← Product Code [GDW](/productcode/GDW) · K181620

# Endoscopic Linear Cutting Staplers and Loading Units for Single Use (K181620)

_Ezisurg (Suzhou) Medical Co., Ltd. · GDW · Sep 18, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181620

## Device Facts

- **Applicant:** Ezisurg (Suzhou) Medical Co., Ltd.
- **Product Code:** [GDW](/productcode/GDW.md)
- **Decision Date:** Sep 18, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4750
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The device is intended for transection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general abdominal, gynecologic, thoracic, and pediatric surgical procedures.

## Device Story

Single-use endoscopic linear cutting stapler; places two, triple-staggered rows of titanium staples while simultaneously dividing tissue along a central line. Manual operation; includes safety mechanism (green button) to prevent mis-firing. Available in multiple lengths (70mm, 160mm) and staple sizes (2.5mm, 3.5mm, 4.1mm, 4.8mm) to accommodate varying tissue thicknesses. Device is reloadable up to 25 times for 1 firing per procedure. Used by surgeons in OR settings for general, gynecologic, thoracic, and pediatric procedures. Provides mechanical tissue closure and transection; benefits include standardized staple formation and hemostasis during surgical resection or anastomosis.

## Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (cytotoxicity, sensitization, irritation, pyrogenicity), package integrity, shelf life, and mechanical performance (firing force, staple formation, staple line pressure/tensile, hemostasis).

## Technological Characteristics

Titanium staples; manual actuation; linear cutting mechanism. Complies with ISO 10993-1/5/10 (biocompatibility), ISO 11137-2 (radiation sterilization), ASTM F 88/F88M-09 (seal strength), and USP <85>/<151> (endotoxins/pyrogens).

## Regulatory Identification

An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Endo GIA Staplers ([K111825](/device/K111825.md))
- ENDOPATH Linear Cutters and Staplers ([K061156](/device/K061156.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 18, 2018

Ezisurg (Suzhou) Medical Co., Ltd. % Ms. Diana Hong General Manager Mid-Link Consulting Co., Ltd. P.O. Box 120-119 Shanghai, 200120 Cn

Re: K181620

Trade/Device Name: Endoscopic Linear Cutting Staplers and Loading Units for Single Use Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: June 7, 2018 Received: June 20, 2018

Dear Ms. Hong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K181620

#### Device Name

Endoscopic Linear Cutting Staplers and Loading Units for Single Use

Indications for Use (Describe)

The device is intended for transection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general abdominal, gynecologic, thoracic, and pediatric surgical procedures.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Tab #7 510(k) Summary

This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.

The assigned 510(k) Number: __K181620

- 1. Date of Preparation: 04/16/2018
- 2. Sponsor Identification

### Ezisurg (Suzhou) Medical Co., Ltd.

Building 16, No.8 Jinfeng Rd., Suzhou National Hi-Tech District, Suzhou 215163 Jiangsu, China

Establishment Registration Number: Not yet registered.

Contact Person: Jingtian Ren Position: VP Quality& Registration Tel: 021-50456176-8026 Fax: 021-50676156 Email: renjingtian@ezisurg.com

- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Ying Xu (Alternative Contact Person)

Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China

Tel: +86-21-22815850, Fax: 3609253199 Email: info@mid-link.net

{4}------------------------------------------------

#### Identification of Proposed Device 4.

Trade Name: Endoscopic Linear Cutting Staplers and Loading Units for Single Use Common Name: Stapler and Reload

Regulatory Information Classification Name: Staple, Implantable Classification: II Product Code: GDW Regulation Number: 21 CFR 878.4750 Review Panel: General& Plastic Surgery

Classification Name: Stapler, Surgical; Classification: I Subsequent Product Code: GAG; Regulation Number: 21CFR 878.4800 Review Panel: General & Plastic Surgery

Intended Use Statement:

The device is intended for transection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general abdominal, gynecologic, thoracic, and pediatric surgical procedures.

#### Device Description

The proposed device places two, triple-staggered rows of titanium staples and simultaneously divides the tissue from central line. The device is available in 70mm, 160mm three lengths. Reloads of Endoscopic Linear Cutting Staplers and Loading Units for Single Use are available in in five staple sizes to accommodate various tissue thickness: 2.5mm, 3.5mm, 4.1mm and 4.8mm. The device may be reloaded and fired up to 25 times for 1 firing in a single procedure.

