K181917 · NuSmile, Ltd. · KIF · Jan 11, 2019 · Dental
Device Facts
Record ID
K181917
Device Name
MTA2.3 MATERIALS
Applicant
NuSmile, Ltd.
Product Code
KIF · Dental
Decision Date
Jan 11, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3820
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The MTA2.3 ROOT & PULP MATERIALS are indicated for dental procedures that contact pulp and periradicular tissues, as well as obturation and sealing of root canals.
Device Story
Hydraulic ceramic cement for dental procedures; used in contact with vital pulp and periradicular tissues. Input: inorganic ceramic powder mixed with water-based gel or organic liquid. Mechanism: hydration of tricalcium/dicalcium silicates in presence of body fluids to form hard, high-pH substance containing hydrated silicates and hydroxides. Output: set cement for sealing/obturation. Used in clinics by dental clinicians. Benefits: provides biocompatible seal for root canal and pulp treatments; radiopaque for visualization. No software or electronic components.
Clinical Evidence
No clinical or animal testing performed. Substantial equivalence supported by bench testing (ISO 6876, ISO 9917-1, ADA 57) verifying radiopacity, solubility, dimensional stability, film thickness, flow, working/setting time, and heavy metal content (arsenic/lead). Biocompatibility testing confirmed non-cytotoxicity.
Technological Characteristics
Hydraulic ceramic cement; primary components: tricalcium silicate, dicalcium silicate, tantalum oxide. Sets in vivo via hydration. Conforms to ISO 6876, ISO 9917-1, and ADA 57. Non-sterile, single-use dose from multi-use container. No software, electronics, or connectivity.
Indications for Use
Indicated for dental procedures contacting pulp and periradicular tissues, including pulp capping, cavity lining, base material, pulpotomies, root-end filling, apexification, perforation repair, root resorption, and obturation (pulpectomy) in primary and secondary dentition.
Regulatory Classification
Identification
A root canal filling resin is a device composed of material, such as methylmethacrylate, intended for use during endodontic therapy to fill the root canal of a tooth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 11, 2019
NuSmile, LTD. Angie Bever Regulatory Affairs & Quality Assurance Specialist 3315 W 12th Street Houston, Texas 77008
Re: K181917
Trade/Device Name: MTA2.3 Root & Pulp Materials Regulation Number: 21 CFR 872.3820 Regulation Name: Root Canal Filling Resin Regulatory Class: Class II Product Code: KIF Dated: October 24, 2018 Received: October 25, 2018
Dear Angie Beyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely. Digitally signed by Marv Mary S. ה Mrary -5.
> Runner -533 - Date: 2019.01.11 10:59:11
> Runner -533 - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Form Approved: OMB No. 0910-0120<br>Food and Drug Administration<br>Expiration Date: 06/30/2020<br>Indications for Use<br>See PRA Statement below.<br>510(k) Number (if known)<br>K181917<br>Device Name<br>MTA2.3 ROOT & PULP MATERIALS<br>Indications for Use (Describe)<br>The MTA2.3 ROOT & PULP MATERIALS are indicated for dental procedures that contact pulp and periradicular<br>tissues, as well as obturation and sealing of root canals.<br>Type of Use (Select one or both, as applicable)<br>Over-The-Counter Use (21 CFR 801 Subpart C)<br>X Prescription Use (Part 21 CFR 801 Subpart D)<br>CONTINUE ON A SEPARATE PAGE IF NEEDED.<br>This section applies only to requirements of the Paperwork Reduction Act of 1995.<br>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*<br>The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to:<br>Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov<br>"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number."<br>FORM FDA 3881 (7/17)<br>Page 1 of 1<br>PSC Publishing Services (301) 443-674 | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
| | | | |
MTA2.3 ROOT & PULP MATERIALS
26 NuSmile, Ltd.
{3}------------------------------------------------
# K181917
# 510(k) Summary
per 21CFR807.92 & 807.93
## I. SUBMITTER
DATE PREPARED: June 30, 2018
PREPARER: Angie Beyer Regulatory & QA Specialist NuSmile, Ltd. 3315 W, 12th Street Houston, TX 77008 USA www.nusmile.com Phone: (713) 861-0033 xtn. 559 Fax: (713) 861-1722 E-mail: regulatory@nusmile.com Web: http://www.nusmile.com/
## II. DEVICE
TRADE OR PROPRIETARY NAME: CLASSIFICATION NAME: PREDICATE DEVICE:
MTA2.3 ROOT & PULP MATERIALS Root Canal Filling Resin, 872.3820 K161239, MTA2.2 material
### III. PREDICATE DEVICES
MTA2.3 ROOT & PULP MATERIALS are substantially equivalent to the following predicate devices with respect to intended use, indications for use, materials, technological characteristics, and device design.
| 510(k) # | Device | Manufacturer |
|----------|--------|--------------------------------------------------------------------------------------|
| K161239 | MTA2.2 | Avalon Biomed Inc.*, |
| | | *assets of Avalon Biomed were acquired by NuSmile, Ltd.<br>(2016), including 510(k)s |
The predicate has not been subject to a recall.
