AI/ML, Software as a Medical Device, Pediatric, 3rd-Party Reviewed
Indications for Use
The Uscan is for non-invasive imaging of the human body and is intended for the following applications: Abdominal. Musculoskeletal, Pediatric, Small Organ, and Peripheral Vessel. Users must have ultrasound training for abdominal, musculoskeletal, pediatric, small organ, and peripheral vessel imaging. The Uscan can also be used to obtain an image of the bladder that is used to automatically determine bladder volume.
Device Story
Hand-held diagnostic ultrasound system; acquires echo data via linear or annular array transducers; displays B-mode images on off-the-shelf tablet (Android OS). Includes automated bladder volume measurement feature. Used in clinical settings by trained healthcare professionals for abdominal, musculoskeletal, pediatric, small organ, and peripheral vessel imaging; assists in needle/catheter placement. Device connects wirelessly; operates via battery. Output provides visual diagnostic information and automated measurements to support clinical decision-making and procedural guidance.
Clinical Evidence
Bench testing only. Performance evaluated via imaging performance and measurement accuracy testing. Compliance with ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-37, IEC 62304, IEC 62366, ISO 14971, and NEMA UD 2-2004 standards. No clinical data presented.
Indicated for non-invasive diagnostic ultrasound imaging of the human body, specifically abdominal, musculoskeletal, pediatric, small organ (prostate), and peripheral vessel applications, including needle/catheter guidance. Also indicated for automated bladder volume measurement. Requires trained clinician use.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 10th, 2018
EchoNous, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Street NW BUFFALO MN 55313
Re: K181574
Trade/Device Name: Uscan Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: June 14, 2018 Received: June 15, 2018
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jeff Rodges
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
### K181574
Device Name Uscan
### Indications for Use (Describe)
The Uscan is for non-invasive imaging of the human body and is intended for the following applications: Abdominal. Musculoskeletal, Pediatric, Small Organ, and Peripheral Vessel. Users must have ultrasound training for abdominal, musculoskeletal, pediatric, small organ, and peripheral vessel imaging.
The Uscan can also be used to obtain an image of the bladder that is used to automatically determine bladder volume.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{3}------------------------------------------------
510(k) Number (if known)
Device Name Uscan
System:
### TABLE 1.3-1: USCAN INDICATIONS FOR USE FORM
Uscan
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | P | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | | | | | | |
| | Small Organ (Specify) | P | | | | | | 1 |
| Fetal Imaging<br>& Other | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | P | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| Cardiac | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | | | | | | 2 |
| | Other (Specify) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
1: Small organ imaging is prostate.
2: Includes imaging to assist in the placement of needles and catheters in vascular structures.
All items marked "P" were previously cleared in 510(k) K160420.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
# Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{4}------------------------------------------------
510(k) Number (if known)
Device Name Uscan
### TABLE 1.3-2: USCAN INDICATIONS FOR USE FORM
| System: | Uscan |
|---------------|-------------------------------------------------------------|
| Transducer: | Uscan Sector Probe (B3-5) |
| Intended Use: | Diagnostic ultrasound imaging of the human body as follows: |
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | P | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | P | | | | | | |
| | Small Organ (Specify) | P | | | | | | 1 |
| | Neonatal Cephalic | | | | | | | |
| Fetal Imaging<br>& Other | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | P | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| Cardiac | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
Additional Comments:
1: Small organ imaging is prostate.
2: Includes imaging to assist in the placement of needles and catheters in vascular structures.
All items marked "P" were previously cleared in 510(k) K160420.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{5}------------------------------------------------
510(k) Number (if known)
Device Name Uscan
### TABLE 1.3-3: USCAN INDICATIONS FOR USE FORM
| System: | Uscan |
|---------------|-------------------------------------------------------------|
| Transducer: | Uscan Linear Probe (L14-6) |
| Intended Use: | Diagnostic ultrasound imaging of the human body as follows: |
| Clinical Application | | Mode of Operation | | | | | | |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Specify) | | | | | | | |
| | Neonatal Cephalic | | | | | | | |
| Fetal Imaging<br>& Other | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal<br>(Conventional) | | | | | | | |
| | Musculo-skeletal<br>(Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| Cardiac | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | | | | | | 2 |
| | Other (Specify) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
X Prescription Use (Part 21 CFR 801 Subpart D)
Additional Comments:
1: Small organ imaging is prostate.
2: Includes imaging to assist in the placement of needles and catheters in vascular structures.
