The Welch Allyn ProBP 2000 Digital blood pressure device is intended for use in measuring blood pressure and heart rate in patients at least 3 years of age or older with arm circumferences between 15 cm to 55 cm (approximately 5.9 to 21.7 inches). The Welch Allyn ProBP 2000 automatically measures systolic and diastolic pressure and pulse rate. The device is intended to be used by clinicians and medically qualified personnel. Contraindications: This device is not intended for use on neonates, infants, or children under the age of 3 years. The effectiveness of this device has not been established in pregnant, including pre-eclamptic, patients.
Device Story
Professional non-invasive blood pressure monitor; uses inflatable arm cuff; oscillometric measurement principle. Input: cuff pressure oscillations during inflation/deflation. Processing: pressure sensor converts mechanical cuff pressure to electrical signals; amplifier boosts signal; A/D converter digitizes; MCU executes oscillometric algorithm to calculate systolic/diastolic BP and pulse rate. Output: BP and pulse rate values on LCD. Used in clinical settings by trained personnel (nurses, clinicians). Benefits: provides automated, accurate BP/pulse measurements for clinical decision-making. Device includes main unit (MCU, pump, valve, sensor, display) and cuff. No wireless transmission functionality enabled.
Clinical Evidence
Clinical validation per ISO 81060-2:2013. Study population: 88 subjects (35 pediatric aged 3-12, 53 adults/adolescents aged 12-75; 39 male, 49 female). Results: mean error and standard deviation of differences for systolic/diastolic pressure within ISO limits; accuracy reported as better than predicate. No adverse effects reported.
Technological Characteristics
Oscillometric measurement; electronic pressure sensor; MCU-based processing. Materials: ABS plastic housing. Dimensions: supports arm circumferences 15-55 cm. Power: 4 AA batteries or AC/DC adapter. Connectivity: None (wireless module disabled). Software: IEC 62304 compliant. Standards: IEC 60601-1, IEC 60601-1-2, IEC 80601-2-30, ISO 10993-1 (cuffs).
Indications for Use
Indicated for patients aged 3 years and older with arm circumferences 15-55 cm. Contraindicated for neonates, infants, children under 3 years, and pregnant/pre-eclamptic patients.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 9, 2019
Guangdong Transtek Medical Electronics Co. Ltd Nicole Hu Certification Commissioner Zone A, No 105 Dongli Road, Torch Development District Zhongshan, 528437 Cn
Re: K181832
Trade/Device Name: Model: ProBP™ 2000, Welch Allyn 901123 Digital Blood Pressure Device Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: March 1, 2019 Received: March 7, 2019
Dear Nicole Hu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "TRANSTEK" in a sans-serif font. The letters are all capitalized and in a light gray color. The letters are evenly spaced and the word is centered in the image. The background is white.
# 510(k) Summary
[per 21 CFR 807.92]
| Submitter / Owner: | Guangdong Transtek Medical Electronics Co., Ltd<br>Address: Zone A, No.105, Dongli Road, Torch Development District,<br>Zhongshan, Guangdong, China 528437 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Nicole Hu, Certification Commissioner<br>Phone: 086-760-85166220 Fax: 086-760-86742877<br>E-mail: wan.hu@lifesense.com |
| Date Prepared: | Apr. 9, 2019 |
| Trade Name: | Model: ProBP™ 2000,<br>Welch Allyn 901123 Digital Blood Pressure Device |
| Common Name: | Blood pressure device |
| Classification Name: | System, Measurement, Blood pressure, Noninvasive<br>(Class II, Cardiovascular, 21 CFR 870.1130, Product Code: DXN) |
| Predicate Device: | Transtek Blood Pressure Monitor, TMB-1591-A<br>510(k) Number: K161886<br>(Class II, Cardiovascular, 870.1130, DXN) |
## Description of the Subject Device:
The device is designed to measure the systolic and diastolic blood pressure and heartbeat rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the arm.
This device is designed for professional use, in other words, it is a professional BP device that only for trained users such as nurses, clinicians, and medically qualified personnel.
The measurement method for determining the systolic and diastolic pressure is similar with the ausculatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating heartbeat rate. This well-known technique for measuring blood pressure is called the "Oscillometric method".
