Portable X-ray System (Model: MiniX-V, Mini X-S)

K181891 · Digimed Co., Ltd. · EHD · Aug 10, 2018 · Dental

Device Facts

Record IDK181891
Device NamePortable X-ray System (Model: MiniX-V, Mini X-S)
ApplicantDigimed Co., Ltd.
Product CodeEHD · Dental
Decision DateAug 10, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.1800
Device ClassClass 2
AttributesPediatric

Indications for Use

The portable x-ray system (Models: MiniX-V, MiniX-S) is intended to be used by trained dentists and dental technicians as extra-oral x-ray source for producing diagnostic x-ray images using intra-oral image receptors. Its use is intended for both adult and pediatric subjects.

Device Story

Portable X-ray system (MiniX-V, MiniX-S) generates diagnostic X-rays for dental imaging. Device comprises X-ray tube, high-voltage transformer, high-frequency inverter, control PCB, and LCD user interface. Operated by dentists/technicians in clinical settings; handheld or mounted on arms/tripods. User selects tube voltage and exposure time via interface; system generates mixed pulse voltage (up to 70kV) to produce X-ray beam. Beam passes through patient's teeth/jaw to intra-oral image receptors (film/digital sensors). Lead-shielded housing and backscatter shield protect users. Output images used by clinicians for dental diagnosis. Benefits include portable, point-of-care diagnostic imaging capability.

Clinical Evidence

Bench testing only. No clinical data required to establish substantial equivalence. Performance evaluation and device inspection reports confirmed suitability for intended use.

Technological Characteristics

Extra-oral X-ray source; 22.2V DC Lithium Polymer battery; mixed pulse waveform; 70kV max (MiniX-V) / 60kV max (MiniX-S); 2mA fixed current; 0.01-2.0s exposure time. Lead-shielded housing. Connectivity: standalone. Standards: IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-65, IEC 62304, EN60601-1-2.

Indications for Use

Indicated for use by trained dentists and dental technicians as an extra-oral x-ray source for producing diagnostic x-ray images of teeth and jaws in adult and pediatric patients using intra-oral image receptors (analog film, CCD/CMOS sensors, or phosphor plates).

