InterSpace Knee is indicated for temporary use (maximum 180 days) as a total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process. InterSpace Knee is applied on the femoral condyles and on the tibial plate following removal of the existing implant and radical debridement. The device is intended for use in conjunction with systemic antibiotic therapy (standard treatment approach to an infection). InterSpace Knee is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.). Because of the inherent mechanical limitations of the device material (gentamicin/polymethylmethacrylate), the is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period InterSpace Knee ATS is indicated for temporary use (maximum 180 days) as an adjunct to total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process, and where a large tibial defect is present. InterSpace Knee ATS is applied on the tibial plate following removal of the existing implant and radical debridement. The device must be combined with the tibial component of InterSpace Knee. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection). InterSpace Knee ATS is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.). Because of the inherent mechanical limitations of the device material (gentamicin/polymethacrylate), the Interspace Knee ATS is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period.
Device Story
Temporary knee spacer system; includes femoral and tibial components (InterSpace Knee) and optional tibial augment (InterSpace Knee ATS). Fabricated from gentamicin-loaded polymethylmethacrylate (PMMA) resin; provides local antibiotic elution and structural support. Used by orthopedic surgeons in two-stage revision procedures following infection; allows natural range of motion and partial weight-bearing while preserving soft tissue. Requires patient use of mobility aids (crutches/walkers) due to material mechanical limitations. Single-use; sterile; maximum 180-day implantation duration. Output is physical joint stabilization and local antimicrobial delivery; aids clinical decision-making by maintaining joint space and soft tissue integrity between infection debridement and permanent implantation.
Clinical Evidence
Bench testing only. No clinical data provided. Performance evaluation included static and fatigue testing of PMMA resin per ISO 5833 and ASTM F2118, device fatigue testing per ASTM F1800, wear debris analysis, and gentamicin elution profile comparison against cleared predicates. Biocompatibility evaluated per ISO 10993.
Indicated for skeletally mature patients undergoing two-stage TKR revision due to septic process. InterSpace Knee ATS indicated for patients with large tibial defects. Contraindicated for patients unable to consistently use mobility assist devices (crutches, walkers, canes) during the 180-day maximum implantation period.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" in bold, followed by "ADMINISTRATION" in a smaller font size. The logo is clean and professional, reflecting the organization's role in regulating food and drugs.
Tecres S.p.A. % Christine Brauer, Ph.D. Regulatory Affairs Consultant Brauer Device Consultants, LLC 7 Trail House Court Rockville, Maryland 20850
Re: K181732
Trade/Device Name: InterSpace Knee Extra-Large Size, InterSpace Knee ATS Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: March 27, 2019 Received: March 28, 2019
Dear Dr. Brauer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, fi applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@tda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181732
Device Name InterSpace Knee Extra-Large Size InterSpace Knee ATS
#### Indications for Use (Describe)
InterSpace Knee is indicated for temporary use (maximum 180 days) as a total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process.
InterSpace Knee is applied on the femoral condyles and on the tibial plate following removal of the existing implant and radical debridement. The device is intended for use in conjunction with systemic antibiotic therapy (standard treatment approach to an infection).
InterSpace Knee is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.). Because of the inherent mechanical limitations of the device material (gentamicin/polymethylmethacrylate), the is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period
InterSpace Knee ATS is indicated for temporary use (maximum 180 days) as an adjunct to total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process, and where a large tibial defect is present.
InterSpace Knee ATS is applied on the tibial plate following removal of the existing implant and radical debridement. The device must be combined with the tibial component of InterSpace Knee. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).
InterSpace Knee ATS is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.). Because of the inherent mechanical limitations of the device material (gentamicin/polymethacrylate), the Interspace Knee ATS is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period.
