K181689 · Meta Biomed Co., Ltd. · EBF · Nov 15, 2018 · Dental
Device Facts
Record ID
K181689
Device Name
Nexcomp
Applicant
Meta Biomed Co., Ltd.
Product Code
EBF · Dental
Decision Date
Nov 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Indications for Use
Nexcomp is intended to restore carious lesions or structural defects in teeth. It is indicated for Class I, II, III, IV, and V restorations, core-buildup to replace missing tooth structure, for splinting and for diastema closure, direct veneers, composite and porcelain repairs.
Device Story
Nexcomp is a light-cured, resin-based dental restorative material used by dentists for direct placement in anterior and posterior teeth. The material consists of a hybrid blend of ultra-fine filler and 40 nm silica particles in a resin matrix. It is applied to cavity preparations and cured using a light source (420-480nm wavelength). The device provides a radiopaque, esthetic restoration that mimics natural tooth structure. By restoring carious lesions or structural defects, it helps maintain tooth integrity and function. It is available in 19 shades corresponding to the Vita Shade Guide.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1, -5, -10, and -11. Physical property testing performed per ISO 4049 (flexural strength, water sorption, solubility), ISO 6507-1 (Vickers hardness), and ISO 3107 (compressive strength). Results confirm compliance with international standards.
Technological Characteristics
Light-cured resin-based composite. Composition: Bis-GMA, UDMA, Bis-EMA, TEGDMA, TMPTMA, GDMA, Camphorquinone. Fillers: Barium glass (0.78 µm) and silica (16nm). Curing: 420-480nm wavelength, 1000-1200 mW/cm² intensity, 20-second duration. Radiopaque. Complies with ISO 4049 (Type 1, Class 2, Group 1), ISO 6507-1, and ISO 3107.
Indications for Use
Indicated for patients requiring restoration of carious lesions or structural tooth defects, including Class I-V restorations, core-buildup, splinting, diastema closure, direct veneers, and composite/porcelain repairs.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The FDA logo is a recognizable symbol of the agency's role in regulating and overseeing food and drug products in the United States.
Meta Biomed Co., Ltd. % April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620
Re: K181689
Trade/Device Name: Nexcomp Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: October 12, 2018 Received: October 17, 2018
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
November 15, 2018
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Marv S. -Mary S. Runner -S3 Runner -S3 12:44:18 -05'00'
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K181689
Device Name Nexcomp
#### Indications for Use (Describe)
Nexcomp is intended to restore carious lesions or structural defects in teeth. It is indicated for Class I, III, IV, and V restorations, core-buildup to replace missing tooth structure, for splinting and for diastema closure, direct veneers, composite and porcelain repairs.
**Type of Use (Select one or both, as applicable)**> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### Submitter
META BIOMED CO., LTD. Suk Song Oh 270. Osongsaengmyeong 1-ro. Osong-eup, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, Korea Email: ef1459@metabiogw.bizmeka.com Phone: +82-43-230-8841 Fax: +82-43-217-1983
## Device Information
#### ● Trade Name: Nexcomp
- Classification Name: Tooth Shade Resin Material ●
- . Product Code: EBF
- Panel: Dental
- Regulation Number: 21 CFR 872.3690
- Device Class: Class II
- Date prepared: 11/08/2018
#### Predicate Devices:
The subject device is substantially equivalent to the following predicate devices:
- K080305, Nexcomp manufactured by Meta Biomed Co., Ltd. .
#### General Description
Nexcomp is a light-cured, esthetic, resin-based dental restorative designed for direct placement in both anterior and posterior restorations. The material is radiopaque and indicated for all types of cavity preparations. The hybrid blend has an ultra-fine filler and 40 nm silica particles. Nexcomp will be marketed in 19 shades which corresponded with the Vita Shade Guide.
#### Indication for Use
Nexcomp is intended to restore carious lesions or structural defects in teeth. It is indicated for Class I, II, III, IV, and V restorations, core-buildup to replace missing tooth structure, for splinting and for diastema closure, direct veneers, composite and porcelain repairs.
