MTS Plazomicin 0.016-256 µg/mL

K181708 · Liofilchem s.r.l. · JWY · Jul 31, 2018 · Microbiology

Device Facts

Record IDK181708
Device NameMTS Plazomicin 0.016-256 µg/mL
ApplicantLiofilchem s.r.l.
Product CodeJWY · Microbiology
Decision DateJul 31, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

The Liofilchem® MTS™ (MIC Test Strip) is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of bacteria. MTS™ consists of specialized paper impregnated with a pre-defined concernment of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in ug/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures. The MTS™ Plazoni.cin at concentrations of 0.016-256 ug/mL should be interpreted at 16-20 hours of incubation. MTSTM PLZ can be used to determine the MC of plazomicin against the microorganisms listed in the table below: Plazomicin Activity According to the FDA Label Clinical and in vitro Enterobacter cloacae Escherichia coli Klebsiella pneumoniae Proteus mirabilis in vitro only Citrobacter freundii Citrobacter koseri Klebsiella (Enterobacter) aerogenes Klebsiella oxytoca Morganella morganii Proteus vulgaris Providencia stuartii Serratia marcescens

Device Story

MTS Plazomicin is a specialized paper strip impregnated with a concentration gradient of the antimicrobial agent plazomicin. Used in clinical microbiology laboratories to determine the Minimum Inhibitory Concentration (MIC) of plazomicin against specific bacteria. The strip is placed on inoculated agar media and incubated for 16-20 hours. The antimicrobial agent diffuses into the agar, creating a concentration gradient. The MIC is determined by observing the point where the bacterial growth inhibition ellipse intersects the strip. Results are read manually by laboratory personnel. This quantitative measurement assists clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections.

Clinical Evidence

Performance evaluated using 538 clinical and challenge Enterobacteriaceae isolates across three U.S. sites. Compared against CLSI reference broth microdilution. Results: 97.0% Essential Agreement (EA) and 96.1% Category Agreement (CA). No major or very major errors observed; 18 minor discrepancies. Trending observed for several species; labeling includes appropriate footnotes.

Technological Characteristics

Specialized paper strip impregnated with a concentration gradient of plazomicin (0.016-256 ug/mL). Principle of operation is antimicrobial diffusion into agar media. Manual reading of MIC endpoint after 16-20 hours of incubation. No electronic components, software, or energy sources.

Indications for Use

Indicated for in vitro determination of antimicrobial susceptibility of non-fastidious Gram-negative bacteria (Enterobacteriaceae) to Plazomicin. Patient population includes those with infections caused by E. coli, K. pneumoniae, E. cloacae, P. mirabilis, C. freundii, C. koseri, K. aerogenes, K. oxytoca, M. morganii, P. vulgaris, P. stuartii, and S. marcescens. Prescription use only.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Submission Summary (Full Text)

