← Product Code [NHA](/productcode/NHA) · K181869

# On1 Universal Abutment (K181869)

_Nobel Biocare AB · NHA · Dec 12, 2018 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181869

## Device Facts

- **Applicant:** Nobel Biocare AB
- **Product Code:** [NHA](/productcode/NHA.md)
- **Decision Date:** Dec 12, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3630
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The On1™ device is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation. The On1 Universal Abutments consist of three major parts. Specifically, the On1 Base, the On1 Universal Abutment, and the mesostructure components make up a multi-piece abutment. The system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.

## Device Story

On1 Universal Abutment is a multi-piece dental abutment system; attaches to On1 Base on Nobel Biocare implants. Input: digital scan data from intra-oral scanners (e.g., 3Shape TRIOS). Process: CAD/CAM workflow using 3Shape Abutment Designer software to design mesostructures; milling via CORiTEC imes-icore unit. Output: custom-designed mesostructure or abutment crown bonded to the abutment. Used in dental clinics by clinicians; facilitates prosthetic rehabilitation by providing a standardized platform for CAD/CAM design. Benefits: simplifies design process; enables patient-specific restorative solutions. Output affects clinical decision-making by providing a stable, predictable interface for final dental restorations.

## Clinical Evidence

Bench testing only. Included dynamic fatigue testing per ISO 14801, biocompatibility testing per ISO 10993-5 (cytotoxicity), and GC-MS analysis for organic leachables/extractables per ISO 10993-12/18. Software verification and validation performed for the design library and workflow integration.

## Technological Characteristics

Materials: Titanium vanadium alloy (ASTM F136). Design: Multi-piece abutment with fixed upper shape and indexing feature for CAD/CAM. Connectivity: Integrates with 3Shape digital workflow. Sterilization: Non-sterile, steam sterilized by end-user (ISO 17665-1/2). Mechanical: Dynamic fatigue testing per ISO 14801. Max angulation: 20°.

## Regulatory Identification

An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)

## Predicate Devices

- On1 Concept ([K161655](/device/K161655.md))

## Reference Devices

- Vita Enamic® Implant Solutions (IS) ([K153645](/device/K153645.md))
- Sirona Dental CAD/CAM System ([K111421](/device/K111421.md))
- Interactive Smartbase Abutments ([K181359](/device/K181359.md))
- 3Shape Abutment Designer Software ([K151455](/device/K151455.md))
- Multilink Hybrid Abutment Cement ([K130436](/device/K130436.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "U.S. Food & Drug Administration" in blue letters.

December 12, 2018

Nobel Biocare AB % Charlemagne Chua Senior Regulatory Affairs Manager Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda. California 92887

Re: K181869

Trade/Device Name: On1™ Universal Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA, PNP Dated: November 8, 2018 Received: November 9, 2018

Dear Charlemagne Chua:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Andrew I. Steen -S

Tina Kiang, Ph.D. for Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K181869

Device Name On1™ Univeral Abutment

#### Indications for Use (Describe)

The Onl™ device is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation. The Onl Universal Abutments consist of three major parts. Specifically, the On1 Base, the On1 Universal Abutment, and the mesostructure components make up a multi-piece abutment.. The system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.

#### Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

A.4.

510(k) Summary

I. SUBMITTER

Nobel Biocare AB Vastra Hamngatan 1 Goteborg, SE-411 17 Sweden

Submitted by: Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, CA 92887

Contact Person: Charlemagne Chua, Senior Regulatory Affairs Manager Phone: (714) 282-4800 x 7830 Fax: (714) 998-9348

Date Prepared: December 11, 2018

II. DEVICE

Name of Device: On1TM Universal Abutment Common or Usual Name: Endosseous Dental Implant Abutment Classification Name: Endosseous Dental Implant (21 CFR 872.3630) Regulatory Class: II Primary Product Code: NHA Secondary Product Code: PNP

III.PREDICATE DEVICE

Primary Predicate K161655, Nobel Biocare AB - On1 Concept

Reference Devices K153645, Vita Zahnfabrik H. Rauter GmbH Co - Vita Enamic® Implant Solutions (IS) K111421. Sirona Dental Systems GmbH - Sirona Dental CAD/CAM System K181359, Implant Direct Sybron Manufacturing Llc - Interactive Smartbase Abutments K151455, 3Shape A/S - 3Shape Abutment Designer Software K130436, Ivoclar Vivadent, Inc. - Multilink Hybrid Abutment Cement

IV. DEVICE DESCRIPTION

{4}------------------------------------------------

The On1 Universal Abutment is a dental implant abutment which attaches to the On1 Base of the On1 Concept (K161655) and is intended to be used with the current Nobel Biocare dental implants that have the existing internal conical connection.

The On1 Universal Abutment features a fixed upper shape with indexing feature that is intended to serve as the platform for either an in-laboratory CAD/CAM system made mesostructure or abutment crown. The fixed upper shape and indexing feature facilitates the use of CAD/CAM systems by providing a known shape that can be imported into the design software, thereby, simplifying the CAD/CAM design process.

