K181633 · Musculoskeletal Transplant Foundation · GAT · Mar 15, 2019 · General, Plastic Surgery
Device Facts
Record ID
K181633
Device Name
MTF Pre-Sutured Tendon
Applicant
Musculoskeletal Transplant Foundation
Product Code
GAT · General, Plastic Surgery
Decision Date
Mar 15, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5000
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MTF Pre-Sutured Tendon is intended for use as a construct in anterior cruciate ligament and posterior cruciate ligament reconstruction. The MTF Pre-Sutured Tendon is for single patient use only.
Device Story
Device consists of single allograft tendon (semitendinosus, anterior/posterior tibialis, or peroneus longus) pre-sutured with UHMWPE non-absorbable surgical suture. Used by orthopedic surgeons in OR for ACL/PCL reconstruction. Pre-sutured construct eliminates need for surgeon to manually suture tendon bundle during surgery. Implanted using traditional clinical methods. Benefits include standardized construct strength and reduced intraoperative time. Device is single-patient use; sterile.
Clinical Evidence
Bench testing and cadaveric study. Bench testing: tensile strength and knot pull strength compared to Arthrex kit (K041553); results equivalent or better. Cadaveric knee study: compared pre-sutured construct to surgeon-sutured quadruple-bundled construct; no statistical difference in biomechanical performance acutely or post-cycling. Endotoxin levels <20 EU/device; USP <71> sterility and USP <85> LAL testing passed.
Technological Characteristics
Construct: Allograft tendon (semitendinosus, tibialis, or peroneus longus) with UHMWPE non-absorbable suture. Sterilization: USP <71> compliant. Biocompatibility: Endotoxin <20 EU/device (USP <85>). Mechanical: Equivalent to USP non-absorbable surgical suture standards. No electronic components or software.
Indications for Use
Indicated for use as a construct in anterior cruciate ligament (ACL) and posterior cruciate ligament (PCL) reconstruction in patients requiring such surgical intervention.
Regulatory Classification
Identification
Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 15, 2019
Musculoskeletal Transplant Foundation Katrina Carroll Regulatory Affairs Manager 125 May Street Edison, New Jersey 08837
Re: K181633
Trade/Device Name: MTF Pre-Sutured Tendon Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable poly(ethylene terephthalate) surgical suture Regulatory Class: Class II Product Code: GAT Dated: February 13, 2019 Received: February 13, 2019
Dear Ms. Carroll:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurence D. Coyne -S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K181633
Device Name MTF Pre-Sutured Tendon
Indications for Use (Describe)
The MTF Pre-Sutured Tendon is intended for use as a construct in anterior cruciate ligament and posterior cruciate ligament reconstruction.
The MTF Pre-Sutured Tendon is for single patient use only.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 1. 510(k) Summary
| Device Trade Name: | MTF Pre-Sutured Tendon |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | Musculoskeletal Transplant Foundation<br>125 May Street<br>Edison, NJ 08837 |
| Contact: | Ms. Katrina Carroll<br>Phone: (732) 661-3154<br>Fax: (732) 661-2189<br>Katrina_Carroll@mtf.org |
| Prepared by: | Mr. Dave McGurl<br>Director, Regulatory Affairs<br>Musculoskeletal Clinical Regulatory Advisers, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20001<br>Phone: 202.552.5797<br>Fax: 202.552.5798<br>dmcgurl@mcra.com |
| Date Prepared: | March 13, 2019 |
| Common Name: | Pre-Sutured Tendon |
| Classification Name: | Nonabsorable poly(ethylene terephthalate) surgical suture |
| Classification: | 21 CFR § 878.5000 |
| Class: | II |
| Product Codes: | GAT |
| Predicate Device: | Allosource ReConnex™ Pre-Sutured Tendon (K170957) Arthex<br>Suture Grafting Kit (K041553) |
| Reference Device: | Riverpoint Medical HS Fiber™ Polyblend Non-Absorbable<br>(K100006)<br>MTF Fascia™ (K120479) |
{4}------------------------------------------------
## Indications for Use:
The MTF Pre-Sutured Tendon is intended for use as a construct in anterior cruciate ligament and posterior cruciate ligament reconstruction.
The MTF Pre-Sutured Tendon is for single patient use only.
## Device Description:
The MTF Pre-Sutured Tendon is a construct consisting of a single tendon pre-sutured with UHWMPE non-absorbable surgical suture. The device may include a semitendinosus tendon. bilateral anterior and posterior tibialis tendon, or bilateral peroneus longus tendon., The Pre-Sutured Tendon passes USP<71> sterility testing and is provided for a single patient use.
