K181867 · El.En Electronic Engineering Spa · GEX · Oct 2, 2018 · General, Plastic Surgery
Device Facts
Record ID
K181867
Device Name
Deka Smartxide2 Trio
Applicant
El.En Electronic Engineering Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 2, 2018
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
CO2 laser: Incision, excision, ablation, vaporization, and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynaecology, neurosurgery, orthopaedics, general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery. 940nm diode laser: Incision, excision, vaporization and coagulation of soft tissues (open surgery), cutting, vaporization, ablation and coagulation of soft tissues (endoscopic surgery) in medical specialties including: plastic surgery, dermatology, ENT, gynaecology, urology, general surgery, gastroenterology and dental procedures. 980nm diode laser: Incision, excision, vaporization ablation and coagulation of soft tissues (open and endoscopic surgery) in medical specialties including plastic surgery, dermatology, ear, nose and throat and oral surgery (otolaryngology), gynaecology, urology, neurosurgery, general and thoracic surgery, gastroenterology and dental procedures.
Device Story
Medical laser system comprising 80W CO2 laser (10600nm) and optional 50W diode laser (940nm or 980nm). CO2 laser energy delivered via articulated arm or flexible silica capillary hollow waveguide; diode laser energy delivered via optical fibers. System operated by physician in clinical settings; parameters selected via front panel; activation via footswitch. Device provides surgical cutting, ablation, and coagulation of soft tissues. Modification adds hollow waveguide delivery for CO2 laser to improve access in specific surgical procedures. Benefits include enhanced surgical precision and versatility in soft tissue management.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical verification and validation activities, including verification of CO2 laser output using the waveguide and software validation.
Technological Characteristics
System includes 80W RF-excited CO2 laser and 50W semiconductor diode laser. Delivery via articulated arm (seven-mirror assembly) or flexible silica capillary hollow waveguide. Dimensions: 240cm x 59cm x 56cm; Weight: 100kg. Electrical: 100-120Vac or 220-230Vac. Includes integrated surgical scanning system (HiScan Surg., Endoscan).
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues across various medical specialties including dermatology, plastic surgery, podiatry, ENT, gynecology, neurosurgery, orthopedics, general/thoracic surgery, and dental/oral surgery. Applicable to both open and endoscopic procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
October 2, 2018
El.En Electronic Engineering Spa Paolo Peruzzi Regulatory Affairs Manager Via Baldanzese 17 Calenzano, 50041 It
Re: K181867
Trade/Device Name: Deka Smartxide2 Trio Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: July 10, 2018 Received: July 12, 2018
Dear Paolo Peruzzi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/ofpmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entiled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K181867
Device Name DEKA SMARTXIDE2 TRIO
#### Indications for Use (Describe)
The DEKA SmartXide2 Trio CO2 laser is indicated for incision, ablation, vaporization, and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynaecology, neurosurgery, orthopaedics, general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery.
The DEKA Smartxide2 Trio 940nm diode laser is incision, excision, vaporization ablation and coagulation of soft tissues (open surgery), cutting, vaporization of soft tissues (endoscopic surgery) in medical specialties including: plastic surgery, dermatology, ENT, gynaecology, urology, general surgery, gastroenterology and dental procedures.
The DEKA SmartXide2 Trio 980mm diode laser is incision, excision, vaporization ablation and coagulation of soft tissues (open and endoscopic surgery) in medical specialties including plastic surgery, dermatology, ear, nose and throat and oral surgery (otolaryngology), gynaecology, neurosurgery, general and thoracic surgery, gastroenterology and dental procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </span> |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary
# DEKA Smartxide2 Trio - Special 510(k)
#### Submitter:
El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy
### Contact:
Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it
### Date Summary Prepared:
July 10, 2018
### Device Trade Name:
Deka Smartxide2 Trio
### Common Name:
Medical Laser system
#### Classification Name:
Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology (GEX)
#### Classification Number:
21 CFR 878.4810
#### Equivalent Devices:
DEKA Smartxide2 (K113504)
#### Device Description:
The DEKA SmartXide2 Trio is a medical laser system equipped with a 80W CO2 laser source and an (optional) 980nm or 940nm 50W diode laser source.
The CO2 laser radiation has a wavelength of 10600nm and is delivered to the treatment area through an articulated arm and a delivery accessory (handpiece/scanner/micromanipulator) connected to its distal end, or through a hollow waveguide.
The articulated arm is an optical assembly that delivers free beam laser radiation. It is made up of seven mirrors placed on rotating knuckles: the mechanical accuracy of the articulated arm allows the CO2 laser beam to travel inside it and along its axis regardless of the arm orientation.
The waveguide consists of a flexible silica capillary whose inside wall has been coated with a durable coating which is highly reflective at the intended wavelength of use.
The waveguides used with DEKA Smartxided Trio are manufactured by Laser Engineering and have been cleared by FDA with K112166.
