K181683 · Microbiologics, Inc. · PMN · Aug 28, 2018 · Microbiology
Device Facts
Record ID
K181683
Device Name
BD MAX CT/GC/TV 20-Day QC Panel
Applicant
Microbiologics, Inc.
Product Code
PMN · Microbiology
Decision Date
Aug 28, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3920
Device Class
Class 2
Indications for Use
The BD MAX™ CT/GC/TV 20-Day QC Panel is intended for use as an external assayed positive quality control material to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Chlamydia trachomatis, Neisseria gonorrhoeae and Trichomonas vaginalis with the BD MAX™ CT/GC/TV Assay on the BD MAX™ System. The controls comprise cultured and inactivated C. trachomatis, N. qonorrhoeae and T. vaqinalis. The BD MAX CT/GC/TV™ 20-Day QC Panel is not intended to replace manufacturer controls provided with the device.
Device Story
The BD MAX™ CT/GC/TV 20-Day QC Panel provides external positive quality control for clinical laboratory nucleic acid testing. The device consists of 20 individually packaged lyophilized pellets containing inactivated Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis. Laboratory personnel process the control pellet like a patient sample within the BD MAX™ System. The system performs extraction, amplification, and detection of the target pathogens. The output is a qualitative result used by the healthcare provider to verify the performance of the assay and the instrument. This monitoring ensures the reliability of patient test results, helping clinicians accurately diagnose these infections.
Clinical Evidence
Precision and reproducibility study conducted across three testing sites using six operators and three lots over five days. Testing performed on BD MAX™ instruments using the BD MAX™ CT/GC/TV assay. Results showed high agreement: C. trachomatis 100% (90/90), N. gonorrhoeae 99% (89/90), and T. vaginalis 100% (90/90). Three unresolved results were resolved upon retesting.
Technological Characteristics
Lyophilized pellet containing inactivated C. trachomatis, N. gonorrhoeae, and T. vaginalis. Packaged in 2.0ml micro-centrifuge tubes. Designed for use with the BD MAX™ System for nucleic acid extraction, amplification, and detection. No negative control included.
Indications for Use
Indicated for use as an external assayed positive quality control material to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Chlamydia trachomatis, Neisseria gonorrhoeae, and Trichomonas vaginalis using the BD MAX™ CT/GC/TV Assay on the BD MAX™ System.
Regulatory Classification
Identification
An assayed quality control material for clinical microbiology assays is a device indicated for use in a test system to estimate test precision or to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. This type of device consists of single or multiple microbiological analytes intended for use with either qualitative or quantitative assays.
Special Controls
An assayed quality control material for clinical microbiology assays must comply with the following special controls: (1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate: (i) Analyte concentration; Expected values: (ii) Analyte source: (iii) (iv) Base matrix; (v) Added components; (vi) Safety and handling information; and, (vii) Detailed instructions for use. (2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including: (i) Description of the process for value assignment and validation. (ii) Description of the protocol(s) used to establish stability. (iii) Line data establishing precision/reproducibility. (iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance. (v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method. (vi) Where applicable, detailed documentation related to studies for surrogate controls. (3) Premarket notification submissions must include an adequate mitigation (e.g., realtime stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling. (4) Your 21 CFR 809.10 compliant labeling must include the following: (i) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following: (A) Assayed control material analyte(s); (B) Whether the material is intended for quantitative or qualitative assays: (C) Stating if the material is a surrogate control; (D)The system(s), instrument(s), or test(s) for which the quality control material is intended. (ii) The intended use in your 21 CFR 809.10(a)(2) and 21 CFR 809.10(b)(2) compliant labeling must include the following statement: "This product is not intended to replace manufacturer controls provided with the device." (iii)A limiting statement that reads "Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements."
*Classification.* Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation and information concerning the composition of the quality control material, including, as appropriate:
(i) Analyte concentration;
(ii) Expected values;
(iii) Analyte source;
(iv) Base matrix;
(v) Added components;
(vi) Safety and handling information; and
(vii) Detailed instructions for use.
(2) Premarket notification submissions must include detailed documentation, including line data as well as detailed study protocols and a statistical analysis plan used to establish performance, including:
(i) Description of the process for value assignment and validation.
(ii) Description of the protocol(s) used to establish stability.
(iii) Line data establishing precision/reproducibility.
