Medtronic StrataMR Valves and Shunts are designed to provide continuous cerebrospinal fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The design enables the physician to noninvasively adjust valve pressure/performance level pre- and post-implantation by using magnetic adjustment tools without the need for radiographic confirmation.
Device Story
StrataMR system manages hydrocephalus by draining CSF from brain ventricles to the peritoneal cavity or right atrium. System includes implantable valve and external magnetic adjustment tools. Physician uses locator, indicator, and adjustment tools to noninvasively read and set valve pressure. New 'guider tool' component added to adjustment kit to ensure proper rotor foot positioning and prevent it from resting on MRI resistance walls. Used in clinical/office settings by neurosurgical staff. Guider tool provides defined distance for rotating adjustment tool, ensuring reliable valve setting without radiographic confirmation. Benefits include reduced risk of improper valve adjustment and improved procedural reliability.
Clinical Evidence
Bench testing and human factors validation only. Bench testing (n=29) confirmed guider tool functionality and reliability (95% confidence/90% reliability). Human factors study (n=14 clinicians, 84 adjustment cycles) using anatomical models validated that users can successfully adjust valve settings without the rotor foot residing on the resistance wall.
Technological Characteristics
System components: locator tool, indicator tool, adjustment tool, guider tool. Materials: Clear Polycarbonate (locator, guider), Aluminum (adjustment). Non-sterile, reusable. Cleaning: warm water and mild detergent. Connectivity: None (mechanical/magnetic). Biocompatibility per ISO 10993-5, 10993-10, 10993-11.
Indications for Use
Indicated for patients requiring continuous CSF flow from brain ventricles to the right atrium or peritoneal cavity for hydrocephalus management. Intended for use by neurosurgical personnel (neurosurgeons, residents, physician assistants, clinical staff) to noninvasively adjust valve pressure settings.
Regulatory Classification
Identification
A central nervous system fluid shunt is a device or combination of devices used to divert fluid from the brain or other part of the central nervous system to an internal delivery site or an external receptacle for the purpose of relieving elevated intracranial pressure or fluid volume (e.g., due to hydrocephalus). Components of a central nervous system shunt include catheters, valved catheters, valves, connectors, and other accessory components intended to facilitate use of the shunt or evaluation of a patient with a shunt.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
October 1, 2018
Medtronic Neurosurgery Xiaojian Sun Regulatory Affairs Manager 125 Cremona Drive Goleta, California 93117
Re: K181622
Trade/Device Name: StrataMR Valves and Shunts (Guider Tool) Regulation Number: 21 CFR 882.5550 Regulation Name: Central Nervous System Fluid Shunt and Components Regulatory Class: Class II Product Code: JXG Dated: July 2, 2018 Received: July 3, 2018
Dear Xiaojian Sun:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to Mav 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Xiaolin Zheng -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181622
Device Name StrataMRTM Valves and Shunts (Guider Tool)
#### Indications for Use (Describe)
Medtronic StrataMR™ Valves and Shunts are designed to provide continuous cerebrospinal fluid (CSF) flow from the ventricles of the brain into the right atrium of the peritoneal cavity. The design enables the physician to noninvasively adjust valve pressure/performance level pre- and post-implantation by using magnetic adjustment tools without the need for radiographic confirmation.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------|--------------------------------------------------------------------------|
| <div> <span>X</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K181622 510(k) Summary
Submitter Name Submitter Address Medtronic Neurosurgery 125 Cremona Drive Goleta, CA 93117-5503 USA
| Contact Person |
|-----------------------|
| Telephone |
| Fax |
| Email |
| Date Summary Prepared |
Hebe Sun, Regulatory Affairs Manager (805) 571-8758 (805) 571-8480 Xiaojian.sun@medtronic.com September 28, 2018
#### Subiect Device
Common Name
Product Code
Classification Name
Trade Name
StrataMRTM Valves and Shunts (Guider Tool) Hydrocephalus Shunt Central nervous system fluid shunt and components 21 CFR 882.5550 Class II JXG
#### Predicate Device K152700
| Trade Name | Medtronic StrataMR TM Valves and Shunts |
|---------------------|------------------------------------------------------------------------------|
| Common Name | Hydrocephalus Shunt |
| Classification Name | Central Nervous System Fluid Shunt and Component<br>21 CFR 882.5550 Class II |
| Product Code | JXG |
#### Subject Device Description
The Medtronic StrataMR™ valves are implantable adjustable valves for the management of hydrocephalus. The valves and their associated catheters drain Cerebrospinal Fluid (CSF) from the ventricles in the brain into the peritoneal cavity or the right atrium of the heart, where it is absorbed by the body. Before and after implantation, the pressure/flow characteristics of the Medtronic StrataMRTM valve can by modified by the StrataMR adjustment tool.