### Performance Testing:
The performance of the MTF Pre-Sutured Tendon was characterized through a comparison study to show that the MTF Pre-Sutured Tendon was comparable to the Arthrex suture grafting kit as described in K041553. Visual characteristics and tensile strength of the pre-sutured tendon were evaluated and are equivalent to or better than the tensile strength data for the predicate device cleared under K041553. The suture does not pull out, or fail when subjected to pull testing. The suture knot pull strength is equivalent to or better than USP non-absorbable Surgical Sutures average knot pull strength.
A cadaveric knee study was conducted to assess clinically relevant performance of the graft in combination with its intended fixation methods. The results of the human cadaveric knee study demonstrate that the construct is equivalent to or better than a quadruple bundled construct sutured by a surgeon at the time of surgery. The biomechanical assessment comparing the sutured tendon bundle constructed in the operating room using an allograft tendon, and a pre-sutured tendon constructed by MTF shows no statistical difference either acutely or post-cyclin. It was also demonstrated that the pre-sutured tendon construct can be implanted using traditional clinical methods by an orthopedic surgeon.
The MTF Pre-Sutured Tendon has an endotoxin level of less than 20 EU/device. Every lot of the MTF Pre-Sutured Tendon must pass a validated limulus amebocyte lysate (LAL) USP <85> Tests. The Pre-Sutured Tendon passes USP<71> sterility testing.
#### Risk Assessment:
MTF has performed a risk assessment in order to compare the risk profile of its Pre-Sutured Tendon to another predicate device that is also a pre-sutured, quadruple-bundled tendon (Allosource's ReConnex, K170957) as well as to a predicate device (Arthrex's Suture Grafting Kit, K041553) which is used to suture a standard allograft tendon into a quadruple-bundled construct in the operating room at the time of surgery. The risk assessment completed provides a sufficient risk/benefit profile to support the equivalence to both Arthrex Suture Grafting Kit and Allosource ReConnex.
{5}------------------------------------------------
## Substantial Equivalence:
The MTF Pre-Sutured Tendon is substantially equivalent with respect to materials, indications, function and performance to the Allosource ReConnex™ Pre-Sutured Tendon (K170957) and Arthex Suture Grafting Kit (K041553).
### Technological Features and Substantial Equivalence:
The MTF Pre-Sutured Tendon and Allosource ReConnex™ Pre-Sutured Tendon (K170957) have identical indications, they are intended for use as a construct in anterior cruciate ligament and posterior cruciate ligament reconstruction. The Arthex Suture Grafting Kit (K041553) is intended for use in soft tissue approximation and or ligation including, but not limited to anterior cruciate ligament and posterior cruciate ligament reconstruction. Like the subject device, the predicate devices incorporate nonabsorbable UHMWPE sutures and are indicated for single patient use only. The MTF Pre-Sutured Tendon is described for the identical patient population, intended use, and indication as the predicates, Allosource ReConnex™ Pre-Sutured Tendon (K170957) and Arthex Suture Grafting Kit (K041553). Any differences do not raise new questions of safety or effectiveness. In addition, comparison testing demonstrates that the MTF Pre-Sutured Tendon is equivalent in ultimate load strength. Therefore, the MTF Pre-Sutured Tendon is substantially equivalent to the Allosource ReConnex™ Pre-Sutured Tendon (K170957) and Arthex Suture Grafting Kit (K041553).
Prior to donation, the donor's medical/social history was screened for medical conditions or disease processes that would contraindicate the donation of tissues in accordance with current policies and procedures approved by the MTF Medical Board of Trustees. Donor blood samples taken at the time of recovery were tested by a facility that is CLIA certified and registered with the FDA. The donor blood samples were tested for: Hepatitis B virus (HBV) surface antigen, HBV core antibody, Hepatitis C virus (HCV) antibody, HIV-1/12 antibody, Syphilis, HIV-1 NAT, HCT NAT, and HBV Nat.
All infectious disease test results passed acceptability for screening. This allograft tissue has been determined to be suitable for transplantation.
The infectious disease test results, consent, current donor medical history interview, physical assessment, available relevant medical records to include previous medical history, laboratory test results, autopsy and coroner reports, if performed, and information obtained from any source or records which may pertain to donor suitability, have been evaluated by an MTF physician and are sufficient to indicate that donor suitability criteria current at the time of procurement, have been met. This tissue is suitable for transplantation. The donor suitability criteria used to screen this donor are in compliance with the FDA regulations published in 21 CFR Part 1271 Human Cells, Tissues, and Cellular and Tissue Based Products, as applicable. All procedures for donor screening, including laboratory testing, meet or exceed current standards established by the American Association of Tissue Banks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.