← Product Code [JTN](/productcode/JTN) · K181700

# HardyDisk AST Plazomicin 30ug (PLZ30) (K181700)

_Hardy Diagnostics · JTN · Jul 10, 2018 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181700

## Device Facts

- **Applicant:** Hardy Diagnostics
- **Product Code:** [JTN](/productcode/JTN.md)
- **Decision Date:** Jul 10, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1620
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

HardyDisk AST Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Pseudomonas spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., and by modified procedures, Haemophilus spp., Neisseria gonorrhoeae, N. meningitidis and Streptococcus spp., including Streptococcus pneumoniae. Use of HardyDisk AST Plazomicin 30μg (PLZ30) for in vitro agar diffusion susceptibility testing is indicated when there is need to determine the susceptibility of bacteria to Plazomicin. Plazomicin has been shown to be active against susceptible isolates of the following bacteria both in vitro and in clinical infections: Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, Enterobacter cloacae. Plazomicin has been shown to be active in vitro against susceptible isolates of the following bacteria: Citrobacter freundii, Citrobacter koseri, Enterobacter aerogenes, Klebsiella oxytoca, Morganella morganii, Proteus vulgaris, Providencia stuartii, Serratia marcenscens. HardyDisk AST Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Pseudomonas spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., and by modified procedures, Haemophilus spp., Neisseria gonorrhoeae, N. meningitidis and Streptococcus spp., including Streptococcus pneumoniae.

## Device Story

HardyDisk AST Plazomicin 30μg (PLZ30) consists of 6-mm filter paper disks impregnated with 30μg of the antimicrobial agent Plazomicin. Used in clinical microbiology laboratories for Kirby-Bauer disk diffusion susceptibility testing. Procedure involves inoculating agar plates with bacterial isolates, placing disks on the surface, and incubating for 16-18 hours. Healthcare providers measure the resulting zone of inhibition diameter and compare it against established interpretive criteria (R/I/S) to determine bacterial susceptibility. This information assists clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections.

## Clinical Evidence

No clinical trials were performed for this device. Substantial equivalence is supported by bench testing, including reproducibility, quality control, and disk-to-MIC correlation studies conducted in accordance with CDER guidance for antimicrobial susceptibility testing. Interpretive criteria and QC ranges were established based on data evaluated by CDER and are consistent with CLSI M100 standards.

## Technological Characteristics

Antimicrobial susceptibility test disc; 30ug Plazomicin concentration; paper disc format; Class II device (21 CFR 866.1620); Product Code JTN.

## Regulatory Identification

An antimicrobial susceptibility test disc is a device that consists of antimicrobic-impregnated paper discs used to measure by a disc-agar diffusion technique or a disc-broth elution technique the in vitro susceptibility of most clinically important bacterial pathogens to antimicrobial agents. In the disc-agar diffusion technique, bacterial susceptibility is ascertained by directly measuring the magnitude of a zone of bacterial inhibition around the disc on an agar surface. The disc-broth elution technique is associated with an automated rapid susceptibility test system and employs a fluid medium in which susceptibility is ascertained by photometrically measuring changes in bacterial growth resulting when antimicrobial material is eluted from the disc into the fluid medium. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE

A. 510(k) Number:

K181700

B. Purpose for Submission:

Addition of Plazomicin Antimicrobial Susceptibility Test Disk

C. Measurand:

Plazomicin 30μg

D. Type of Test:

Antimicrobial Susceptibility Test Disks

E. Applicant:

Hardy Diagnostics

F. Proprietary and Established Names:

HardyDisk AST Plazomicin 30μg (PLZ30)

G. Regulatory Information:

1. Regulation section:

21 CFR 866.1620 Antimicrobial Susceptibility Test Disc

2. Classification:

Class II

3. Product code:

JTN

4. Panel:

83, Microbiology

1

{1}

# H. Intended Use:

# 1. Intended use(s):

HardyDisk AST Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Pseudomonas spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., and by modified procedures, Haemophilus spp., Neisseria gonorrhoeae, N. meningitidis and Streptococcus spp., including Streptococcus pneumoniae.

# 2. Indication(s) for use:

Use of HardyDisk AST Plazomicin 30μg (PLZ30) for in vitro agar diffusion susceptibility testing is indicated when there is need to determine the susceptibility of bacteria to Plazomicin.