The original StrataMR adjustment tools, including handheld locator, and adjustment tool, are designed to allow the user to determine the pressure/performance level setting of StrataMR valves and adjust the setting when needed. In this 510(k) submission, Medtronic is proposing to add an additional component, guider tool, to the StrataMR adjustment tools, to improve the method to reliably adjust the StrataMR valve, cleared in K152700.
# Intended Use/Indications for Use
Medtronic StrataMR Valves and Shunts are designed to provide continuous cerebrospinal fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The design enables the physician to noninvasively adjust valve pressure/performance level pre- and post-implantation by using magnetic adjustment tools without the need for radiographic confirmation.
{4}------------------------------------------------
# Summary of Technological Characteristics of the Subject Device Compared to the Predicate Device
The previously cleared StrataMR valve and shunt system, K152700, includes StrataMR valves, shunts, and StrataMR adjustment tools which are designed to read and adjust the pressure/ performance level setting of StrataMR. There is no change to these previously cleared StrataMR valves or shunts.
The difference between the proposed StrataMR adjustment tools and the currently available StrataMR adjustment tools is the following modification: the addition of a fourth component called the guider tool. This tool is intended to improve the ability of users to reliably adjust the StrataMR valve. This change does not affect the intended use of the system.
| Feature | K181622<br>Subject Device | K152700<br>Predicate Device | Discussion |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | Medtronic StrataMR™ Valves<br>and Shunts are designed to<br>provide continuous cerebrospinal<br>fluid (CSF) flow from the<br>ventricles of the brain into the<br>right atrium of the heart or the<br>peritoneal cavity. The design<br>enables the physician to<br>noninvasively adjust valve<br>pressure/performance level pre-<br>and post-implantation by using<br>magnetic adjustment tools<br>without the need for radiographic<br>confirmation. | Medtronic StrataMR™ Valves<br>and Shunts are designed to<br>provide continuous<br>cerebrospinal fluid (CSF) flow<br>from the ventricles of the brain<br>into the right atrium of the heart<br>or the peritoneal cavity. The<br>design enables the physician to<br>noninvasively adjust valve<br>pressure/performance level pre-<br>and post-implantation by using<br>magnetic adjustment tools<br>without the need for<br>radiographic confirmation. | Identical |
| Components | 4 – locator tool, indicator tool,<br>adjustment tool, guider tool | 3 – locator tool, indicator tool,<br>adjustment tool | Similar.<br>The only difference<br>between the modified<br>device and the predicate<br>is that an additional<br>guider Tool is added to<br>the system, which is<br>intended to be used<br>together with the locator<br>Tool and adjustment Tool<br>during the adjustment<br>process, in order to<br>reduce the risk of rotor<br>foot being positioned on<br>top of the wall. Risk<br>assessment shows<br>introduction of the guider<br>tool does not introduce<br>new unacceptable risk.<br>Design verification and<br>validation studies<br>demonstrated that the<br>proposed modified device<br>meets the performance<br>requirement. |
| | K181622<br>Subject Device | K152700<br>Predicate Device | Discussion |
| Feature | | | |
| Patient -<br>contacting<br>materials | Locator Tool: Clear<br>Polycarbonate Indicator Tool:<br>polycarbonate<br>Adjustment Tool: Aluminum<br>Guider tool: Clear Polycarbonate | Locator Tool: Clear<br>Polycarbonate Indicator Tool:<br>polycarbonate<br>Adjustment Tool: Aluminum | Similar.<br>The only difference<br>between the modified<br>device and the predicate<br>is that an additional<br>guider tool is added to the<br>system. However, the<br>patient contacting<br>material from the guider<br>tool is identical with the<br>patient contacting<br>material of the locator<br>tool in formulation,<br>processing, geometry, and<br>no other chemicals<br>have been added.<br>Therefore, the modified<br>StrataMR adjustment tool<br>is substantially equivalent<br>to the original device,<br>from a biocompatibility<br>perspective. |
| Packaging | Reusable, magnetically shielded<br>case. Foam insert is updated to<br>accommodate the guider tool | Reusable, magnetically shielded<br>case with foam insert | Similar.<br>The only difference is the<br>addition of the guider<br>tool to the package, and<br>the foam insert update to<br>accommodate the guider<br>tool.<br>Transportation study was<br>conducted to demonstrate<br>that the packaging can<br>protect the device from<br>damage, and product<br>passed functional<br>requirements after<br>transportation simulation. |
