Pressure Sensor | Semi-conductor | Semi-conductor | Identical |
| Irregular Heart Beat detection | Yes | Yes | Identical |
| Body movement detection | Yes | Yes | Identical |
| Hypertension indicator | Yes | Yes | Identical |
| Average of latest three<br>measurements | Yes | Yes | Identical |
{8}------------------------------------------------
#### X3
# GMC0尚钧医疗
GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China
| Model name | | Predicate Device<br>Omron HEM-7311 (K133379) | Subject Device<br>GMC X3 | Comparison |
|----------------|---------------------|-----------------------------------------------------|---------------------------------|---------------------------------------------------------------------------------------------|
| Multiple Users | | Yes(2) | Yes(2) | Identical |
| Power Source | | Battery as well as AC adaptor | Battery as well as AC adaptor | Identical |
| Dimension | | 183 X 230 X 99mm (Lx W x H) | 96 X 139.7 X 63.2mm (L x W x H) | Similar<br>Size is not a factor in function of the<br>device |
| Weight | | Approx. 640g (1 lbs6 5/8 oz)<br>(without batteries) | (without batteries)<br>246g | Similar<br>Weight is not a factor in function of the<br>device |
| Operating | Temperature | 10~40℃ (50-104°F) | 10~40℃ (50-104°F) | Identical |
| Environment | Humidity | 15%~90% RH | 15%~85% RH | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
| | Ambient<br>pressure | N/A | 700-1060 hpa | Different<br>This feature does not impact safety and<br>effectiveness of the device |
{9}------------------------------------------------
#### X3
## GMC0尚钧医疗
GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China
| Model name | Predicate Device<br>Omron HEM-7311 (K133379) | | Subject Device<br>GMC X3 | Comparison |
|------------------------|----------------------------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Storage<br>Environment | Temperature | -20~60°C (-4-140°F) | -10~60 ℃(14-140°F) | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
| | Humidity | 10%~95% RH | 10%~90% RH | Similar<br>This minor difference does not impact<br>safety and effectiveness of the device. |
| | Ambient<br>pressure | 700-1060 hpa | 700-1060 hpa | Identical |
| Compliance Standards | | ANSI/AAMI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 80601-2-30<br>ISO81060-2<br>ISO 10993-1 | ANSI/AAMI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-1-11<br>IEC 80601-2-30<br>ISO81060-2<br>ISO 10993-1 | Identical |
{10}------------------------------------------------
GMC Inc. No. 686, Su Chu Rd., Chuzhou, Anhui, China
The subject device is viewed as substantially equivalent to the predicate devices Omron HEM-7311 because they have same intended use and use the well known oscillometric method within the software algorithm to determine the systolic and diastolic blood pressure and pulse rate. There are no significant diffect the safety or effectiveness of the intended device as compared to the predicate devices
{11}------------------------------------------------
Image /page/11/Picture/3 description: The image shows the logo for GMC+. The logo consists of the letters "GMC" in a dark gray sans-serif font, followed by a red circle with a white plus sign inside. To the right of the circle are two Chinese characters, also in dark gray. The overall design is simple and modern, with a focus on the company's initials and a medical symbol.
No. 6
## Performance Testing:
Bench testing was conducted to demonstrate that the device meets its requirements and specification.
The following performance tests were completed:
- #19-4
ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety
and essential performance (IEC 60601-1:2005, MOD)
- #19-8 •
IEC 60601-1-2 Edition 4.0 2014-02
Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- #19-14
IEC 60601-1-11 Edition 2.0 2015-01
Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- #3-123
IEC 80601-2-30 Edition 1.1 2013-07
Medical electrical equipment - Part 2-30: Particular requirements for the basic safety and essential performance of automated non-invasive sphygmomanometers
The GMC Upper Arm Automatic Blood Pressure Monitor, Model X3 meets all applicable requirements.
{12}------------------------------------------------
Image /page/12/Picture/3 description: The image shows the logo for GMC+. The logo consists of the letters "GMC" in a dark gray sans-serif font, followed by a red circle with a white plus sign inside. To the right of the circle is a Chinese character, also in dark gray.
No. 6
## Biocompatibility
The materials of construction are widely used in the same or similar applications; therefore, demonstrating the material's use in other medical devices and supporting existing biocompatibility.
It is evaluated according to #2-220 ISO 10993-1 Fourth edition 2009-10-15 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process.
## Clinical testing
Testing to insure clinical accuracy of the device in accordance with ISO 81060-2 as documented in Clinical Investigation Report. 85 patients (39 males and 46 females) were invited for the study. Standard auscultation method was used as the reference blood pressure monitor measuring in the left arm. Blood pressure measurements were repeated alternatively with the device and auscultation in the same arm according to the sequence in #3-122 ISO 81060-2 Second edition 2013-05-01 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type.
## Conclusion
Based on the evaluations completed, there are no new safety or effectiveness issues introduced with this new medical device. Therefore, the GMC Upper Arm Automatic Blood Pressure Monitor, Model X3, is substantially equivalent to the predicate devices Omron HEM-7311 (K133379).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.