| Type of Use (Select one or both, as applicable) |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span style="font-size: 10pt;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
# Applicant Information
| Applicant: | Tecres S.p.A.<br>Via A. Doria 6<br>37066 Sommacampagna<br>Verona - Italy |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent: | Christine Brauer<br>US Agent for Tecres, S.p.A.<br>Brauer Device Consultants, LLC<br>7 Trail House Court<br>Rockville, MD 20850<br>Telephone: 301 545-1990<br>Email: chris.brauer@comcast.net |
| Date of Preparation: | March 27, 2019 |
| 510(k) Number: | K181732 |
Device Information
| Trade/Proprietary Name: | InterSpace Knee (XL size)<br>InterSpace Knee ATS |
|----------------------------|---------------------------------------------------------------------------------------------------|
| Common/Usual Name: | Temporary Knee Spacer with Gentamicin<br>Temporary Augmented Tibial Stem Spacer with gentamicin |
| Classification Name: | Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis |
| Classification Regulation: | 21 CFR 888.3560 |
| Class: | II |
| Product Code: | JWH - Prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer/metal/polymer |
| Panel: | Orthopedic |
{4}------------------------------------------------
### Device Description
The InterSpace Knee, cleared by FDA via K101356 and K062274, is a temporary spacer device available in three sizes (S-M-L). With this 510(k) submission, Tecres is adding the additional XL size to the FDA cleared range of InterSpace Knee sizes.
The InterSpace Knee ATS is a temporary spacer device available in four sizes, intended to be combined with the tibial component of InterSpace Knee. Specific sizes of the InterSpace Knee ATS are compatible with certain InterSpace Knee sizes.
The devices, as single device (InterSpace Knee, K101356 and K062274) or once InterSpace Knee is combined with the InterSpace Knee ATS, provide patients with a temporary, complete knee implant that allows for a natural range of motion and partial weight-bearing during treatment of the infection, and preserves the soft tissue to prevent further complications, such as muscular contraction, to facilitate the subsequent joint replacement procedure after systemic treatment of the underlying infection.
The InterSpace Knee XL size and InterSpace Knee ATS components are sterile, single-use device intended for temporary use (maximum 180 days) as a partial joint replacement. The devices are made of fully formed polymethylmethacrylate (PMMA), which is radio-opaque and contains gentamicin. The mass used to fill the molds (the unformed PMMA resin) is prepared from powder and liquid components. The liquid component consists of methylmethacrylate (MMA), N,N- dimethyl-p-toluidine and hydroquinone; the powder component consists of PMMA, barium sulphate, benzoyl peroxide and gentamicin sulphate.
### Indication for Use
### InterSpace Knee
InterSpace Knee is indicated for temporary use (maximum 180 days) as a total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process.
InterSpace Knee is applied on the femoral condyles and on the tibial plate following removal of the existing implant and radical debridement. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).
InterSpace Knee is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.). Because of the inherent mechanical limitations of the device material (gentamicin/polymethylmethacrylate), Interspace Knee is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period.
{5}------------------------------------------------
## InterSpace Knee ATS
InterSpace Knee ATS is indicated for temporary use (maximum 180 days) as an adjunct to total knee replacement (TKR) in skeletally mature patients undergoing a two-stage procedure due to a septic process, and where a large tibial defect is present.
InterSpace Knee ATS is applied on the tibial plate following removal of the existing implant and radical debridement. The device must be combined with the tibial component of InterSpace Knee. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).
InterSpace Knee ATS is not intended for use for more than 180 days, at which time it must be explanted and a permanent device implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion etc.). Because of the inherent mechanical limitations of the device material (gentamicin/ polymethylmethacrvlate), the Interspace Knee ATS is only indicated for patients who will consistently use traditional mobility assist devices (e.g. crutches, walkers, canes) throughout the implantation period.
### InterSpace Knee (XL size): Substantial Equivalence Comparison
For purposes of the substantial equivalence comparisons, the following predicate device was selected:
- Tecres S.p.A. InterSpace Knee which was cleared through 510(k) applications . K062274 and K101356.
Intended Use: The InterSpace Knee is intended to be used for patients undergoing a two-stage revision procedure for an infection of a total knee implant. The intended use of the XL size is exactly the same as for the other FDA cleared sizes (S-M-L) (see Table 1).
