K181825 · Pie Medical Imaging BV · LLZ · Dec 4, 2018 · Radiology
Device Facts
Record ID
K181825
Device Name
CAAS MR Solutions
Applicant
Pie Medical Imaging BV
Product Code
LLZ · Radiology
Decision Date
Dec 4, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
CAAS MR Solutions is a modular software product intended to be used by or under supervision of a cardiologist or radiologist in order to aid in reading and interpreting cardiovascular MR images and to visualize and evaluate blood flow in cardiovascular structures to support clinical decision making.
Device Story
CAAS MR Solutions is a standalone software package for PC/Windows; imports DICOM MR images from local/network storage or PACS. It performs semi-automatic, automatic, or manual segmentation of cardiovascular structures; quantifies functional/regional heart ventricle parameters; analyzes myocardial signal intensity; quantifies MR parametric maps (T1, T2, T2*); and visualizes/quantifies blood flow velocity and direction. Used by cardiologists or radiologists in clinical settings to aid image interpretation and clinical decision-making. Results are displayed, printed, or exported to PACS/network. Benefits include standardized quantification of cardiovascular MR data to support diagnosis.
Clinical Evidence
Bench testing only. Verification and validation confirmed system requirements and risk controls. Regression testing demonstrated numerical equivalence for workflows inherited from predicate devices. Validation of new functionalities (myocardial signal intensity, T1 mapping) confirmed accuracy and reproducibility. Usability testing performed per IEC 62366.
Technological Characteristics
Standalone software for Windows PC. DICOM-compliant (NEMA PS 3.1-3.20). Implements semi-automatic, automatic, and manual segmentation algorithms. Supports cardiovascular MR image processing, flow quantification, and parametric mapping. Developed under ISO 13485, ISO 14971, and IEC 62304 standards.
Indications for Use
Indicated for segmentation of cardiovascular structures, analysis of blood flow in heart and large vessels using multi-slice, multi-phase, and velocity-encoded MR images, signal intensity analysis of the myocardium, and quantification of MR parametric maps (T1, T2, T2* relaxation). Not for use as the sole, irrefutable basis for clinical decision making.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Pie Medical Imaging B.V. Annemiek Bouts Regulatory Affairs Coordinator Philipsweg 1 Maastricht, Limburg 6227 AJ THE NETHERLANDS
December 3, 2018
Re: K181825
Trade/Device Name: CAAS MR Solutions Regulation Number: 21 CFR 892.2050 Regulation Name: Picture Archiving And Communications System Regulatory Class: Class II Product Code: LLZ Dated: October 4, 2018 Received: October 9, 2018
Dear Annemiek Bouts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181825
Device Name CAAS MR Solutions
# Indications for Use (Describe)
CAAS MR Solutions features segmentation of cardiovascular structures and enables the analysis of blood flow in the heart and large vessels based on multi-phase and velocity encoded MR images as well as measurement and reporting tools by providing the following functionality:
- · Segmentation of cardiovascular structures and calculation of quantitative results
- · Support signal intensity analysis for the myocardium
- · Quantification of MR parametric maps (such as T1, T2, T2* relaxation)
- · Visualization and quantification of blood flow velocity and directions
When the results provided by CAAS MR Solutions are used in a clinical setting to support diagnoses, the results are explicitly not to be regarded as the sole, irrefutable basis for clinical decision making.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K181825
## 510(k) Summary
### CAAS MR Solutions
Pie Medical Imaging BV
| Submitter/Owner Name | Pie Medical Imaging BV | | |
|----------------------|---------------------------------------------------------------------------|-------------------------|--|
| Address | Philipsweg 1, 6227 AJ Maastricht, The Netherlands | | |
| Phone Number | +31 43 32 81 328 | | |
| Fax Number | +31 43 32 81 329 | | |
| Contact Person | Annemiek Bouts, Regulatory Affairs Coordinator | | |
| Email Address | reg@pie.nl | | |
| Preparation Date | 27-Sep-18 | | |
| Trade Name | CAAS MR Solutions | | |
| Common Name | CAAS MR Solutions | | |
| Classification | Regulation Name: | Image Processing System | |
| | Regulation Class: | Class II | |
| | Regulation number: | 21 CFR 892.2050 | |
| | Classification Product Code: | LLZ | |
| Predicate Device | - CAAS MRV (K162112, Image Processing System, 21 CFR 892.2050, LLZ) | | |
| | - CAAS MR Flow (K090155, Image Processing System, 21 CFR 892.2050, LLZ) | | |
| | - CAAS MR 4D Flow (K162376, Image Processing System, 21 CFR 892.2050 LLZ) | | |
- MR-CT VVA (K140587, Image Processing System, 21 CFR 892.2050, LLZ)
#### Device Description
CAAS MR Solutions is an image post-processing software package which offers functionality to import DICOM MR images, to segment cardiovascular structures in these images, to visualize and evaluate blood flow in cardiovascular structures, to analyze and quantify these cardiovascular structures and to present the analysis results in different formats.