| Operating<br>Principal | Guider Tool provides a defined<br>distance for the users to rotate<br>adjustment tool in order to<br>prevent the rotor foot from being<br>positioned on top of wall.<br>Indicator Tool is used with<br>locator tool to couple with valve<br>magnet to read the valve setting.<br>Adjustment Tool is used to<br>couple with valve magnet and<br>rotate to the designed valve<br>setting. | Indicator Tool is used with<br>locator tool to couple with valve<br>magnet to read the valve setting.<br>Adjustment Tool is used to<br>couple with valve magnet and<br>rotate to the designed valve<br>setting.<br>At the end of the adjustment,<br>indicator tool is used with<br>locator tool to confirm the valve<br>setting. | The operation principle to<br>adjust the valve setting is<br>similar. The operation<br>principle for the<br>StrataMR valve setting<br>reading and adjustment is<br>equivalent between the<br>predicate and proposed<br>device.<br>The difference is the<br>supplementary steps after<br>the pressure level setting<br>steps, which includes<br>turning the adjustment<br>tool (which is attached to<br>the guider tool) clockwise |
| Feature | K181622<br>Subject Device | K152700<br>Predicate Device | Discussion |
| | At the end of the adjustment,<br>indicator tool is used with locator<br>tool to confirm the valve setting. | | and counter-clockwise at<br>a certain distance above<br>the valve to remove the<br>rotor foot from the MRI<br>resistance wall, if the<br>rotor foot was left on top<br>of the wall from the<br>previous adjusting steps.<br>The distance above the<br>valve is controlled by the<br>guider tool. This<br>supplementary step does<br>not change the<br>fundamental operation<br>principle. Design<br>verification and<br>validation studies<br>demonstrated that the<br>modified StrataMR<br>adjustment tools meet the<br>functional requirement of<br>adjusting the pressure<br>level of StrataMR valves. |
| Cleaning &<br>Sterilization | Non-sterile and reusable. Tools<br>can be cleaned with warm water<br>and mild detergent (5% Dawn<br>soap/95% DI water) before use. | Non-sterile and reusable. Tools<br>can be cleaned with warm water<br>and mild detergent (5% Dawn<br>soap/95% DI water) before use. | Similar.<br>The only difference is the<br>addition of the guider<br>Tool. Design verification<br>study was conducted to<br>demonstrate that the<br>guider Tool can<br>withstand the cleaning<br>with warm water and<br>mild detergent, without<br>exhibiting cracking or<br>removal of marking. |
{5}------------------------------------------------
{6}------------------------------------------------
# Summary of Design Performance Testing
### Non-Clinical
Results of verification and validation testing met pre-established acceptance criteria: 1) that when the rotor foot was placed on top of an MRI resistance wall, use of the guider tool can remove the rotor foot from the wall and position it to a pressure-level-setting well after the adjustment procedure and 2) use of guider tool will not inadvertently change the pressure level setting. This was done by varying the configuration to represent a worst-case scenario involving misalignment and off-centering, which demonstrates that the modified StrataMR adjustment tool met the performance specification.
In order to reach 95% confidence and 90% reliability interval, 29 guider tools were used together with locator tool and adjustment tool to conduct the test on 29 StrataMR valves.
{7}------------------------------------------------
Design verification and validation studies demonstrated that the modified StrataMR adjustment tool meet the functional requirement of adjusting the pressure level setting of StrataMR: use of the guider tool did not inadvertently change the setting.
#### Human Factors
The design validation study is intended to validate that under simulated use conditions, users can adjust the setting of the StrataMR valve with the new guider tool together with the existing adjustment tools, and the rotor foot does not reside on an MRI resistance wall at the end of the adjustment procedure.
Testing was conducted in an environment similar to an office setting where post-operative adjustment may occur. The evaluators for this validation study were neurosurgical personnel (attending neurosurgeons, residents, physician assistants, or clinical staff) with experience managing StrataMR or Strata-type valves. Two plastic anatomical model heads with imitation skin were used to simulate the clinical use, with the valve positioned under the imitation skin so that it was not visible to the evaluator. 14 clinician users were selected as evaluators and each evaluator was requested to perform 6 adjustments, and the total number of adjustment cycles was 84.
#### Biocompatibility
No new biocompatibility testing was conducted on the new Guider Tool because it was fabricated from similar materials as the predicate StrataMR Adjustment Tools. Therefore, the following biocompatibility testing was leveraged from the predicate device (K152700).
- . ISO 10993-5 Third edition 2009-06-01 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10 Third Edition 2010-08-01 ● Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
- . ISO 10993-11 Second edition 2006-08-15 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
# Conclusion
The performance testing demonstrates that the device is safe, as effective, and performs as well as or better than the predicate. The proposed modification of the StrataMR adjustment tools does not affect the intended use of the device. Therefore, it is concluded that the proposed modification of the StrataMR adjustment tool is substantially equivalent to the original StrataMR adjustment tool.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.