Type of Use (Select one or both, as applicable)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{6}------------------------------------------------
### 510(k) Summary
#### 1. Submitter:
EchoNous, Inc. 8310 154th Ave NE Bldg B. Ste 200 Redmond, WA 98052 USA
#### 2. Contact Person:
Trish Liau Regulatory Affairs Manager Telephone: (425) 402-4044 E-mail: patricia.liau@echonous.com
#### 3. Date Prepared:
March 30, 2018
- 4. Proprietary/Marketed Names: Uscan (subject to change)
#### ട. Common/Usual Name: Diagnostic ultrasound system with accessories
#### Classification: 6.
Regulatory Class: II Review Category: Tier II Classification Panel: Radiology
| Name | 21 CFR Number | Product Code |
|---------------------------------------|---------------|--------------|
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
#### 7. Predicate Devices:
Primary predicate: Uscan Ultrasound System (K160420) Reference predicate: SonoSite SII Ultrasound System (K162045)
#### Device Description: 8.
The Uscan is a hand-held, diagnostic ultrasound system with an on-screen display. Its purpose is to acquire ultrasound echo data and display it in B-Mode on an off-the-shelf display. Automated bladder volume measurements are supported.
{7}------------------------------------------------
#### 9. Intended Use:
The Uscan is for non-invasive imaging of the human body and is intended for the following applications: Abdominal, Musculoskeletal, Pediatric, Small Organ, and Peripheral Vessel. Users must have ultrasound training for abdominal, musculoskeletal, pediatric, small organ, and peripheral vessel imaging.
The Uscan can also be used to obtain an image of the bladder that is used to automatically determine bladder volume.
#### 10. Technological Characteristics:
The Uscan and SII ultrasound systems are Track 3 devices that employ the same fundamental scientific technology. A comparison table is provided below.
| Feature | Uscan Ultrasound System<br>(This submission) | Uscan Ultrasound System<br>(K160420) | SonoSite SII Ultrasound<br>System<br>(K162045) |
|---------------------------------|----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Diagnostic ultrasound imaging<br>of the human body | Diagnostic ultrasound imaging<br>of the human body | Diagnostic ultrasound imaging<br>or fluid flow analysis of the<br>human body |
| Indications for<br>Use | Abdominal<br>Pediatric<br>Small Organ (prostate) | Abdominal<br>Pediatric<br>Small Organ (prostate) | Opthalmic<br>Fetal - OB/GYN<br>Abdominal<br>Pediatric<br>Small Organ (breast, thyroid,<br>testicle, prostate)<br>Neonatal Cephalic<br>Adult Cephalic<br>Trans-Rectal<br>Trans-Vaginal |
| | Musculo-skeletal<br>(Conventional) | Musculo-skeletal<br>(Conventional) | Musculo-skeletal<br>(Conventional)<br>Musculo-skeletal (Superficial)<br>Cardiac Adult<br>Cardiac Pediatric<br>Peripheral Vessel<br>Needle guidance |
| | Peripheral Vessel<br>Needle guidance | | |
| Transducer<br>Types | Linear Array | Annular Array | Linear Array<br>Curved Linear Array<br>Intracavitary<br>Phased Array |
| Transducer<br>Frequency | 3.0 - 14.0 MHz | 3.0 - 7.0 MHz | 1.0 - 15.0 MHz |
| Modes of<br>Operation | 2D / B-mode | 2D / B-mode | 2D / B-mode<br>Tissue Harmonic Imaging<br>M-mode<br>Color M-mode<br>Color Power Doppler<br>Combination Modes<br>Velocity Color Doppler |
| PW Doppler | Not available | Not available | Not available |
| CW Doppler | Not available | Not available | Not available |
| Feature | Uscan Ultrasound System<br>(This submission) | Uscan Ultrasound System<br>(K160420) | SonoSite SII Ultrasound<br>System<br>(K162045) |
| Patient<br>Contact<br>Materials | Transducers:<br>ABS+PC (acrylonitrile<br>butadiene styrene +<br>polycarbonate) | Transducers: | Transducers:<br>Acrylonitrile-butadiene-<br>styrene (ABS) |
| | Cycoloy | Cycoloy | Cycoloy<br>Polycarbonate (PC) |
| | Polymethyl Pentene based<br>Olefin Copolymer | Polymethyl Pentene based<br>Olefin Copolymer | Polymethyl Pentene based<br>Olefin Copolymer |
| | Silicone Rubber RTV<br>Thermoplastic polyurethane | Thermoplastic polyurethane | Polysulfone<br>Poly-Vinyl-Chloride (PVC)<br>Silicone Rubber<br>Silicone Rubber RTV |