{0} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: K181708 B. Purpose for Submission: To obtain a substantial equivalence determination for Plazomicin (PLZ) at concentrations of 0.016-256 µg/mL for susceptibility testing of non-fastidious Gram-negative organisms C. Measurand: Plazomicin 0.016-256 µg/mL D. Type of Test: Quantitative Antimicrobial Susceptibility Test growth-based detection E. Applicant: Liofilchem s.r.l. F. Proprietary and Established Names: MTS Plazomicin 0.016-256 µg/mL G. Regulatory Information: 1. Regulation section: 866.1640 Antimicrobial Susceptibility Test Powder 2. Classification: II 3. Product code: JWY - Manual Antimicrobial Susceptibility Test Systems 4. Panel: 83 – Microbiology 1 {1} ### H. Intended Use: #### 1. Intended use(s): The Liofilchem MTS (MIC Test Strip) is a quantitative method intended for the in vitro determination of antimicrobial susceptibility of bacteria. MTS consists of specialized paper impregnated with a pre-defined concentration gradient of an antimicrobial agent, which is used to determine the minimum inhibitory concentration (MIC) in \( \mu \) g/mL of antimicrobial agents against bacteria as tested on agar media using overnight incubation and manual reading procedures. The MTS Plazomicin at concentrations of 0.016-256 \( \mu \) g/mL should be interpreted at 16-20 hours of incubation. MTS PLZ can be used to determine the MIC of plazomicin against the microorganisms listed in the table below: | Plazomicin Activity According to the FDA Label | | | --- | --- | | Clinical and in vitro | in vitro only | | Escherichia coliKlebsiella pneumoniaeEnterobacter cloacaeProteus mirabilis | Citrobacter freundiiCitrobacter koseriKlebsiella (Enterobacter) aerogenesKlebsiella oxytocaMorganella morganiiProteus vulgarisProvidentia stuartiiSerratia marcescens | #### 2. Indication(s) for use: Same as Intended Use #### 3. Special conditions for use statement(s): - For prescription use - The ability of the MTS to detect resistant isolates with the following drug/bacterial species combinations is unknown because resistant isolates were either not available or an insufficient number was encountered at the time of comparative testing. Plazomicin: Citrobacter freundii, Citrobacter koseri, Klebsiella (Enterobacter) aerogenes, Klebsiella oxytoca, Providencia stuartii, Proteus vulgaris, Serratia marcescens - Characterization of 16S rRNA methyltransferases, aminoglycoside modifying enzymes (AMEs), altered efflux and loss of outer membrane porins was not 2 {2} available for organisms at the time of comparative testing, and therefore the performance of MTS Plazomicin for non-fastidious Gram-negative bacilli with these resistance mechanisms is unknown for the following: Enterobacteriaceae - The safety and efficacy of plazomicin in treating clinical infections due to Gram-negative organisms other than E. coli, K. pneumoniae, E. cloacae and P. mirabilis may not have been established in adequate and well-controlled clinical trials. The clinical significance of susceptibility information in such instances is unknown. 4. Special instrument requirements: Manual reading only # I. Device Description: The Plazomicin MIC Test Strip (MTS) consists of specialized paper impregnated with a predefined concentration gradient of Plazomicin across 15 two-fold dilutions similar to dilutions used by conventional MIC methods. One side of the strip is labelled with the Plazomicin code (PLZ) and the MIC reading scale in µg/mL. When the MIC Test Strip is applied onto an inoculated agar surface, the preformed exponential gradient of antimicrobial agent is immediately transferred to the agar matrix. After 16-20 hours incubation, a symmetrical inhibition ellipse centered along the strip is formed. The MIC is read directly from the scale in terms of µg/mL at the point where the edge of the inhibition ellipse intersects the MIC Test Strip. Since MTS strip generates MIC values which fall between two-fold dilutions for interpretation, the MIC value read is recorded to the next two-fold dilution value. # J. Substantial Equivalence Information: 1. Predicate device name(s): Liofilchem MTS, vancomycin 2. Predicate 510(k) number(s): K153687 3 {3} ### 3. Comparison with predicate: Table 1: Comparison with the Predicate Device | Similarities | | | | --- | --- | --- | | Item | DeviceLiofilchem MTS, Plazomicin (K181708) | PredicateLiofilchem MTS, vancomycin (K153687) | | Intended Use | Quantitative susceptibility to antimicrobial agents | Same | | Media | Mueller Hinton agar | Same | | Inoculation | Isolated colonies from culture in suspension equivalent to 0.5 McFarland. Inoculum is applied manually using the manual plate inoculation method or plate rotator for even distribution of inoculum | Same | | Reading | Manual; the point where the edge of inhibition ellipse intersects the MIC Test Strip | Same | | Result | MIC | Same | | Differences | | | | --- | --- | --- | | Item | DeviceLiofilchem MTS, Plazomicin (K181708) | PredicateLiofilchem MTS, vancomycin (K153687) | | Antibiotic | Plazomicin code (PLZ) | Vancomycin code (VA) | | Incubation | \( 35 \pm 2^{\circ}C \) for 16 - 20hrs | \( 35 \pm 2^{\circ}C \) for 24 hours | ## K. Standard/Guidance Document Referenced: - Guidance for Industry and FDA - Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems – August 28, 2009. - CLSI M07-A10 “Methods for Dilution Antimicrobial Susceptibility Tests for Bacteria That Grow Aerobically; Approved Standard, Tenth Edition January 2015”. - CLSI M100-S26 “Performance Standards for Antimicrobial Susceptibility Testing; Twenty-Fifth Informational Supplement, January 2016”. ## L. Test Principle: MTS are made of specialized paper impregnated with a predefined concentration gradient of antibiotic, across 15 two-fold dilutions similar to dilutions used by conventional MIC methods. When the MIC Test Strip is applied onto an inoculated agar surface, the preformed exponential gradient of antimicrobial agent is immediately transferred to the agar matrix. After 16-20 hours incubation, a symmetrical inhibition ellipse centered along the strip is formed. The MIC is read directly from the scale in terms of \(\mu \mathrm{g} / \mathrm{mL}\) at the point where the edge of the inhibition ellipse intersects the strip MIC Test Strip. 4 {4} Growth along the entire gradient (i.e., no inhibition ellipse) indicates that the MIC value is greater than or equal to (≥) the highest value on the scale. An inhibition ellipse that intersects below the lower end of the scale is read as less than (<) the lowest value. An MIC of 0.125μg/mL is considered to be the same as 0.12μg/mL for reporting purposes. An MTS MIC value which falls between standard two-fold dilutions must be rounded up to the next standard upper two-fold value before categorization. # M. Performance Characteristics: # 1. Analytical performance: a. Precision/Reproducibility: Reproducibility testing was conducted at three sites using ten Gram-negative organisms. Each isolate was tested in triplicate over three days. The reproducibility panel included three E. coli, three K. pneumoniae, two E. cloacae, one P. mirabilis, and one P. vulgaris isolate. The mode MIC values were pre-determined and the reproducibility was calculated based on the number of MIC values that fell within ±1 doubling dilution of the mode MIC values. All MIC results were on scale. The testing resulted in overall reproducibility of greater than 95%. The results were acceptable. b. Linearity/assay reportable range: Not applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): # Quality Control (QC) Testing: The CLSI recommended QC strains, namely E. coli ATCC 25922 and P. aeruginosa ATCC 27853 were tested a sufficient number of times (i.e., at least 20/site) at each testing site using both MTS and reference methods. The results are summarized in Table 2 below. The quality control results are acceptable. 