The On1 Universal Abutment is available for the Nobel Biocare Narrow Platform (NP), Regular Platform (RP) and Wide Platform (WP) for the internal conical connection. The On1 Universal Abutment is made of titanium vanadium alloy.

The digital workflow requires the use of the following equipment:

- Scanner: 3Shape intra oral scanner Trios (3Shape A/S) ।
- -Design Software: 3Shape Abutment Designer Software (3Shape A/S) -K151455 where the Implant Libraries are obtained via the 3Shape server in the software
- -Restorative Material: Enamic (Vita Zahnfabrik H. Rauter GmbH Co) -K153645
- Milling Unit: CORiTEC, imes-icore milling unit -

| Restorative design specifications for Universal Base |                                                 |
|------------------------------------------------------|-------------------------------------------------|
| Parameter                                            | Specification                                   |
| Angle from axis of the implant                       | 20° Max                                         |
| Wall Thickness Circular                              | 0.8mm min.                                      |
| Wall Thickness Margin                                | 0.275mm min.                                    |
| Post Height                                          | 5.2mm min.                                      |
| Maximum Length, width and Height                     | EM-14 blank 12x14x18mm<br>EM-10 blank 8x10x15mm |

The following restorative design specifications is required:

# V. INDICATIONS FOR USE

The On1™ device is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation. The On1 Universal Abutments consists of three major parts. Specifically, the On1 Base, the On1 Universal Abutment, and the mesostructure components make up a 

{5}------------------------------------------------

multi-piece abutment. The system integrates multiple components of the digital dentistry workflow: scan files from Intra-Oral Scanners, CAD software, CAM software, ceramic material, milling machine and associated tooling and accessories.

{6}------------------------------------------------

#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS VI.

