EliA M2 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to M2 in human serum and plasma (Li-heparin, EDTA) to aid in the clinical diagnosis of primary biliary cirrhosis in conjunction with other laboratory and clinical findings. EliA M2 uses the EliA IgG method on the instrument Phadia 2500/5000.
Device Story
EliA M2 is an in vitro diagnostic immunoassay for detecting IgG antibodies against M2 protein in human serum or plasma. The device uses polystyrene wells coated with native pyruvate dehydrogenase complex and recombinant M2-antigen. Patient samples are processed on the Phadia 2500/5000 automated instrument system. The assay follows an enzyme-linked immunosorbent assay (ELISA) principle: patient antibodies bind to the coated antigen; enzyme-labeled anti-IgG conjugate is added; after washing, a development solution is added to produce a fluorescent signal proportional to the concentration of specific IgG. The system automatically measures fluorescence and compares results against a calibration curve. Used in clinical laboratories by trained personnel to aid in the diagnosis of primary biliary cirrhosis. Results are interpreted by clinicians alongside other laboratory and clinical findings to support patient management.
Clinical Evidence
Bench testing only. Precision study (n=84 replicates per sample) showed total imprecision (%CV) ranging from 6.2% to 15.8%. Linearity was confirmed across the reportable range (0.5–220 U/mL). Instrument comparison (n=103 samples) between Phadia 250 and Phadia 2500/5000 showed high correlation (Passing-Bablok slope 1.04, R=0.995) and 99% total percent agreement. Reference range study (n=400 healthy donors) established a 95th percentile of 1.9 U/mL.
Technological Characteristics
Polystyrene wells coated with native pyruvate dehydrogenase complex and recombinant M2-antigen. Automated immunofluorescence assay. Uses Phadia 2500/5000 instrument. Reagents include EliA Sample Diluent, IgG Conjugate (beta-galactosidase labeled), calibrators, and curve controls. Connectivity: Automated system with integrated barcode reader and belt-driven reagent loading. Software-controlled processing.
Indications for Use
Indicated for in vitro semi-quantitative measurement of IgG antibodies to M2 in human serum and plasma to aid in clinical diagnosis of primary biliary cirrhosis. For prescription use only. Not for point-of-care use.
Regulatory Classification
Identification
An antimitochondrial antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the antimitochondrial antibodies in human serum. The measurements aid in the diagnosis of diseases that produce a spectrum of autoantibodies (antibodies produced against the body's own tissue), such as primary biliary cirrhosis (degeneration of liver tissue) and chronic active hepatitis (inflammation of the liver).
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM
A. 510(k) Number:
K181556
B. Purpose for Submission:
Adding a previously cleared assay on a new instrument platform (Phadia 2500/5000)
C. Measurand:
IgG autoantibodies specific to native pyruvate dehydrogenase complex from mitochondria and recombinant M2-antigen (M2)
D. Type of Test:
Automated semi-quantitative solid-phase fluoroimmunoassay
E. Applicant:
Phadia US, Inc.
F. Proprietary and Established Name:
EliA M2 Immunoassay
G. Regulatory Information:
1. Regulation section:
21 CFR §866.5090, Antimitochondrial antibody immunological test system
2. Classification:
Class II
3. Product code:
DBM, Antimitochondrial Antibody, Indirect Immunofluorescent, Antigen, Control
4. Panel:
Immunology (82)
{1}
H. Intended Use:
1. Intended use:
EliA M2 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to M2 in human serum and plasma (Li-heparin, EDTA) to aid in the clinical diagnosis of primary biliary cirrhosis in conjunction with other laboratory and clinical findings. EliA M2 uses the EliA IgG method on the instrument Phadia 2500/5000.
2. Indication for use:
Same as intended use
3. Special conditions for use statement:
For prescription use only
4. Special instrument requirements:
For use on the Phadia 2500 and Phadia 5000 instruments
I. Device Description:
EliA uses a modular reagent system. The assay-specific, method-specific and general reagents are packaged and purchased as separate units. The reagents on Phadia 2500 and Phadia 5000 are identical; they are only filled in different containers.
