← Product Code [HIH](/productcode/HIH) · K181545

# Hysteroscope System (K181545)

_Shanghai AnQing Medical Instrument Co., Ltd. · HIH · May 10, 2019 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181545

## Device Facts

- **Applicant:** Shanghai AnQing Medical Instrument Co., Ltd.
- **Product Code:** [HIH](/productcode/HIH.md)
- **Decision Date:** May 10, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1690
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

The Hysteroscope System is intended to be used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures. Note: Hysteroscopes are used as tools to access the uterine cavity and of themselves, a method of surgery. Generally recognized indications for diagnostic hysteroscopy include: - Abnormal bleeding - · Infertility and pregnancy wastage - · Evaluation of abnormal hysterosalpingogram - · Intrauterine foreign body - · Amenorrhea - · Pelvic pain Generally recognized indications for operative hysteroscopy include: - · Directed endometrial biopsy - · Polypectomy - Submucous myomectomy - · Transection of intrauterine adhesions - · Transection of intrauterine septa - · Endometrial ablation

## Device Story

Hysteroscope System consists of sterile, single-use rigid hysteroscope and video processor. CMOS chip at distal end of scope collects anatomical images; transmitted via processor to external monitor for physician viewing. Used in clinical settings by physicians to visualize cervical canal and uterine cavity during diagnostic and surgical procedures. Provides real-time visual guidance for interventions like biopsy, polypectomy, and myomectomy. Benefits include direct visualization of intrauterine pathology, facilitating minimally invasive gynecological procedures.

## Clinical Evidence

Bench testing only. Biocompatibility (ISO 10993-5, -10), sterilization validation (ISO 11135), electrical safety/EMC (IEC 60601-1, -2-18, -1-2), optical performance (ISO 8600 series), and mechanical performance (leaking, bending, irrigation) met predefined acceptance criteria.

## Technological Characteristics

Rigid hysteroscope with CMOS sensor; LED illumination. Models: GSA01-B (straight, no operative channel), GSB04-B (straight, operative channel), GSB14-B (curved tip, operative channel). Working length 260mm. Power: 5V. Sterilization: Ethylene Oxide (EO). Software: Moderate level of concern.

## Regulatory Identification

A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.

## Predicate Devices

- LiNA OperaScope ([K171113](/device/K171113.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".

May 10, 2019

Shanghai AnQing Medical Instrument CO., Ltd. Lina Fei Quality Manager 150 Cailun Road, Zhangjiang High-Tech Park Shanghai, 201210 China

Re: K181545

> Trade/Device Name: Hysteroscope System Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH Dated: April 3, 2019 Received: April 10, 2019

Dear Lina Fei:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Sharon M. Andrews Assistant Division Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181545

Device Name Hysteroscope System

Indications for Use (Describe)

The Hysteroscope System is intended to be used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.

Note: Hysteroscopes are used as tools to access the uterine cavity and of themselves, a method of surgery.

Generally recognized indications for diagnostic hysteroscopy include:

- Abnormal bleeding
- · Infertility and pregnancy wastage
- · Evaluation of abnormal hysterosalpingogram
- · Intrauterine foreign body
- · Amenorrhea
- · Pelvic pain

Generally recognized indications for operative hysteroscopy include:

- · Directed endometrial biopsy
- · Polypectomy
- Submucous myomectomy
- · Transection of intrauterine adhesions
- · Transection of intrauterine septa
- · Endometrial ablation

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)    </span> |
|------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)    </span>                 |

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary – K181545

# I. Submitter

Device submitter: Shanghai AnQing Medical Instrument CO., Ltd. Room 202, 2nd Floor, No.7 Building & Room 202, 2nd Floor No.4 Building, 150 Cailun Road, Zhangjiang High-Tech Park, 201210, Shanghai, China

Contact person: Wen Shi Quality Manager Phone: +86 135 8580 5802 Fax: +86 21 5019 1132 – 8004 Email: wenshi@innovexmed.com.cn

Date prepared: May 9, 2019

### II. Device

Trade Name of Device: Hysteroscope System Common Name: Hysteroscope (And Accessories) Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH Product Code Name: Hysteroscope and Accessories

#### III. Predicate Devices

| Trade name:             | LiNA OperaScope                              |
|-------------------------|----------------------------------------------|
| Common name:            | Hysteroscope                                 |
| Classification:         | 21 CFR 884.1690 Hysteroscope and Accessories |
| Product Code:           | HIH, Hysteroscope and Accessories            |
| Premarket Notification: | K171113                                      |

The predicate device has not been subject to a design-related recall.