| Characteristic | InterSpace Knee (XL size) - Subject<br>Device | InterSpace Knee - Predicate Device<br>(K101356, K062274) |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification<br>Regulation | 888.3560 | 888.3560 |
| Product Code | JWH | JWH |
| Class | II | II |
| Indication for Use | ...is indicated for temporary use<br>(maximum 180 days) as a total knee<br>replacement (TKR) in skeletally<br>mature patients undergoing a two-<br>stage procedure due to a septic<br>process... | ...is indicated for temporary use<br>(maximum 180 days) as a total knee<br>replacement (TKR) in skeletally mature<br>patients undergoing a two-stage<br>procedure due to a septic process... |
| Primary Function | Provide a temporary implant to allow<br>for natural range of motion and partial<br>weight-bearing | Provide a temporary implant to allow for<br>natural range of motion and partial<br>weight-bearing |
| Secondary Function | Preserve soft tissue structure<br>To release antibiotics (gentamicin) in<br>the local area | Preserve soft tissue structure<br>To release antibiotics (gentamicin) in the<br>local area |
Table 1. Summary of Intended Use for the InterSpace Knee XL Size Compared to the Predicate Device
{6}------------------------------------------------
| Characteristic | InterSpace Knee (XL size) - Subject<br>Device | InterSpace Knee - Predicate Device<br>(K101356, K062274) |
|--------------------------------|----------------------------------------------------|----------------------------------------------------------|
| Target Patient<br>Population | Patients with an infected total knee<br>prosthesis | Patients with an infected total knee<br>prosthesis |
| | Intended Users | Orthopedic surgeons |
| Intended Surgical<br>Procedure | Two-stage knee revision procedure | Two-stage knee revision procedure |
| | Body Contact and<br>Duration | Bone – Maximum of 180 days |
| Prescription Device | | Yes |
Comparison of the Technological Characteristics: All the InterSpace Knee sizes (S-M-L-XL) share the same design and technological characteristics (see Table 2).
| Characteristic | Tecres S.p.A. -<br>InterSpace Knee (XL size) -<br>Subject Device | Tecres S.p.A. -<br>InterSpace Knee - Predicate<br>Device (K101356, K062274) |
|---------------------------------------------|---------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Main Components | Polymethylmethacrylate<br>(PMMA)<br>Methylmethacrylate (MMA)<br>Barium Sulphate | Polymethylmethacrylate (PMMA)<br>Methylmethacrylate (MMA)<br>Barium Sulphate |
| Other Components | Benzoyl peroxide<br>N,N-Dimethyl-p-toluidine<br>Hydroquinone | Benzoyl peroxide<br>N,N-Dimethyl-p-toluidine<br>Hydroquinone |
| Antibiotic | Gentamicin Sulphate | Gentamicin Sulphate |
| Design (shape) | Femoral + Tibial | Femoral + Tibial |
| Sizes | 1 size of femoral + tibial<br>component (XL) | 3 sizes of femoral + tibial<br>component (S, M, L) |
| X-ray visibility | Yes | Yes |
| Single-use device | Yes | Yes |
| Provided Sterile | Yes | Yes |
| Spacer Sterilization Method | Ethylene Oxide | Ethylene Oxide |
| Sterility Assurance Level (SAL)<br>– Powder | 10-6 | 10-6 |
| Shelf Life | 5 years | 5 years |
Table 2: Summary of Technological Characteristics between the InterSpace Knee (XL size) and InterSpace Knee Other Sizes (S-M-L)
Performance Data: This 510(k) notification provided performance evaluation to establish the substantial equivalence of the InterSpace Knee (XL size) to the other sizes already FDA cleared (K06274 and K101356).
Sterilization and Shelf Life: The devices are sterilized using standard methods and the sterilization cycles have been validated following international standards. The sterilization validation complies with: UNI EN ISO 11135:2014; EC 1-2011 UNI EN ISO 11737-1:2006, UNI EN ISO 11737-2:2010, UNI EN ISO 10993-7:2009, UNI EN 556-1:2002, EP current
{7}------------------------------------------------
Edition. The sterilization cvcle was validated to a sterility assurance level of 10-6. The shelf life was established through real-time and accelerated aging studies. All demonstrate a shelf life of 5 years.
Biocompatibility: Biocompatibility evaluation has been performed to show the device materials are safe, biocompatible and suitable for their intended use. Both ISO 10993 and FDA Draft Guidance "Use of International Standard ISO 10993. Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process" have been taken into account to evaluate the biocompatibility of the device materials. All the Interspace Knee spacer devices (including the XL size) are manufactured with same materials using the same manufacturing process: they share the same tissue contact and duration of contact.
Performance Testing: Performance testing was performed to characterize the InterSpace Knee (XL size):
- Static performances of the resin (ISO 5833): Resin was tested and shown to meet the ● minimum requirements of ISO 5833.
