# SuperCable® Iso-Elastic™ Cerclage System (K181749)

_Kinamed, Incorporated · JDQ · Aug 1, 2018 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181749

## Device Facts

- **Applicant:** Kinamed, Incorporated
- **Product Code:** [JDQ](/productcode/JDQ.md)
- **Decision Date:** Aug 1, 2018
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3010
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Kinamed SuperCable Iso-Elastic Cerclage System is intended to be used in the following: repair of long bone fractures due to trauma or reconstruction; reattachment of the greater in total hip arthroplasty, surface replacement arthroplasty, or other procedures involving trochanteric osteotomy; sternotomy closure; and sublaminar and intrafacet wiring of the spinal column.

## Device Story

System consists of implantable cable and locking clasp (clip and wedge). Cable features monofilament nylon core with braided UHMWPE jacket. Locking clasp secures cable. Used by surgeons in OR for orthopedic and spinal procedures. Manual instrumentation applies cable to bone or tissue structures to provide fixation. Device provides mechanical stabilization of fractures or reattached bone segments. Benefits include secure fixation during healing. Subject device introduces metal injection molding (MIM) for clasp manufacturing; otherwise identical to predicate.

## Clinical Evidence

Bench testing only. Testing included ISO 10993 biocompatibility, bacterial endotoxin (LAL) levels, static construct tensile testing, and fatigue construct tensile testing.

## Technological Characteristics

Implantable cerclage system. Cable: monofilament nylon core with braided UHMWPE jacket. Locking clasp: Ti-6Al-4V alloy (ASTM F136/ASTM F2885). Manufacturing: Metal Injection Molding (MIM) for clasp. Manual application. Non-electronic, non-software.

## Regulatory Identification

A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.

## Predicate Devices

- SuperCable® Iso-Elastic™ Cerclage System ([K030256](/device/K030256.md))

## Reference Devices

- SuperCable® Iso-Elastic™ Cerclage System ([K102834](/device/K102834.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is blue and white.

August 1. 2018

Kinamed, Incorporated Heather Neely Senior Director of Quality Assurance and Regulatory Compliance 820 Flynn Road Camarillo. California 93012

Re: K181749

Trade/Device Name: SuperCable® Iso-Elastic™ Cerclage System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ Dated: June 25, 2018 Received: July 2, 2018

Dear Heather Neely:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Image /page/1/Picture/6 description: The image shows a digital signature. The signature is by Vesa Vuniqi -S. The date of the signature is 2018.08.01. The time of the signature is 17:08:45 -04'00'.

For: Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181749

Device Name

SuperCable® Iso-Elastic™ Cerclage System

#### Indications for Use (Describe)

The Kinamed SuperCable Iso-Elastic Cerclage System is intended to be used in the following: repair of long bone fractures due to trauma or reconstruction; reattachment of the greater in total hip arthroplasty, surface replacement arthroplasty, or other procedures involving trochanteric osteotomy; sternotomy closure; and sublaminar and intrafacet wiring of the spinal column.

| Type of Use (Select one or both, as applicable)                                                 |  |
|-------------------------------------------------------------------------------------------------|--|
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |  |
| <span style="text-decoration: underline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  |  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Kinamed Incorporated. The logo features a green square with a stylized white "K" inside, with a gradient effect on the lower part of the "K". To the right of the square, the word "KINAMED" is written in bold, black letters, with the word "INCORPORATED" written in smaller, lighter letters below it.

K181749

Page 1 of 3

## Special 510(k) Summary

| Manufacturer:   | KINAMED® Incorporated<br>820 Flynn Road<br>Camarillo, CA 93012-8701<br>USA |
|-----------------|----------------------------------------------------------------------------|
|                 | Phone (805) 384-2748<br>FAX (805) 384-2792                                 |
| Contact Person: | Heather Neely, RAC                                                         |

Senior Director of Quality Assurance and Regulatory Compliance Kinamed®, Inc. Phone (805) 886-1069 and (805) 384-2748 x210 FAX (805) 384-2792 Email: hneely@kinamed.com

Date Prepared: June 25, 2018

## DEVICE INFORMATION

Trade/Proprietary Name: SuperCable® Iso-Elastic™ Cerclage System

| Common Name:         | Bone fixation cerclage |
|----------------------|------------------------|
| Classification Name: | Cerclage, Fixation     |

21 CFR 888.3010 Class II Device Product Code: JDQ

K030256 (SuperCable® Iso-Elastic™ Cerclage Predicate Device: System)

### Product Description:

The Kinamed SuperCable® Iso-Elastic™ Cerclage System consists of an implantable cable and locking clasp (the locking clasp consists of a clip and a wedge). The cable is comprised of a monofilament core of nylon with a jacket of ultra-high molecular-weight-polyethylene (UHMWPE) fibers braided around the nylon core. Manual instrumentation is used for applying the cable.