Plazomicin has been shown to be active against susceptible isolates of the following bacteria both in vitro and in clinical infections:

Escherichia coli

Klebsiella pneumoniae

Proteus mirabilis

Enterobacter cloacae

Plazomicin has been shown to be active in vitro against susceptible isolates of the following bacteria:

Citrobacter freundii

Citrobacter koseri

Enterobacter aerogenes

Klebsiella oxytoca

Morganella morganii

Proteus vulgaris

Providencia stuartii

Serratia marcenscens

HardyDisk AST Disks are used for semi-quantitative in vitro susceptibility testing by the agar diffusion test procedure (Kirby-Bauer) of rapidly growing and certain fastidious bacterial pathogens. Standardized methods for agar diffusion testing have been described for Enterobacteriaceae, Staphylococcus spp., Pseudomonas spp., Acinetobacter spp., Listeria monocytogenes, Enterococcus spp., and by modified procedures, Haemophilus spp., Neisseria gonorrhoeae, N. meningitidis and Streptococcus spp., including Streptococcus pneumoniae.

2

{2}

3. Special conditions for use statement(s):

For prescription use only

4. Special instrument requirements:

N/A

# I. Device Description:

HardyDisk AST Disks utilize 6-mm diameter white filter paper disks. The disks are prepared by impregnating absorbent paper with a known concentration of 30μg Plazomicin. The disks are marked with the code PLZ30 on both sides.

HardyDisk AST Disks are supplied in plastic cartridges containing 50 disks each. They are also packaged as one cartridge per vial with desiccant or five cartridges per vial with desiccant.

# J. Substantial Equivalence Information:

1. Predicate device name(s):

HardyDisk Tigecycline 15μg

2. Predicate 510(k) number(s):

K062245

3. Comparison with predicate:

3

{3}

Table 1: Comparison with the Predicate

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | DeviceK181700HardyDiskPlazomicin 30μg | PredicateK062245HardyDiskTigecycline 15μg  |
|  Test Method | Antimicrobial Susceptibility Testing using paper disks impregnated with an antimicrobial agent | Same  |
|  Methodology | Kirby-Bauer Disk Diffusion Susceptibility Test Protocol requires the user to determine categorical interpretations (S/I/R) using the measured zone diameters. | Same  |
|  Inoculum | Prepared from pure isolated colonies to match the turbidity equivalent of a 0.5 McFarland in Tryptic Soy Broth. | Same  |
|  Inoculation Method | Dip a sterile swab into the prepared inoculum and streak an appropriate agar plate's surface three times. Add the disks impregnated with the antimicrobial agent to the surface of the plate. Incubate the agar plate agar side up in a \( 35 \pm 2^{\circ}C \) incubator for 16-18 hours. | Same  |
|  Reading Method | The user will interpret the zone diameters according established interpretive criteria for the drug. | Same  |

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Product Name | HardyDisk Plazomicin 30μg (PLZ30) | HardyDisk Tigecycline  |
|  Antimicrobial Agent | Plazomicin | Tigecycline  |
|  Concentration | 30μg | 15μg  |

### K. Standard/Guidance Document Referenced (if applicable):

CLSI M100 28 \( ^{th} \) Edition, Performance Standards for Antimicrobial Susceptibility Testing

### L. Test Principle:

The HardyDisk AST Disk is based on the agar diffusion (Kirby-Bauer) methodology. It utilizes dried filter paper disks impregnated with a known concentration of an antimicrobial agent that are placed onto the test medium surface. Mueller Hinton agar is recommended for agar diffusion testing of non-fastidious organisms and Mueller Hinton with 5% Sheep Blood

4

{4}

is recommended for Streptococcus spp. Three to five similar colonies are transferred to 4-5 mL of a suitable broth medium. The broth is incubated at 35°C for 2-6 hours to develop a turbidity that exceeds or is equivalent to a 0.5 McFarland standard. Alternatively, a direct broth or saline suspension of colonies may be prepared from an overnight culture. The final inoculum density should be equivalent to a 0.5 McFarland turbidity standard. The inoculum density may also be standardized photometrically.