- Fatigue performances of the resin (ASTM F2118): Resin was tested and shown to meet ● the acceptance criteria according to ASTM F2118.
- . Fatigue test: the bigger size (XL) has been demonstrated to be a better case compared to the other InterSpace Knee sizes (S-M-L).
- Wear test: an evaluation of the debris has been compared with another temporary spacer ● already cleared by FDA.
- Surface Roughness: surface roughness is the same of the other sizes already approved
- Antibiotic (gentamicin) elution testing: The gentamicin elution profile demonstrated . substantial equivalence with gentamicin elution of another temporary spacer already cleared by FDA.
The performance data demonstrate that the new device is substantially equivalent to the predicate device.
### InterSpace Knee ATS: Substantial Equivalence Comparison
For purposes of the substantial equivalence comparisons, the following predicate device was selected:
- Knee Modular Spacer which was submitted by Garventis, LLC and cleared through 510(k) application K112470.
The InterSpace Knee ATS when combined with the InterSpace Knee provides a total temporary knee spacer for patients with a large tibial defect; this is analogous to the Knee Modular Spacer.
The following device was selected as a reference device:
- . Tecres S.p.A. InterSpace Knee which was cleared originally through 510(k) applications K062274 and K101356.
{8}------------------------------------------------
Intended Use: The InterSpace Knee ATS is intended to be used with the InterSpace Knee for patients undergoing a two-stage revision procedure for an infection of a total knee implant who have a large tibial defect. This is the same intended use as the predicate Knee Modular Spacer (see table below). Both devices provide the same primary function; i.e., allowing for natural range of motion and partial weight bearing and preserving the soft tissue structure. Thus, the InterSpace Knee ATS has the same intended use as the predicate device.
| Characteristics | InterSpace Knee ATS Plus<br>InterSpace Knee | Knee Modular Spacer - Predicate<br>Device |
|--------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Classification Regulation | 888.3560 | 888.3360 |
| Product Code | JWH | JWH |
| Class | II | II |
| Primary Function | Provide a temporary implant to allow<br>for natural range of motion and partial<br>weight-bearing | Provide a temporary implant to allow<br>for natural range of motion and partial<br>weight-bearing |
| | Preserve soft tissue structure | Preserve soft tissue structure |
| Secondary Function | To release antibiotic (gentamicin) in<br>the local area | To release antibiotic (gentamicin) in<br>the local area |
| Target Patient Population | Patients with an infected knee<br>prosthesis | Patients with an infected knee<br>prosthesis |
| Intended Users | Orthopedic surgeons | Orthopedic surgeons |
| Intended Surgical<br>Procedure | Two-stage knee revision procedure | Two-stage knee revision procedure |
| Body Contact and<br>Duration | Bone – Maximum of 180 days | Bone – Maximum of 180 days |
| Prescription Device | Yes | Yes |
Table 3. Summary of Intended Use for the InterSpace Knee ATS plus InterSpace Knee Compared to the Predicate Device
Comparison of the Technological Characteristics: The combined InterSpace Knee ATS plus InterSpace Knee share the same design and technological characteristics as the Knee Modular Spacer (see table). Both temporary spacers are made from gentamicin-loaded PMMA bone cement. Both consist of modular components with a tibial insert. Both are provided sterile for single-use.