{4}------------------------------------------------

K181749

Page 2 of 3

The SuperCable Iso-Elastic Cerclage System that is the subject of this submission is a line extension to the predicate SuperCable Iso-Elastic Cerclage System (K030256) that was cleared on October 21, 2003.

Additional sizes of the SuperCable implant were previously added to the System via a Special 510(k) line extension (K102834) that was cleared on January 12, 2011.

The subject of the present Special 510(k) submission is a line addition of a cable implant whose locking clasp is manufactured using a metal injection molding (MIM) process. The cable portion of the implant is the exact same as in the predicate system. This line addition is intended to improve manufacturing efficiency and is not motivated by any clinical or performance factors associated with the SuperCable Iso-Elastic Cerclage System. An overview of the changes being proposed is described in Figure 5-1.

| Figure 5-1                                                 |  |
|------------------------------------------------------------|--|
| Description of Proposed Changes for the present submission |  |

| 510(k) Designation           | Locking Clasp<br>Material of Construction  | Locking Clasp<br>Manufacturing Process |
|------------------------------|--------------------------------------------|----------------------------------------|
| K030256<br>(predicate)       | ASTM F1361<br>Titanium<br>(implant-grade)  | Machined from Wrought                  |
| K181749<br>(this submission) | ASTM F28852<br>Titanium<br>(implant-grade) | Metal Injection Molded<br>(MIM)        |

There are no changes to the polymer cable itself or to the product's intended use, indications for use, functionality, shelf life, biocompatibility, packaging, and sterilization method.

The SuperCable Iso-Elastic System locking clasp that is the subject of this submission meets the specifications of ASTM F2885 (Type 1, Densified).

### Indications for Use (for this submission):

The Kinamed SuperCable Iso-Elastic Cerclage System is intended to be used in the following: repair of long bone fractures due to trauma or reconstruction; reattachment of the greater trochanter in total hip arthroplasty, surface

<sup>1</sup> ASTM F136. Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401)

<sup>2</sup> ASTM F2885, Standard Specification for Metal Injection Molded Titanium-6Aluminum-4Vanadium Components for Surgical Implant Applications

{5}------------------------------------------------

Page 3 of 3

replacement arthroplasty, or other procedures involving trochanteric osteotomy; sternotomy closure; and sublaminar and intrafacet wiring of the spinal column.

With the exception of the addition of Kinamed's "SuperCable" brand name, there are no changes to the indications for use or intended use from the previously cleared SuperCable Iso-Elastic Cerclage System predicate (K030256). Form FDA 3881, Indications for Use, can be found in Section 4 of this Special 510(k) submission. The addition of the "SuperCable" brand name to the Indications for Use statement is not considered a change because the product line has been branded "SuperCable" since it was launched in the USA in 2004.

### Performance Testing

Performance testing of the SuperCable Iso-Elastic Cerclage System includes ISO 10993 testing, Bacterial endotoxin levels (LAL), Static construct tensile testing, and Fatigue construct tensile testing.

## Basis of Substantial Equivalence

The SuperCable Iso-Elastic Cerclage System that is the subject of this submission shares the following similarities with the predicate SuperCable Iso-Elastic Cerclage System cleared under K030256:

- same polymer cable
- same intended use .
- same indications for use ●
- same raw material alloy (Ti-6Al-4V) for locking clasp ●
- . same design
- similar size and dimensions
- . same type of mating components
- same shelf life .
- same biocompatibility .
- same sterilization and packaging methods .

### Conclusion:

The data and information provided in this submission support the conclusion that the SuperCable Iso-Elastic Cerclage System that is the subject of this 510(k) submission is substantially equivalent to its predicate device (K030256), and the proposed change does not significantly affect the safety or effectiveness of the device. This line addition is being proposed to improve manufacturing efficiency and is not motivated by any clinical or performance factors.

---

**Source:** [https://fda-staging.innolitics.com/device/K181749](https://fda-staging.innolitics.com/device/K181749)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).