eLens Rigid Gas Permeable Contact Lens for Daily Wear, eLens Rigid Gas Permeable Contact Lens for Daily Wear Ortho-K
Applicant
E&E Optics, Inc.
Product Code
HQD · Ophthalmic
Decision Date
Jul 26, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5916
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear are indicated for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic and not aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a gas permeable contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lenses may be disinfected using a chemical disinfection system only. The eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear Ortho-K is indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. The lenses may be disinfected using a chemical disinfection system only.
Device Story
Lathe-cut rigid gas permeable (RGP) contact lenses; fabricated from hydrophobic fluoro-silicone acrylate materials (hexafocon A, hexafocon B, or oprifocon A). Lenses act as refractive medium to focus light on retina; orthokeratology lenses temporarily reshape cornea to reduce myopia. Prescribed by eye care professionals for daily wear. Patients use chemical disinfection systems. Benefits include vision correction and management of irregular corneal conditions; risks consistent with standard RGP lens wear.
Clinical Evidence
No clinical data provided; safety and effectiveness established by reference to predicate devices.
Technological Characteristics
Lathe-cut rigid gas permeable contact lenses; materials: hexafocon A, hexafocon B, or oprifocon A (FDA Group #3 fluoro-silicone acrylate); water content <1%; oxygen permeability (Dk) ranges 85-161 x 10^-11; includes color additives (D&C Green No. 6, C.I. Solvent Yellow No. 18, D&C Violet No. 2, FD&C Red No. 17); chemical disinfection required.
Indications for Use
Indicated for correction of refractive ametropia (myopia, hyperopia, astigmatism, presbyopia) in aphakic/non-aphakic persons with non-diseased eyes; management of irregular corneal conditions (keratoconus, pellucid marginal degeneration, post-penetrating keratoplasty, post-refractive surgery); and temporary reduction of myopia up to 5.00D via orthokeratology in non-diseased eyes.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
Predicate Devices
Boston XO (hexafocon A) Contact Lens for Daily Wear Orthokeratology (K001960)
Boston Equalens II (oprifocon A) Contact Lens for Daily Wear Orthokeratology (K003933)
Boston XO2 (hexafocon B) Contact Lens for Daily Wear (Orthokeratology) (K071266)
Boston XO (Hexafocon A) Rigid Gas Permeable Contact Lens (K071043)
Acuity 85 (Oprifocon A) Rigid Gas Permeable Contact Lens (K170001)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo, which features a stylized human figure. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 26, 2018
E&E Optics Inc. % Bret Andre Principal Consultant EyeReg Consulting, Inc. 6119 Canter Ln West Linn, OR 97068
Re: K181579
Trade/Device Name: eLens (hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lens for Daily Wear. eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lens for Daily Wear Ortho-K Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lens Regulatory Class: Class II Product Code: HOD. MUW Dated: June 13, 2018 Received: June 15, 2018
Dear Bret Andre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# J Angelo Green -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181579
### Device Name
eLens (hexafocon A. hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear; eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear Ortho-K
### Indications for Use (Describe)
The eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear are indicated for the correction of refractive ametropia (myopia, astigmatism and presbyopia) in aphakic and not aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a gas permeable contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lenses may be disinfected using a chemical disinfection system only.
The eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear Ortho-K is indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. The lenses may be disinfected using a chemical disinfection system only.
Note: To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;">☑ Recreation Use (Part 91 CFR 201.5(b))</span> |
|-------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☐ On-The-Spot Use (91 CFR 201.5(c))</span> |
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K181579
#### I. SUBMITTER
- July 11th, 2018 Date Prepared:
- Name: E&E Optics Inc. Address: 1795 E. Holt Blvd, Unit 104 Ontario, CA 91761 United States Contact Person: Bonnie Wan President (909) 391-8330 Phone number: Consultant: Bret Andre EyeReg Consulting, Inc. 6119 Canter Ln.
West Linn, OR 97068 Phone number: (503) 372-5226
#### II. DEVICE
| Trade Name: | eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable<br>Contact Lenses for Daily Wear;<br>eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable<br>Contact Lenses for Daily Wear Ortho-K | |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Common<br>Name: | Daily wear rigid gas permeable (hydrophobic) contact lens;<br>Daily wear rigid gas permeable (hydrophobic) contact lens (Orthokeratology) | |
| Classification<br>Name: | Rigid gas permeable contact lens. (21 CFR 886.5916) | |
| Regulatory<br>Class: | Class II | |
| Product Code: | HQD; MUW | |
{4}------------------------------------------------
#### III. PREDICATE DEVICE
The eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear and eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear Ortho-K are substantially equivalent to the following predicate devices:
- . "Boston XO (hexafocon A) Contact Lens for Daily Wear Orthokeratology" By Bausch + Lomb Inc. 510(k) number; K001960
- . "Boston Equalens II (oprifocon A) Contact Lens for Daily Wear Orthokeratology" By Bausch + Lomb Inc. 510(k) number: K003933
- . "Boston XO2 (hexafocon B) Contact Lens for Daily Wear (Orthokeratology)" By Bausch + Lomb Inc. 510(k) number: K071266
- 트 "Boston XO (Hexafocon A) Rigid Gas Permeable Contact Lens" By Bausch + Lomb Inc. 510(k) number; K071043
- . "Acuity 85 (Oprifocon A) Rigid Gas Permeable Contact Lens" By Acuity Polymers Inc. 510(k) number; K170001