## Official Correspondent
Withus Group Inc April Lee 106 Superior. Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
{4}------------------------------------------------
# Comparison of technological Characteristics with Predicate Devices:
| | Specification | Subject Device | Primary Predicate Device |
|----------------------------------------|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | - | META BIOMED CO., LTD. | META BIOMED CO., LTD. |
| Device Name | - | Nexcomp | Nexcomp |
| 510(k) Number | - | K181689 | K080305 |
| Classification Name | - | Tooth Shade Resin Material | Tooth Shade Resin Material |
| Product Code | - | EBF | EBF |
| Regulation Number | - | 21 CFR 872.3690 | 21 CFR 872.3690 |
| Indications for Use | - | Nexcomp is intended to restore carious lesions or structural<br>defects in teeth. It is indicated for Class I, II, III, IV, and V<br>restorations, core-buildup to replace missing tooth structure, for<br>splinting and for diastema closure, direct veneers, composite<br>and porcelain repairs. | Nexcomp is intended to restore carious lesions or structural<br>defects in teeth. It is indicated for Class I, II, III, IV, and V<br>restorations, core-buildup to replace missing tooth structure, for<br>splinting and for diastema closure, direct veneers, composite<br>and porcelain repairs. |
| Principle of operation | - | Nexcomp is visual light curing polymer based restorative<br>product as suitable for restorations involving occlusal surfaces<br>classed by Type 1, Class 2, Group 1 on ISO 4049. | Nexcomp is visual light curing polymer based restorative<br>product as suitable for restorations involving occlusal surfaces<br>classed by Type 1, Class 2, Group 1 on ISO 4049. |
| General Composition | - | - Bis-Gma[Bispenol A Glycerolate<br>(1glycerol/ Phenol)dimethacrylate]<br>- UDMA[Diurethane dimethacrylate]<br>- Bis-Ema [Bisphenol A ethoxylate dimethacrylate]<br>- TEGDMA[Tri(ethylene glycol)dimethacrylate]<br>- TMPTMA[Trimethylol propane trimethacrylate]<br>- GDMA[Glycerol dimethacrylate]<br>- Camphorquinone | - Bis-Gma[Bispenol A Glycerolate<br>(1glycerol/ Phenol)dimethacrylate]<br>- UDMA[Diurethane dimethacrylate]<br>- Bis-Ema [Bisphenol A ethoxylate dimethacrylate]<br>- TEGDMA[Tri(ethylene glycol)dimethacrylate]<br>- TMPTMA[Trimethylol propane trimethacrylate]<br>- GDMA[Glycerol dimethacrylate]<br>- Camphorquinone |
| Sensitivity of ambient<br>light | Remain<br>physically<br>homogeneous | Physical homogeneity | Physical homogeneity |
| Compressive strength | Should be more<br>than 5MPa. | Average 232.76 MPa | Average 288.68MPa |
| Flexural strength | > 80 MPa | Average 87MPa | 91.82MPa |
| Intensity for curing | - | 1000-1200 mV/cm² | - |
| Wavelength for curing | - | 420-480nm | - |
| Depth of cure | > 1.0 mm | 2.88mm | 4.96 mm |
| Filler particle size<br>distribution | - | Barium glass : 0.78 µm<br>Silica :16nm | - |
| Surface hardness<br>(Vickers hardness) | - | Average 56KHN | Average 39.36KHN |
| Radio-opacity | > 1mm | More than 1mm | More than 1mm |
| Water sorption | < 40 $\mu g/mm^3$ | 2.97 $\mu g/mm^3$ | 13.78 $\mu g/mm^3$ |
| Solubility | < 7.5 $\mu g/mm^3$ | 0.0 $\mu g/mm^3$ | 0.23 $\mu g/mm^3$ |
| Release profile | | This device did not contain fluoride or nitrate ions. | |
| Working time | | This device is a light curing type product.<br>(Working time test is applied for self-cured resin.) | |
| Curing time | - | 20sec | - |
| Setting time | | This device is a light curing type product.<br>(Setting time test is applied for self-cured resin.) | |
| Shade and color<br>stability | - | Consistency | Consistency |
| Shelf Life | - | 3 years | 3 years |
{5}------------------------------------------------
Both subject and predicate devices have indications for use, principle of operation, Sensitivity of ambient light, radio opacity and shelf life. The physical properties of subject and predicate devices such as compressive strength, flexural strength, and solubility are all similar.
As depicted above, the values for the depth of cure, surface hardness and water sorption between the subject device and predicate device are different but the test results are within the range of the criteria of ISO standards such as ISO 4049 and ISO 6507 this difference does not affect the substantial equivalence.
## Non-clinical Testing
The subject device was tested to the following standards:
- Biocompatibility Testing according to ISO 10993-1:2009, ISO 10993-5:2009, ISO 10993--10:2010, and ISO 10993-11:2006.
- Physical properties testing and shelf life test according to ISO 4049:2009 -
- -Surface hardness test according to ISO 6507-1:2005
- Compressive Strength Test according to ISO 3107:2011 -
The data analysed from the biocompatibility evaluation of the subject device, and the results of the biocompatibility tests according to ISO 10993 substantiate that the Nexcomp is substantially equivalent to the predicate device. The subject device is different from the predicate devices in that it contains the same key biocompatible ingredients, but in different proportions. The biocompatibility test report supports this difference and it does not affect the substantial equivalence.
{6}------------------------------------------------
# Conclusion
The Subject Device, Nexcomp has been compared with its predicate, Nexcomp with regard to indications, performance data and chemical composition. The comparison shows that Nexcomp is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.