# IV. Device description

The Hysteroscope System consists of a sterile single-use disposable Rigid Hysteroscope and video processor for clinical image processing. The Rigid Hysteroscope (available in models GSA01-B, GSB04-B and GSB14-B) are used to permit direct viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. Model GSA01-B is intended for diagnostic hysteroscopy only, and has a rigid 

{4}------------------------------------------------

shaft without operative channel (injection channel only). Models GSB04-B and GSB14-B are intended for both diagnostic and operative hysteroscopy, and have an operative (insertion) channel. GSB04-B has a rigid shaft, and GSB14-B has a curved tip. The specifications for the three models are as follows:

| Model   | Shape                | Scope diameter        | Working length | Minimum<br>insertion<br>channel<br>diameter | Injection<br>channel<br>Diameter | Tip Curve  | Curved Tip<br>length |
|---------|----------------------|-----------------------|----------------|---------------------------------------------|----------------------------------|------------|----------------------|
| GSA01-B | Straight             | Φ4.0mm<br>+0.0/-0.3mm | 260mm±8mm      | N/A                                         | 0.8mm<br>+0.2/-0.0mm             | N/A        | N/A                  |
| GSB04-B | Straight             | Φ6.4mm<br>+0.0/-0.3mm | 260mm±8mm      | 2.3mm<br>+0.2/-0.0mm                        | 1.1mm<br>+0.2/-0.0mm             | N/A        | N/A                  |
| GSB14-B | Curved<br>distal tip | Φ6.4mm<br>+0.0/-0.3mm | 260mm±8mm      | 2.3mm<br>+0.2/-0.0mm                        | 1.1mm<br>+0.2/-0.0mm             | 22°<br>±5° | 25mm<br>±1.3mm       |

The video processor provides power and processes the images from the Rigid Hysteroscopes. Anatomical images are collected via a CMOS chip at the distal end of the hysteroscope and are transmitted through the video processor to a monitor.

#### V. Indications for use

The Hysteroscope System is intended to be used to permit viewing of the cervical canal and uterine cavity for the purpose of performing diagnostic and surgical procedures.

Note: Hysteroscopes are used as tools to access the uterine cavity and are not, in and of themselves, a method of surgery.

Generally recognized indications for diagnostic hysteroscopy include:

- Abnormal bleeding
- Infertility and pregnancy wastage
- Evaluation of abnormal hysterosalpingogram
- Intrauterine foreign body
- Amenorrhea
- Pelvic pain

{5}------------------------------------------------

Generally recognized indications for operative hysteroscopy include:

- · Directed endometrial biopsy
- · Polypectomy
- · Submucous myomectomy
- Transection of intrauterine adhesions
- · Transection of intrauterine septa
- Endometrial ablation

#### VI. Comparison of technological characteristics with the predicate devices

The Hysteroscope System has the same intended use as the predicate device. The Hysteroscope System and the predicate device are designed to provide real-time images to the physician in order to facilitate diagnostic and therapeutic procedures in the cervical canal and uterus. As summarized in the table below, there are differences in technological characteristics between the subject and predicate device. The technological differences between the Hysteroscope System and predicate device do not raise different questions of safety or effectiveness.