MTA2.3 ROOT & PULP MATERIALS have been modified from MTA2.2 MATERIAL to allow for more radiopacity and varied clinical handling characteristics.
### IV. DEVICE DESCRIPTION:
The MTA2.3 ROOT & PULP MATERIALS are designed and developed for dental clinicians to use in contact with vital pulp tissue and periradicular tissue, including sealing and obturation of root canals. The materials can be used for dental procedures contacting pulpal or periradicular tissue such as: pulp capping, cavity lining, base material in a cavity, pulpotomies, root-end filling, apexification, perforation repair, root resorption, and obturation (pulpectomy) including root canal sealing.
Premarket Notification
MTA2.3 ROOT & PULP MATERIALS
28 NuSmile, Ltd.
#### MTA 2.3 ROOT & PULP MATERIALS
{4}------------------------------------------------
The mechanism of action is for the subject and predicate devices is setting of the inorganic, ceramic cement in the presence of the water-based body fluids present in teeth.
MTA2.3 ROOT & PULP MATERIALS were developed in 2 forms (designs): powder with water-based gel, and pastes. The dentist may choose to mix the powder and water-based gel of MTA2.3 (design 1), or choose to use a paste form of MTA2.3 (design 2). The MTA2.3 ROOT & PULP MATERIALS are placed in the space created by the procedure and set in vivo. The procedure may be part of caries treatment, root canal treatment, or periapical surgery. Such procedures are performed on primary and secondary dentition. MTA2.3 ROOT & PULP MATERIALS may be white or tinted vellow, pink or grav. MTA2.3 ROOT & PULP MATERIALS are color stable over time.
## V. INDICATIONS FOR USE:
The MTA2.3 ROOT & PULP MATERIALS are intended for dental procedures that contact pulp and periradicular tissues, as well as obturation and sealing of root canals.
These indications are identical to the predicate K161239, MTA2.2 MATERIAL.
## VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
TECHNOLOGICAL CHARACTERISTICS: The MTA2.3 ROOT & PULP MATERIALS are primarily hydraulic ceramic powders, as is the MTA2.2 MATERIAL (K161239). The ceramic powders set in vivo in contact with body fluids. The technology is unchanged.
The hydrophilic ceramic powders are known to set with water into a hard substance, containing a high pH, calcium hydroxide reaction product, dispersed among the hydrated particles of tri/dicalcium silicates. Both the predicate and the subject materials are radiopaque and contain the same ceramic oxide for radiopacity.
COMPARISONS TO PREDICATE: We believe the MTA2.3 ROOT & PULP MATERIALS are substantially equivalent to the MTA2.2 material (K161239) when used for contact with pulp tissue and periapical tissue, as well as obturation and sealing of root canals.
The modifications to the predicate MTA2.2 material (K161239) are an increase in the radiopaque component, and modified liquids for the convenience delivery into the tooth by the clinician. These modifications do not change the intended use, indications for use, or technological characteristics. The modifications do not raise any new questions of safety or effectiveness.
Similarities: The predicate MTA2.2 material and the modified MTA2.3 ROOT & PULP MATERIALS are based on the inorganic ceramic powders composed of primarily tricalcium silicate, dicalcium silicate, and tantalite for radiopacity. The predicate and the MTA2.3 ROOT & PULP MATERIALS rely on water-based fluids in the body to hydrate the hydraulic ceramic phases, and cause setting into a hard substance containing hydrated silicates and hydroxides. MTA2.3 ROOT & PULP MATERIALS may contain the same inorganic ceramic pigments, at minor levels, as were cleared for MTA2.2 materials.
Premarket Notification
MTA2.3 ROOT & PULP MATERIALS
29 NuSmile, Ltd.
{5}------------------------------------------------
Both the MTA2.3 ROOT & PULP MATERIALS and the predicate MTA2.2 material are similar because they are:
- o ceramic cements,
- used in root canals or on vital pulp tissue O
- used for sealing and obturation of root canals O
- suitable for use in primary dentition for pulpal treatment O
- are based on ceramic powders that set with water O
- are radiopaque O
- do not discolor o
- set in vivo with contact with water-based body fluids. o
The new and predicate materials have similar compositions and properties. Both MTA2.3 ROOT & PULP MATERIALS and the predicate MTA2.2 material create a high pH environment, which induces the precipitation of hydroxyapatite-like crystals in synthetic body fluid when the tri/dicalcium silicate powder component is hydrated.
Differences: MTA2.3 ROOT & PULP MATERIALS are more radiopaque than MTA2.2 material.
MTA2.3 ROOT & PULP MATERIALS may be packaged with a modified water-based liquid for mixing, or premixed with an organic liquid.
The similarities and differences are summarized in Table VI-1. Two design modifications are being submitted. The two designs are described in Table VI-2.