The device includes the main unit and cuff unit. The main unit consists of the microprocessor, the pressure sensor, the operation key, the pump, the electromagnetic deflation control valve and the LCD. ABS is the plastic used in the outer housing. The device comes with one standard Reusable-11 Welch Allyn FlexiPort cuff. Other size cuffs can be ordered as needed for arm circumferences ranging from 15cm to 55cm. This device is equipped with the same wireless radio (Bluetooth) as supplied with the predicate device cleared in K161886, TMB-1591-A.
The subject device does not incorporate any software to support the BT wirelessly transmit function. In other word, ProBP 2000 has BT module but disabled the Bluetooth functionality. And Welch Allyn does not provide an application to receive data utilizing BT
The subject device does not have any externally-facing electronic interfaces.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "TRANSTEK" in a stylized, sans-serif font. The letters are in a light gray color, and the overall design has a modern and slightly futuristic appearance. The letters are evenly spaced and appear to be slightly italicized.
ProBP™ 2000
## The device's block diagram
Image /page/3/Figure/4 description: The image is a block diagram of a blood pressure monitoring system. The system consists of a cuff, pressure sensor, amplifier, A/D converter, MCU, key, valve, motor pump, and display. The cuff is connected to the pressure sensor, motor pump, and valve. The pressure sensor is connected to the amplifier, which is connected to the A/D converter, which is connected to the MCU. The key, valve, motor pump, and display are also connected to the MCU.
Refer to above chart, the operational processes as following.
- * The patient prepared according to the device's guidance. The user (trained operator, such as nurse) places the CUFF wrapped on the proper position of the patient's upper arm.
- * The user presses the Key to power the device for measuring.
- * The MCU sends a signal to activate Motor Pump to inflate the CUFF.
- * During the inflation, the Pressure sensor measures the pulse waveform of the pressure change.
- * The waveform is amplified by the Amplifier for subsequent processing.
- * The analog signal of waveform is converted into a digital signal through the A/D (Analog/Digital) module.
- * The MCU analyzes and calculates the digital signal by a specific algorithm and outputs the value of blood pressure and pulse rate.
- * After the measurement is completed, the Motor Pump is controlled by MCU to stop inflating. And a signal from MCU is sent to the Valve to deflate the CUFF.
- * The device stores the measurement data in the ROM, and directly shows the Systolic/Diastolic blood pressure (Arabic numerals + mmHg) & Pulse rate (Arabic numerals + Pul/min) on the LCD Display, available for user to read and record.
- * The device is powered by either four AA alkaline batteries or an optional AC to DC power supply. These options are identical to the predicate device TMB-1591-A (K161886).
## Principle and Measurement Algorithm of the Device:
The subject device uses the Oscillometric Measuring method to detect blood pressure. The device detects blood pressure measurement on the inflation. Before every measurement, the unit establishes a "zero pressure" equivalent to the air pressure. Then it starts inflating the cuff with the inflation rate 7mmHg/s, meanwhile, the device detects pressure oscillations generated by beat-to-beat pulsatile. Then the measurement finishes, and the device starts deflation occurs in step intervals based on the cuff pressure. The deflation step range is about 6-12 mmHg. Higher cuff pressure corresponds to the larger pressure deflation step.
The oscillometric algorithm in the device measures pulse heights at various cuff pressures throughout the cycle. Those pulse heights are used to build a patient specific pulse envelope as taught by standard oscillometric technique. That pulse envelope is used to calculate systolic and diastolic blood pressure.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "TRANSTEK" in a stylized, sans-serif font. The letters are all capitalized and appear to be a light gray color. The font has a futuristic or technological feel due to its sharp angles and geometric shapes.
The algorithm in the device detect erroneous readings by checking for inconsistencies in the data measured. This can be strange shapes to the pulse profile, large changes in rate of inflation, and wide variances in pulse heights and rates. If one of those conditions is detected, the device switches to the deflation algorithm. The deflation algorithm uses comparisons between multiple pulses to filter out erroneous pulses caused by motion. It also ensures that the pulse profile is of a reasonable shape.
The signal is picked up by the transducer then converted to a digital value via the Analog-to-Digital converter. The digital value is fed into the MCU will calculate the blood pressure and heart rate measurement values according to built-in algorithms.
