K181646 · Canon Medical Systems Corporation · JAK · Nov 16, 2018 · Radiology
Device Facts
Record ID
K181646
Device Name
Celesteion, PCA-9000A/3, V6.5
Applicant
Canon Medical Systems Corporation
Product Code
JAK · Radiology
Decision Date
Nov 16, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
The device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT component produces crosssectional images of the body by computer reconstruction of x-ray transmission data. The PET component images the distribution of PET radiopharmaceuticals in the patient body. The PET component utilizes CT images for attenuation correction and anatomical reference in the fused PET and CT images. This device is to be used by a trained health care professional to gather metabolic and functional information from the distribution of the radiopharmaceutical in the body for the assessment of metabolic and physiologic functions. This information can assist in the evaluation, detection, diagnosis, therapeutic planning and therapeutic outcome assessment of (but not limited to) cancer, cardiovascular disease and brain dysfunction. Additionally, this device can be operated independently as a whole body multi-slice CT scanner.
Device Story
Celesteion, PCA-9000A/3, V6.5 is a large-bore, time-of-flight (TOF) PET-CT system; combines multi-slice helical CT scanner with high-throughput PET detector (450 ps temporal resolution). Inputs: X-ray transmission data for CT; positron-emitting radiopharmaceuticals for PET. PET component uses CT data for attenuation correction and anatomical fusion. Operated by trained healthcare professionals in clinical settings. Features include respiratory gating, PET ECG gating, Variable Bed Time (vBT), and Clear Adaptive Low-noise Method (CaLM) reconstruction. Output: cross-sectional and fused metabolic/functional images. Clinicians use images for diagnosis, therapeutic planning, and outcome assessment of cancer, cardiovascular disease, and brain dysfunction. Benefits include improved workflow, reduced image noise, and preserved detail/contrast.
Clinical Evidence
Bench testing only. Phantoms used to validate CaLM reconstruction, demonstrating reduced noise with preserved detail and contrast. Representative clinical images acquired to demonstrate functionality of PET ECG gating and new features.
Technological Characteristics
Large-bore PET-CT system; 900 mm gantry aperture; 700 mm scan field. TOF PET detector (450 ps resolution). Multi-slice helical CT. Connectivity: networked. Standards: IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-28, IEC 60601-2-44, IEC 60825-1, IEC 62304, IEC 62366, NEMA PS 3.1-3.18, NEMA XR-25, NEMA XR-26, NEMA NU-2. Radiation safety: 21 CFR 1010/1020.
Indications for Use
Indicated for use by trained healthcare professionals to gather metabolic and functional information via PET radiopharmaceutical distribution and CT imaging for assessment of metabolic/physiologic functions. Assists in evaluation, detection, diagnosis, therapeutic planning, and outcome assessment for conditions including cancer, cardiovascular disease, and brain dysfunction. Can operate as a standalone whole-body multi-slice CT scanner.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
November 16, 2018
Canon Medical Systems Corporation % Orlando Tadeo Sr. Manager, Regulatory Affairs 2441 Michelle Drive TUSTIN CA 92780
Re: K181646
Trade/Device Name: Celesteion, PCA-9000A/3, V6.5 Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: Class II Product Code: KPS, JAK Dated: October 17, 2018 Received: October 18, 2018
Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hol 2. Nils
for Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
K 181646
510(k) Number (if known)
Device Name
Celesteion, PCA-9000A/3, v6.5
#### Indications for Use (Describe)
The device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT component produces crosssectional images of the body by computer reconstruction of x-ray transmission data. The PET component images the distribution of PET radiopharmaceuticals in the patient body. The PET component utilizes CT images for attenuation correction and anatomical reference in the fused PET and CT images.
This device is to be used by a trained health care professional to gather metabolic and functional information from the distribution of the radiopharmaceutical in the body for the assessment of metabolic and physiologic functions. This information can assist in the evaluation, detection, diagnosis, therapeutic planning and therapeutic outcome assessment of (but not limited to) cancer, cardiovascular disease and brain dysfunction. Additionally, this device can be operated independently as a whole body multi-slice CT scanner.
Type of Use (Select one or both, as applicable)
ك Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
1.