EliA Assay-specific reagent:
- EliA M2 Wells are coated with native pyruvate dehydrogenase complex from mitochondria and recombinant M2-antigen–4 carriers (12 wells each), ready to use
EliA Method-specific reagents:
- EliA IgG Calibrator Strips: Human IgG (0, 4, 10, 20, 100, 600 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide – 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use
- EliA IgG Curve Control Strips: Human IgG (20 µg/L) in PBS containing BSA, detergent and 0.095% (w/v) sodium azide – 5 strips, 6 single-use vials per strip, 0.3 mL each, ready to use
- EliA Sample Diluent: PBS containing BSA, detergent and 0.095% (w/v) sodium azide – 6 bottles, 48 mL each, ready to use; or 6 bottles, 400 mL each, ready to use
- EliA IgG Conjugate 50 or 200: β-Galactosidase labeled anti-IgG (mouse monoclonal antibodies) in PBS containing BSA and 0.06% (w/v) sodium azide – 6 wedge shaped bottles, 5 mL each, ready to use; or 6 wedge shaped bottles, 19 mL each, ready to use
- EliA IgG Calibrator Well: Coated with mouse monoclonal antibodies – 4 carriers (12 wells each), ready to use
{2}
EliA General reagents:
- Development Solution (0.01 %4-Methylumbelliferyl-β-D-galacto-side, <0.0010% preservative), ready for use
- Stop Solution (4% Sodium Carbonate), ready for use
- Dilution Wells (high density polyethylene wells), 50 carriers, ready to use
- Pipette Tips in Racks (polyethylene tips), 24 racks × 160 tips, ready for use
- Washing Solution (information in separate Washing Solution package insert)
Phadia 2500 and Phadia 5000 are identical instruments except for sample throughput. The Phadia 2500 consists of one process module (two process lines), whereas Phadia 5000 consists of two process modules (2 x 2 process lines). Instrument operation is handled by onboard Instrument Software (ISW). Data output is administered by Information Data Manager (IDM). All steps of an assay are performed within a single process line. Thus, study protocols used for Phadia 2500 are also valid for Phadia 5000.
J. Substantial Equivalence Information:
1. Predicate device name:
EliA M2 Immunoassay on Phadia 250
2. Predicate 510(k) number:
K141375
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device
EliA M2 on Phadia 2500/5000 | Predicate
EliA M2 on Phadia 250 |
| Intended Use/
Indications for Use | EliA M2 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to M2 in human serum and plasma (Li-heparin, EDTA) to aid in the clinical diagnosis of primary biliary cirrhosis in conjunction with other laboratory and clinical findings.
EliA M2 uses the EliA IgG method on the instrument Phadia 2500/5000. | EliA M2 is intended for the in vitro semi-quantitative measurement of IgG antibodies directed to M2 in human serum and plasma (heparin, EDTA) to aid in the clinical diagnosis of primary biliary cirrhosis in conjunction with other laboratory and clinical findings.
EliA M2 uses the EliA IgG method on the instrument Phadia 250. |
| Assay Type | Automated fluoroenzyme immunoassay | Same |
| Assay Format | Semi-quantitative | Same |
| Assay Set-up | Random access | Same |
| Solid Phase | Polystyrene microwells | Same |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | DeviceEliA M2 on Phadia 2500/5000 | PredicateEliA M2 on Phadia 250 |
| Measurand | Autoantibodies to M2 | Same |
| Coating Antigens | Native pyruvate dehydrogenase complex from mitochondria and recombinant M2-antigen | Same |
| Sample Volume | 90 μL (20 μL of undiluted (sample)) | Same |
| Incubation Temperature | 37°C (controlled) | Same |
| Incubation Times | Diluted patient samples: 30 min. Conjugate: 28 min. Development Solution: 39 min. | |
| Time to 1st result | ~2 h | Same |
| Detection Antibody (Conjugate) Volume | 90 μL | Same |
| Substrate/Chromogen (Development) Volume | 90 μL | Same |
| Stop Solution Volume | 200 μL | |
| Calibration Strip | Human IgG (0, 4, 10, 20, 100, 600 μg/L) | Same |
| Calibration Frequency | 28 days | Same |
| Assay Cut-off/ Results Interpretation | < 4 U/mL: Negative4 - 6 U/mL: Equivocal> 6 U/mL: Positive | Same |
| Result Calculation Software | Phadia Information Data Manager (IDM) | Same |
| Differences | | |
| --- | --- | --- |