The diode laser source can be provided in two alternative wavelengths: 940nm and 980mm.
{4}------------------------------------------------
The diode laser radiation is delivered to the treatment area through optical fibers, which are guided to the target tissue with the aid of handpieces. The spot size is effectively the diameter of the fiber being connected to the system.
Emission parameters are selected on the front panel while laser emission is activated by a footswitch. The on-off switch and emergency switch are also located on the front panel of the svstem.
A warning light is located on the top cover, close to the control panel. Light ON state indicates that the system is enabled and ready.
Overall weight of the device is 100 kg, and the size is 240 cm x 59 cm x 56 cm (H x W x D). Electrical requirement is 100-120Vac 50/60Hz, 220-230Vac 50Hz, 16A.
The modification to the device is the addition of a hollow waveguide delivery system as an alternative to the articulated arm for the CO2 laser beam . It allows easier delivery of laser energy to the targeted tissue in some surgical procedures.
The intended use of the modified devices, as described in the labelling, has not changed as a result of the modification.
## Intended Use:
#### CO2 laser
Incision, excision, ablation, vaporization, and coagulation of body soft tissues in medical specialties including aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (ENT), gynaecology, neurosurgery, orthopaedics, general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery.
#### 940nm diode laser
Incision, excision, vaporization and coagulation of soft tissues (open surgery), cutting. vaporization, ablation and coagulation of soft tissues (endoscopic surgery) in medical specialties including: plastic surgery, dermatology, ENT, gynaecology, urology, general surgery, gastroenterology and dental procedures.
#### 980nm diode laser
Incision, excision, vaporization ablation and coagulation of soft tissues (open and endoscopic surgery) in medical specialties including plastic surgery, dermatology, ear, nose and throat and oral surgery (otolaryngology), gynaecology, urology, neurosurgery, general and thoracic surgery, gastroenterology and dental procedures.
#### Substantial equivalence discussion:
The DEKA Smartxide Trio is substantially equivalent to the DEKA Smartxide (K113504).
| Feature | Proposed 510(k) Device<br>DEKA SmartXide2 Trio | Predicate Device<br>K113504<br>DEKA SmartXide2<br>Laser System and delivery<br>accessories |
|-----------------|------------------------------------------------|--------------------------------------------------------------------------------------------|
| CO2 laser | | |
| Laser type | CO2 laser - sealed off - RF excited | CO2 laser - sealed off - RF excited |
| Wavelength | 10.6 µm | 10.6 µm |
| Delivery system | Articulated arm, Waveguide | Articulated arm |
| Output power | 0.1 W to 80 W | 0.1 W to 80 W |
{5}------------------------------------------------
| Feature | Proposed 510(k) Device<br>DEKA SmartXide2 Trio | Predicate Device<br>K113504<br>DEKA SmartXide2<br>Laser System and delivery<br>accessories |
|-----------------------------|----------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| Spot size | 0.125 mm - 1.5 mm (articulated arm)<br>0.3 -0.5 mm (waveguide) | 0.125 mm - 1.5 mm (articulated arm) |
| Pulse length | 0.04ms to 0.9s | 0.04ms to 0.9s |
| Pulse Rate | 5 to 800 Hz | 5 to 800 Hz |
| Aiming beam | 635 nm, 5mW max | 635 nm, 5mW max |
| Surgical scanning<br>system | Integrated<br>(HiScan Surg., Endoscan) | Integrated<br>(HiScan Surg., Endoscan) |
| Diode laser | | |
| Laser type | Semiconductor Diode Laser | Semiconductor Diode Laser |
| Wavelength (nm) | 940, 980 nm | 940, 980 nm |
| Output power | 0.5W - 50W | 0.5W - 50W |
| Spot size (mm) | 0.2, 0.3, 0.6 mm | 0.2, 0.3, 0.6 mm |
| Laser Operating<br>Modes | CW, pulsed | CW, pulsed |
| Pulsed mode Ton | 5ms to 2s | 5ms to 2s |
| Delivery system | Optical Fiber | Optical Fiber |
| Aiming beam | 630-670mm, 3mW max | 630-670mm, 3mW max |
The DEKA Smartxide Trio has the same indications for use as the abovementioned predicate device, with same principle of operation and same performances.
#### Clinical Performance Data:
None
#### Non-Clinical Performance Data:
The following verification and validation activities have been performed on the modified devices:
- Verification of CO2 laser output using the waveguide. The test methods, acceptance criteria and test results are documented in the Design History File of Smartxide2 Trio.
- Verification and validation of modified software. The test methods, acceptance criteria and test results are documented in the Design History File of Smartxide2 Trio.
#### Conclusion:
We can conclude that the DEKA Smartxide Trio Laser system is substantially equivalent to the predicate device.
#### Additional Information:
None
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.