- SUBMITTER'S NAME: Canon Medical Systems 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
## 2. OFFICIAL CORRESPONDENT: Naofumi Watanabe Senior Manager, Regulatory Affairs and Vigilance
## 3. ESTABLISHMENT REGISTRATION: 9614698
## 4. CONTACT PERSON:
Orlando Tadeo, Jr. Sr. Manager, Regulatory Affairs Canon Medical Systems USA 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459
#### 5. Date Prepared:
June 21, 2018
#### 6. TRADE NAME(S):
Celesteion, PCA-9000A/3, V6.5
#### 7. COMMON NAME:
System, X-ray, Computed Tomography System, Emission Computed Tomography
#### 8. DEVICE CLASSIFICATION:
Class II (per 21 CFR 892.1750, Computed Tomography X-ray System and 21 CFR §892.1200, Emission Computed Tomography System)
PHONE: 800-421-1968 2441 Michelle Drive, Tustin, CA 92780
{4}------------------------------------------------
#### 9. PRODUCT CODE / DESCRIPTION:
90JAK / Computed Tomography X-Ray System 90KPS / Emission Computed Tomography System
#### 10. PERFORMANCE STANDARD:
This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]
### 11. PREDICATE DEVICE:
| Product | Marketed<br>by | Regulation<br>Number | Regulation<br>Name | Product<br>Code | 510(k)<br>Number | Clearance<br>Date |
|-------------------------------------|------------------------------------|----------------------|----------------------------------------------|-----------------|------------------|-------------------|
| Celesteion,<br>PCA-9000A/3,<br>v6.4 | Canon<br>Medical<br>Systems<br>USA | 21 CFR<br>892.1200 | Emission<br>Computed<br>Tomography<br>System | KPS | K170490 | June 1, 2017 |
## 12. REASON FOR SUBMISSION:
Modification to existing device
## 13. DEVICE DESCRIPTION:
Celesteion, PCA-9000A/3, V6.5, is a large bore, TOF, PET-CT system, which combines a high-end CT system with a high-throughput PET system. The high-end CT system is a multi-slice helical CT scanner with a gantry aperture of 900 mm and a maximum scanning field of 700 mm. The highthroughput PET system has a time of flight (TOF) detector with temporal resolution of 450 ps. Celesteion, PCA-9000A/3, V6.5 is intended to acquire PET images of any desired region of the whole body and CT images of the same region (to be used for attenuation correction or image fusion), to detect the location of positron emitting radiopharmaceuticals in the obtained images. This device is used to gather the metabolic and functional information from the distribution of radiopharmaceuticals in the body for the assessment of metabolic and physiologic functions. This information can assist research, diagnosis, therapeutic planning, and therapeutic outcome assessment. This device can also function independently as a whole body multi-slice CT scanner.
## 14. INDICATIONS FOR USE:
The device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and Xray Computed Tomography (CT) systems. The CT component produces cross- sectional images of the body by computer reconstruction of x-ray transmission data. The PET component images the distribution of PET radiopharmaceuticals in the patient body. The PET component utilizes CT images for attenuation correction and anatomical reference in the fused PET and CT images.
This device is to be used by a trained health care professional to gather metabolic and functional information from the distribution of the radiopharmaceutical in the body for the assessment of metabolic and physiologic functions. This information can assist in the evaluation, detection,
{5}------------------------------------------------
diagnosis, therapeutic planning and therapeutic outcome assessment of (but not limited to) cancer, cardiovascular disease and brain dysfunction. Additionally, this device can be operated independently as a whole body multi-slice CT scanner.
## 15. SUBSTANTIAL EQUIVALENCE:
Celesteion, PCA-9000A/3, V6.5, is substantially equivalent to Celesteion, PCA-9000A/3, V6.4, which received premarket clearance under K170490 and is marketed by Canon Medical Systems USA. The Celesteion, PCA-9000A/3, V6.5, incorporates modifications to the cleared device including a modification to the existing respiratory gating system to improve workflow, implementation of a new PET ECG gating system, and two new features, Variable Bed Time (vBT) and Clear Adaptive Low-Noise Method (CaLM) Reconstruction. These changes do not affect the safety or efficacy of the cleared device, as demonstrated in performance testing. The method of operation and manufacturing process for the Celesteion remain unchanged from the cleared device. See below for a brief comparison of the technological characteristics between the subject and the predicate device:
| Item | Celesteion, PCA-9000A/3, V6.5 | Celesteion, PCA-9000A/3, V6.4 |
|----------------------------------------------------------|------------------------------------|-------------------------------|
| 510(k) Number | This submission | K170490 |
| Variable bed time (vBT) | Available | N/A |
| CaLM (Clear adaptive Low-noise<br>Method Reconstruction) | Available | N/A |
| PET Respiratory Gating System<br>(NKRS-001A) | Available<br>-Workflow Improvement | Available |
| PET ECG gating<br>(NHEG-001A) | Available | N/A |
#### 16. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the following standards IEC60601-1-2, IEC60601-1-2, IEC60601-1-6, IEC60601-1-6, IEC60601-2-28, IEC60601-2-44, IEC60825-1, IEC62304, IEC62366, NEMA PS 3.1-3.18, NEMA XR-25, NEMA XR-26 and NEMA NU-2. Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.
#### 17. TESTING
Risk analysis and verification/validation testing conducted through bench testing are included in this submission which demonstrates that the established specifications for the device have been met.
{6}------------------------------------------------
Bench testing utilizing phantoms were conducted and it was determined that use of CaLM Reconstruction resulted in images with reduced noise while preserving detail and contrast. Representative clinical images were acquired to demonstrate that the subject device is capable of obtaining multi-phase, PET ECG gated data and that the new feature performs as intended.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
Additionally, testing of the subject device was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and CT Systems.
#### 18. CONCLUSION
Celesteion, PCA-9000A/3, V6.5, performs in a manner that is similar to and is intended for the same use as the predicate device, as indicated in product labeling. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.