5 {5} Table 2: Plazomicin MTS QC Results | Organism | Concentration (μg/mL) | Reference | MTS | | --- | --- | --- | --- | | *E. coli* ATCC 25922 Expected Result: 0.25-2 μg/mL | 0.12 | | | | | 0.25 | | 29 | | | 0.5 | 40 | 30 | | | 1 | 20 | 2 | | | 2 | 1 | | | | 4 | | | | *P. aeruginosa* ATCC 27853 Expected Result: 1-4 μg/mL | 0.5 | | | | | 1 | 8 | 1 | | | 2 | 50 | 47 | | | 4 | 3 | 13 | | | 8 | | | # **Inoculum Density Check:** The inoculum was prepared to achieve turbidity equivalent to a 0.5 McFarland standard. Colony counts were performed periodically at each site for all QC replicates, from one replicate of each reproducibility isolate on each of the three days of testing, and from a minimum of 10% of the clinical and challenge strains tested. Inoculum density checks were performed and the colony counts obtained for each isolate were within the recommended range of approximately 1x 10⁸ CFU/mL. d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: Results obtained with Liofilchem MIC Test Strip (MTS) with Plazomicin were compared to results obtained from frozen reference MIC panels. Reference panels were prepared and interpreted as outlined in the recommendations in CLSI document M7-A10. Isolated colonies from an overnight blood agar plate were suspended in saline to achieve a 0.5 McFarland standard turbidity (approximately 10⁸ CFU/mL). Testing 6 {6} conditions consisted of incubation of the inoculated Mueller Hinton agar plates in an inverted position at 35°C ± 2° for 16-20 hours. At the end of incubation, the MIC value at which the edge of the inhibition ellipse intersected the strip was compared to MIC results obtained with the reference method. # **Growth Rate:** The growth rate for the Liofilchem MIC Test Strip (MTS) with Plazomicin was 100%. # **Clinical:** Clinical testing was performed at three U.S. sites. A total of 452 clinical Enterobacteriaceae isolates were tested which included 15 C. freundii, 15 C. koseri, 30 K. aerogenes, 45 E. cloacae, 120 E. coli, 15 K. oxytoca, 111 K. pneumoniae, 15 M. morganii, 35 P. mirabilis, 15 P. stuartii, 21 P. vulgaris, and 15 S. marcescens isolates. Of the clinical isolates, 61.7% were tested within 6 months of isolation. # **Challenge:** Challenge testing was performed at one internal site. A total of 86 challenge Enterobacteriaceae isolates were tested which included 2 C. freundii, 2 C. koseri, 5 K. aerogenes, 10 E. cloacae, 16 E. coli, 10 K. oxytoca, 12 K. pneumoniae, 4 M. morganii, 17 P. mirabilis, 2 P. stuartii, 5 P. vulgaris, and 1 S. marcescens isolate. The performance for the total 538 clinical and challenge isolates is summarized in Table 3 below. Table 3: Overall Performance of Clinical and Challenge Isolates (Combined) | Plazomicin | EA Tot | EA N | EA % | Eval. EA Tot | Eval. EA N | Eval. EA% | CA N | CA% | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | Enterobacteriaceae (all species) Clinical | 452 | 437 | 96.7 | 449 | 434 | 96.7 | 434 | 96.0 | 10 | 18 | 0 | 0 | | Challenge | 86 | 85 | 98.8 | 75 | 74 | 98.7 | 83 | 96.5 | 44 | 3 | 0 | 0 | | Combined | 538 | 522 | 97.0 | 524 | 508 | 96.9 | 517 | 96.1 | 54 | 21 | 0 | 0 | EA – Essential Agreement CA – Category Agreement EVAL – Evaluable isolates R – Resistant isolates min – minor errors maj – major errors vmj – very major errors Essential Agreement (EA) is when the Liofilchem MIC Test Strip (MST) results agree exactly or within one doubling dilution of the reference broth microdilution results. Category Agreement (CA) is when the Liofilchem MIC Test Strip (MST) result interpretation agrees exactly with the reference broth microdilution result interpretation. The overall performance of all Enterobacteriaceae isolates is acceptable with 97.0% EA and 96.1% CA. There were 18 minor discrepancies (3.9%) and no major or very major errors. 7 {7} When the performance was evaluated individually by species, it was noted that the CA for both P. mirabilis and P. stuartii was less than 90%. However, all categorical errors for these species were minor and within essential agreement and therefore, acceptable. ## Resistance Mechanisms: Molecular characterization for resistance to Plazomicin was not evaluated for all organisms as this information was not available of the time of testing. This was addressed by adding the following footnote in the labeling: "Characterization of 16S rRNA methyltransferases, aminoglycoside modifying enzymes (AMEs), altered efflux and loss of outer membrane porins was not available for organisms at the time of comparative testing, and therefore the performance of MTS Plazomicin for non-fastidious Gram-negative bacilli with these resistance mechanisms is unknown for the following: Enterobacteriaceae" ## Trending: Trending was observed (Table 4) for MIC values for