← Product Code [FEB](/productcode/FEB) · K181865

# Acu-sInQ Complete Endoscope Cleaning Aid System (K181865)

_STERIS Corporation · FEB · Dec 14, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181865

## Device Facts

- **Applicant:** STERIS Corporation
- **Product Code:** [FEB](/productcode/FEB.md)
- **Decision Date:** Dec 14, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Acu-sInQ Complete Endoscope Cleaning Aid System is intended to assist manual flexible endoscope cleaning according to the instructions provided by the endoscope manufacturer, by facilitating performance of a leak test, accurate detergent dosing with water temperature monitoring, channel flushing, rinsing, and process documentation.

## Device Story

Acu-sInQ is a 4-in-1 system for manual flexible endoscope cleaning; used in hospital/endoscopy center reprocessing rooms by technicians. Inputs include barcode scans of endoscope serial numbers and technician IDs; system pre-loaded with endoscope inventory. Device performs automated scope inflation for leak testing, precision chemical dosing via peristaltic pump, and automated channel flushing with pulsating fluid technology. System monitors water temperature. Outputs include process data/chemical usage recorded via USB or printed reports. Software enables data transfer to customer PC. Device assists manual cleaning process by providing step-by-step sequences, ensuring adherence to manufacturer instructions, and documenting the cleaning process for quality management.

## Clinical Evidence

Bench testing only. Testing confirmed consistent fluid flushing volume (open and connected to endoscope), detergent dosing accuracy (±5%), and water temperature monitoring accuracy (±5%). Electrical safety and electromagnetic compatibility verified per IEC 61010-1 and IEC 61326-1.

## Technological Characteristics

Microprocessor-controlled unit; internal air pump for leak testing; peristaltic pump for detergent dosing; flush pump for channel irrigation. Connectivity via USB port for data export. Electrical safety per IEC 61010-1; EMC per IEC 61326-1.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- US Endoscopy Fluid Pump CP-3 ([K914524](/device/K914524.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG' in blue, with 'ADMINISTRATION' written below in a smaller font.

December 14, 2018

STERIS Corporation Gregory Land Senior Regulatory Affairs Specialist 5960 Heisley Road Mentor, Ohio 44060

Re: K181865

Trade/Device Name: Acu-sInQ Complete Endoscope Cleaning Aid System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope And Accessories Regulatory Class: Class II Product Code: FEB Dated: November 13, 2018 Received: November 14, 2018

Dear Gregory Land:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Clarence W. Murray lii III -S

For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K181865

### Device Name

Acu-sInQ Complete Endoscope Cleaning Aid System

### Indications for Use (Describe)

The Acu-sInQ Complete Endoscope Cleaning Aid System is intended to assist manual flexible endoscope cleaning according to the instructions provided by the endoscope manufacturer, by facilitating performance of a leak test, accurate detergent dosing with water temperature monitoring, channel flushing, rinsing, and process documentation.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for STERIS. The word STERIS is in all capital letters and in a serif font. Below the word STERIS is a graphic of several horizontal wavy lines in blue.

## 510(k) Summary For Aqu-sInQ Complete Endoscope Cleaning Aid System

STERIS Corporation 5960 Heisley Road Mentor, OH 44060

Contact: Gregory Land Senior Regulatory Affairs Specialist Tel: 440-392-7424 Fax: 440-357-9198

November 9, 2018 Summary Date:

Premarket Notification Number: K181865

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{4}------------------------------------------------

#### 1. Device Name

| Trade Name:            | Acu-sInQ Complete Endoscope Cleaning Aid<br>System |
|------------------------|----------------------------------------------------|
| Device Class:          | II                                                 |
| Common/usual Name:     | Circulating pump                                   |
| Classification Name:   | Accessories, Cleaning, For Endoscope               |
| Classification Number: | 21 CFR 876.1500                                    |
| Product Code:          | FEB                                                |

#### 2. Predicate Device

US Endoscopy Fluid Pump CP-3, cleared under K914524.

#### 3. Description of Device

The Acu-sInQ Complete Endoscope Cleaning Aid System is used in the reprocessing room of a hospital or endoscopy center, and aids technicians by providing a step-by-step cleaning sequence of flexible endoscopes for process management and reports.