(iv) Where applicable, assessment of matrix effects and any significant differences between the quality control material and typical patient samples in terms of conditions known to cause analytical error or affect assay performance.
(v) Where applicable, identify or define traceability or relationship to a domestic or international standard reference material and/or method.
(vi) Where applicable, detailed documentation related to studies for surrogate controls.
(3) Premarket notification submissions must include an adequate mitigation (e.g., real-time stability program) to the risk of false results due to potential modifications to the assays specified in the device's 21 CFR 809.10 compliant labeling.
(4) Your 21 CFR 809.10 compliant labeling must include the following:
(i) The intended use of your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following:
(A) Assayed control material analyte(s);
(B) Whether the material is intended for quantitative or qualitative assays;
(C) Stating if the material is a surrogate control; and
(D) The system(s), instrument(s), or test(s) for which the quality control material is intended.
(ii) The intended use in your 21 CFR 809.10(a)(2) and (b)(2) compliant labeling must include the following statement: “This product is not intended to replace manufacturer controls provided with the device.”
(iii) A limiting statement that reads “Quality control materials should be used in accordance with local, state, federal regulations, and accreditation requirements.”
{0}
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
A. 510(k) Number:
K181683
B. Purpose for Submission:
To obtain a substantial equivalence determination for the BD MAX CT/GC/TV 20-Day QC Panel.
C. Measurand:
Nucleic acids from inactivated *Chlamydia trachomatis*, *Neisseria gonorrhoeae*, and *Trichomonas vaginalis*
D. Type of Test:
The BD MAX CT/GC/TV 20-Day QC Panel is an external assayed positive quality control material designed to monitor the performance of *in vitro* laboratory nucleic acid testing procedures for the qualitative detection of *Chlamydia trachomatis*, *Neisseria gonorrhoeae*, and *Trichomonas vaginalis* in genitourinary specimens.
E. Applicant:
Microbiologics, Inc.
F. Proprietary and Established Names:
Trade Name: BD MAX CT/GC/TV 20-Day QC Panel
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3920, Assayed quality control material for clinical microbiology assays
2. Classification:
Class II (Special Controls)
3. Product code:
PMN
1
{1}
4. Panel:
(83) Microbiology
# H. Intended Use:
1. Intended use:
The BD MAX CT/GC/TV 20-Day QC Panel is intended for use as an external assayed positive quality control material to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Chlamydia trachomatis, Neisseria gonorrhoeae and Trichomonas vaginalis performed with the BD MAX CT/GC/TV Assay on the BD MAX System. This product is not intended to replace manufacturer controls provided with the device. The controls comprise cultured and inactivated C. trachomatis, N. gonorrhoeae and T. vaginalis organisms. The BD MAX CT/GC/TV 20-Day QC Panel is not intended to replace manufacturer controls provided with the device.
2. Indication(s) for use:
Same as Intended Use.
3. Special conditions for use statement(s):
For in vitro diagnostic use only
For prescription use only
3. Special instrument requirements:
The BD MAX CT/GC/TV 20-Day QC Panel is intended for use on the BD MAX Instrument.
# I. Device Description:
The BD MAX CT/GC/TV 20-Day QC Panel is a quality control material provided to the customer as 20 individually packaged positive control pellets. Each pellet consists of inactivated Chlamydia trachomatis, Neisseria gonorrhoeae and Trichomonas vaginalis microorganisms. The organisms are purified and suspended in a buffered aqueous matrix (Base Matrix), followed by inactivation by gamma radiation. The organisms are then quantified by a by quantitative PCR assay and spiked into the Base Matrix at predefined target concentrations. Aliquots of the suspension are lyophilized and packaged. Each production lot is tested for purity and is assayed for performance on the BD MAX instrument.