| Device Feature                                              | Hysteroscope<br>System (subject<br>device)                                                                                                                                                                                                                                                                                                                                                                                                                             | LiNA OperaScope<br>K171113                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Discussion                                                                                                                                                        |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>use                                      | The Hysteroscope<br>System is intended<br>to be used to permit<br>viewing of the<br>cervical canal and<br>uterine cavity for the<br>purpose of<br>performing<br>diagnostic and<br>surgical procedures.<br>Note: Hysteroscopes<br>are used as tools to<br>access the uterine<br>cavity and are not, in<br>and of themselves, a<br>method of surgery.<br><br>Generally<br>recognized<br>indications for<br>diagnostic<br>hysteroscopy<br>include:<br>• Abnormal bleeding | The LiNA OperaScope is<br>intended for use in<br>visualization of the<br>cervical canal and uterine<br>cavity during diagnostic<br>and therapeutic<br>gynecological<br>procedures. The types of<br>procedures where the<br>OperaScope could offer<br>visualization include:<br>• Assessment of<br>abnormal bleeding, pelvic<br>pain, amenorrhea and<br>abnormal findings from<br>hysterosalpingogram;<br>• Assessment of Infertility<br>and pregnancy wastage;<br>• Confirmation of the<br>presence of intrauterine<br>foreign body;<br>• Assist in locating<br>submucosal fibroids and<br>polyps targeted for | Similar; the difference<br>in wording of the<br>indications does not<br>alter the intended use.                                                                   |
|                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                   |
|                                                             | • Infertility and<br>pregnancy wastage<br>• Evaluation of<br>abnormal<br>hysterosalpingogram<br>• Intrauterine foreign<br>body<br>• Amenorrhea<br>• Pelvic pain<br>Generally<br>recognized<br>indications for<br>operative<br>hysteroscopy<br>include:<br>• Directed<br>endometrial biopsy<br>• Polypectomy<br>• Submucous<br>myomectomy<br>• Transection of<br>intrauterine<br>adhesions<br>• Transection of<br>intrauterine septa<br>• Endometrial<br>ablation       | removal;<br>• Provide visual guidance<br>during directed biopsy,<br>submucosal<br>myomectomy,<br>transection of intrauterine<br>adhesions and septa.                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                   |
| System<br>Components                                        | The Hysteroscope<br>System consists of a<br>sterile single-use<br>disposable Rigid<br>Hysteroscope and<br>video processor for<br>clinical image<br>processing.                                                                                                                                                                                                                                                                                                         | Sterile/Single-use<br>Handpiece<br>(Hysteroscope) with<br>CMOS camera, LED<br>illumination, LCD display<br>unit, channel for infusion<br>of irrigating fluid, channel<br>for fluid outflow,<br>microelectronics,<br>firmware and single use<br>batteries for powering the<br>device.<br>Non-Sterile/Reusable<br>Recording Module with<br>software,<br>microelectronics and<br>controls for recording<br>pictures/video, HDMI<br>ports (2) for connection to                                                                                                                                                    | Similar;<br>the basic system<br>composition is<br>identical (single use<br>scope with video<br>processor).                                                        |
|                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                   |
|                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                   |
| Optical Image                                               | CMOS                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | handpiece<br>(hysteroscope) and<br>external monitor and<br>input port for external 9V<br>DC power supply.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Identical                                                                                                                                                         |
| Image Resolution                                            | 100,000 pixels                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Digital CMOS<br>Technology<br>CMOS camera consists<br>of approximately 160,000<br>pixels                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Different; the<br>difference does not<br>raise different<br>questions of safety or<br>effectiveness.                                                              |
| Illumination<br>Light Source                                | LED                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | LED at tip of cannula                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                                                         |
| Inflow and<br>outflow channel<br>for saline<br>instillation | Inflow and outflow<br>channels separately                                                                                                                                                                                                                                                                                                                                                                                                                              | Inflow: Inflow channel<br>and working channel<br>combined in one channel<br>above the camera.<br>Outflow: One outflow<br>channel with two<br>entrances placed on top<br>and bottom of distal tip.                                                                                                                                                                                                                                                                                                                                                                                                              | Different; The<br>difference does not<br>raise different<br>questions of safety or<br>effectiveness.                                                              |
| Cannula tip<br>design                                       | Angled shaft<br>proximal to tip:<br>Straight: 0°<br>Curved tip:22°                                                                                                                                                                                                                                                                                                                                                                                                     | Rounded polymer tip.<br>Angled shaft proximal to<br>tip (~ 20°C). Steerable by<br>rotation knob.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Different; the<br>predicate device is<br>steerable by rotation<br>knob. The difference<br>does not raise<br>different questions of<br>safety or<br>effectiveness. |
| Image<br>Transmission                                       | Transmits images<br>are transmitted to<br>the user by the video<br>processor with a<br>CMOS chip at the<br>distal end of the<br>endoscope and the<br>image showing on a<br>monitor.                                                                                                                                                                                                                                                                                    | Transmits images from<br>CMOS video camera to<br>local LCD display unit on<br>the handpiece and<br>remote monitor.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Identical                                                                                                                                                         |
| Maximum                                                     | 4.0 to 6.4mm                                                                                                                                                                                                                                                                                                                                                                                                                                                           | 4.3 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Different; the                                                                                                                                                    |
|                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                   |
| Insertion<br>Diameter (Tip)                                 | depending on model                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | difference in diameter<br>does not raise<br>different questions of<br>safety or<br>effectiveness.                                                                 |
| Shaft/Cannula<br>Length                                     | 260mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | 240mm (24cm)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Different; the<br>difference in length<br>does not raise<br>different questions of<br>safety and<br>effectiveness.                                                |
| Energy Source                                               | Hysteroscope: 5V<br>Video Processor: 5V<br>2A                                                                                                                                                                                                                                                                                                                                                                                                                          | Handpiece: 3V (2 AA<br>batteries)<br>Recording Module: 9V<br>DC external power<br>adapter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Similar;<br>Different output<br>formats do not raise<br>different safety or<br>effectiveness.                                                                     |
| Sterilization                                               | The Rigid<br>Hysteroscope is<br>provided sterilized<br>by EO. No<br>re-sterilization is<br>permitted.<br><br>Video Processor:<br>Non-Sterile                                                                                                                                                                                                                                                                                                                           | Handpiece<br>(Hysteroscope):<br>Ethylene Oxide (EO)<br><br>Handpiece Batteries:<br>Radiation (Gamma)<br><br>Recording Module:<br>Non-Sterile                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Identical                                                                                                                                                         |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

