K181669 · Brightwater Medical · FGE · Mar 14, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K181669
Device Name
ConvertX Biliary Stent System
Applicant
Brightwater Medical
Product Code
FGE · Gastroenterology, Urology
Decision Date
Mar 14, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ConvertX Biliary Stent System with releasable drainage catheter is delivered percutaneously and is intended for internal drainage of the bile ducts, for splinting of the bile ducts during healing, or for providing bile duct patency in a stricture or past a stone while maintaining external access to the stent, as well as providing external drainage. For patients for whom external drainage is not, or no longer desirable, the releasable drainage catheter may be removed, leaving the stent to provide internal drainage from the bile duct to the duodenum.
Device Story
Percutaneous biliary stent system; provides internal drainage, duct splinting, or stricture/stone patency; enables optional external drainage. Components: double pigtail biliary stent, releasable drainage catheter with hub. Hub contains pull tabs for loop suture and release wire; allows detachment of catheter from stent. Radiopaque throughout. Used in clinical settings by physicians; inserted via percutaneous access. Stent remains in situ after catheter removal. Benefits: allows transition from external to internal drainage without secondary intervention; maintains duct patency; facilitates healing.
Clinical Evidence
Bench testing only. Performance verified via ASTM F1828-97 (ureteral stents), ASTM F2129-17 (corrosion), ASTM F623-99 (Foley catheters), ISO 10555-1 (catheters), and ISO 10993 (biocompatibility). MRI compatibility assessed per ASTM F2182-11a, F2119-07, F2052-15, and F2503-13. Testing included stent/catheter dimensions, loop recovery, flow rates, tensile strength, kink resistance, and simulated use. All criteria met.
Technological Characteristics
Sterile, single-use, non-vascular biliary stent system. Materials: biocompatible polymers/metals (ISO 10993). Features: double pigtail design, releasable drainage catheter, hub with pull tabs for suture/wire release. Radiopaque. Sterilization: EtO. Standards: ASTM F1828-97, F2129-17, F623-99, F640-12; ISO 10555-1. Mechanical operation: percutaneous insertion, suture-based loop closure, manual detachment of catheter.
Indications for Use
Indicated for patients requiring internal bile duct drainage, splinting during healing, or maintenance of patency in strictures or past stones. Allows for optional external drainage; catheter is removable when external drainage is no longer required.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Boston Scientific Advanix™ Double Pigtail (K101786)
Boston Scientific Expel™ Drainage Catheter (K141335)
Reference Devices
BrightWater Medical ConvertX Nephroureteral Stent System (K161277)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 14, 2019
BrightWater Medical Bob Smouse, MD CEO/CMO 42580 Rio Nedo Road Temecula, CA 92590
Re: K181669
> Trade/Device Name: ConvertX Biliary Stent System Regulation Number: 21 CFR§ 876.5010 Regulation Name: Biliary Catheter and Accessories Regulatory Class: II Product Code: FGE Dated: February 11, 2019 Received: February 12, 2019
Dear Bob Smouse:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Daniel G. Walter Jr -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K181669
Device Name ConvertX Biliary Stent System
#### Indications for Use (Describe)
The ConvertX Biliary Stent System with releasable drainage catheter is delivered percutaneously and is intended for internal drainage of the bile ducts, for splinting of the bile ducts during healing, or for providing bile duct patency in a stricture or past a stone while maintaining external access to the stent, as well as providing external drainage. For patients for whom external drainage is not, or no longer desirable, the releasable drainage catheter may be removed, leaving the stent to provide internal drainage from the bile duct to the duodenum.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for BrightWater Medical. The logo consists of a stylized water droplet made up of three curved lines in teal, light green, and yellow. To the right of the droplet are the words "BrightWater" in a teal sans-serif font, with the word "Medical" in a smaller font size below it.
# 510(K) SUMMARY
In accordance with 21 CFR §807.92 the following summary of information is provided:
| Submitted By | BrightWater Medical<br>42580 Rio Nedo Rd<br>Temecula, CA 92590<br>U.S.A. |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Bob Smouse, M.D.<br>Chief Executive Officer / Chief Medical Officer<br>Phone: 1-650-210-8100<br>Email: bsmouse@brightwatermed.com |
| Date Prepared | June 22, 2018 |
| Proprietary Name | ConvertX™ Biliary Stent System |
| Common Name | Biliary Stent |
| Classification | Class II |
| Regulation Number | 21 CFR 876.5010 |
| Regulation Name | Biliary catheter and accessories |
| Classification Panel | Gastroenterology/Urology |
| Product Code | FGE |
| Predicate Devices | Boston Scientific Advanix™ Double Pigtail (K101786), Boston Scientific Expel™ Drainage Catheter (K141335) |
| Reference Device | BrightWater Medical ConvertX Nephroureteral Stent System (K161277) |
| Intended Use /<br>Indications for Use | The ConvertX Biliary Stent System with releasable drainage catheter is delivered percutaneously and is intended for internal drainage of the bile ducts, for splinting of the bile ducts during healing, or for providing bile duct patency in a stricture or past a stone while maintaining external access to the stent, as well as providing external drainage. For patients for whom external drainage is not, or no longer desirable, the releasable drainage catheter may be removed, leaving the stent to provide internal drainage from the bile duct to the duodenum. |
{4}------------------------------------------------
# Device Description
The ConvertX Biliary Stent System is a sterile, single-use non-vascular interventional device inserted using percutaneous access techniques to re-establish internal drainage and is intended for internal drainage of the bile ducts, for splinting of the bile ducts during healing, or for providing bile duct patency in a stricture or past a stone. The ConvertX Biliary System allows for external drainage and access as needed.