The pellets are contained in a vial and enclosed in a vacuum sealed foil pouch. The user opens the pouch, uncaps the tube and transfers the pellet into the BD MAX UVE Sample
2
{2}
Buffer Tube. The mixture is vortexed and processed like a patient sample according to the test instructions for the BD MAX CT/GC/TV assay. Negative controls are not included in the panel.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Bio-Rad Amplichek II
2. Predicate 510(k) number(s):
DEN150058
3. Comparison with predicate:
| Characteristic | BD MAX™ CT/GC/TV 20-Day QC Panel | Predicate Device – Bio-Rad Amplichek II (DEN 150058) |
| --- | --- | --- |
| Intended Use | The BD MAX CT/GC/TV 20-Day QC Panel is intended for use as an external assayed positive quality control material to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Chlamydia trachomatis, Neisseria gonorrhoeae and Trichomonas vaginalis with the BD MAX CT/GC/TV Assay on the BD MAX System. The controls comprise cultured and inactivated C. trachomatis, N. gonorrhoeae and T. vaginalis. The BD MAX CT/GC/TV 20-Day QC Panel is not intended to replace manufacturer controls provided with the device. | Amplichek II is intended for use as an external assayed quality control material to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative detection of Methicillin Resistant Staphylococcus aureus, Methicillin Sensitive Staphylococcus aureus, Clostridium difficile and Vancomycin-resistant Enterococci performed on Cepheid GeneXpert Systems. This product is not intended to replace manufacturer controls provided with the device. This product is only for use with assays and instruments listed in the Representative Results Chart in this labeling. |
| Physical Format | Lyophilized pellet | Ready-to-use liquid |
| Composition | Inactivated microorganisms | Inactivated microorganisms |
| Analytes | Chlamydia trachomatis Neisseria gonorrhoeae Trichomonas vaginalis | Methicillin Resistant Staphylococcus aureus Methicillin Sensitive Staphylococcus aureus Clostridium difficile Vancomycin-resistant Enterococci |
3
{3}
| Test System | BD MAX System | Cepheid GeneXpert System |
| --- | --- | --- |
| Directions for Use | Process like patient sample | Process like patient sample |
| Assay Steps Monitored | Extraction, amplification, and detection | Extraction, amplification, detection |
| Number of Targets monitored in one assay | Multiple | Multiple |
### K. Standard/Guidance Document Referenced (if applicable):
CLSI. Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline. CLSI Document EP25-A. Wayne, PA: Clinical and Laboratory Standards Institute; 2009.
### L. Test Principle:
Not applicable; this is control material to monitor performance of a test.
### M. Performance Characteristics (if/when applicable):
#### 1. Analytical performance:
##### a. Reproducibility:
The reproducibility of results obtained with the BD MAX 20-Day QC Panel was evaluated in a study performed at three sites over five days, with two operators per site. Three lots of the control material were evaluated by testing with the BD MAX CT/GC/TV Assay on three BD MAX instruments. Each operator tested three lyophilized control pellets on each day of the study (3 sites x 5 days x 2 operators x 3 replicates (lots) = 90 measurements in total). Controls associated with incomplete instrument runs (2) or that produced unresolved results (1) were retested using a new control pellet according to the Instructions For Use. A Negative Control consisting of molecular grade water was also tested at least once a day on each instrument to monitor for contamination. The results of the study are summarized below. All Negative Controls produced the expected results.
| Analyte | Agreement (%) by Test Site/BD MAX System | | | |
| --- | --- | --- | --- | --- |
| | \( Site 1^1 \) | \( Site 2^1 \) | \( Site 3^2 \) | Overall |
| C. trachomatis | 30/30(100%) | 30/30(100%) | 30/30(100%) | 90/90(100%) |
| N. gonorrhoeae | 30/30(100%) | 30/30(100%) | 29/30(96.7%) | 89/90(98.9%) |
| T. vaginalis | 30/30(100%) | 30/30(100%) | 30/30(100%) | 90/90(100%) |
\( ^{1} \) Three Unresolved results were obtained; in all cases a new control was retested and the expected results were obtained
\( ^{2} \) One unresolved result was obtained, no retest was performed.