| Item | DeviceEliA M2 on Phadia 2500/5000 | PredicateEliA M2 on Phadia 250 |
| Instrument Platform | Phadia 2500/5000 | Phadia 250 |
| Measuring Range | 0.5 – 276 U/mL | 0.8 – 220 U/mL |
| Barcode Reader | On the Phadia 2500/5000, the reagents are on a moving belt which conveys them past the barcode reader. The lot-specific information will be read automatically by the instrument during loading. | The Phadia 250 instrument has a built-in barcode reader at the front of the instrument, but the operator needs to scan the barcodes manually by showing the reagents to the barcode reader. Alternatively, the operator can also enter the characters below the barcode manually. |
| Loading of EliA Carriers | The Phadia 2500/5000 instruments do not have a Loading Tray. The EliA carriers are loaded into racks which are directly transferred to the | The EliA carriers are loaded manually on the Loading Tray from where they can be processed directly or transferred to the cooled |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Device
EliA M2 on Phadia 2500/5000 | Predicate
EliA M2 on Phadia 250 |
| | cooled storage compartment. | storage compartment. |
| Process Time/Time to Patient Result | Phadia 2500/5000 instruments process two Wells in parallel in 48 seconds. Phadia 2500/5000 provides the results at a 24 seconds interval. | Phadia 250 needs 1 minute to process one Well. Phadia 250 provides the results at a one minute interval. |
| Daily Sample Throughput | Phadia 2500: ~2500 tests/day
Phadia 5000: ~5000 tests/day | ~250 tests/day |
| Sample Dilution | Use disposable Pipette Tips in Racks for pipetting samples in Dilution Well | Use steel pipette to dilute the samples in Dilution Plates |
| Carry-over due to Reagent Re-use | No carry-over from samples to conjugate because Phadia 2500/5000 instruments use disposable tips for pipetting samples and a separate pipette for the conjugate. | Conjugate should not be reused due to risk of carryover contamination. |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A3; Evaluation of Precision Performance of Quantitative Measurement Methods; September 2014
CLSI EP6-A Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; April 2003
CSLI EP17-A2: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline
L. Test Principle:
EliA tests are fluorescence immunoassays for the detection and measurement of human antibodies based on EliA solid-phase components, which contain specific antigens for the antibodies to be measured.
EliA M2: Polystyrene EliA wells are coated with native pyruvate dehydrogenase complex from mitochondria and recombinant M2-antigen. If present in the patient's specimen, antibodies to M2 bind to their specific antigen. After washing away non-bound antibodies, enzyme-labeled antibodies against human IgG antibodies (EliA IgG Conjugate) are added to form an antibody-conjugate complex. After incubation, non-bound conjugate is washed away and the bound complex is incubated with a Development Solution. After stopping the reaction, the fluorescence in the reaction mixture is measured. The higher the value of fluorescent signal, the higher the amount of antibody bound and detected in the sample tested. To evaluate test results, the response for patient samples is compared directly to the response for calibrators.
{5}
Phadia 2500/5000: The Phadia 2500 and Phadia 5000 instruments are automated platforms for EliA test procedures from sample and reagent handling up to processing of results.
## M. Performance Characteristics:
### 1. Analytical performance:
All results presented below met the manufacturer's pre-determined acceptance criteria for all analytical performance studies.
#### a. Precision/Reproducibility:
The precision of the assay was assessed in a study with a total of 21 runs, one run/day over a period of seven days. Each sample was tested in four replicates/run giving a total of 84 replicates per sample (3 instruments × 7 runs x 4 replicates = 84). The data was calculated against the calibration curve from Day 1. Five native patient serum samples were tested: one negative, two in the equivocal range, and two positive specimens. All samples were tested on all three Phadia 2500/5000 instruments.