{5}------------------------------------------------

#### ડ. Identification of Predicate Device

Predicate Device 1 510(k) Number: K111825 Product Name: Endo GIATM Staplers

Predicate Device 2 510(k) Number: K061156 Device Name: ENDOPATH Linear Cutters and Staplers

- Non-Clinical Test Conclusion 6.
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:

- > ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process.
- > ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity.
- > ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and delayed-type hypersensitivity
- > USP 38-NF 33 <85> Bacterial Endotoxins Tests
- > ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materials;
- > ISO 11137-2:2013 Sterilization of health care products -Radiation- Part 2: Establishing the sterilization dose
- > USP 39-NF34:2016 <151> Pyrogen Test

The tests provided in this submission include package integrity test, bacteria endotoxin test, biocompatibility test, shelf life test and performance test were performed on both proposed device and predicate device to determine substantial equivalence, the performance test include following items

- > Firing Force Test
- > Staple Formation Test
- > Staple Line Pressure Test
- > Staple Line Tensile Test
- > Hemostasis Evaluation Test
- > Close Staple Height Test

Biocompatibility tests included cytotoxicity, sensitization, intracutaneous irritation and pyrogenicity.

{6}------------------------------------------------

Shelf life testing which included package integrity and performance evaluations were performed.

- 7. Clinical Test Conclusion
No clinical study is included in this submission.

{7}------------------------------------------------

#### Substantially Equivalent (SE) Comparison 8.

| Table 1 Comparison for Endoscopic Linear Cutting Staplers and Loading Units for Single Use |                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                              |
|--------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item                                                                                       | Proposed Device                                                                                                                                                                                                                          | Predicate Device 1                                                                                                                                                                                                                                                                                                       | Predicate Device 2                                                                                                                                                                                                                                                                                                                           |
|                                                                                            |                                                                                                                                                                                                                                          | K111825                                                                                                                                                                                                                                                                                                                  | K061156                                                                                                                                                                                                                                                                                                                                      |
| Product Code                                                                               | GDW                                                                                                                                                                                                                                      | GDW                                                                                                                                                                                                                                                                                                                      | GDW                                                                                                                                                                                                                                                                                                                                          |
| Regulation Number                                                                          | 21 CFR 878.4750                                                                                                                                                                                                                          | 21 CFR 878.4750                                                                                                                                                                                                                                                                                                          | 21 CFR 878.4750                                                                                                                                                                                                                                                                                                                              |
| Intended Use                                                                               | The device are intended for transection, resection, and/or creation of anastomoses. The instruments have application in multiple open or minimally invasive general abdominal, gynecologic, thoracic, and pediatric surgical procedures. | The Endo GIATM Staplers have applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of pancreas. | The ENDOPATH Endocutter 60 Endoscopic Linear Cutter is intended for transection, resection, and/or creation of anastomoses. The instrument has application in multiple open minimally invasive general, gynecologic, urologic, thoracic, and pediatric surgical procedures. It can be used with staple line or tissue buttressing materials. |
| Cutting Mechanism                                                                          | Linear                                                                                                                                                                                                                                   | Linear                                                                                                                                                                                                                                                                                                                   | Linear                                                                                                                                                                                                                                                                                                                                       |
| Operation Principle                                                                        | Manual                                                                                                                                                                                                                                   | Manual                                                                                                                                                                                                                                                                                                                   | Manual                                                                                                                                                                                                                                                                                                                                       |
| Safety Mechanism                                                                           | Green button is used for preventing from mis-firing.                                                                                                                                                                                     | Green button is used for preventing from mis-firing.                                                                                                                                                                                                                                                                     | Green button is used for preventing from mis-firing.                                                                                                                                                                                                                                                                                         |
| Suture Length                                                                              | 30mm, 45mm, 60mm                                                                                                                                                                                                                         | 30mm, 45mm, 60mm                                                                                                                                                                                                                                                                                                         | 60mm                                                                                                                                                                                                                                                                                                                                         |
| Staple height                                                                              | 2.5mm, 3.5mm, 3.8mm, 4.1mm, 4.8mm                                                                                                                                                                                                        | 2.0mm, 2.5mm, 3.5mm, 4.8mm                                                                                                                                                                                                                                                                                               | 2.5mm, 3.5mm, 3.8mm, 4.1mm                                                                                                                                                                                                                                                                                                                   |
| Closed staple form                                                                         | Image: staple                                                                                                                                                                                                                            | Image: staple                                                                                                                                                                                                                                                                                                            | Image: staple                                                                                                                                                                                                                                                                                                                                |
| Endotoxin Limit                                                                            | 20 EU                                                                                                                                                                                                                                    | 20EU                                                                                                                                                                                                                                                                                                                     | 20EU                                                                                                                                                                                                                                                                                                                                         |
| Labeling                                                                                   | Conforms with 21 CFR 801                                                                                                                                                                                                                 | Conforms with 21 CFR 801                                                                                                                                                                                                                                                                                                 | Conforms with 21 CFR 801                                                                                                                                                                                                                                                                                                                     |

## Table 1 Comparison for Endoscopic Linear Cutting Staplers and Loading Units for Single Use

{8}------------------------------------------------

- 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K181620](https://fda-staging.innolitics.com/device/K181620)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
