← Product Code [IYN](/productcode/IYN) · K181783

# LOGIQ P9; LOGIQ P7 (K181783)

_GE Healthcare · IYN · Sep 20, 2018 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181783

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Sep 20, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transrectal; Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular). When Pinpoint™ GT Technology, from C.R. Bard, Inc., is included with the system, the Indications for Use include: Pinpoint GT Technology is intended to provide clinicians with visual tools for passive magnetic tracking of a needle with respect to ultrasound image data.

## Device Story

The LOGIQ P9/P7 is a mobile, console-based diagnostic ultrasound system for clinical use by physicians. It acquires, processes, and displays ultrasound data via various transducers. The system supports B-mode, M-mode, PW/CW Doppler, Color/Power Doppler, and harmonic imaging. New features include shear wave elastography (migrated from LOGIQ S8), HDlive, and connectivity to 'LOGIQ P Apps' for smart device integration (image capture, remote UI control). The system provides visual tools for needle tracking when using Pinpoint GT Technology. Output is displayed on a 21.5-inch LCD monitor for clinical review and decision-making. It aids in diagnostic imaging and fluid flow analysis across multiple anatomical regions, benefiting patients through non-invasive diagnostic assessment.

## Clinical Evidence

No clinical studies were required to support substantial equivalence. Evidence is based on non-clinical performance testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing. System-level verification and validation were performed.

## Technological Characteristics

Mobile console (53x74x157 cm) with 21.5-inch LCD and 10.4-inch touch screen. Employs ultrasonic pulsed Doppler/echo imaging. Supports B, M, PW/CW, Color/Power Doppler, and harmonic modes. Connectivity via DICOM and LOGIQ P Apps. Biocompatible patient-contact materials. Complies with AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, ISO10993-1, and NEMA UD 2.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- LOGIQ P9; LOGIQ P7 Diagnostic Ultrasound System ([K163596](/device/K163596.md))

## Reference Devices

- LOGIQ S8 Diagnostic Ultrasound System ([K170445](/device/K170445.md))
- Voluson P6 And Voluson P8 ([K160162](/device/K160162.md))
- LOGIQ E10 ([K173555](/device/K173555.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

September 20, 2018

GE Healthcare Bryan Behn RA Director 9900 Innovation Drive WAUWATOSA, WI 53226

Re: K181783

Trade/Device Name: LOGIO P9; LOGIO P7 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: July 2, 2018 Received: July 3, 2018

Dear Bryan Behn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Hse 2. Nils

for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181783

Device Name LOGIQ P9; LOGIQ P7

#### Indications for Use (Describe)

The device is intended for use by a qualified physician for ultrasound evaluation of Fetal; Abdominal; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transvaginal; Transesophageal and Intraoperative (abdominal, thoracic, vascular).

When Pinpoint GT Technology, from C.R. Bard, Inc., is included with the system, the Indications for Use include: Pinpoint GT Technology is intended to provide clinicians with visual tools for passive magnetic tracking of a needle with respect to ultrasound image data.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. The circle is further embellished with decorative, swirling elements around its perimeter, giving it a classic and recognizable appearance.

# Indications for Use Forms

The following forms represent indications with clinical applications and exam types along with the modes of operation for the LOGIQ P9 and LOGIQ P7 system. Combinations identified "P" for the transducers represents those previously cleared with LOGIQ P9 and LOGIQ P7 and *P for those cleared on another GE Ultrasound system and those identified and "N" are new. Please see section 11 Table 11.2.1 for information on previous clearance information on these transducers.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by eight curved, white shapes that resemble stylized water droplets or clouds. The overall design is simple, clean, and recognizable as the GE brand.