{6}------------------------------------------------
| | Predicate Device<br>MTA2.2 Material | Subject Device<br>MTA2.3 ROOT & PULP MATERIALS |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K161239 | K18XXXX |
| Manufacturer | Avalon Biomed Inc. | NuSmile, Ltd.1 |
| Device Name | MTA2.2 | MTA2.3 |
| Description | Root canal filling resin | Root canal filling resin |
| Medical Specialty | Dental | Dental |
| Product Code | KIF | KIF |
| Reg Number | 872.3820 | 872.3820 |
| Class | 2 | 2 |
| Indications for<br>Use | The MTA2.2 MATERIAL is indicated<br>for dental procedures that contact<br>pulp and periradicular tissues, as<br>well as obturation and sealing of<br>root canals. | The MTA2.3 ROOT & PULP MATERIALS<br>are indicated for dental procedures that<br>contact pulp and periradicular tissues,<br>as well as obturation and sealing of root<br>canals. |
| Sterility | Non-Sterile | Non-Sterile |
| Utility | Single-use Only- for each dose<br>dispensed from a multi-use<br>container | Single-use Only- for each dose<br>dispensed from a multi-use container |
| Materials | Powder: Tri/dicalcium silicate with<br>tantalum oxide.<br>Liquid: Water-based gel with<br>polymers. | Powder: Tri/dicalcium silicate with<br>more tantalum oxide. One design has<br>additional ceramic powders.<br>Liquid: Water-based gel with thickeners<br>& proprietary additives<br>OR an organic liquid as described for<br>each design. |
Table VI-1: Comparison of Subiect MTA2.3 to Predicate (MTA2.2 Materials, K161239)
Table VI-2: Description of Predicate (MTA2.2 K161239) and Subject MTA2.3 Designs
| Predicate design | Powder based on water-setting cements |
|--------------------------|---------------------------------------------------------------------------------|
| | Radiopaque with tantalite |
| | Water-based gel containing polymers |
| | Powder and gel to be mixed prior to use by clinician |
| Design<br>modification-1 | Powder contains same components as predicate with more tantalite |
| | Water-based gel with more thickening agents & other additives than<br>predicate |
| | Powder and gel to be mixed prior to use by clinician |
| Design<br>modification-2 | Powder contains same components as predicate with more tantalite. |
| | Powder contains additional ceramic additives for handling. |
| | Powder supplied as a paste, or a powder to the clinician. |
<sup>1</sup> NuSmile, Ltd. purchased the assets of Avalon Biomed Inc. in 2016, including the prior 510(k) licenses.
{7}------------------------------------------------
## VII. PERFORMANCE DATA
NON-CLINICAL PERFORMANCE: Bench testing of the physical properties of the MTA2.3 ROOT AND PULP MATERIALS was performed to verify conformance to ISO 6876, ADA 57, and ISO 9917-1, as they apply. We believe that the performance data provided herein demonstrate that MTA2.3 ROOT AND PULP MATERIALS are substantially equivalent to the predicate MTA2.2 material (K161239) in design, principle of performance, technology, and composition. We believe the MTA2.3 ROOT & PULP MATERIALS perform as well as or better than the predicate device.
As part of design validation, MTA2.3 ROOT & PULP MATERIALS met the ISO 6876 standard for radiopacity, solubility, dimensional stability, film thickness, and flow. The ISO 6876 tests were also performed to measure the working time and setting time of MTA2.3 ROOT & PULP MATERIALS. Tests for arsenic and lead were performed to ensure conformance to ISO 9917-1 requirements. We believe that the performance data provided herein demonstrate that MTA2.3 ROOT AND PULP MATERIALS are substantially equivalent to the predicate MTA2.2 material (K161239) in design, principle of performance, technology, and composition. We believe the MTA2.3 ROOT & PULP MATERIALS perform as well as or better than the predicate device.
BIOCOMPATIBILITY: The MTA2.3 ROOT & PULP MATERIALS were evaluated for biocompatibility as part of design validation. The materials were not cytotoxic. See Appendix C and Section 16.
CLINICAL PERFORMANCE: No animal or human clinical tests were performed in the development of the MTA2.3 ROOT & PULP MATERIALS.
MTA2.3 ROOT & PULP MATERIALS have a similar composition to the predicate device. All of the components found in the MTA2.3 ROOT & PULP MATERIALS have been used in legally marketed devices.
## VIII. CONCLUSIONS
No substantial differences in terms of technology are known between the MTA2.3 ROOT & PULP MATERIALS and the predicate device. The main difference in the modified material is additional radiopaque component. Secondary differences include a modified gel for one design, and use of an organic liquid in the second design. The compositional changes do not raise any new issues of risk, safety and effectiveness. Therefore, we believe that the information provided herein demonstrates that the MTA2.3 ROOT & PULP MATERIALS are substantially equivalent to the predicate device in technology, design, principles of performance, and intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.