One lot of EliA M2 reagents was used in the study because inter-lot-variation was determined in the predicate device K141375. The results for all three instruments are summarized in the table below:
| Mean (U/mL) | Within-Run | | Between-Run | | Between-Instrument | | Total Imprecision | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | SD | %CV | SD | %CV | SD | %CV | SD | %CV |
| 1.7 | 0.1 | 7.3 | 0.1 | 4.8 | 0.2 | 13.1 | 0.3 | 15.8 |
| 4.0 | 0.2 | 4.3 | 0.1 | 2.9 | 0.3 | 6.6 | 0.3 | 8.4 |
| 5.9 | 0.2 | 3.7 | 0.1 | 2.5 | 0.3 | 4.2 | 0.4 | 6.2 |
| 74.8 | 2.3 | 3.1 | 1.7 | 2.2 | 4.0 | 5.3 | 4.9 | 6.5 |
| 175.9 | 8.0 | 4.5 | 6.8 | 3.9 | 12.1 | 6.9 | 16.0 | 9.1 |
#### b. Linearity/assay reportable range:
Four patient serum samples were diluted in EliA Sample Diluent and tested in triplicates with one lot of EliA M2 reagents and one set of system reagents on Phadia 2500/5000. The results of the dilutions were compared with their expected values. The ratio observed/expected (O/E) was calculated. Mean observed value was used in the calculation. The results of the linear regression analysis are summarized below:
| Sample | Dilution range (U/mL) | Slope (95% CI) | Intercept (95% CI) | R² | %CV Range |
| --- | --- | --- | --- | --- | --- |
| 1 | 0.7 – 48.3 | 0.99 (0.97 to 1.01) | -0.32 (-0.78 to 0.14) | 1.00 | 1.2 – 5.4 |
{6}
| Sample | Dilution range (U/mL) | Slope (95% CI) | Intercept (95% CI) | R² | %CV Range |
| --- | --- | --- | --- | --- | --- |
| 2 | 2.1 – 211.3 | 1.02 (0.99 to 1.05) | 1.90 (-1.01 to 4.81) | 1.00 | 0.5 – 3.5 |
| 3 | 5.7 – 253.2 | 1.03 (0.97 to 1.08) | 2.36 (-3.78 to 8.5) | 1.00 | 2.1 – 6.6 |
| 4 | 0.5 – 16.6 | 1.02 (0.98 to 1.07) | 0.13 (-0.18 to 0.44) | 1.00 | 1.0 – 4.1 |
The reportable range (Limit of Detection to upper limit) is from 0.5 to 220 U/mL. The measuring range (Limit of Quantitation to upper limit) is from 0.8 to 220 U/mL.
The following statements are included in the package insert: "Please note that concentration values between LoD and LoQ may show a higher uncertainty" and "Please note that due to differing binding characteristics of the antibodies in patient samples, not all sera can be diluted linearly within the measuring range."
High dose hook effect: The hook effect was previously reviewed in K141375.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
i) Traceability:
The EliA IgG calibration method and traceability was previously reviewed in K141375.
Controls:
Controls are sold separately from the assay. The EliA M2 Positive is prepared from selected pooled human sera and contains IgG antibodies to M2. The EliA IgG/IgM/IgA Negative Control is prepared from selected pooled sera from normal, healthy donors. The controls are pre-diluted and ready for use. Each EliA Control package contains a Control Certificate listing predefined acceptance criteria for the EliA products the Controls can be used with.
ii) Kit Stability:
Shelf-life stability, open-stability and on-board stability were previously reviewed in K141375.
d. Detection limit:
The limit of blank (LoB) is based on 72 blank replicates (1 blank sample tested in 12 replicates × one run/day × six days). The limit of detection (LoD) is based on 72 replicates (12 replicates × one run/day × six days) for each of the five low-level serum samples tested. The limit of quantitation (LoQ) is based on the concentration of one of the five low-level serum samples tested with a %CV closest but below the precision target level of 20% CV.
{7}
The results are summarized in the table below:
| | LoB | LoD | LoQ |
| --- | --- | --- | --- |
| EliA M2 (U/mL) | 0.3 | 0.5 | 0.8 |
e. Analytical specificity:
i) Endogenous Interference:
Interference by endogenous substances was previously reviewed in K141375.
ii) Cross-reactivity:
Cross-reactivity was previously reviewed in K141375.
iii) Carry-over:
The Phadia 2500/5000 instruments use disposable tips for pipetting samples and a separate pipette for the conjugate, therefore carry-over from samples to conjugate was not evaluated.
f. Assay cut-off:
The assay cut-offs are the same as in the predicate device (refer to K141375) and are summarized in the table below:
| Assay cut-off | Interpretation |
| --- | --- |
| <4 U/mL | Negative |
| 4 – 6 U/mL | Equivocal |
| >6 U/mL | Positive |
2. Comparison studies:
a. Method comparison with predicate device:
Instrument comparison was performed with the predicate device, see 2c. below.