| Technological<br>characteristics | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Primary Predicate                                                                                                                                                                | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                           |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                  | On1 Universal Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | On1 Concept<br>(K161655)                                                                                                                                                         | VITA ENAMIC IS<br>(K153645)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Sirona Dental CAD/CAM-<br>System (K111421)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | InterActive SMARTBase<br>Abutments<br>(K181359)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                           |
|                                  | Compatible<br>Implant<br>Platform                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Nobel Biocare Internal<br>Conical Connection<br>- Narrow Platform (NP)<br>- Regular Platform (RP)<br>- Wide Platform (WP)                                                        | Nobel Biocare Internal<br>Conical Connection<br>- Narrow Platform (NP)<br>- Regular Platform (RP)<br>- Wide Platform (WP)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | The InCoris mesostructure<br>and TiBase two-piece<br>abutment is compatible with<br>the following implant<br>systems: Nobel Biocare<br>Replace (K020646), Nobel<br>Biocare Branemark<br>(K022562), Friadent Xive<br>(K013867), Biomet 3i<br>Osseotite (K980549), Astra<br>Tech Osseospeed<br>(K091239), Zimmer<br>Tapered Screw-Vent<br>(K061410), Straumann<br>Synocta (K061176),<br>Straumann Bone Level<br>(K053088), Biomet 3i<br>Certain (K014235), Nobel<br>Biocare Active (K071370).                                                                                                                                                                                                                                                                                                                                                 | The InCoris mesostructure<br>and TiBase two-piece<br>abutment is compatible with<br>the following implant<br>systems: Nobel Biocare<br>Replace (K020646), Nobel<br>Biocare Branemark<br>(K022562), Friadent Xive<br>(K013867), Biomet 3i<br>Osseotite (K980549), Astra<br>Tech Osseospeed<br>(K091239), Zimmer<br>Tapered Screw-Vent<br>(K061410), Straumann<br>Synocta (K061176),<br>Straumann Bone Level<br>(K053088), Biomet 3i<br>Certain (K014235), Nobel<br>Biocare Active (K071370).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | InterActive SMARTBase<br>abutments are compatible<br>at the implant level with<br>InterActive (3.0mm and<br>3.4mm Platform) and<br>SwishActive (3.0mm and<br>3.4mm Platform) system<br>implants.                                                                                                                                                                                          |
|                                  | Ti/ Abutment<br>Material                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | On1 Base (K161655), On1<br>Universal Abutment -<br>Titanium vanadium alloy<br>(ASTM F136)                                                                                        | On1 Base, On1 Esthetic<br>Abutment Titanium -<br>Titanium vanadium alloy<br>(ASTM F136)<br>On1 Esthetic Abutment<br>Zirconia<br>- Y-TZP zirconium oxide<br>(ISO<br>6872, ISO 13356)                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Titanium vanadium alloy<br>(ISO 5832-3)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Tibase - Titanium 6AL4V                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Titanium                                                                                                                                                                                                                                                                                                                                                                                  |
| Design Features                  | Mesostructure<br>/<br>Crown<br>Material                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Enamic (K153645)                                                                                                                                                                 | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | VITA ENAMIC IS, identical<br>in composition to Enamic<br>(K153645)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Cercon HT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Zirconia                                                                                                                                                                                                                                                                                                                                                                                  |
|                                  | Mesostructure<br>/Crown<br>Fixation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Bonded                                                                                                                                                                           | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Bonded                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Bonded                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Bonded                                                                                                                                                                                                                                                                                                                                                                                    |
|                                  | Abutment<br>Design                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 2 piece abutment on On1<br>Base - Enamic (K153645)                                                                                                                               | 1 piece abutment on On1<br>Base - Esthetic Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 2 piece - VITA ENAMIC IS<br>bonded to Sirona TiBase                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | The TiBase is a<br>premanufactured prosthetic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Abutment body consisting<br>of a titanium base and                                                                                                                                                                                                                                                                                                                                        |
| Technological<br>characteristics |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Subject Device                                                                                                                                                                   | Primary Predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Reference Device                                                                                                                                                                                                                                                                                                                                                                          |
|                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | On1 Universal Abutment                                                                                                                                                           | On1 Concept<br>(K161655)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | VITA ENAMIC IS<br>(K153645)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Sirona Dental CAD/CAM-<br>System (K111421)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | InterActive SMARTBase<br>Abutments<br>(K181359)                                                                                                                                                                                                                                                                                                                                           |
|                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | bonded to the On1<br>Universal Abutment and<br>mounted on to the On1<br>Base (K161655). On1 Base<br>is mounted on to the<br>implant and fixed with a<br>screw.                   | Titanium is mounted on to<br>the On1 Base. On1 Base is<br>mounted on to the implant<br>and fixed with a screw.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | (K111421) mounted on to<br>the implant and fixed with a<br>screw                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | component directly<br>connected to endosseous<br>dental implants with a<br>screw and is intended for<br>use as an aid in prosthetic<br>rehabilitation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | supplied with a fixation<br>screw. The bases are<br>provided with straight,<br>angled, and modified<br>zirconia tops for patient<br>specific devices. The<br>devices are also provided<br>without a zirconia top and a<br>superstructure or hybrid<br>crown or bridge can be<br>milled to fit the bases<br>intended to be<br>manufactured at Implant<br>Direct Manufacturing<br>facility. |
|                                  | Abutment<br>Fixation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Screwed                                                                                                                                                                          | Screwed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Screw or Cement<br>Retained                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Screw or Cement<br>Retained                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Screw Retained                                                                                                                                                                                                                                                                                                                                                                            |
|                                  | Maximum<br>Angulation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 20°                                                                                                                                                                              | 0°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 20°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 20°                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 30°                                                                                                                                                                                                                                                                                                                                                                                       |
|                                  | Design<br>Workflow                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 3Shape intra oral scanner<br>Trios (3Shape A/S),<br>3Shape Abutment Designer<br>Software (3Shape A/S) -<br>K151455                                                               | Traditional workflow                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Sirona software - inlab<br>15.0 and above<br>Sirona software - CEREC<br>4.4 and above                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Sirona software - inlab<br>15.0 and above<br>Sirona software - CEREC<br>4.4 and above                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 3M Tru-Definition, ITero<br>Scanner<br>3Shape Abutment Designer<br>Software (3Shape A/S) -<br>K151455                                                                                                                                                                                                                                                                                     |
|                                  | Manufacturing<br>Workflow                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | CORITEC milling unit<br>(imes-icore)                                                                                                                                             | Traditional workflow                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Sirona CEREC milling<br>systems                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Sirona CEREC milling<br>systems                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Wieland-Zenotec Select &<br>Zenotec CAM                                                                                                                                                                                                                                                                                                                                                   |
|                                  | Mechanical<br>Testing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Dynamic Fatigue<br>Testing per ISO<br>14801                                                                                                                                      | Dynamic Fatigue<br>Testing per ISO<br>14801                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Dynamic Fatigue<br>Testing per ISO<br>14801                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | -                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Dynamic Fatigue<br>Testing per ISO<br>14801                                                                                                                                                                                                                                                                                                                                               |
|                                  | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Primary Predicate                                                                                                                                                                | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Reference Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                           |
| Technological<br>characteristics | On1 Universal Abutment                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | On1 Concept<br>(K161655)                                                                                                                                                         | VITA ENAMIC IS<br>(K153645)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Sirona Dental CAD/CAM-<br>System (K111421)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | InterActive SMARTBase<br>Abutments<br>(K181359)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                           |
| Indications for Use              | The On1™ device is a<br>premanufactured prosthetic<br>component directly<br>connected to an<br>endosseous implant and it<br>is intended for use in<br>prosthetic rehabilitation.<br>The On1 Universal<br>Abutments consist of three<br>major parts. Specifically,<br>the On1 Base, the On1<br>Universal Abutment, and<br>the mesostructure<br>components make up a<br>multi-piece abutment.<br>The system integrates<br>multiple components of the<br>digital dentistry workflow:<br>scan files from Intra-Oral<br>Scanners, CAD software,<br>CAM software, ceramic<br>material, milling machine<br>and associated tooling and<br>accessories | The On1™ device is a<br>premanufactured prosthetic<br>component directly<br>connected t…

---

**Source:** [https://fda-staging.innolitics.com/device/K181869](https://fda-staging.innolitics.com/device/K181869)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com