CAAS MR Solutions is designed as a stand-alone software package to run on a PC with a Windows operating system. It can read DICOM MR images from an accessible file system, hard disk (local directory), or (indirectly) received from the MR or PACS system. CAAS MR Solutions provides the functionality to import the DICOM MR images and to organize the found DICOM MR images into patients, studies and series.
CAAS MR Solutions is composed out of analysis workflows of its previously cleared devices: CAAS MRV, CAAS MR Flow and CAAS MR 4D Flow. CAAS MR Solutions comprises their respective functionalities for viewing and quantification of cardiovascular MR images. In addition, CAAS MR Solutions offers new functionality for signal intensity analysis of the myocardium and for MR parametric maps for T1 relaxation. The CAAS MR Solutions product has a moderate level of concern.
Results can be displayed on the screen, printed or saved in a variety of formats to hard disk, network, PACS system or CD. Results and clinical images with overlay can also be printed as a hardcopy and exported in various electronic formats.
#### Intended Use
CAAS MR Solutions is a modular software product intended to be used by or under supervision of a cardiologist or radiologist in order to aid in reading and interpreting cardiovascular MR images and to visualize and evaluate blood flow in cardiovascular structures to support clinical decision making.
#### Indications for Use
CAAS MR Solutions features segmentation of cardiovascular structures, enables the analysis of blood flow in the heart and large vessels based on multi-slice, multi-phase and velocity encoded MR images as well as measurement and reporting tools by providing the following functionality:
- Segmentation of cardiovascular structures and calculation of quantitative results
- Support signal intensity analysis for the myocardium ●
- Quantification of MR parametric maps (such as T1, T2, T2* relaxation)
- Visualization and quantification of blood flow velocity and directions
{4}------------------------------------------------
When the results provided by CAAS MR Solutions are used in a clinical setting to support diagnoses, the results are explicitly not to be regarded as the sole, irrefutable basis for clinical decision making.
#### Substantial equivalence
Four devices have been selected as predicates devices for CAAS MRV, CAAS MRV, CAAS MR Flow, CAAS MR 4D Flow and MR-CT VVA. All selected devices have technological features and characteristics comparable to CAAS MR Solutions and are intended to be used or under supervision of a cardiologist to support clinical decision making of cardiovascular conditions.
CAAS MR Solutions provides functionalities for segmentation of cardiovascular structures (i.e. the heart) semiautomatically, automatically or manually and for quantification of the functional and regional parameters of the heart ventricles. Further it offers functionality to quantify MR parametric maps for T2 and T2* relaxation values. All these functionalities are also supported by the predicate device CAAS MRV.
CAAS MR Solutions features functionalities for segmentation of the blood vessels and analysis and quantification of the blood flow in vessels and through heart valves in phase-contrast MR images semi-automatically or manually. These functionalities are supported by the predicate devices CAAS MR Flow for 2D MR images (single plane, velocity-encoded) and CAAS MR 4D Flow for 4D MR images (multi-slice, multi-phase, velocity-encoded).
CAAS MR Solutions provides functionality for signal intensity analysis of the myocardium which is supported by MR-CT VVA. Further it offers functionality to quantify MR parametric maps for T1 relaxation values which is also supported by MR-CT VVA.
The basic features and technology of the new CAAS MR Solutions product are the same in terms of intended use and indications for use and have the same technological characteristics as the predicate devices CAAS MRV (K162112), CAAS MR Flow (K090155), CAAS MR 4D Flow (K162376) and MR-CT VVA (K140587). All software applications use the same types of data and operating principles for the user and technology regarding data import, contour definition, image display and storage of results.
## Performance Data
Verification and validation of the CAAS MR Solutions showed that the system requirements – derived from the intended use and indications for use – as well as risk control measures were implemented correctly and that the device meets its specifications including conformance to the following standards:
- · ISO 14971:2007, Medical devices Application of risk management to medical devices
- · NEMA PS 3.1 3.20 (2016), Digital Imaging and Communication in Medicine (DICOM)
- · IEC 62304 Ed 1.1, 2015-06, Medical device software Software life cycle processes
- IEC 62336-1 Ed 1.0 2015-02, Medical devices Application of usability engineering to medical devices
- ISO 13485:2016, Medical devices Quality Management Systems
For each analysis workflow in CAAS MR Solution approach is created and the proper functioning of the algorithms is validated.
- . Equivalence in numerical results for the analysis workflows already available in the predicate devices CAAS MRV, CAAS MR Flow and CAAS MR 4D Flow was demonstrated with regression testing.
- The validation of the new functionalities for signal intensity analysis workflow and the T1 mapping quantification demonstrated that the results meet the accuracy and reproducibility requirements.
- Usability testing is performed in accordance with IEC62366 which demonstrated that the user is able to use CAAS MR Solutions for the purpose it was developed for.
The verification and validation results demonstrate the safety and effectiveness of CAAS MR Solution to its intended use and therefore CAAS MR Solutions can be considered as safe and effective as its predicate devices.
#### Conclusion
Based on the application of risk management and performance testing inherent to PMI's QA system (compliant with recognized standards as stated above) we conclude that CAAS MR Solutions is as safe and effective as its predicate devices in terms of intended use, indications for use, technological characteristics, measurements and operating environment and does not raise any new issues related to safety and effectiveness compared to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.