easyLCTM Linear Cutter Staplers and Loading Units fot Single Use; easyCSTM Circular Stapler for Single Use
K181657 · Ezisurg Medical Co., Ltd. · GDW · Feb 1, 2019 · General, Plastic Surgery
Device Facts
Record ID
K181657
Device Name
easyLCTM Linear Cutter Staplers and Loading Units fot Single Use; easyCSTM Circular Stapler for Single Use
Applicant
Ezisurg Medical Co., Ltd.
Product Code
GDW · General, Plastic Surgery
Decision Date
Feb 1, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The easyLC™ Linear Cutter Staplers and Loading Units for Single Use have application in gastrointestinal for transection, resection, and/or creation of anastomoses. The easyCS™ Circular Stapler for Single Use has applications throughout the alimentary for end to end, end to side and side to side anastomoses.
Device Story
Manual surgical staplers for gastrointestinal/alimentary procedures; linear cutter and circular stapler variants. Linear cutter places two double staggered rows of titanium staples while simultaneously cutting tissue; available in 60mm/80mm lengths with 2.5mm/3.8mm/4.8mm staple sizes; reloadable up to 8 times. Circular stapler places a double staggered circular row of titanium staples and uses a circular knife to resect excess tissue; available in 21mm-33mm diameters with 3.5mm/4.8mm staple sizes. Used by surgeons in clinical settings to facilitate tissue joining and resection. Devices provide mechanical staple formation and tissue division; benefit includes standardized, efficient anastomosis creation. Output is physical tissue closure and resection.
Clinical Evidence
No clinical data. Bench testing only. Tests included pressure resistance, closed staple dimension, staple formation, and firing force on porcine stomach and intestine. Biocompatibility testing (cytotoxicity, irritation, sensitization, pyrogenicity) performed.
Technological Characteristics
Manual surgical staplers; titanium staples. Linear cutter: 60-80mm length, 2.5-4.8mm staple sizes. Circular stapler: 21-33mm diameter, 3.5-4.8mm staple sizes. Standards: ISO 10993-1/5/10 (biocompatibility), ISO 11137-2 (radiation sterilization), ASTM F 88/F88M-09 (seal strength), USP <85>/<151> (endotoxins/pyrogens).
Indications for Use
Indicated for patients requiring gastrointestinal or alimentary tract surgery involving transection, resection, or creation of anastomoses (end-to-end, end-to-side, side-to-side).
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
ENDOPATH™ and PROXIMATE™ Linear Cutters and Staplers (K020779)
PROXIMATE Curved and Straight Intraluminal Staplers (K983536)
PROXIMATE Curved and Straight Intraluminal Staplers (K100723)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, with the letters "FDA" in a blue box. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 1, 2019
Ezisurg Medical Co., Ltd. % Ms. Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, CN 200120
Re: K181657
Trade/Device Name: easyLCTM Linear Cutter Staplers and Loading Units for Single Use; easyCS™ Circular Stapler for Single Use
Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW, GAG Dated: June 12, 2018 Received: June 22, 2018
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
$$\textbf{David }\textbf{Krausse}\text{-}\textbf{S}_{\begin{subarray}{c}\textbf{S}_{\begin{subarray}{c}\textbf{Diag}\\0\le\textbf{W}\end{subarray}}\text{-}\textbf{S}_{\begin{subarray}{c}\textbf{S}_{\begin{subarray}{c}\textbf{Diag}\\0\le\textbf{W}\end{subarray}}\text{-}\textbf{M}\end{subarray}}$$
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K181657
## Device Name
easyLCTM Linear Cutter Staplers and Loading Units for Single Use, easyCSTM Circular Stapler for Single Use
## Indications for Use (Describe)
The easyLCTM Linear Cutter Staplers and Loading Units for Single Use have application in gastrointestinal for transection, resection, and/or creation of anastomoses.
The easyCSTM Circular Stapler for Single Use has applications throughout the alimentary for end to side and side to side anastomoses.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <span> <span style="padding-right: 20px;"> <b> <span style="font-size: 14px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </b></span> </span> <span> <span style="font-size: 14px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{3}------------------------------------------------
# Exhibit #2 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K181657
- 1. Date of Preparation: 12/29/2018
- 2. Sponsor Identification
## Ezisurg Medical Co., Ltd.
Rm.103, Bldg.2,No.1690 Cailun Road, China (Shanghai) Pilot Free Trade Zone, 201203 Shanghai, China.
Establishment Registration Number: Not yet registered.
Contact Person: Liang Xia Position: Non-active Device Senior Regulatory Affairs Supervisor Tel: +86-21-50456176 Fax: +86-21-50676156 Email: xialiang@ezisurg.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 3609253199 Email: info@mid-link.net
{4}------------------------------------------------
- Identification of Proposed Device 4.
Trade Name: easyLCTM Linear Cutter Staplers and Loading Units for Single Use easyCSTM Circular Stapler for Single Use Common Name: Stapler and Reload
Regulatory Information
Classification Name: Staple, Implantable Classification: II Product Code: GDW Regulation Number: 21 CFR 878.4750 Review Panel: General & Plastic Surgery
Classification Name: Stapler, Surgical; Classification: I Subsequent Product Code: GAG; Regulation Number: 21CFR 878.4800 Review Panel: General & Plastic Surgery
Intended Use Statement:
The easyLC™ Linear Cutter Staplers and Loading Units for Single Use have application in gastrointestinal for transection, resection, and/or creation of anastomoses.