K181545 Page 6 of 7

# VII. Performance data

The following performance data were provided in support of the substantial equivalence determination.

# Biocompatibility testing

Biocompatibility of the Hysteroscope System was evaluated in accordance with ISO 10993-1:2009 for the body contact category of "Surface – Mucosal Membrane" with a contact duration of "Limited (< 24 hours)". The following tests were performed: Cytotoxicity per ISO 10993-5:2009, Irritation and Sensitization per ISO 10993-10:2010. The device was demonstrated to be non-cytotoxic, non-sensitizing and non-irritating.

# Sterilization and shelf life testing

The ethylene oxide sterilization method has been validated to ISO 11135:2014. The shelf life of the Rigid Hysteroscope was demonstrated through a stability study which included

{9}------------------------------------------------

testing of aged devices to ensure devices maintained specifications and that device packaging maintained sterility.

#### Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the Hysteroscope System. The system complies with the IEC 60601-1:2005/(R)2012, IEC60601-2-18 Edition 3, and IEC 60601-1-2:2014.

#### Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for a "moderate" level of concern.

#### Optical performance testing

Optical performance testing was conducted on the Hysteroscope System. The optical performance of the system complies with ISO 8600 series. The system was demonstrated to meet predefined acceptance criteria.

#### Mechanical performance testing

Mechanical characteristics including leaking, bending, articulating and irrigation tests were performed. Devices were demonstrated to meet predefined acceptance criteria.

#### VIII. Conclusion

The subject device has the same intended use as the predicate, and differences in technological characteristics between the subject and predicate devices do not raise different questions of safety or effectiveness. The non-clinical testing demonstrates that the subject device is as safe and as effective as the legally marketed predicate device. Therefore, the subject device is substantially equivalent to the predicate.

---

**Source:** [https://fda-staging.innolitics.com/device/K181545](https://fda-staging.innolitics.com/device/K181545)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