b. Matrix comparison:
Matrix comparison between serum and plasma (Li-Heparin, and EDTA) was reviewed in K141375.
c. Instrument comparison:
The purpose of this study was to evaluate conformance and show comparability of EliA M2 on the Phadia 250 instrument versus the Phadia 2500/5000 instrument. A
{8}
total of 103 samples (84 positive, 12 negative and 7 equivocal) were run in singlicate on one Phadia 250 instrument and one Phadia 2500/5000 instrument for comparison. Only samples inside the measuring range were included in the calculations. Results were analyzed by Passing-Bablok regression. The results are summarized below:
| Range tested (U/mL) | Slope (95% CI) | Intercept (95% CI) | *R |
| --- | --- | --- | --- |
| 0.9 – 211.6 | 1.04 (1.02 to 1.06) | -0.14 (-0.46 to 0.03) | 0.995 |
| *Pearson correlation coefficient | | | |
No sample switched from negative to positive or vice versa. One sample in the equivocal range (5.7 U/mL) tested positive (6.3 U/mL), and one negative sample (3.8 U/mL) yielded test result in the equivocal range (4.4 U/mL) on the Phadia 2500/5000 instrument. The results are summarized below:
| | EliA M2 on Phadia 250 | | | Total | |
| --- | --- | --- | --- | --- | --- |
| | | Positive >6 | Equivocal:4 6 | | Negative ≤ 4 |
| EliA M2 on Phadia 2500/5000 | Positive: >6 | 84 | 1 | 0 | 85 |
| | Equivocal: 4 6 | 0 | 6 | 1 | 7 |
| | Negative: <4 | 0 | 0 | 11 | 11 |
| Total | | 84 | 7 | 12 | 103 |
Positive percent agreement, negative percent agreement, and total percent agreement were evaluated with equivocal results considered positive in the table below:
| Equivocal EliA M2 results considered as positive | | | | |
| --- | --- | --- | --- | --- |
| | EliA M2 on Phadia 250 | | Total | |
| | | Positive: ≥ 4 | | Negative: < 4 |
| EliA M2 on Phadia 2500/5000 | Positive: ≥ 4 | 91 | 1 | 92 |
| | Negative: < 4 | 0 | 11 | 11 |
| Total | | 91 | 12 | 103 |
| Positive percent agreement: 100% (91/91) 95% CI: 96.0 – 100%
Negative percent agreement: 91.7% (11/12) 95% CI: 61.5 – 99.8%
Total percent agreement: 99.0% (103/103) 95% CI: 94.7 – 100% | | | | |
{9}
Results when equivocal results are considered negative are presented in the table below:
| Equivocal EliA M2 results considered as negative | | | | |
| --- | --- | --- | --- | --- |
| | EliA M2 on Phadia 250 | | Total | |
| | | Positive: >6 | | Negative: ≤6 |
| EliA M2 on Phadia 2500/5000 | Positive: > 6 | 84 | 1 | 85 |
| | Negative: ≤ 6 | 0 | 18 | 18 |
| Total | | 84 | 19 | 103 |
| Positive percent agreement: 100% (84/84) 95% CI: 95.7 – 100%
Negative percent agreement: 95.5% (18/19) 95% CI: 77.2 – 99.9%
Total percent agreement: 99.0% (102/103) 95% CI: 94.9 – 100% | | | | |
3. Clinical studies:
Clinical performance was previously reviewed in K141375.
a. Clinical sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
4. Clinical cut-off:
See assay cut-off
5. Expected values/Reference range:
A total of 400 apparently healthy blood donor samples from a Caucasian population matched by age and gender were tested on the Phadia 2500/5000 instrument to evaluate expected values in the normal population. Seven samples were found in the equivocal range and two were found with positive results. There was no significant difference between male and female. Expected values may vary depending on the population tested. The results are summarized below:
| | U/mL |
| --- | --- |
| Mean | 1.1 |
| Median | 0.9 |
| Rang7e | 0.2 33.8 |
| 95th percentile | 1.9 |
| 99th percentile | 5.2 |
{10}
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
11
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.