### Diagnostic Ultrasound Indications for Use Form

#### LOGIQ P9; LOGIQ P7 Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Ophthalmic                                         |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]                              | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 5,6              |
| Abdominal[1]                                       | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6,10         |
| Pediatric                                          | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6            |
| Small Organ[2]                                     | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6,10         |
| Neonatal Cephalic                                  | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 5,6              |
| Adult Cephalic                                     | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 5,6              |
| Cardiac Adult & Pediatric                          | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              |                  |
| Peripheral Vascular                                | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6            |
| Musculo-skeletal Conventional                      | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6,10         |
| Musculo-skeletal Superficial                       | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6,9,10       |
| Other[4]                                           | P | P | P             | P             | P                | P                  | P                | P                  | P                   |                | 3,5,6            |
| Exam Type, Means of Access                         |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transesophageal                                    | N | N | N             | N             | N                | N                  | N                | N                  | N                   | N              |                  |
| Transrectal[8]                                     | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6            |
| Transvaginal                                       | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6            |
| Transuretheral                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative[8]                                  | P | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6            |
| Intraoperative Neurological                        |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intravascular                                      |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Laparoscopic                                       |   |   |               |               |                  |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Needle guidance with PinpointTM GT needle technology

[10] Shear Wave Elastography

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The letters and the frame are both blue, and the background is white. The logo is simple and recognizable, representing the well-known multinational corporation.

### Diagnostic Ultrasound Indications for Use Form

#### LOGIQ P9; LOGIQ P7with L12n-RS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Ophthalmic                                         | P | P |               |               | P                |                    | P                | P                  | P                   |                |                  |
| Fetal / Obstetrics[7]                              |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Abdominal[1]                                       | P | P |               |               | P                |                    | P                | P                  | P                   |                |                  |
| Pediatric                                          | P | P |               |               | P                |                    | P                | P                  | P                   |                |                  |
| Small Organ[2]                                     | P | P |               |               | P                |                    | P                | P                  | P                   |                |                  |
| Neonatal Cephalic                                  | P | P |               |               | P                |                    | P                | P                  | P                   |                |                  |
| Adult Cephalic                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Cardiac Adult & Pediatric                          |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Peripheral Vascular                                | P | P |               |               | P                |                    | P                | P                  | P                   |                |                  |
| Musculo-skeletal Conventional                      | P | P |               |               | P                |                    | P                | P                  | P                   |                |                  |
| Musculo-skeletal Superficial                       | P | P |               |               | P                |                    | P                | P                  | P                   |                | [9]              |
| Other[4]                                           |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Exam Type, Means of Access                         |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transesophageal                                    |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transrectal[8]                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transvaginal                                       |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transuretheral                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative[8]                                  | P | P |               |               | P                |                    | P                | P                  | P                   |                |                  |
| Intraoperative Neurological                        |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intravascular                                      |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Laparoscopic                                       |   |   |               |               |                  |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

- [2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.

- [4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[9] Needle guidance with PinpointTM GT needle technology;

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The letters and the frame are both blue, and the background is white. The logo is simple and recognizable, representing the well-known multinational corporation.

### Diagnostic Ultrasound Indications for Use Form

#### LOGIQ P9; LOGIQ P7 with P6D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Abdominal[1]                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Pediatric                                          |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Small Organ[2]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Adult Cephalic                                     |                   |   | P             | P             |                  |                    |                  |                    |                     |                |                  |
| Cardiac Adult & Pediatric                          |                   |   | P             | P             |                  |                    |                  |                    |                     |                |                  |
| Peripheral Vascular                                |                   |   | P             | P             |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Other[4]                                           |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transrectal[8]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative[8]                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |

N = new indication; *P = previously cleared by FDA

[1] Abdominal includes Renal, GYN/Pelvic. Notes:

- [2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.

- [4] Other use includes Urology/Prostate
[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The logo is colored in blue, with the letters and the frame having the same shade of blue.