The Acu-sInQ Complete Endoscope Cleaning Aid System is a 4-in-1 system with automated scope inflation for performance of a leak test, precision chemical dosing, and automated channel flushing with pulsating fluid technology for main and auxiliary flush ports. The System also provides:

- Data management via printout with every cycle
- USB drive and port output for capture and recording of process data and chemical usage
- Software to enable transfer of data between system and Customer's PC
- Barcode scanner to input endoscope serial number and technician ID which aids in documentation process
- Pre-loading with Customer's endoscope inventory

#### Indications for Use 4.

The Acu-sInQ Complete Endoscope Cleaning Aid System is intended to assist manual flexible endoscope cleaning according to the instructions provided by the endoscope manufacturer, by facilitating performance of a leak test, accurate

{5}------------------------------------------------

detergent dosing with water temperature monitoring, channel flushing, rinsing, and process documentation.

#### 5. Technical Characteristics Comparison Table

A comparison of technical characteristics versus the predicate is summarized in Table 5-1

| Feature                                     | Acu-sInQ Complete Endoscope<br>Cleaning Aid<br>(proposed)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Fluid Pump CP-3 Predicate<br>(K914524)                                                                                                                                                                                                                                                                                                                                                                                                                                           | Comparison                                                                                                                                                                                                                                                                                                                                                      |                                               |                                                                                                         |                                                                                                         |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Indications<br>for Use                      | The Acu-sInQ Complete<br>Endoscope Cleaning Aid<br>System is intended to assist<br>manual flexible endoscope<br>cleaning according to the<br>instructions provided by the<br>endoscope manufacturer, by<br>facilitating performance of a<br>leak test, accurate detergent<br>dosing with water temperature<br>monitoring, channel flushing,<br>rinsing, and process<br>documentation.                                                                                                                                                                                                                                                         | The Fluid pump CP #3 is<br>intended to pump fluid media<br>thru the channels of endoscope<br>and accessories with lumens.<br>The various fluid media that<br>are compatible are soapy<br>water, enzyme/water solution,<br>water, 70% alcohol, and liquid<br>disinfectants. The pump will<br>also pump minimal amounts of<br>air to help purge the channels<br>or lumens of droplets of fluid<br>media. Consult with<br>instrument manufacturer for<br>fluid media compatibility. | Both devices are<br>intended to flush<br>fluid through the<br>channel of an<br>endoscope to aid<br>in the manual<br>cleaning process.<br>The subject<br>device has the<br>added<br>functionality to<br>assist the user<br>with additional<br>steps of the<br>endoscope<br>cleaning process,<br>leak test,<br>detergent dosing<br>and temperature<br>monitoring. |                                               |                                                                                                         |                                                                                                         |
| Feature                                     | Acu-sInQ Complete Endoscope<br>Cleaning Aid<br>(proposed)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Fluid Pump CP-3 Predicate<br>(K914524)                                                                                                                                                                                                                                                                                                                                                                                                                                           | Comparison                                                                                                                                                                                                                                                                                                                                                      |                                               |                                                                                                         |                                                                                                         |
| Operating<br>Principles/<br>Technology      | A microprocessor controlled<br>unit operates internal pumps<br>to aid in the cleaning of<br>endoscopes. An internal air<br>pump pressurizes an<br>endoscope through a dedicated<br>leak test tubing to perform a<br>leak test.  A detergent tube is<br>placed in a container of<br>detergent and a peristaltic<br>pump pulls the detergent<br>through the tubing, the<br>detergent is then expelled<br>through a discharge tube into<br>the cleaning solution. A<br>tubing set pulls fluid from the<br>cleaning solution through a<br>peristaltic pump and discharge<br>tubing set where it is expelled<br>through the endoscope<br>channels. | Flushing fluids are placed into<br>a container, pre-cleaning basin<br>or sink. A tubing set pulls<br>fluid from the container, basin,<br>or sink into and through an<br>internal fluid pump.  The fluid<br>is then expelled from the<br>pumps through discharge<br>tubing sets which are<br>connected to the endoscope<br>channels.  The fluid is then<br>pumped through the endoscope<br>channels.                                                                              | Both devices<br>utilize<br>electronically<br>controlled flush<br>pumps to flush<br>fluid through the<br>channels of an<br>endoscope. The<br>predicate device<br>is controlled<br>through analog<br>circuits while the<br>subject device<br>uses software.                                                                                                       |                                               |                                                                                                         |                                                                                                         |
| Software<br>Controlled<br>User<br>Interface | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | The subject<br>device uses<br>software to<br>control the unit<br>whereas the<br>predicate device<br>uses analog<br>circuits to control<br>to device.                                                                                                                                                                                                            |                                               |                                                                                                         |                                                                                                         |
| Fluid<br>Delivery<br>Flow Rate              | Meet or exceed the endoscope<br>manufacturer's requirements<br>for fluid delivery through<br>endoscope channels                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Meet or exceed the endoscope<br>manufacturer's requirements<br>for fluid delivery through<br>endoscope channels                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                            |                                               |                                                                                                         |                                                                                                         |
| Mechanism<br>of Action                      | Software controlled air pump,<br>peristaltic pump and flush<br>pump                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Internal flush pumps with<br>built-in timer                                                                                                                                                                                                                                                                                                                                                                                                                                      | Like the<br>predicate device<br>the subject device<br>has a flush pump.<br>Additionally the<br>subject device<br>uses an air pump<br>and peristaltic<br>pump for leak<br>testing and<br>detergent dosing.                                                                                                                                                       |                                               |                                                                                                         |                                                                                                         |
| Feature                                     | Acu-sInQ Complete Endoscope<br>Cleaning Aid<br>(proposed)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Fluid Pump CP-3 Predicate<br>(K914524)                                                                                                                                                                                                                                                                                                                                                                                                                                           | Comparison                                                                                                                                                                                                                                                                                                                                                      |                                               |                                                                                                         |                                                                                                         |
| Amount of<br>Cleaning<br>Liquid             | Amount sufficient to meet or<br>exceed the endoscope<br>manufacturer's requirements<br>for volume of fluid delivery<br>through endoscope channels.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Amount sufficient to meet or<br>exceed the endoscope<br>manufacturer's requirements<br>for volume of fluid delivery<br>through endoscope channels.                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                                                            |                                               |                                                                                                         |                                                                                                         |
|                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                 | Time for<br>Cleaning<br>Liquid<br>Circulation | Variable, the length of the<br>cleaning liquid circulation<br>time is determined by setting a<br>timer. | Variable, the length of the<br>cleaning liquid circulation time<br>is determined by setting a<br>timer. |

