The Panta 2 Arthrodesis Nail System is intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. Indications include: Post-traumatic and degenerative arthritis involving both ankle and subtalar joints Rheumatoid arthritis Revision of failed ankle arthrodesis with subtalar involvement or with insufficient talar body Revision of failed total ankle arthroplasty with subtalar intrusion Talar deficiency conditions (requiring a tibiocalcaneal arthrodesis) Avascular necrosis of the talus Neuroarthropathy or neuropathic ankle deformity Severe deformity as a result of talipes equinovarus, cerebral vascular accident, paralysis or other neuromuscular disease Severe pilon fractures with trauma to the subtalar joint.
Device Story
Panta 2 Arthrodesis Nail System is a retrograde intramedullary nail system for hindfoot arthrodesis. Device consists of a central nail, screws, and end caps. Used by orthopedic surgeons in clinical settings to achieve bone fusion, re-align the foot on the weight-bearing axis, correct coronal/sagittal plane deformities, and provide rotational stability and axial compression. Instrumentation allows multi-planar screw fixation in the tibia, talus, and calcaneus. Output is mechanical stabilization of the hindfoot to facilitate fusion and treat trauma. Benefits include improved structural alignment and stability for patients with severe hindfoot pathology.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing per ASTM F1264 and ASTM F543 standards, including static/fatigue bending, torsional strength, and axial pullout testing.
Technological Characteristics
Intramedullary fixation rod system. Materials: Titanium alloy (ASTM F136). Components: Retrograde nail, screws, end caps. Mechanical principle: Axial compression and multi-planar screw fixation. Sterilization: Not specified. Connectivity: None (mechanical device).
Indications for Use
Indicated for patients requiring tibiotalocalcaneal arthrodesis or treatment of hindfoot/distal tibia trauma, including those with post-traumatic/degenerative arthritis, rheumatoid arthritis, failed ankle arthrodesis/arthroplasty, talar deficiency, avascular necrosis, neuroarthropathy, severe deformities (talipes equinovarus, paralysis, CVA), or severe pilon fractures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
Newdeal Panta® Nail (10 mm and Panta® Nail XL) System (K091788)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
November 2, 2018 Integra Lifesciences Corporation Blesson Abraham Senior Regulatory Affairs Associate 11101 Metric Blvd. Austin, Texas 78758
Re: K181639
Trade/Device Name: Panta 2 Arthrodesis Nail System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary Fixation Rod Regulatory Class: Class II Product Code: HSB Dated: October 10, 2018 Received: October 12, 2018
Dear Blesson Abraham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
## Vincent J. Devlin -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
#### K181639
Device Name Panta 2 Nail Arthrodesis System
The Panta 2 Nail Arthrodesis System is intended for use in tibiotalocalcaneal arthrodesis and treatment of trauma to the hindfoot and distal tibia. Indications include:
- · Post-traumatic and degenerative arthritis involving both ankle and subtalar joints
- · Rheumatoid arthritis
- · Revision of failed ankle arthrodesis with subtalar involvement or with insufficient talar body
- · Revision of failed total ankle arthroplasty with subtalar intrusion
- · Talar deficiency conditions (requiring a tibiocalcaneal arthrodesis)
- Avascular necrosis of the talus
- · Neuroarthropathy or neuropathic ankle deformity
- · Severe deformity as a result of talipes equinovarus, cerebral vascular accident, paralysis or other neuromuscular disease
- · Severe pilon fractures with trauma to the subtalar joint.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for Integra. The word "INTEGRA" is written in gray, block letters. Below the word "INTEGRA" is the phrase "LIMIT UNCERTAINTY" in smaller, gray letters. To the right of the word "INTEGRA" is a design of green squares that are different sizes.
