K181910 · American Orthodontics · KLE · Oct 15, 2018 · Dental
Device Facts
Record ID
K181910
Device Name
BracePaste MTP Light Cure Primer
Applicant
American Orthodontics
Product Code
KLE · Dental
Decision Date
Oct 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
Orthodontic Primers are intended to be used in the bonding procedure (direct or indirect) of brackets and tubes by preparing the enamel during orthodontic treatment to increase adhesion to wet or dry, normal and atypical enamel surfaces. The Primer wets the bonding surface of the object to be bonded (i.e. enamel, composite materials, etc.) to promote adhesion.
Device Story
BracePaste MTP Light Cure Primer is a liquid, light-curing orthodontic primer supplied in a dropper bottle. Used by orthodontists during direct or indirect bonding of brackets and tubes to enamel. Primer wets bonding surfaces (enamel, composite) to promote adhesion; hydrophilic formulation allows bonding to wet or dry surfaces. Applied to enamel; polymerized via external curing light to form bond between etched enamel and bracket adhesive. Benefits patient by facilitating secure orthodontic appliance attachment in varied moisture conditions.
Clinical Evidence
No clinical performance testing conducted. Bench testing only. Performance evaluated via in vitro cytotoxicity assays, stability testing, and shear bond strength testing per ISO 29022:2013. Comparative testing against predicate Transbond MIP Primer showed average bond strengths of 38.94 N/mm² for the subject device versus 38.70 N/mm² for the predicate.
Technological Characteristics
Liquid, light-cured hydrophilic primer. Composition: Ethyl alcohol, Bisphenol A Diglycidyl Ether Dimethacrylate, 2-Hydroxyethyl Methacrylate, 2-Hydroxy-1,3-Dimethacryloxypropane, Diurethane Dimethacrylate, and water. Delivery: 6mL dropper bottle. Curing: Light-activated polymerization. Adhesion testing per ISO 29022:2013.
Indications for Use
Indicated for use in orthodontic bonding procedures (direct or indirect) of brackets and tubes to prepare enamel and increase adhesion to wet or dry, normal or atypical enamel surfaces.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
American Orthodontics Andre Leak Regulatory Affairs Coordinator 3524 Washington Avenue Sheboygan, Wisconsin 53081
Re: K181910
Trade/Device Name: BracePaste MTP Light Cure Primer Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE, DYH Dated: July 17, 2018 Received: July 17, 2018
Dear Andre Leak:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
October 15, 2018
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mary S. Runner -S3
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181910
Device Name BracePaste MTP Light Cure Primer
#### Indications for Use (Describe)
Orthodontic Primers are intended to be used in the bonding procedure (direct or indirect) of brackets and tubes by preparing the enamel during orthodontic treatment to increase adhesion to wet or dry, normal and atypical enamel surfaces. The Primer wets the bonding surface of the object to be bonded (i.e. enamel, composite materials, etc.) to promote adhesion.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| | <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
| | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized, blue font, with a triangle shape forming the "A". To the right of the letters, the words "AMERICAN ORTHODONTICS" are written in a smaller, sans-serif font, also in blue.
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
Page 5-1
# 510(k) Summarv
21CFR807.92
Preparation Date: July 13, 2018
#### Company Information:
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081 Phone: 920-457-5051 Fax: 920-457-5773
#### Submitter Information:
Trang Adams / Regulatory Affairs Specialist
#### Device Information:
Trade Name: BracePaste MTP Light Cure Primer Common Name: Orthodontic Primer Classification Name: Agent, Tooth Bonding, Resin 510(k) Number: unknown Product Code: KLE Regulation Number (21CFR): 872.3200
#### Predicate Device Information:
Product/Trade Name: Transbond MIP Primer Classification Name: Agent, Tooth Bonding, Resin 510(k) Number: K962785 Product Code: KLE Regulation Number (21CFR): 872.3200
#### Description of the Device:
American Orthodontics' BracePaste MTP Light Cure Primer is a light-curing primer that is intended to be used in direct or indirect bonding procedures of brackets and tubes. The primer prepares the enamel during orthodontic treatment to increase adhesion to wet or dry, normal and atypical enamel surfaces. The primer "wets" the bonding surface of the object to be bonded [i.e. enamel, composite materials, etc.) to promote adhesion. The primer will be available in liquid form in a dropper bottle.
The use of light-curing primers/adhesives has been around since 1970's. In the early 1980's, visible light-cured restorative materials were introduced. Hydrophilic primers were introduced in 1998. These primers were hydrophilic and bonded well to wet or dry enamel.
The primer contains Bisphenol a Diglycidyl Ether Dimethacrylate and 2-Hydroxyethyl Methacrylate resin compounds. These resin compounds are mixed with ethyl alcohol and water to form an aqueous solution. This solution has the hydrophilic characteristics necessary for the primer to adhere under wet conditions. Using a curing light, the irradiation causes the methylacrylate resins to undergo polymerization. This polymerization forms a bond between the etched enamel surface and the bracket adhesive
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a stylized font, followed by the words "AMERICAN ORTHODONTICS" in a smaller, sans-serif font. The letters and words are all in a dark blue color. The logo is simple and modern, and it is likely used to represent the company's brand.