#### DEVICE DESCRIPTION IV.
The eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear and eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear Ortho-K are lathe cut from one of the following hydrophobic. FDA Group #3 fluoro-silicone acrylate materials: hexafocon A, hexafocon B, or oprifocon A-with the following properties:
| | HEXAFOCON A | HEXAFOCON B | OPRIFOCON A |
|----------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|------------------------------------------------------------------|
| Refractive Index | 1.414 | 1.424 | 1.423 |
| Water Content | <1% | <1% | <1% |
| Specific Gravity | 1.266 | 1.19 | 1.24 |
| Wetting Angle | 49° | 38° | 30° |
| Oxygen Permeability<br>(Dk) ISO/FATT Method<br>(cm2/sec) (ml O2/ml x mm Hg @ 35°C) | $100 \times 10^{-11}$ | $141 \times 10^{-11}$ (edge corrected)<br>$161 \times 10^{-11}$ (non-edge<br>corrected) | $85 \times 10^{-11}$ |
| Contain one or more of the following color<br>additives conforming to:<br>21 CFR Part 73 & 74, Subpart D | D&C Green No. 6, C.I. Solvent<br>Yellow No. 18, and D&C Violet<br>No. 2 | D&C Green No. 6, C.I. Solvent<br>Yellow No. 18, and D&C Violet<br>No. 2 | D & C Green No. 6, FD & C Red<br>No. 17,<br>CI Solvent Yellow 18 |
{5}------------------------------------------------
The eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear are available in the following lens parameters:
| Parameter | Range | Tolerance |
|-------------------|------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Base Curve | 4.00mm to 9.00mm | ± 0.05 mm |
| Center Thickness | 0.20mm to 0.50mm | ± 0.02 mm |
| Diameter | 7.00mm to 21.00mm | ± 0.10mm |
| Spherical Power | -20.00 D to +20.00 D<br>(in 0.25D steps) | ± 0.12 (0 to = 5D)<br ± 0.18 (5 to = 10.0D)<br ± 0.25 (10 to = 15D)<br ± 0.37 (15 to = 20D)<br ± 0.50 (over 20D) |
| Cylindrical Power | Up to -9.00 D<br>(in 0.25 D steps) | ± 0.25 (0 to = 2D)<br ± 0.37 (2 to = 4D)<br ± 0.50 (over 4D) |
| Cylindrical Axis | 1° to 180° (in 1º steps) | ± 5° |
| Multifocal Power | +1.00 D to 3.75 D<br>(in 0.25 D steps) | ± 0.25D |
The eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear Ortho-K are available in the following lens parameters:
| Parameter | Range | Tolerance |
|----------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Base Curve | 6.50mm to 11.00mm | ± 0.05 mm |
| Center Thickness | 0.10mm to 0.30mm (low minus lens)<br>0.20mm to 0.70mm (plus lens) | ± 0.02 mm |
| Diameter | 6.50mm to 11.50mm | ± 0.10mm |
| Secondary Curves | 0.10mm to 2.00mm (flatter or<br>steeper than base curve) | ± 0.10mm |
| Peripheral Curves | 0.10mm to 2.00mm (flatter or<br>steeper than base curve) | ± 0.10mm |
| Spherical Power | -10.00 D to +30.00 D<br>(in 0.25D steps) | ± 0.12 (0 to = 5D)<br ± 0.18 (5 to = 10.0D)<br ± 0.25 (10 to = 15D)<br ± 0.37 (15 to = 20D)<br ± 0.50 (over 20D) |
| Aspheric Lens Eccentricity | -1.5 to 1.5 (oblate, prolate, or<br>tangent conic) | |
{6}------------------------------------------------
#### V. INDICATIONS FOR USE
The eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear are indicated for the correction of refractive ametropia (myopia, hyperopia, astigmatism and presbyopia) in aphakic and not aphakic persons with non-diseased eyes. Also, the lenses may be prescribed in otherwise non-diseased eyes that require a gas permeable contact lens for the management of irregular corneal conditions such as keratoconus, pellucid marginal degeneration, or following penetrating keratoplasty or refractive (e.g., LASIK) surgery. The lenses may be disinfected using a chemical disinfection system only.
The eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear Ortho-K is indicated for daily wear in an orthokeratology fitting program for the temporary reduction of myopia of up to 5.00 diopters in non-diseased eyes. The lens may be disinfected using a chemical disinfection system only.