The values calculated by the algorithm are displayed to the correct section of the screen for the values. Digits are displayed on individual 7 segment areas specifically designed to display digits. The areas where the readings are displayed are labeled as to which reading is being displayed in those segments.
The accuracy of the device's algorithm has been validated through bench testing including testing to the blood pressure standard, IEC 80601-2-30. The device was also clinically tested and met the accuracy requirements of ISO 81060-2.
## Indications for Use:
The Welch Allyn ProBP 2000 Digital blood pressure device is intended for use in measuring blood pressure and heart rate in patients at least 3 years of age or older with arm circumferences between 15 cm to 55 cm (approximately 5.9 to 21.7 inches).
The Welch Allyn ProBP 2000 automatically measures systolic and diastolic pressure and pulse rate. The device is intended to be used by clinicians and medically qualified personnel.
#### Contraindications:
This device is not intended for use on neonates, infants, or children under the age of 3 years. The effectiveness of this device has not been established in pregnant, including pre-eclamptic, patients.
#### Technological Characteristics:
The subject device is modified based on the predicate device. Model 2000 essentially the same device and has the same technical characteristics as the marketed devices cleared in K161886. They use the same oscillometric blood pressure algorithm. The form factor and device material are the same.
The differences between the subject device and predicate device are as following.
- 1) The proposed device is for use by clinically trained professionals (Rx), rather than by patients for self-monitoring (OTC) as stated in the cleared device indications;
- 2) The subject device is intended for patients at least 3 years of age or older with arm circumferences between 15 cm to 55 cm (approximately 5.9 to 21.7 inches), while the predicate device is intended for use in adult patient populations with arm circumferences between 15 cm to 54 cm (approximately 6 to 21 inches);
- 3) The subject device BP hose is not detachable from the device. The cuff connection is Welch Allyn FlexiPort design which is inherently compliant to IEC 80369-5 because the inner-fluid pathway of the FlexiPort connector is greater than 8.5 mm. Model 2000 is exempt from the requirements of IEC 80369-1:2010 – small-bore connectors for liquids and gasses in healthcare applications because there are no connectors accessible to the user in normal use with a diameter less than 8.5mm.
- 4) The subject device modified the range of arm circumferences from 15~55cm;
- 5) The proposed device uses Welch Allyn FlexiPort Blood Pressure Cuffs (K070060);
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "TRANSTEK" in a sans-serif font. The letters are in a light gray color. The letters are bolded and have a slight 3D effect.
- 6) Differs from the predicate device, ProBP 2000 device does not have wirelessly transmission functionality.
- 7) The subject device is not contraindicated for "...any person who is connected to a wearable or implantable electronic device or instrument, such as a pacemaker or defibrillator." We have no evidence to support this patient demographic should be contraindicated. We performed a literature review and found no evidence that any person who is connected to a wearable or implantable electronic device or instrument, such as a pacemaker or defibrillator should be contraindicated. We did however add a note in the directions for use stating the device has not been evaluated for this demographic. "Note This device has not been evaluated for any person who is connected to a wearable or implantable electronic device or instrument, such as a pacemaker or defibrillator."
#### Performance Tests
We conducted design verification and design validation activities to demonstrate that these proposed devices met all design specifications as are Substantially Equivalent to the predicate device.
The results show that the subject devices conform to the following standards:
### Biocompatibility Testing
The Welch Allyn FlexiPort Blood Pressure Cuffs used with these devices were tested for biocompatibility in compliance to ISO 10993-1 and cleared in 510(k) (K070060).
#### Software Verification and Validation Testing
The software of proposed devices was verified and validated in compliance to IEC 62304 Edition 1.1 2015-06. Medical Device Software - Software life cycle processes.
#### Clinical Validation
This device was tested to ISO 81060-2 Second edition 2013-05-01 Non-invasive sphygmomanometers - Part 2: Clinical validation of automated measurement type.
The study population consisted of 88 qualified healthy subjects ranging in age of 3~12 years (35 pediatric) & 12~75 years (53 adults / adolescents), and gender of 39 males & 49 females, of any racial / ethnic background. All data's mean error and standard deviation of differences for systolic, diastolic pressure is not over the limits of ISO 81060-2:2013, and the accuracy of subject device is better than predicate device. No adverse effect and/or complication is found in this study.
