Lead-free Chemical Indicators for Steam Sterilization
K181788 · Kem Medical Products Corp. · JOJ · Mar 1, 2019 · General Hospital
Device Facts
Record ID
K181788
Device Name
Lead-free Chemical Indicators for Steam Sterilization
Applicant
Kem Medical Products Corp.
Product Code
JOJ · General Hospital
Decision Date
Mar 1, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
The Kem Medical Lead-free Chemical Indicators for Steam Sterilization are designed for use by a health care provider to demonstrate that the unit or load has been exposed to a steam sterilization process, and to distinguish between processed and unprocessed units or loads. Use the Kem Medical Lead-free Chemical Indicators for Steam Sterilization in the validated steam sterilization processes described below: *Gravity: 121 C/250 F - 30 minutes (wrapped/porous) *Gravity: 132 C/270 F - 3 minutes (unwrapped/nonporous) *Gravity: 132 C/270 F - 15 minutes (wrapped/porous) *Gravity: 135 C/275 F - 3 minutes (unwrapped/nonporous) * Gravity: 135 C/275 F - 10 minutes (wrapped/porous or unwrapped/nonporous, mixed load) * Vacuum assisted (prevacuum): 132 C/270 F - 3 minutes (unwrapped/nonporous) * Vacuum assisted (prevacuum): 132 C/270 F - 4 minutes (wrapped/porous) * Vacuum assisted (prevacuum): 134 C/273 F - 4 minutes (wrapped/porous) * Vacuum assisted (prevacuum): 135 C/275 F - 3 minutes (wrapped/nonporous, mixed load)
Device Story
Chemical indicator device for steam sterilization; consists of bismuth sulfide-based indicator ink printed on paper substrates (dots, strips, or cards). Used by healthcare providers to verify exposure to steam sterilization cycles; provides immediate visual confirmation via color change from original state to dark brown/black. Distinguishes processed from unprocessed units/loads. Operates via chemical reaction to steam exposure conditions. Benefits patient by ensuring sterilization process completion for medical instruments. No electronic or software components.
Clinical Evidence
Bench testing only. Independent study by Lexamed confirmed compliance with ISO 11140-1:2014. Testing included BIER steam sterilizer cycles, full gravity and pre-vacuum autoclave cycles, dry heat, color change stability, and transference/migration testing. Results met all pass/fail criteria for endpoint specifications and shelf-life.
Technological Characteristics
Chemical indicator ink (bismuth sulfide-based) printed on paper substrates (with or without adhesive). Type 1 chemical indicator per ISO 11140-1:2014. Passive device; no energy source, connectivity, or software.
Indications for Use
Indicated for use by healthcare providers to monitor steam sterilization processes and distinguish between processed and unprocessed loads in gravity and prevacuum steam sterilizers.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
3M Comply Lead Free Process Indicators for Steam (K101528)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG" are written in blue, with the word "ADMINISTRATION" written in a smaller font size below.
March 1, 2019
Kem Medical Products Corp. Douglas Kruger President/CEO 400 Broadhollow Road, Suite 2 Farmingdale, New York 11735
Re: K181788
Trade/Device Name: Lead-free Chemical Indicators for Steam Sterilization Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: Class II Product Code: JOJ Dated: January 17, 2019 Received: January 18, 2019
Dear Douglas Kruger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Clarence W. Murray lii III -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181788
Device Name
Lead-free Chemical Indicators for Steam Sterilization
#### Indications for Use (Describe)
The Kem Medical Lead-free Chemical Indicators for Steam Sterilization are designed for use by a health care provider to demonstrate that the unit or load has been exposed to a steam sterilization process, and to distinguish between processed and unprocessed units or loads. Use the Kem Medical Lead-free Chemical Indicators for Steam Sterilization in the validated steam sterilization processes described below:
*Gravity: 121 C/250 F - 30 minutes (wrapped/porous)
*Gravity: 132 C/270 F - 3 minutes (unwrapped/nonporous)
*Gravity: 132 C/270 F - 15 minutes (wrapped/porous)
*Gravity: 135 C/275 F - 3 minutes (unwrapped/nonporous)
* Gravity: 135 C/275 F - 10 minutes (wrapped/porous or unwrapped/nonporous, mixed load)
* Vacuum assisted (prevacuum): 132 C/270 F - 3 minutes (unwrapped/nonporous)
* Vacuum assisted (prevacuum): 132 C/270 F - 4 minutes (wrapped/porous)
* Vacuum assisted (prevacuum): 134 C/273 F - 4 minutes (wrapped/porous)
* Vacuum assisted (prevacuum): 135 C/275 F - 3 minutes (wrapped/nonporous, mixed load)
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------|-------------------------------------------------------------|
| <span></span> ☐ Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> ☒ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Kem Medical Products Corp. The logo is blue and features a stylized "K" on the left side. The words "Kem Medical Products Corp." are written in a bold, sans-serif font to the right of the "K" symbol.