| | | | Urethane<br>Needle Guides:<br>Acetal copolymer<br>Acrylonitrile-butadien-styrene<br>(ABS) |
| System<br>Characteristics | Uscan:<br>Handheld tablet (off-the-shelf)<br>9.7" Liquid Crystal Display<br>(LCD)<br>Operating system: Android | Uscan:<br>Handheld tablet (off-the-shelf)<br>7.86" Liquid Crystal Display<br>(LCD)<br>Operating system: Android | SII:<br>Handheld display and control<br>12.1" Liquid Crystal Display<br>(LCD)<br>Operating system: Windows<br>CE |
| | Uscan ultrasound software<br>running as an "app" on tablet | Uscan ultrasound software<br>running as an "app" on tablet | |
| | 1 Micro USB port | 1 Micro USB port | 3 USB ports |
| | Dimensions: 9.34"(H) x<br>6.65"(W) x 0.24"(L)<br>Weight: 0.39 lbs | Dimensions: 8.15"(H) x<br>5.39"(W) x 0.31"(L)<br>Weight: 0.77 lbs | Dimensions: 4.8"(H) x<br>11.5"(W) x 17.6"(L)<br>Weight: 12.6 lbs |
| | System operates via battery | System operates via battery | System operates via battery or<br>AC power |
| | Input: 100-240 VAC, 50/60<br>Hz<br>Output: 5VDC, 2 A max | Input: 100 - 240 VAC, 50/60<br>Hz<br>Output: 12VDC, 1.5 A max | Input: 100 - 240V options,<br>50/60 Hz<br>Output: 15VDC |
| | Distance calculations | Distance, volume calculations | Various obstetrical, cardiac,<br>volume, and M-mode<br>measurement and calculation<br>packages |
| | Wireless networking | Wireless networking | Wireless networking |
| 510(k) Track | Track 3 | Track 3 | Track 3 |
{8}------------------------------------------------
{9}------------------------------------------------
#### 11. Basis for Substantial Equivalence:
### Summary of Non-Clinical Tests
The Uscan has been designed and evaluated to comply with the following FDA recognized standards.
| Reference No. | Title |
|----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-1 | AAMI / ANSI / ISO 10993-1:2009/(R)2013, Biological evaluation of<br>medical devices – Part 1: Evaluation and testing within a risk management<br>process |
| IEC 60601-1 | AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012<br>and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment -<br>Part 1: General requirements for basic safety and essential performance<br>(IEC 60601-1:2005, MOD) |
| IEC 60601-1-2 | AAMI / ANSI / IEC 60601-1-2:2014, Medical electrical equipment - Part<br>1-2: General requirements for basic safety and essential performance -<br>Collateral standard: Electromagnetic compatibility - Requirements and<br>tests |
| IEC 60601-1-6 | IEC 60601-1-6 Edition 3.1 2013-10, Medical electrical equipment - Part<br>1-6: General requirements for basic safety and essential performance -<br>Collateral standard: Usability |
| IEC 60601-2-37 | IEC 60601-2-37:2015 Edition 2.1, Medical electrical equipment – Part 2-<br>37: Particular requirements for the basic safety and essential performance<br>of ultrasonic medical diagnostic and monitoring equipment |
| IEC 62304 | AAMI / ANSI / IEC 62304:2006, Medical device software - Software life<br>cycle processes |
| IEC 62366 | AAMI / ANSI / IEC 62366-1:2015, Medical devices – Part 1: Application<br>of usability engineering to medical devices |
| ISO 14971 | ISO 14971:2007, Medical devices - Application of risk management to<br>medical devices |
| NEMA UD 2-2004 | NEMA UD 2-2004 (R2009), Acoustic Output Measurement Standard for<br>Diagnostic Ultrasound Equipment |
The Uscan has been bench tested for imaging performance and measurement accuracy, with tests showing the Uscan imaging performance and measurement accuracy to be substantially equivalent to the predicate devices.
Verification and validation reports, traceability, and risk analysis demonstrate the Uscan operates as intended and risks mitigated have been verified.
The conclusion from the testing is the device is safe and effective for its intended use, and performs as well or better than the predicate devices.
The Uscan system and predicates meet FDA requirements for Track 3 devices, share indications for use, have biosafety equivalence, and conform to applicable electromedical device safety standards. EchoNous, Inc. believes the Uscan is substantially equivalent with regard to safety and effectiveness to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.