4
{4}
The calculated average Ct score value, the associated standard deviation, and % coefficient of variation were calculated for each organism in the panel (based on results from two testing sites, the third site did not have software allowing access to the signal generated by the instrument):
| | **C. trachomatis** | **N. gonorrhoeae** | **T. vaginalis** | **IPC** |
| --- | --- | --- | --- | --- |
| **N** | 60 | 60 | 60 | 60 |
| **Mean** | 30.2 | 33.6 | 30.8 | 30.5 |
| **SD** | 0.472 | 1.086 | 0.475 | 0.365 |
| **%CV** | 1.56 | 3.23 | 1.55 | 1.09 |
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
Not applicable
Stability:
1. Shelf life stability was initially determined through an accelerated study with three lots of the controls. The product was placed into chambers at elevated temperatures: 43°C, 53°C and 63°C. Each lot was tested in 4 replicates at each time point: Day 0, Day 14 and Day 42. The stability of the product was evaluated by calculating the mean Ct value at each time point for CT and for NG and comparing that to the mean Ct value obtained at Day 0. The data showed that there was no product degradation under the conditions tested. The change in Ct score, across the three lots ranged from -0.2 to 1.3 for C. trachomatis, from -0.6 to 1.6 for N. gonorrhoeae, and from -0.5 to 0.6 for T. vaginalis. The data is summarized below.
| **C. trachomatis** | | Lot 1 | | | Lot 2 | | | Lot 3 | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | 43°C | 53°C | 63°C | 43°C | 53°C | 63°C | 43°C | 53°C | 63°C |
| | Day | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score |
| Mean | 0 | 30.1 | 30.1 | 30.1 | 30.1 | 30.1 | 30.1 | 29.8 | 29.8 | 29.8 |
| Mean Change in Ct score | 14 | -0.1 | 0.8 | 0.5 | 0.3 | 0.4 | 1.0 | 0.6 | 1.3 | 1.0 |
| | 28 | 1.1 | -0.2 | 1.0 | 0.2 | 0.0 | 0.8 | 0.8 | 0.5 | 1.1 |
| | 42 | 0.6 | 0.1 | 0.2 | 0.9 | 0.3 | 0.6 | 0.6 | 0.4 | 0.5 |
5
{5}
| N. gonorrhoeae | | Lot 1 | | | Lot 2 | | | Lot 3 | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | 43°C | 53°C | 63°C | 43°C | 53°C | 63°C | 43°C | 53°C | 63°C |
| | Day | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score |
| Mean | 0 | 34.0 | 34.0 | 34.0 | 34.0 | 34.0 | 34.0 | 33.8 | 33.8 | 33.8 |
| Mean Change in Ct score | 14 | 0.0 | 0.0 | -0.2 | 0.6 | 0.4 | 0.2 | 1.3 | 0.7 | 0.4 |
| | 28 | 0.7 | 0.3 | -0.4 | 0.2 | 0.7 | 0.8 | 0.1 | 0.2 | -0.4 |
| | 42 | 0.6 | -0.3 | -0.3 | 1.6 | 0.5 | 0.9 | 0.4 | 0.8 | -0.6 |
| T. vaginalis | | Lot 1 | | | Lot 2 | | | Lot 3 | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | 43°C | 53°C | 63°C | 43°C | 53°C | 63°C | 43°C | 53°C | 63°C |
| | Day | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score | Ct score |
| Mean | 0 | 31.3 | 31.3 | 31.3 | 31.0 | 31.0 | 31.0 | 30.7 | 30.7 | 30.7 |
| Mean Change in Ct score | 14 | -0.3 | 0.3 | -0.1 | 0.5 | 0.2 | 0.4 | 0.5 | 0.6 | 0.1 |
| | 28 | 0.4 | 0.3 | 0.3 | 0.2 | 0.3 | 0.4 | 0.2 | 0.6 | 0.3 |
| | 42 | 0.0 | -0.2 | 0.0 | 0.4 | 0.1 | -0.5 | 0.6 | 0.6 | -0.2 |
A real-time shelf life stability study is in process with a goal of 24 months stability claim.
2. An in-use stability of the dissolved pellets was evaluated by keeping the hydrated pellets at room temperature (23°C) for six hours. The testing was conducted with 4 replicates at 3 and at 6 hours. The calculated % CV (based on Ct values) across all three analytes ranged from 0.16% to 1.95%, demonstrating no degradation of the product once hydrated for up to 5 hours at ambient laboratory temperature.
### Expected Values:
The BD MAX CT/GC/TV 20-Day QC Panel is a qualitative positive control expected to produce positive results for C. trachomatis, N. gonorrhoeae, and T. vaginalis.
d. Detection limit:
Not applicable.
e. Analytical Specificity:
Not applicable.
f. Assay cut-off:
Not applicable.
7
{6}
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable.
b. Clinical specificity:
Not applicable.
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable.
4. Clinical cut-off:
Not Applicable.
# N. Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
# O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.