www.kemmed.com
## 510(k) Summary For Kem Medical Lead-free Chemical Indicators for Steam Sterilization
### K181788
Date of Submission: February 28, 2019
Submission Number: K181788
Contact/Owner: Douglas Kruger President/CEO Kem Medical Products Corp. 400 Broadhollow Road, Suite 2 Farmingdale, NY 11735 Phone: (631) 454-6565 Fax: (631) 454-8083 Email: dougk@kemmed.com
Trade Name: Lead-free Chemical Indicators for Steam Sterilization
Common Name: Sterilization Process Indicator for Steam
Classification Name: Indicator, Physical/Chemical Sterilization Process (21 CFR800.2800, Product Code JOJ)
Predicate Device: 3M Comply Lead Free Process Indicators for Steam, K101528
**There have not been any prior submissions for this device. **
{4}------------------------------------------------
www.kemmed.com
Device Summary: The Kem Medical Lead-free Chemical Indicators for Steam Sterilization provide immediate confirmation of the sterilization process. The indicator ink is lead-free and can be printed onto suitable paper substrates. As an example, the ink is printed onto adhesive-coated paper for autoclave labels, or "dots", which are removed from a poly or paper backing for use.
Technological Comparison: The predicate and our premarket device consist of chemical indicator ink printed onto paper with and without adhesive. The Kem Medical Lead-free Chemical Indicators for Steam Sterilization have the same technological characteristics of the predicate device (3M Comply Lead Free Process Indicators for Steam cleared under K101528), in design, material, chemical composition, and performance characteristics. A stated purpose of the predicate device submission was to replace the lead ink in the Comply Autoclave Tape.
| Table 1- Comparison of the New Device to the Predicate | | |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ELEMENT | NEW DEVICE<br>K181788 | PREDICATE (3M)<br>K101528 |
| Intended use: | process indicator for steam sterilization | process indicator for steam sterilization |
| Device design Aspects: | Paper dot/strip/card printed with bismuth<br>sulfide based chemical indicator ink | Paper tape/card printed with bismuth<br>sulfide based chemical indicator ink |
| Indicator agent: | Lead-free Steam Indicator Ink | Lead-free Steam Indicator Ink |
| Sterilization method and cycles: | *Gravity: 121 C/250 F - 30 minutes<br>(wrapped/porous)<br>*Gravity: 132 C/270 F - 3 minutes<br>(unwrapped/nonporous)<br>*Gravity: 132 C/270 F - 15 minutes<br>(wrapped/porous)<br>*Gravity: 135 C/275 F - 3 minutes<br>(unwrapped/nonporous)<br>*Gravity: 135 C/275 F - 10 minutes<br>(wrapped/porous or unwrapped/nonporous,<br>mixed load)<br>*Vacuum assisted (prevacuum): 132 C/270<br>F - 3 minutes (unwrapped/nonporous)<br>*Vacuum assisted (prevacuum): 132 C/270<br>F - 4 minutes (wrapped/porous)<br>*Vacuum assisted (prevacuum): 134 C/273<br>F - 4 minutes (wrapped/porous)<br>*Vacuum assisted (prevacuum): 135 C/275<br>F - 3 minutes (wrapped/porous or<br>unwrapped/nonporous, mixed load) | *Gravity: 121 C/250 F - 30 minutes<br>(wrapped)<br>*Gravity: 132 C/270 F - 3 minutes<br>(unwrapped)<br>*Gravity: 132 C/270 F - 15 minutes<br>(wrapped)<br>*Gravity: 135 C/275 F - 3 minutes<br>(unwrapped)<br>*Gravity: 135 C/275 F - 10 minutes<br>(wrapped)<br>*Vacuum assisted (prevacuum): 132 C/270 F - 3 minutes (unwrapped)<br>*Vacuum assisted (prevacuum): 132 C/270 F - 4 minutes (wrapped)<br>*Vacuum assisted (prevacuum): 134 C/273 F - 3.5 minutes (unwrapped)<br>*Vacuum assisted (prevacuum): 134 C/273 F - 4 minutes (wrapped) |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Kem Medical Products Corp. The logo consists of a blue icon on the left and the company name on the right. The icon is a combination of vertical lines and a triangle pointing to the right. The company name is written in a bold, blue font.