Regulatory Classification

Identification

An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 10, 2018 DIGIMED Co., Ltd. % Mr. Yong Park President INNODEN LLC 212 Wells Ave.S #102 RENTON WA 98057 Re: K181891 Trade/Device Name: Portable X-rav System (Models: MiniX-V. MiniX-S) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: EHD Dated: July 13, 2018 Received: July 13, 2018 Dear Mr. Park: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); and Part 809; medical {1}------------------------------------------------ device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Hsl 2. Nils Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) #### K181891 Device Name Portable x-ray system (Model: MiniX-V, MiniX-S) #### Indications for Use (Describe) The device is a diagnostic x-ray system, which is intended to be used by trained dentists and dental technicians as an extra-oral x-ray source for producing diagnostic x-ray images using intra-oral image receptors. its use is intended for both adults and pediatric subjects. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary of Portable X-ray System (MiniX-V & MiniX-S) #### Date Prepared: Mar. 12, 2018 Company Name: DIGIMED CO., LTD. 309~311, 318-ho, 145, Gasan digital 1-ro, Geumcheon-gu, Address: Seoul, 08506 Rep. of Korea Telephone number: +82-2-2624-1551 Fax: +82-2-2624-1553 E-mail: digimed@digimed.co.kr Contact: Youngbae Kwon, CEO ## 1. Company and Correspondent Making the Submission: ## 2. US Agent for FDA Contact: | Name | Mr. Yong Park | |-------------------|--------------------------------------------| | Company Name: | INNODEN LLC. | | Address: | 212 Wells Ave. S #102,<br>Renton, WA 98057 | | Telephone number: | 425-572-0283 | | Fax: | 425-988-1256 | | E-mail: | drpark@innoden.com | ## 3. Device Information: | Proprietary-Trade Name: | Portable X-Ray System (Models: MiniX-V, MiniX-S) | |-------------------------|--------------------------------------------------------------------------------| | Classification Name: | Extra-oral Source X-Ray System: EHD, Class II per<br>regulation 21CFR 872.1800 | | Common/Usual Name: | Portable X-Ray System | ## 4. Predicate Devices: | Manufacturer: | DIGIMED CO., LTD. | |-----------------|--------------------------------------------------------------------------------| | Device: | Portable X-Ray System<br>(Model: MiniX-V, MiniX-S) | | Classification: | Extra-oral Source X-Ray System: EHD, Class II per<br>regulation 21CFR 872.1800 | | 510(k) Number: | K152859 | {4}------------------------------------------------ ## 5. Indications for Use (Intended Use): The portable x-ray system (Models: MiniX-V, MiniX-S) is intended to be used by trained dentists and dental technicians as extra-oral x-ray source for producing diagnostic x-ray images using intra-oral image receptors. Its use is intended for both adult and pediatric subjects. Compatible intra-oral image receptors are described as below. - 1) Analog dental x-ray films - 2) CCD and CMOS digital sensors - 3) Digital phosphor plates - 4) Most of the intra-oral image receptors which intended for dental x-ray examination ## 6. Description of Device The portable x-ray system (Models: MiniX-V, MiniX-S) is an x-ray generating device which is mainly designed for dental examination. The device has an x-ray tube for generating x-rays, a high voltage transformer for generating high voltages, a high voltage rectification circuit for transforming and boosting AC voltage to mixed pulse voltage, a high voltage divide circuit for lowering high voltage to measure and calibrate high voltages, as well as a high voltage tube tank, a high frequency inverter circuit for generating high voltages, a control P.C.B for controlling, saving and displaying the data, a power P.C.B for supplying power to the circuit and apparatus in housing, user interface (LCD screen) and beam limiting part (x-ray emitting cone). The apparatuses above can be embedded in one or several cases, and except for the radiation opening in the x-ray system, all units are completely shielded by lead or high-density materials, protecting patients and users from unnecessary exposure of radiation. The package includes a remote control switch, a battery charger, and a backscatter shield. And the remote control switch can be used when the device is mounted on fixable stands as optional arms in user manual or digital camera tripods. Operating principle is that the high voltage electricity is getting into the x-ray tube and generates x-ray source. This x-ray source comes out from the emitting cone and goes through teeth and jaw of patient, and makes x-ray images onto image receptors as a chemical film or a digital sensor. Once the user enters the desired tube voltage data in the numerical value (%) by means of the user interface and press the x-ray exposure button, the system generates mixed pulse voltage (MiniX-V: 70kV max, MiniX-S: 60kV max) from the transformed and boosted AC voltage through high voltage generator. So the user is able to obtain x-ray images of patients' teeth and jaws from the image receptors, and diagnose about the images. Detailed information of the device is described in "13. Device Description" of this submission. ### 7. Safety and Effectiveness, Comparison to Predicate Device: Safety and effectiveness of the subject device is considered with the latest version of test regulations. The subject device was shown to provide an equivalent level of safety and performance as compared to the predicate devices. "Clinical images were provided however they were not necessary in order to establish substantial equivalence with the predicate devices" {5}------------------------------------------------ ### 8. Safety, EMC and Performance Data: The subject device complies with the safety and performance standards listed in the chart below. Test reports were provided to demonstrate conformance. All required documents and reports are submitted to the appropriate oversight agency to establish compliance with the applicable requirements. ### 9. The differences between the subject device and the predicate devices The subject device has little differences with its design, size, and user interface compare to the predicate devices. Detailed differences can be identified from "Substantial Equivalence Chart". ### 10. FDA Guidance Documents utilized; - 1) Radiation Safety Considerations for X-ray Equipment Designed for Hand-Held Use, issued December 24, 2008 - 2) Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued May 11, 2005 - 3) Pediatric Information for X-ray Imaging Device Premarket Notifications, issued November 28, 2017 - 4) Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, issued October 2, 2014 {6}------------------------------------------------ # 11. Substantial Equivalence Chart | FEATURE | DIGIMED<br>MiniX-V, MiniX-S<br>(K152859) | DIGIMED<br>MiniX-V, MiniX-S<br>(K181891) | |-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | INTENDED USE | All three systems are intended as extraoral x-ray sources to be used with intraoral image receptors for diagnostic imaging by dentists or dental technicians. | | | MECHANICAL | | | | Size | 4.8" x 8.7" x 5.3" | 4.8" x 8.7" x 5.3" | | Source to<br>skin distance | 20 cm | 20 cm | | Cone diameter | 5.3 cm | 5.3 cm | | User interface | Up-down buttons for exposure time<br>selection with timer display.<br>Additionally, several user-selectable<br>preset times with patient size, image-<br>receptor type, and tooth selection icons<br>on an LCD display | Up-down buttons for exposure time<br>selection with timer display.<br>Additionally, several user-selectable<br>preset times with patient size, image-<br>receptor type, and tooth selection icons<br>on an LCD display | | Backscatter<br>radiation protection | 6.3" dia. Pb-filled acrylic plastic scatter<br>shield | 6.3" dia. Pb-filled acrylic plastic scatter<br>shield | | Exposure switch | Exposure button at front cover on right<br>hand side,<br>or a remote control switch | Exposure button at front cover on right<br>hand side,<br>or a remote control switch | | Tubehead mounting | Handheld, on optional arms or on a<br>tripod | Handheld, on optional arms or on a<br>tripod | | Half-value layer | MiniX-V: 2.3 mm Al<br>MiniX-S: 2.0 mm Al | MiniX-V: 2.3 mm Al<br>MiniX-S: 2.0 mm Al | | ELECTRICAL | | | | Energy source | Rechargeable 22.2 V DC Lithium<br>Polymer battery pack | Rechargeable 22.2 V DC Lithium<br>Polymer battery pack | | Exposure time | 0.01~2.0 seconds in 0.01 increments | 0.01~2.0 seconds in 0.01 increments | | Time Accuracy | $\pm$ (10% +1 ms) | $\pm$ (10% +1 ms) | | mA | 2 mA fixed | 2 mA fixed | | kVp | MiniX-V: 70kV<br>MiniX-S: 60kV | 70 kV max | | Waveform | Constant Potential (DC) | Mixed pulse | | Duty Cycle | 1:60 | 1:60 | | EMI standards | EN60601-1-2, IEC60601-1-2<br>CISPR 11, IEC61000-3-2<br>IEC61000-3-3 | EN60601-1-2, IEC60601-1-2<br>CISPR 11, IEC61000-3-2<br>IEC61000-3-3 | | X-RAY PERFORMANCE | | | | Performance<br>standards | IEC 60601-1<br>IEC 60601-1-3<br>IEC 60601-2-65 | IEC 60601-1<br>IEC 60601-1-3<br>IEC 60601-1-6<br>IEC 60601-2-65<br>IEC 62304 | | Difference | X-ray waveform | | - 1) Predicate device (Model: MiniX-V, MiniX-S) from DIGIMED {7}------------------------------------------------ ## 12. Conclusion In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided above comparison table, the new 510k submission of the portable xray system (Models: MiniX-V, MiniX-S) has difference with its x-ray waveform compare to previous submission (K152829). Except the waveform, the system has the same specification as the previous submission. Performance evaluation (test) reports and device inspection report confirmed that the Portable X-ray System is suitable for its intended use and user instruction of the device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...