| Table 4: Summary of Technological Characteristics between the InterSpace Knee ATS plus InterSpace | | | |
|---------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|----------------------------------------------------------------------------------------------|--|
| Knee and the Predicate Device | | | |
| Characteristics | InterSpace Knee ATS Plus InterSpace<br>Knee | Knee Modular Spacer – Predicate<br>Device | |
| Total Knee<br>Components | Femoral component<br>Tibial component<br>ATS component (optional) | Femoral component<br>Tibial component<br>Tibial insert (optional) | |
| Component<br>Materials | Antibiotic loaded PMMA | Antibiotic loaded PMMA with color<br>additive (details not specified in the 510k<br>Summary) | |
| Antibiotic | Gentamicin | Gentamicin | |
| Sizes | 4 InterSpace Knee (femoral + tibial)<br>4 InterSpace Knee ATS | Femoral and tibial components<br>Optional tibial insert | |
{9}------------------------------------------------
| Characteristics | InterSpace Knee ATS Plus InterSpace Knee | Knee Modular Spacer - Predicate Device |
|-------------------------|------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| | | Available in a range of sizes; further details unspecified in 510(k) Summary |
| Modularity | Yes. InterSpace Knee ATS matches in size to be compatible with specific InterSpace Knee sizes. | Yes. Femoral component can be combined with tibial component; tibial component can be combined with optional tibial insert |
| Augment | ATS component | Tibial insert |
| Function of the Augment | Optional<br>Increase the thickness | Optional<br>Increase the thickness |
| Radiopaque | Yes | Yes |
| Single-use device | Yes | Yes |
| Provided Sterile | Yes | Yes |
For a comparison of certain technological characteristics, the InterSpace Knee ATS is compared to the InterSpace Knee as a reference device. The InterSpace Knee ATS shares many of the same technological characteristics as the InterSpace Knee, including the identical materials, and same PMMA bone cement formulation including antibiotic concentration. The technological characteristics are the same between the proposed device and the reference device (see table).
| Characteristics | InterSpace Knee ATS | InterSpace Knee |
|------------------------------------------------|------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| Main Components | Polymethylmethacrylate (PMMA)<br>Methylmethacrylate (MMA)<br>Barium Sulphate | Polymethylmethacrylate (PMMA)<br>Methylmethacrylate (MMA)<br>Barium Sulphate |
| Other Components | Benzoyl peroxide<br>N,N-Dimethyl-p-toluidine<br>Hydroquinone | Benzoyl peroxide<br>N,N-Dimethyl-p-toluidine<br>Hydroquinone |
| Antibiotic | Gentamicin Sulphate | Gentamicin Sulphate |
| Design (shape) | Tibial Augment for use with the<br>InterSpace Knee | Femoral + Tibial |
| Sizes | 4 sizes matching the InterSpace Knee<br>tibial components | 4 sizes |
| X-ray visibility | Yes | Yes |
| Single-use device | Yes | Yes |
| Provided Sterile | Yes | Yes |
| Spacer Sterilization<br>Method | Ethylene Oxide | Ethylene Oxide |
| Sterility Assurance<br>Level (SAL) -<br>Powder | 10-6 | 10-6 |
| Shelf Life | 5 years | 5 years |
Table 5: Summary of Technological Characteristics between the InterSpace Knee ATS and the InterSpace Knee
{10}------------------------------------------------
Performance Data: This 510(k) notification provided performance data to establish the substantial equivalence of the InterSpace Knee ATS to the predicate device or the reference device. The following is a summary of the performance data.
Sterilization and Shelf Life: The devices are sterilized using standard methods and the sterilization cycles have been validated following international standards. The sterilization validation complies with: UNI EN ISO 11135:2014; EC 1-2011 UNI EN ISO 11737-1:2006. UNI EN ISO 11737-2:2010, UNI EN ISO 10993-7:2009, UNI EN 556-1:2002, EP current Edition. The sterilization cycle was validated to a sterility assurance level of 106. The shelf life was established through real-time and accelerated aging studies. All demonstrate a shelf life of 5 years.
Biocompatibility evaluation has been performed to show the device Biocompatibility: materials are safe, biocompatible and suitable for their intended use. Both ISO 10993 and FDA Draft Guidance "Use of International Standard ISO 10993. Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a Risk Management Process" have been taken into account to evaluate the biocompatibility of the device materials. The device materials of the InterSpace Knee ATS are identical to the InterSpace Knee, which share the same tissue contact and duration of contact.
Performance Testing: Performance testing was performed to characterize the InterSpace Knee ATS. Performance testing was performed on either the InterSpace Knee ATS or the InterSpace Knee and included an evaluation of the following:
- Static performances of the resin (ISO 5833): Resin was tested and shown to meet the ● minimum requirements of ISO 5833.
- Fatigue performances of the resin (ASTM F2118): Resin was tested and shown to meet the acceptance criteria according to ASTM F2118.
- Fatigue test on the device has been performed according to ASTM F1800. ●
- Antibiotic (gentamicin) elution testing: The gentamicin elution profile demonstrated substantial equivalence with gentamicin elution of another temporary spacer already cleared by FDA.
The performance data demonstrate that the new device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.