The ConvertX Biliary Stent System has two primary components: a double pigtail Biliary Stent and a Releasable Drainage Catheter with Hub.
The ConvertX Biliary Stent System is a single lumen device that has a detachable Biliary Stent with proximal and distal loops (pigtails). Both proximal and distal loops (pigtails) contain drainage holes. The proximal pigtail forms in the bile duct, while the distal pigtail forms in the duodenum. The distal pigtail has a tapered, atraumatic tip. Drainage can occur internally (from the bile duct to the duodenum) or externally (from the bile duct to the outside of the patient). A suture ("Loop Suture") is threaded through the proximal pigtail enabling the pigtail, straightened for delivery, to be closed in the bile duct.
The ConvertX Biliary Stent System's Hub has three ports which are used for both ends of the Loop Suture, external drainage of bile and the release wire. Ends of the Loop Suture and the Release Wire are connected to separate pull tabs within the Hub. The Releasable Drainage Catheter and Loop Suture may be detached from the Biliary Stent when external drainage is no longer necessary, leaving the stent in place. While still attached, the Releasable Drainage Catheter may be used to withdraw the Biliary Stent. The ConvertX Biliary Stent System is radiopaque throughout its length and has a marker on the stent between junction and proximal loop
The ConvertX Biliary Stent System is comprised of the following functional components:
- . Biliary Stent
- . Releasable Drainage Catheter with Hub
The ConvertX Biliary Stent System is supplied with the following accessories:
- . Metal Stiffening Cannula
- . Luer Cap
- . Plastic Stiffening Cannula
- . Loop Straightener
## Summary of Non-Clinical Testing
Design verification / performance testing for the ConvertX Biliary Stent System was completed in accordance with applicable sections of ASTM F1828-97 - Standard Specification for Ureteral Stents (using simulated bile), ASTM F2129-17- "Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices": X2.5 Simulated Bile (Human Simulated Bile).
{5}------------------------------------------------
ASTM F623-99 - Standard Performance Specification for Foley Catheters, ISO 10555-1: 2013 -Intravascular catheters - Sterile and single-use catheters, Part 1: General requirements, and the FDA Guidance for the Content of Premarket Notifications for Ureteral Stents. Biological testing was performed in accordance with applicable sections of ISO 10993 – Biological Evaluation of Medical Devices. MRI Compatibility was evaluated according to ASTM F2182-11a - Standard Test Method for Measurement of Radio Frequency Heating On or Near Passive Implants During Magnetic Resonance Imaging, ASTM F2119-07 - Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants, ASTM F2052-15 - Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment, and ASTM F2503-13 - Standard Practice for Marking Medical Devices and Other Items for Safety in Magnetic Resonance Environment. Radiopacity of the predicate ConvertX Nephroureteral Stent System was evaluated in accordance with ASTM F640-12 - Standard Test Methods for Determining Radiopacity for Medical Use.
- Stent / Catheter Dimensions
- · Stiffener Dimensions
- · Straightener to Stent Compatibility
- Guidewire Compatibility
- · Loop Recovery
- Flow Recovery Post Kinking
- · Stent-Catheter Interface Strength
- · Kink Resistance
- · Stent Shaft Tensile Strength
- Hub to Shaft Tensile Strength
- Hub to Stiffener Strength
- · Stiffener Tip Strength
- · Loop Retention / Removal Force
- Pull Wire to Pull Tab Tensile Strength
- Loop Suture to Pull Tab Tensile Strength
- · Frictional Force
- · Loop Suture Removal Force
- Lock Suture to Hub Tensile Strength
- · Decoupling Force
- · Resistance to Deformation
- · Bile Compatibility Stent
- · Bile Compatibility- System
- Resistance to Liquid Leakage Under Pressure and During Aspiration/ Vacuum
- · Flow Rate
- · Shelf Life
- · MRI Compatibility Conditional
- · Sterilization
- · Biocompatibility
- · Radiopacity
- Pouch Seal Strength
- · Corrosion Testing
- · Simulated Use Testing
The Convert X Biliary Stent System, including its packaging, met the predetermined acceptance criteria. No new safety or performance issues were raised during testing.