Note: To maintain the orthokeratology effect of myopia reduction, lens wear must be continued on a prescribed wearing schedule.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE VI. DEVICE
The following matrix illustrates the production method, intended use and materials of the eLens (hexafocon A. hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear and eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear Ortho-K, as well as the predicate device.
| | eLens RGP<br>Contact Lenses | Boston XO RGP<br>Contact Lens | Boston XO2 RGP<br>Contact Lens | Boston Equalens II<br>RGP Contact Lens |
|----------------------|-----------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|
| | Subject Device | Predicate Device<br>(K001960) | Predicate Device<br>(K071266) | Predicate Device<br>(K003933) |
| Classification | Same as predicate | Class II<br>Lenses, Rigid Gas Permeable,<br>Daily Wear<br>21 CFR 886.5916 | Class II<br>Lenses, Rigid Gas Permeable,<br>Daily Wear<br>21 CFR 886.5916 | Class II<br>Lenses, Rigid Gas Permeable,<br>Daily Wear<br>21 CFR 886.5916 |
| Product Code | Same as predicate | HQD; MUW | HQD; MUW | HQD; MUW |
| FDA Group # | Same as predicate | Group # 3 Fluoro<br>Silicone Acrylate | Group # 3 Fluoro<br>Silicone Acrylate | Group # 3 Fluoro<br>Silicone Acrylate |
| USAN | Same as predicate | hexafocon A | hexafocon B | oprifocon A |
| Production<br>Method | Same as predicate | Lathe-Cut | Lathe-Cut | Lathe-Cut |
# Substantial Equivalence Matrix
{7}------------------------------------------------
| Actions | Same as predicate | The contact lenses act as a<br>refractive medium that<br>focus light rays from near<br>and distant objects on the<br>retina. (orthokeratology) | The contact lenses act as a<br>refractive medium that<br>focus light rays from near<br>and distant objects on the<br>retina. (orthokeratology) | The contact lenses act as a<br>refractive medium that<br>focus light rays from near<br>and distant objects on the<br>retina. (orthokeratology) |
|--------------------|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Same as predicate | Daily Wear | Daily Wear | Daily Wear |
| Indication for Use | Same as predicate | Indicated for the correction of<br>refractive ametropia (myopia,<br>hyperopia, astigmatism and<br>presbyopia) in aphakic and not<br>aphakic persons with non-<br>diseased eyes. Also, the lenses<br>may be prescribed in otherwise<br>non-diseased eyes that require<br>a gas permeable contact lens<br>for the management of<br>irregular corneal conditions<br>such as keratoconus, pellucid<br>marginal degeneration, or<br>following penetrating<br>keratoplasty or refractive (e.g.,<br>LASIK) surgery. Also,<br>indicated for daily wear in an<br>orthokeratology fitting<br>program for the temporary<br>reduction of myopia of up to<br>5.00 diopters in non-diseased<br>eyes. | Indicated for the correction of<br>refractive ametropia (myopia,<br>hyperopia, astigmatism and<br>presbyopia) in aphakic and<br>not aphakic persons with non-<br>diseased eyes. Also, the<br>lenses may be prescribed in<br>otherwise non-diseased eyes<br>that require a gas permeable<br>contact lens for the<br>management of irregular<br>corneal conditions such as<br>keratoconus, pellucid<br>marginal degeneration, or<br>following penetrating<br>keratoplasty or refractive<br>(e.g., LASIK) surgery. Also.<br>indicated for daily wear in an<br>orthokeratology fitting<br>program for the temporary<br>reduction of myopia of up to<br>5.00 diopters in non-diseased<br>eyes. | Indicated for the correction of<br>refractive ametropia (myopia,<br>hyperopia, astigmatism and<br>presbyopia) in aphakic and<br>not aphakic persons with non-<br>diseased eyes. Also, the<br>lenses may be prescribed in<br>otherwise non-diseased eyes<br>that require a gas permeable<br>contact lens for the<br>management of irregular<br>corneal conditions such as<br>keratoconus, pellucid<br>marginal degeneration, or<br>following penetrating<br>keratoplasty or refractive<br>(e.g., LASIK) surgery. Also,<br>indicated for daily wear in an<br>orthokeratology fitting<br>program for the temporary<br>reduction of myopia of up to<br>5.00 diopters in non-diseased<br>eyes. |
#### VII. PERFORMANCE DATA
## ~ Non-Clinical Studies ~
The non-clinical performance data to establish the safety and effectiveness of contact lenses manufactured from hexafocon A, hexafocon B, and oprifocon A has been addressed by reference to the predicate devices.
# ~ Clinical Studies ~
Clinical performance data to demonstrate the safety and effectiveness of contact lenses manufactured from hexafocon A. hexafocon B. and oprifocon A has been addressed in previous applications.
# VIII. CONCLUSIONS
### Substantial Equivalence
Information presented in this Premarket Notification establishes that eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear and eLens (hexafocon A, hexafocon B and oprifocon A) Rigid Gas Permeable Contact Lenses for Daily Wear Ortho-K are as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the proposed indication.
{8}------------------------------------------------
# Risks and Benefits
The risks of the subject device are the same as those normally attributed to the wearing of rigid gas permeable (RGP) daily wear contact lenses. The benefits to the patient are the same as those for other RGP contact lenses.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.