Within 15 minutes of inoculum preparation, the Mueller Hinton agar plate is streaked with an inoculated swab to obtain an even inoculation of organism. Disks are aseptically placed onto the agar surface with a disk dispenser and the disks are pressed down with a sterile needle or forceps to make contact with the agar surface. Agar plates are incubated in an ambient air incubator at 35±2°C for 16 - 18 hours. Fastidious organisms are tested using appropriate media incubated in an atmosphere enriched with 5% CO₂, as recommended in the CLSI M02 approved standard document.

After incubation the agar medium is examined for a zone of inhibition around the disks. The zones of inhibition are measured to the nearest millimeter and compared to recognized zone size ranges for the antimicrobial agent being tested.

### M. Performance Characteristics (if/when applicable):

Descriptive characteristics were sufficient for the HardyDisk Plazomicin 30μg (PLZ30) disk based on extensive data from several microbiology disk studies evaluated by CDER which were used to generate the breakpoints and quality control (QC) expected ranges used for this subject device. The disk data used to support this submission included data from testing organisms within the spectrum of activity of plazomicin. Data was obtained from reproducibility, quality control and disk to MIC correlation studies and was generated in accordance with the CDER Clinical/Antimicrobial guidance, Microbiology Data for Systemic Antibacterial Drugs- Development, Analysis, and Presentation to ensure precise, accurate, and reproducible results.

For this review, the interpretative criteria are applied broadly to include the Enterobacteriaceae family. The list of bacterial species has been expanded and is not limited only to the indicated species.

The following statements are added as footnotes to the plazomicin interpretative criteria table for Enterobacteriaceae in the HardyDisk AST package insert:

The statement below is added to the package insert and is consistent with the FDA STIC website:

- The safety and efficacy of plazomicin in treating clinical infections due to organisms other than Escherichia coli, Klebsiella pneumoniae, Proteus mirabilis, and Enterobacter cloacae may not have been established in adequate and well-controlled clinical trials. The clinical significance of susceptibility information in such instances is unknown.

5

{5}

Regarding resistance marker/enzyme characterization, Enterobacteriaceae isolates harboring extended spectrum βeta-Lactamases (ESBL) including CTX-M, SHV, TEM; AmpC; carbapenemases (KPC, NDM, OXA, VIM); aminoglycoside modifying enzymes (AMEs) including AAC, AAD, ANT, APH; 16S rRNA methyltransferases (plazomicin resistant RMTB and armA) were tested with the HardyDisk Plazomicin. Information regarding the performance of Plazomicin with isolates that exhibit overexpression of efflux pumps or lower expression of porins is provided in the following footnote:

o The performance of HardyDisk AST Plazomicin 30μg (PLZ30) is unknown for Enterobacteriaceae with the following resistance mechanisms: overexpression of efflux pumps (e.g., acrAB-tolC) or lower expression of porins (e.g., ompF or ompK36).

# 1. Analytical performance:

a. Precision/Reproducibility:

Not applicable

b. Linearity/assay reportable range:

Not applicable

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Not applicable

d. Detection limit:

Not applicable

e. Analytical specificity:

Not applicable

f. Assay cut-off:

Not applicable

# 2. Comparison studies:

a. Method comparison with predicate device:

Not applicable

b. Matrix comparison:

6

{6}

Not applicable

### 3. Clinical studies:

a. Clinical Sensitivity:

Not applicable

b. Clinical specificity:

Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not applicable

### 4. Clinical cut-off:

Not applicable

### 5. Expected values/Reference range:

The plazomicin interpretative criteria for disk diffusion in the approved pharmaceutical drug package insert is located at the FDA STIC website and shown in Table 2 below.

Table 2: Interpretative Criteria for Plazomicin Disk

|  Indications For Use Organism(s) | Interpretative Criteria  |   |   |
| --- | --- | --- | --- |
|   |  Zone Diameter (mm)  |   |   |
|   |  R | I | S  |
|  Enterobacteriaceae | ≤13 | 14-15 | ≥16  |

The QC isolates and expected ranges are the same as recommended by the current (28 \( ^{th} \) ) edition of CLSI M100.

## N. Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

## O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

7

---

**Source:** [https://fda-staging.innolitics.com/device/K181700](https://fda-staging.innolitics.com/device/K181700)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com