## Predicate Device Comparison
| Performance /<br>Technological<br>Characteristic | Convert X Biliary<br>Stent System<br>Submitted | Boston<br>Scientific<br>Advanix™<br>Double Pigtail<br>K101786 | Boston Scientific<br>Expel Drainage<br>Catheter<br>K141335 | ConvertX<br>Nephroureteral<br>Stent System<br>K161277<br>(Reference Device) |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The ConvertX Biliary<br>Stent System with | Advanix Double<br>Pigtail Biliary | The drainage<br>catheter is intended | The ConvertX<br>Nephroureteral Stent |
| | releasable drainage<br>catheter is delivered<br>percutaneously and is<br>intended for internal<br>drainage of bile ducts,<br>for splinting of the bile<br>ducts during healing,<br>or for providing bile<br>duct patency in a<br>stricture or past a stone<br>while maintaining<br>external access to the<br>stent, as well as<br>providing external<br>drainage.<br>For patients for whom<br>external drainage is<br>not, or no longer<br>desirable, the<br>releasable drainage<br>catheter may be<br>removed, leaving the<br>stent to provide<br>internal drainage from<br>the bile duct to the<br>duodenum | Stent is intended<br>for drainage of the<br>bile ducts, for<br>splinting of a bile<br>duct during<br>healing, or for<br>providing bile<br>duct patency in a<br>stricture or past a<br>stone. | to provide<br>percutaneous<br>drainage of abscess<br>fluid and biliary<br>collections | System with<br>releasable drainage<br>catheter is delivered<br>percutaneously and is<br>intended to establish<br>internal drainage from<br>the ureteropelvic<br>junction to the bladder<br>while maintaining<br>external access to the<br>stent, as well as<br>providing external<br>drainage.<br>For patients for whom<br>external drainage is<br>not, or no longer<br>desirable, the<br>releasable drainage<br>catheter may be<br>removed, leaving the<br>stent to provide<br>internal drainage from<br>the ureteropelvic<br>junction to the<br>bladder. |
| Common Name | Biliary Stent | Biliary Stent | Biliary Catheter | Ureteral Stent |
| Classification | Class II, Performance<br>Standards | Class II,<br>Performance<br>Standards | Class II,<br>Performance<br>Standards | Class II, Performance<br>Standards |
| Regulation<br>Number | 21 CFR 876.5010 | 21 CFR 876.5010 | 21 CFR 876.5010 | 21 CFR 876.5010 |
| Product Code | FGE | FGE | FGE | FAD |
| Single Use Only | Yes | Yes | Yes | Yes |
| Provided Sterile | Yes | Yes | Yes | Yes |
| Sterilization<br>Method | EtO | EtO | EtO | EtO |
| Diameters | 10.3 F | 7, 8.5 & 10F | 8F to 14F | 10.3F |
| Stent Lengths | 5,7,10, and 15 cm | 3 cm to 15 cm | 20 to 40 cm | 20 to 28 cm |
| Guidewire<br>Compatibility | 0.038 | | | |
{6}------------------------------------------------
{7}------------------------------------------------
| Radiopaque | Yes | Yes | Yes | Yes |
|----------------------------------|-----------------------------------------------------------------------------------------------|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| MRI Compatible | Yes | Yes | Yes | Yes |
| Bile Compatible | Yes | Yes | Yes | Yes |
| Maximum<br>Indwelling Time | Catheter: Not to<br>exceed 30-days<br>Stent: Not to exceed<br>90-days | Not to exceed 90-<br>days | Not to exceed 90-<br>days | Not to exceed 30-days |
| Delivery Method | Percutaneous | Percutaneous | Percutaneous | Percutaneous |
| Method of<br>conversion to stent | Disconnect | N/A | N/A | Disconnect |
| Biocompatibility | Meets requirements of<br>10993-1 | Meets<br>requirements of<br>10993-1 | Meets requirements<br>of 10993-1 | Meets requirements of<br>10993-1 |
| Accessory Devices | Metal Stiffening<br>Cannula<br>Plastic Stiffening<br>Cannula<br>Loop Straightener<br>Luer Cap | NaviFlex™RX<br>Delivery System | Cannulas<br>Trocars<br>Connecting Tubes<br>Plugs / Caps<br>Pigtail Straightener<br>Facial Dilators<br>Guidewires<br>Dressing<br>Catheter Cuff<br>Introducers | Stiffening Cannula<br>Luer Cap |
## Conclusion
The ConvertX™ Biliary Stent System has the same intended use and similar technological characteristics as the Advanix Double Pigtail Stent System and Expel Drainage Catheter marketed by Boston Scientific. Moreover, performance testing contained in this submission demonstrate that any differences in their technological characteristics do not raise any new questions of safety or effectiveness. Thus, the ConvertX Biliary Stent System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.