C. freundii, C. koseri, K. aerogenes, E. coli, M. morganii, P. stuartii, P. vulgaris and S. marcescens which tended to be in exact agreement or lower when compared to the reference method. The difference between higher and lower dilutions for these organisms was ≥30%. The following footnote was included in the labeling to indicate this trending: "The MTS Plazomicin MIC values tended to be in exact agreement or at least one doubling dilution lower when testing C. freundii, C. koseri, K. aerogenes, E. coli, M. morganii, P. stuartii, P. vulgaris and S. marcescens compared to the CLSI reference broth microdilution." Table 4. Trending for Enterobacteriaceae by Species | Total | ≥2 dil. lower | 1 dil. lower | Exact | 1 dil. higher | ≥2 dil. higher | | --- | --- | --- | --- | --- | --- | | **C. freundii^{a}** | | | | | | | 17 | 0 | 6 | 10 | 1 | 0 | | | (35.29%) | | (58.82%) | (5.88%) | | | **C. koseri^{b}** | | | | | | | 17 | 0 | 7 | 10 | 0 | 0 | | | (41.18%) | | (58.82%) | (0%) | | | **K. aerogenes^{c}** | | | | | | | 35 | 1 | 16 | 17 | 1 | 0 | | | (48.57%) | | (48.57%) | (2.86%) | | | **E. cloacae^{d}** | | | | | | | 51 | 0 | 13 | 36 | 2 | 0 | | | (25.49%) | | (70.59%) | (3.92%) | | | **E. coli^{e}** | | | | | | | 132 | 12 | 59 | 61 | 0 | 0 | | | (53.79%) | | (46.21%) | (0%) | | 8 {8} | K. oxytoca\( ^{f} \) | | | | | | | --- | --- | --- | --- | --- | --- | | 25 | 0 | 7 | 17 | 1 | 0 | | | (28%) | | (68%) | (4%) | | | K. pneumoniae\( ^{g} \) | | | | | | | 119 | 0 | 19 | 86 | 14 | 0 | | | (15.97%) | | (72.27%) | (11.76%) | | | M. morganii\( ^{h} \) | | | | | | | 19 | 0 | 9 | 10 | 0 | 0 | | | (47.37%) | | (52.63%) | (0%) | | | P. mirabilis\( ^{i} \) | | | | | | | 52 | 0 | 15 | 37 | 0 | 0 | | | (28.85%) | | (71.15%) | (0%) | | | P. stuartii\( ^{j} \) | | | | | | | 17 | 0 | 7 | 9 | 1 | 0 | | | (41.18%) | | (52.94%) | (5.88%) | | | P. vulgaris\( ^{k} \) | | | | | | | 26 | 1 | 9 | 16 | 0 | 0 | | | (38.46%) | | (61.54%) | (0%) | | | S. marcescens\( ^{l} \) | | | | | | | 16 | 0 | 5 | 11 | 0 | 0 | | | (31.25%) | | (68.75%) | (0%) | | | All Enterobacteriaceae\( ^{m} \) | | | | | | | 526 | 16 | 170 | 320 | 20 | 0 | | | (35.36%) | | (60.84%) | (3.8%) | | \( ^{a} \) Difference between the higher and lower dilutions for C. freundii is: -29.41%; 95% C.I. (-53.31% to -1.69%) \( ^{b} \) Difference between the higher and lower dilutions for C. koseri is: -41.18%; 95% C.I. (-63.99% to -14.30%) \( ^{c} \) Difference between the higher and lower dilutions for K. aerogenes is: -45.71%; 95% C.I. (-61.75% to -26.25%) \( ^{d} \) Difference between the higher and lower dilutions for E. cloacae is: -21.57%; 95% C.I. (-35.24% to -7.96%) \( ^{e} \) Difference between the higher and lower dilutions for E. coli is: -53.79%; 95% C.I. (-62.07% to -44.84%) \( ^{1} \) Difference between the higher and lower dilutions for K. oxytoca is: -24%; 95% C.I. (-43.85% to -3.27%) \( ^{g} \) Difference between the higher and lower dilutions for K. pneumoniae is: -4.2%; 95% C.I. (-13.12% to 4.71%) \( ^{h} \) Difference between the higher and lower dilutions for M. morganii is: -47.37%; 95% C.I. (-68.29% to -21.21%) \( ^{1} \) Difference between the higher and lower dilutions for P. mirabilis is: -28.85%; 95% C.I. (-42.27% to -16.28%) \( ^{j} \) Difference between the higher and lower dilutions for P. stuartii is: -35.29%; 95% C.I. (-58.62% to -6.52%) \( ^{k} \) Difference between the higher and lower dilutions for P. vulgaris is: -38.46%; 95% C.I. (-57.47% to -17.90%) \( ^{1} \) Difference between the higher and lower dilutions for S. marcescens is: -31.25%; 95% C.I. (-55.60% to -5.43%) \( ^{m} \) Difference between the higher and lower dilutions for all Enterobacteriaceae is: -31.56%; 95% C.I. (-35.94% to -27.12%) #### b. Matrix comparison: Not Applicable 9 {9} 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range: The FDA susceptibility interpretive criteria for Plazomicin are as listed in Table 5. Table 5: FDA Interpretive Criteria for Plazomicin ( \( \mu \) g/mL) | Organisms | S | I | R | | --- | --- | --- | --- | | Enterobacteriaceae | \( \leq 2 \) | 4 | \( \geq 8 \) | ### N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. ### O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 10
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...