### 510(k) Summary
| Sponsor | Integra Lifesciences Corp.<br>Ascension Orthopedics, Inc.<br>11101 Metric Blvd.<br>Austin, TX 78758 |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment<br>Number | 3014207283 |
| Point of Contact | Blesson Abraham<br>Senior Regulatory Affairs Specialist<br>11101 Metric Blvd.<br>Austin, TX 78758<br>512-368-1423 |
| Date | July 10, 2018 |
| Trade Name | Panta 2 Arthrodesis Nail System |
| Classification<br>Name | Intramedullary Fixation Rod |
| Classification<br>Panel | Orthopedic |
| Classification | Class II |
| Regulation | 21 CFR 888.3020 |
| Product Code | HSB |
| Predicate Device | K091788<br>Newdeal Panta® Nail (10 mm and Panta® Nail XL) System, currently marketed<br>under Integra as the Panta Arthrodesis Nail System. |
| Device<br>Description | The Panta 2 Arthrodesis Nail System (Panta 2) is used to perform a tibio-talo-<br>calcaneal arthrodesis, a procedure undertaken for hindfoot problems that affect<br>both the ankle and subtalar joints. The aim is to achieve fusion of the bones using<br>a retrograde Nail that allows re-alignment of the foot on the weight-bearing axis,<br>correction of coronal and sagittal plane deformities, rotational stability as well as<br>axial compression. The Panta 2 System includes instrumentation which allows<br>placement of screws through the central Nail and enables application of<br>compression along with multi-planar screw fixation in the tibia, the talus, and the<br>calcaneus. |
{4}------------------------------------------------
## Integra Lifesciences Corporation – Special 510(k)
Integra Panta 2 Arthrodesis Nail System
| Intended Use/<br>Indications for Use | The Panta 2 Arthrodesis Nail System is intended for use in tibiotalocalcaneal<br>arthrodesis and treatment of trauma to the hindfoot and distal tibia. Indications<br>include: Post-traumatic and degenerative arthritis involving both ankle and subtalar<br>joints Rheumatoid arthritis Revision of failed ankle arthrodesis with subtalar involvement or with<br>insufficient talar body Revision of failed total ankle arthroplasty with subtalar intrusion Talar deficiency conditions (requiring a tibiocalcaneal arthrodesis) Avascular necrosis of the talus Neuroarthropathy or neuropathic ankle deformity Severe deformity as a result of talipes equinovarus, cerebral vascular<br>accident, paralysis or other neuromuscular disease Severe pilon fractures with trauma to the subtalar joint. | | | |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|------------|---------------------------------------------------------------------------------------------------------------------|
| Nonclinical<br>Performance Data | The Panta 2 Arthrodesis Nail System implants were subjected to verification<br>testing per required standards to establish substantial equivalent performance in<br>comparison to the predicate device. | | | |
| | No. | Testing Description | Standard | Results |
| | 1 | Nail - Bending<br>Static Comparison<br>Test | ASTM F1264 | The Panta 2 System Nail<br>performed substantially<br>equivalent when compared to the<br>predicate system Nail. |
| | 2 | Nail - Bending<br>Fatigue Life<br>Comparison Test | ASTM F1264 | |
| | 3 | Nail - Static<br>Torsional Strength<br>Comparison Test | ASTM F1264 | |
| | 4 | Screws - Bending<br>Fatigue Strength<br>Test | ASTM F1264 | The Panta 2 System screws<br>performed substantially<br>equivalent when compared to the<br>predicate system screws. |
| | 5 | Screws - Torsional<br>Properties Test | ASTM F543 | The Panta 2 System screws<br>performed to requirements of the<br>identified standards. |
| | 6 | Screws – Axial<br>Pullout Strength<br>Test | ASTM F543 | |
| Clinical | Clinical performance data is not required to demonstrate substantial equivalence | | | |
{5}------------------------------------------------
# Integra Lifesciences Corporation – Special 510(k)
Integra Panta 2 Arthrodesis Nail System
| Substantial<br>Equivalence<br>Conclusion | Basis of substantial equivalence of the modified device and predicate device<br>established on the following:<br><br>The Panta 2 and Panta 1 Systems have the following similarities:<br>Both systems have the same Indications for Use. The intended patient<br>population and indications for use of the Panta 1 Arthrodesis Nail<br>System are unchanged. Both systems operate using the same fundamental scientific technology. Both systems incorporate the same basic implant design. Both systems use the same methods of sterilization. Both systems use the same operational principles for the surgical<br>implantation of the IM Nails, Screws, and End Caps. The Panta 2 and Panta 1 have the following differences: Conformance to implant material from ISO 5832 to ASTM 136 Minor dimensional differences in the nail design Additional screw sizes added for fully threaded screws Reduce available sizes of partially threaded screws Minor change to the anodization and dimensional characteristics of the<br>screws Reduce available end cap sizes and modify anodization of the end cap Evaluation of the risks and performance data based on the differences between<br>the subject device and predicate does not raise any new issues or concerns related<br>to safety or effectiveness. It is concluded that the modified device, the Panta 2<br>Arthrodesis Nail System, is as safe and effective as the predicate device for its<br>intended use, and is substantially equivalent to the legally marketed predicate<br>device, the Newdeal Panta® Nail (10 mm and Panta® Nail XL) System. |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.