The easyCS™ Circular Stapler for Single Use has applications throughout the alimentary for end to end, end to side and side to side anastomoses.
Device Description
The easyLCTM Linear Cutter Staplers and Loading Units for Single Use place two double staggered rows of titanium staples and simultaneously cut and divides tissue between the two double rows. The device is available in 60mm, 80mm length with reload in 2.5mm, 3.8mm and 4.8mm staple size to accommodate various tissue thicknesses. The device may be reloaded and fired up to 8 times in a single procedure.
The easyCS™ Circular Stapler for Single Use places a double staggered, circular row of titanium staples upon activation, which was achieved by squeezing the handles firmly as far as they could go. Immediately after formation of the staples, the excess tissue will be resect by the circular knife, and then a circular anastomosis is created. The stapler is available in 21mm, 29mm, 31mm and 33mm
{5}------------------------------------------------
five specifications. Two staple sizes (3.5mm and 4.8mm) are available to accommodate various tissue thicknesses.
- ર . Identification of Predicate Devices
Predicate Device 1 510(k) Number: K020779 Product Name: ENDOPATH™ and PROXIMATE™ Linear Cutters and Staplers
Predicate Device 2 510(k) Number: K983536 Device Name: PROXIMATE Curved and Straight Intraluminal Staplers
Predicate Device 3 510(k) Number: K100723 Device Name: PROXIMATE Curved and Straight Intraluminal Staplers
- Non-Clinical Test Conclusion 6.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process.
- > ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity.
- > ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
- > USP 38-NF 33 <85> Bacterial Endotoxins Tests
- > ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materials;
- > ISO 11137-2:2013 Sterilization of health care products -Radiation- Part 2: Establishing the sterilization dose
- > USP 39-NF 34 <151> Pyrogen
Bench test was conducted on porcine stomach and intestine for both proposed device and predicate device to determine substantially equivalence. The bench tests include following tests
- > Pressure Resistance Test
- > Closed Staple Dimension Test
- > Staple Formation Test
{6}------------------------------------------------
> Force Required to Fire Stapler Test
Biocompatibility test was conducted on the proposed device, the test include cytotoxicity, irritation, skin sensitization and pyrogenicity.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
- 8. Substantially Equivalent (SE) Comparison
| Item | Proposed Device | Predicate Device 1 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GDW | GDW |
| Regulation Number | 21 CFR 878.4750 | 21 CFR 878.4750 |
| Intended Use | The easyLCTM Linear Cutter Staplers<br>and Loading Units for Single Use have<br>application in gastrointestinal for<br>transection, resection, and/or creation of<br>anastomoses. | The PROXIMATE Linear and Vascular<br>Linear Cutters with Safety Lockout<br>have application in gastrointestinal,<br>gynecologic, thoracic and pediatric<br>surgery for transection, resection, and/or<br>creation of anastomoses and can be used<br>with staple line or tissue buttressing<br>materials, such as bovine pericardium |
| Cutting Mechanism | Linear | Linear |
| Operation Principle | Manual | Manual |
| Safety Mechanism | Safety release for prevent from<br>mis-firing | Safety release for prevent from<br>mis-firing |
| Suture Length | 60mm, 80mm, 100mm | 55, 75mm |
| Closed Staple Height | 1.0mm, 1.5mm, 2.0mm | 1.0mm, 1.5mm, 1.8mm, 2.0mm |
| Closed staple form | Image: staple form | Image: staple form |
| Endotoxin Limit | 20 EU | Same |
| Labeling | Conforms with 21 CFR 801 | Conforms with 21 CFR 801 |
Table 1 Comparison for easyLC™ Linear Cutter Staplers and Loading Units for Single Use
{7}------------------------------------------------
| Item | Proposed Device | Predicate Device 2<br>K983536 | Predicate Device 3<br>K100723 |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GDW | GDW | GDW |
| Regulation Number | 21 CFR 878.4750 | 21 CFR 878.4750 | 21 CFR 878.4750 |
| Intended Use | The easyCSTM Circular Stapler for<br>Single Use has applications throughout<br>the alimentary for end to end, end to<br>side and side to side anastomoses. | The PROXIMATER Curved and<br>Straight Intraluminal Staplers have<br>application throughout the alimentary<br>for end to end, end to side and side to<br>side anastomoses. | The Disposable Circular Stapler has<br>application throughout the alimentary<br>tract for end-to-end, end-to-side anc<br>side-to-side anastomoses. |
| Cutting Mechanism | Circular Knife | Circular Knife | Circular Knife |
| Operation Principle | Manual | Manual | Manual |
| Safety Mechanism | Safety Release is used for preventing<br>from mis-firing. | Safety Release is used for preventing<br>from mis-firing. | Safety Release is used for preventing<br>from mis-firing. |
| Diameter | 21mm, 25mm, 29mm, 31mm, 33mm | 21mm, 25mm, 29mm, 33mm | 25.5mm, 29.5mm |
| Closed Staple Height | 1.0~2.5mm | 1.0~2.5mm | 1.5~2.0mm |
| Closed staple form | Image: staple form | Image: staple form | Image: staple form |
| Endotoxin Limit | 20 EU | 20EU | 20EU |
| Labeling | Conforms with 21 CFR 801 | Conforms with 21 CFR 801 | Conforms with 21 CFR 801 |
| | Table 2 Comparison for easy CSTM Circular Stapler for Single Use |
|--|------------------------------------------------------------------|
|--|------------------------------------------------------------------|
{8}------------------------------------------------
- 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.