### Diagnostic Ultrasound Indications for Use Form

#### LOGIQ P9; LOGIQ P7 with L4-12t-RS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | B | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes) |
|----------------------------------------------------|---|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------|
| Ophthalmic                                         |   |   |            |            |               |                 |               |                 |                  |             |               |
| Fetal / Obstetrics[7]                              |   |   |            |            |               |                 |               |                 |                  |             |               |
| Abdominal[1]                                       |   |   |            |            |               |                 |               |                 |                  |             |               |
| Pediatric                                          | P | P | P          |            | P             | P               | P             | P               | P                | P           | [3,6]         |
| Small Organ[2]                                     | P | P | P          |            | P             | P               | P             | P               | P                | P           | [3,6]         |
| Neonatal Cephalic                                  | P | P | P          |            | P             | P               | P             | P               | P                | P           | [3,6]         |
| Adult Cephalic                                     |   |   |            |            |               |                 |               |                 |                  |             |               |
| Cardiac Adult & Pediatric                          |   |   |            |            |               |                 |               |                 |                  |             |               |
| Peripheral Vascular                                | P | P | P          |            | P             | P               | P             | P               | P                | P           | [3,6]         |
| Musculo-skeletal Conventional                      | P | P | P          |            | P             | P               | P             | P               | P                | P           | [3,6]         |
| Musculo-skeletal Superficial                       | P | P | P          |            | P             | P               | P             | P               | P                | P           | [3,6]         |
| Other[4]                                           |   |   |            |            |               |                 |               |                 |                  |             |               |
| Exam Type, Means of Access                         |   |   |            |            |               |                 |               |                 |                  |             |               |
| Transesophageal                                    |   |   |            |            |               |                 |               |                 |                  |             |               |
| Transrectal[8]                                     |   |   |            |            |               |                 |               |                 |                  |             |               |
| Transvaginal                                       |   |   |            |            |               |                 |               |                 |                  |             |               |
| Transurethral                                      |   |   |            |            |               |                 |               |                 |                  |             |               |
| Intraoperative[8]                                  |   |   |            |            |               |                 |               |                 |                  |             |               |
| Intraoperative Neurological                        |   |   |            |            |               |                 |               |                 |                  |             |               |
| Intravascular                                      |   |   |            |            |               |                 |               |                 |                  |             |               |
| Laparoscopic                                       |   |   |            |            |               |                 |               |                 |                  |             |               |

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The logo is colored in blue, with the letters and the frame having the same shade of blue.

### Diagnostic Ultrasound Indications for Use Form

#### LOGIQ P9; LOGIQ P7 with L10-22-RS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |            |            |               |                 |               |                 |                  |             |               |
|----------------------------------------------------|-------------------|---|------------|------------|---------------|-----------------|---------------|-----------------|------------------|-------------|---------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW Doppler | CW Doppler | Color Doppler | Color M Doppler | Power Doppler | Combined Modes* | Harmonic Imaging | Coded Pulse | Other [Notes] |
| Ophthalmic                                         |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Fetal / Obstetrics[7]                              |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Abdominal[1]                                       |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Pediatric                                          |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Small Organ[2]                                     | P                 | P | P          |            | P             | P               | P             | P               | P                | P           |               |
| Neonatal Cephalic                                  | P                 | P | P          |            | P             | P               | P             | P               | P                | P           |               |
| Adult Cephalic                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Cardiac Adult & Pediatric                          |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Peripheral Vascular                                |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Musculo-skeletal Conventional                      | P                 | P | P          |            | P             | P               | P             | P               | P                | P           |               |
| Musculo-skeletal Superficial                       | P                 | P | P          |            | P             | P               | P             | P               | P                | P           |               |
| Other[4]                                           |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Exam Type, Means of Access                         |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transesophageal                                    |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transrectal[8]                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transvaginal                                       |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Transuretheral                                     |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Intraoperative[8]                                  |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Intraoperative Neurological                        |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Intravascular                                      |                   |   |            |            |               |                 |               |                 |                  |             |               |
| Laparoscopic                                       |                   |   |            |            |               |                 |               |                 |                  |             |               |

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The letters and the frame are both blue, and the background is white. The logo is simple and recognizable, representing the well-known multinational corporation.

### Diagnostic Ultrasound Indications for Use Form

#### LOGIQ P9; LOGIQ P7 with L3-9i-RS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Ophthalmic                                         |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]                              |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Abdominal[1]                                       |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Pediatric                                          |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Small Organ[2]                                     | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              |                  |
| Neonatal Cephalic                                  |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Adult Cephalic                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Cardiac Adult & Pediatric                          |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Peripheral Vascular                                | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              |                  |
| Musculo-skeletal Conventional                      | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              |                  |
| Musculo-skeletal Superficial                       | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              |                  |
| Other[4]                                           |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Exam Type, Means of Access                         |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transesophageal                                    |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transrectal[8]                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transvaginal                                       |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transuretheral                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative[8]                                  | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              |                  |
| Intraoperative Neurological                        |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intravascular                                      |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Laparoscopic                                       |   |   |               |               |                  |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

- [2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The logo is colored in blue, with the letters and the frame having the same shade of blue.