#### Non-Clinical Tests
These devices were tested / assessed to evaluate its safety and effectiveness based on below standards:
| Standard | Version | Title |
|----------------|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1 | 3ED 2005,<br>AMD 1 2012 | Medical electrical equipment - Part 1: general requirements for<br>basic safety and essential performance |
| IEC 60601-1-2 | 4 ED 2014 | Medical electrical equipment - Part 1-2: general requirements for<br>basic safety and essential performance - collateral standard:<br>electromagnetic compatibility - requirements and tests |
| IEC 80601-2-30 | 1.1 ED 2013 | Medical electrical equipment - Part 2-30: particular requirements<br>for the basic safety and essential performance of automated non-<br>invasive sphygmomanometers |
| EN/ISO 14971 | 2007 | Medical devices - Application of risk management to medical<br>devices |
{6}------------------------------------------------
| IEC 60601-1-6 | 2013 | Medical electrical equipment - Part 1-6: General requirements for<br>basic safety and essential performance - Collateral standard:<br>Usability |
|---------------|------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 62366 | 2014 | Medical devices - Application of usability engineering to medical<br>devices |
## Device Comparison:
The subject device is very similar to the predicate device (K161886). They have the identical technology principle, blood pressure algorithm, energy source; and have the same appearance and mechanical architecture. Also, they have very similar intended use, performance specifications, software and functions.
The subject device's clinical use is modified to professional use that different from the home use of the predicate device. The indications for use and contraindications for the subject devices have been modified to be suitable for a trained clinician, and to match the target patient populations, at least 3 years of age or older. This subject device is set-up to use Welch Allyn FlexiPort BP Cuffs (K070060) for arm circumferences from 15~54 to 15~55 cm. Correspondingly, the hose connection fitting (FlexiPort) was designed to be applied for these Cuffs.
The subject device ProBP 2000 does not provide wireless transmission functionality.
For above mentioned modifications / differences, we implemented as much as required performance tests, validations, evaluation, to ensure subject device is safe and effective. These bench tests' results proved that the subject device met all requirements of statutory standards. The clinical validation shows the subject device complied with the specific standard, and the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate device.
In addition, the predicate device's contraindication of "This device is contraindicated for any person who is connected to a wearable or implantable electronic device or instrument" was added when the parent predicate device embedded wireless module. At that time we did not find any evidence of risk to those persons who are connected to wearable or implantable electronic device or instrument. We added this contraindication only because of precautionary motive. From our experiences, industry practices, and those market wireless devices' 510(k) records, still no risk be found, and this (similar) contraindication is not required. Thus, we remove it in this submission and added a note in the DFU, Directions For Use, indicating that "This device has not been evaluated for any person who is connected to a wearable or implantable electronic device or instrument, such as a pacemaker or defibrillator."
## Conclusion:
Based on the information presented in this 510(k) premarket notification submission, the subject device is considered substantially equivalent. This subject device is as safe and effective as the currently marketed predicate device TMB-1591-A (K161886). The differences between the indications for use and contraindications, hardware, and mechanical aspects of the proposed device and the predicate device were successfully tested with the relevant standards and FDA guidance, and do not affect equivalent safety and effectiveness or raise new questions of safety or effectiveness.
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K181832
Device Name
Welch Allyn 901123 Digital Blood Pressure Device ProBP 2000
Indications for Use (Describe)
The Welch Allyn ProBP 2000 Digital blood pressure device is intended for use in measuring blood pressure and heart rate in patients at least 3 years of age or older with arm circumferences between 15 cm to 55 cm (approximately 5.9 to 21.7 inches).
The Welch Allyn ProBP 2000 automatically measures systolic and diastolic pressure and pulse rate. The device is intended to be used by clinicians and medically qualified personnel.
Contraindications:
This device is not intended for use on neonates, infants, or children under the age of 3 years. The effectiveness of this device has not been established in pregnant, including pre-eclamptic, patients.
| Type of Use (Select <i>one or both</i> , as applicable) |
|---------------------------------------------------------|
|---------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
EF
PSC Publishing Services (301) 443-6740
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.