Solutions for safe work environments
www.kemmed.com
| | | *Vacuum assisted (prevacuum): 135<br>C/275 F - 3 minutes (wrapped or<br>unwrapped) |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Endpoint specifications: | Dark brown/black color change | Dark brown/black color change |
| Shelf-life: | 1 year from date of manufacture | 18 months from date of manufacture |
| Performance Standards: | An independent study by Lexamed®<br>confirms that Kem Medical Lead-free<br>Chemical Indicators met the requirements<br>defined in ISO 11140-2014-1;2014<br>Sterilization of healthcare products-<br>Chemical Indicators - Part 1 General<br>Requirements, including:<br>Data supporting endpoints with<br>pass/fail criteria Data supporting endpoint color<br>stability Data supporting integrator<br>parameters Data supporting shelf-life | Testing that was conducted met the<br>process indicator requirements of the<br>FDA's Premarket Notification (510K)<br>Submissions for Chemical Indicators:<br>Guidance for Industry and FDA Staff, Dec.<br>19, 2003 and ANSI/AAMI/ISO 11140-<br>1:2005 Sterilization of health care<br>products-Chemical Indicators, Part 1:<br>General Requirements. |
Indications for Use: The Kem Medical Lead-free Chemical Indicators for Steam Sterilization are designed for use by a health care provider to demonstrate that the unit or load has been exposed to a steam sterilization process, and to distinguish between processed and unprocessed units or loads. Use the Kem Medical Lead-free Chemical Indicators for Steam Sterilization in the validated steam sterilization processes described below:
*Gravity: 121 C/250 F - 30 minutes (wrapped/porous)
*Gravity: 132 C/270 F - 3 minutes (unwrapped/nonporous)
*Gravity: 132 C/270 F - 15 minutes (wrapped/porous)
*Gravity: 135 C/275 F - 3 minutes (unwrapped/nonporous)
*Gravity: 135 C/275 F - 10 minutes (wrapped/porous or unwrapped/nonporous, mixed load)
* Vacuum assisted (prevacuum): 132 C/270 F - 3 minutes (unwrapped/nonporous)
* Vacuum assisted (prevacuum): 132 C/270 F - 4 minutes (wrapped/porous)
* Vacuum assisted (prevacuum): 134 C/273 F - 4 minutes (wrapped/porous)
* Vacuum assisted (prevacuum): 135 C/275 F - 3 minutes (wrapped/porous or unwrapped/nonporous, mixed load)
{6}------------------------------------------------
www.kemmed.com
## Summary for Non-Clinical Testing:
The Kem Medical Lead-free Chemical Indicators for Steam Sterilization meet all ISO 11140-1:2014 performance criteria for Type 1 steam sterilization indicators. All BIER Steam Sterilizer cycles resulted in acceptable pass/fail criteria for all temperatures and times specified. Full autoclave cycles for both Gravity and Pre-vacuum were performed showing acceptable results for all temperatures and times specified. Dry heat, Color Change Stability, and Transference/Migration testing also resulted in acceptable results for all performance criteria of 1140-1:2014. The condition of the CI color change after exposure to a steam sterilization process remained unchanged for a period of not less of 6 months from the date of exposure.
The Lexamed® study results demonstrated the Kem Medical Lead-free Type 1 Chemical Indicator can be used as an external or internal chemical indicator for use in monitoring exposure to conditions of steam sterilization for the indicated gravity and pre-vacuum sterilizer cycles and comply with all applicable requirements defined in ISO 11140-1:2014, Sterilization of healthcare products- Chemical Indicators-Part 1 General Requirements.
Conclusion: The conclusions drawn from the nonclinical tests demonstrate that Kem Medical Lead-free Chemical Indicators for Steam Sterilization are as safe, as effective, and performs as well as or better than the legally marketed devices cleared under K101528.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.