### Diagnostic Ultrasound Indications for Use Form

#### LOGIQ P9; LOGIQ P7 with E8CS-RS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Ophthalmic                                         |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]                              | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,6]            |
| Abdominal[1]                                       | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,6]            |
| Pediatric                                          |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Small Organ[2]                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Neonatal Cephalic                                  |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Adult Cephalic                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Cardiac Adult & Pediatric                          |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Peripheral Vascular                                |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Conventional                      |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Superficial                       |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Other[4]                                           | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,6]            |
| Exam Type, Means of Access                         |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transesophageal                                    |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transrectal[8]                                     | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,6]            |
| Transvaginal                                       | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,6]            |
| Transurethral                                      |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative[8]                                  |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative Neurological                        |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intravascular                                      |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Laparoscopic                                       |   |   |               |               |                  |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

- [2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The letters and the frame are both blue, and the background is white. The logo is simple and recognizable, representing the well-known multinational corporation.

### Diagnostic Ultrasound Indications for Use Form

#### LOGIQ P9; LOGIQ P7 with BE9CS-RS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Abdominal[1]                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Pediatric                                          |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Small Organ[2]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Cardiac Adult & Pediatric                          |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Other[4]                                           | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,6]            |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transrectal[8]                                     | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [3,6]            |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative[8]                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

- [2] Small organ includes breast, testes and thyroid
[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The letters and the frame are both blue, and the background is white. The logo is simple and recognizable, representing the well-known multinational corporation.

### Diagnostic Ultrasound Indications for Use Form

#### LOGIQ P9; LOGIQ P7 with RIC5-9A-RS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Anatomy/Region of Interest | B | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
|----------------------------------------------------|---|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Ophthalmic                                         |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]                              | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [5,6]            |
| Abdominal[1]                                       | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [5,6]            |
| Pediatric                                          |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Small Organ[2]                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Neonatal Cephalic                                  |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Adult Cephalic                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Cardiac Adult & Pediatric                          |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Peripheral Vascular                                |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Conventional                      |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Superficial                       |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Other[4]                                           | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [5,6]            |
| Exam Type, Means of Access                         |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transesophageal                                    |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transrectal[8]                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transvaginal                                       | P | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [5,6]            |
| Transuretheral                                     |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative[8]                                  |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative Neurological                        |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intravascular                                      |   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Laparoscopic                                       |   |   |               |               |                  |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a circular frame. The logo is colored in blue, with the letters and the frame having the same shade of blue.

### Diagnostic Ultrasound Indications for Use Form

#### LOGIQ P9; LOGIQ P7 with 12S-RS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal / Obstetrics[7]                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Abdominal[1]                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              |                  |  |
| Small Organ[2]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Neonatal Cephalic                                  | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              |                  |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac Adult & Pediatric                          |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Peripheral Vascular                                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Other[4]                                           |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transrectal[8]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative[8]                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |

N = new indication; P = previously cleared by FDA

Notes: [1] Abdominal includes Renal, GYN/Pelvic.

[2] Small organ includes breast, testes and thyroid

[3] Elastography Imaging - Elasticity.

[4] Other use includes Urology/Prostate

[5] 3D/4D Imaging mode

[6] Needle guidance imaging

[7] Includes infertility monitoring of follicle development

[8] Intraoperative includes abdominal, thoracic (cardiac), and vascular (PV)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" in a stylized, cursive font, enclosed within a blue circle. The circle is surrounded by eight curved, white shapes that resemble stylized water droplets or clouds. The overall design is simple, clean, and recognizable as the GE brand.

### Diagnostic Ultrasound Indications for Use Form

#### LOGIQ P9; LOGIQ P7 with C1-5-RS Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal / Obstetrics[7]                              | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 5,6              |  |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6,<br>10     |  |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6            |  |
| Small Organ[2]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac Adult & Pediatric                          |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Peripheral Vascular                                | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6            |  |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Other[4]                                           | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | 3,5,6            |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transrectal[8]                                     |                   |   |               |               |                  |                    |…

---

**Source:** [https://fda-staging.innolitics.com/device/K181783](https://fda-staging.innolitics.com/device/K181783)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com