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
Page 5-2
#### Indications for Use:
Orthodontic Primers are intended to be used in the bonding procedure (direct or indirect) of brackets and tubes by preparing the enamel during orthodontic treatment to increase adhesion to wet or dry, normal and atypical enamel surfaces. The Primer wets the bonding surface of the object to be bonded (i.e. enamel, composite materials, etc.) to promote adhesion.
#### Substantial Equivalence Discussion:
The Indications for Use statement for American Orthodontics' BracePaste MTP Light Cure Primer is not identical to the predicate in that the subject device's intended use is more descriptive. However, the differences do not alter the intended use of the device relative to the predicate. Both devices have the same intended use – to prepare the enamel for orthodontic treatment for bonding of brackets and tubes to increase adhesion.
Both devices have the same technological characteristics through delivery system, chemical characteristics, curing method, and incorporation of similar materials such as ethyl alcohol, Bisphenol a Diglycidyl Ether Dimethacrylate, 2-Hydroxyethyl Methacrylate, 2-Hydroxy-1,3-Dimethacryloxypropane, Diurethane Dimethacrylate, and Water.
| | Device Name / Manufacturer | | |
|----------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Element | Device: Transbond MIP Primer<br>Manufacturer: 3M Dental Products Laboratory | Device: BracePaste MTP Light Cure Primer<br>Manufacturer: American Orthodontics | |
| 510(k) Number | K962785 | unknown | |
| Classification<br>Code/ Regulation<br>Number | KLE<br>872.3200 | KLE<br>872.3200 | |
| Intended Use | This product is intended for use in orthodontic<br>treatment. | Orthodontic Primers are intended to be used in the<br>bonding procedure (direct or indirect) of brackets<br>and tubes by preparing the enamel during<br>orthodontic treatment to increase adhesion to wet<br>or dry, normal and atypical enamel surfaces. The<br>Primer wets the bonding surface of the object to be<br>bonded (i.e. enamel, composite materials, etc.) to<br>promote adhesion. | |
| Delivery System | 6mL Dropper Bottle | 6mL Dropper Bottle | |
| Material<br>Composition | · Ethyl Alcohol<br>· Bisphenol a Diglycidyl Ether Dimethacrylate<br>· 2-Hydroxyethyl Methacrylate<br>· 2-Hydroxy-1,3-Dimethacryloxypropane<br>· Diurethane Dimethacrylate<br>· Water | · Ethyl Alcohol<br>· Bisphenol a Diglycidyl Ether Dimethacrylate<br>· 2-Hydroxyethyl Methacrylate<br>· 2-Hydroxy-1,3-Dimethacryloxypropane<br>· Diurethane Dimethacrylate<br>· Water | |
| Chemical<br>Characteristics | Hydrophilic | Hydrophilic | |
| Curing Method | Light Activated | Light Activated | |
| Consistency | Liquid | Liquid | |
| Bonding Type | Direct/Indirect | Direct/Indirect | |
The table below outlines the comparison of the predicate device (Transbond MIP Primer) and subject device (BracePaste MTP Light Cure Primer) to show substantial equivalence.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for American Orthodontics. The logo consists of the letters "AO" in a large, bold, sans-serif font, followed by the words "AMERICAN ORTHODONTICS" in a smaller, sans-serif font. The letters and words are all in the same blue color. The logo is simple and modern.
American Orthodontics 3524 Washington Avenue Sheboygan, WI 53081
www.americanortho.com
1 800 558 7687 +1 920 457 5051 office +1 920 457 1485 fax
Page 5-3
## Performance Testing:
Clinical Performance Testing No clinical performance testing has been conducted.
## Non-Clinical Performance Testing
The following non-clinical performance tests were conducted:
- In vitro Cytotoxicity Assay: Direct Cell Contact Test with MTP Light Curable Moisture Tolerant ● Primer (Study No. 176263)
- Interaction with Accessories: MTP Light Curable Moisture Tolerant Primer Compatibility ●
- Stability Test: MTP Light Curable Moisture Tolerant Primer Compatibility
- Test Report # 1701006; DIN 13990-2 Primer Testing ●
- Test Report # 1706013: DIN 13990-2: Bond Strength of MTP Primer with BracePaste
- Test Report # 1708011: Moisture Tolerant Primer Adhesive Strength Testing
- Performance Testing in accordance with ISO 29022:2013 Dentistry-Adhesion-Notched-edge shear bond strength test
- . Test Protocol: Transbond MIP, 3M Unitek
- Test Protocol: MTP Light Curable Moisture Tolerant Primer .
The combination of in-house testing and side-by-side comparison performed by the original manufacturer has demonstrated the efficacy or suitability to the intended purpose of the BracePaste MTP Light Cure Primer. Results of bench testing indicate that BracePaste MTP Light Cure Primer performs as well as the predicate Transbond MIP Primer.
#### Test Method Summary:
| Test Method | BracePaste MTP Light Cure Primer | Transbond MIP Primer |
|---------------------------|----------------------------------|----------------------|
| Average Bond Strength | 38.94/mm² | 38.70 N/mm² |
| Refractive Index (20.0°C) | 1.4284 | 1.4441 |
## Conclusion:
In summary, BracePaste MTP Light Cure Primer has the following similarities to the legally marketed predicate Transbond MIP Primer (K962785):
- Same intended use, and ●
- Same technological characteristics through delivery system, chemical characteristics, curing ● method, consistency, bonding type (direct/indirect) and incorporation of similar materials.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.