Table 5-1. Physical Description and Technological Properties vs the Predicate Device.

{6}------------------------------------------------

{7}------------------------------------------------

#### 6. Summary of Nonclinical Tests

The non-clincal performance testing demonstrated that the subject device met the acceptance criteria for each test that is identified and summarized in Table 5-2 below.

| Test                             | Acceptance Criteria                                                                     | Conclusion |
|----------------------------------|-----------------------------------------------------------------------------------------|------------|
| Fluid Flushing<br>(open)         | Device provides consistent fluid flushing volume<br>when not connected to an endoscope. | PASS       |
| Fluid Flushing<br>(Endoscope)    | Device provides consistent fluid flushing volume<br>when connected to an endoscope.     | PASS       |
| Dosing                           | Accuracy of dose of $\pm$ 5%                                                            | PASS       |
| Temperature<br>Accuracy          | Accuracy of temperature of $\pm$ 5%                                                     | PASS       |
| Electrical Safety                | Compliant with IEC 61010-1                                                              | PASS       |
| Electromagnetic<br>Compatibility | Compliant with IEC 61326-1                                                              | PASS       |

Table 5-2. Summary of Non-clinical Testing

#### 7. Conclusion

Based on the conclusions from the non-clinical performance data, the subject device is as safe, as effective and performs as well as or better than the legally marketed predicate device (K914524), Class II (21 CFR 876.1500), product code FEB.

---

**Source:** [https://fda-staging.innolitics.com/device/K181865](https://fda-staging.innolitics.com/device/K181865)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
