K181544 · Ad-Tech Medical Instrument Corporation · GZL · Aug 9, 2018 · Neurology
Device Facts
Record ID
K181544
Device Name
Anchor Bolt (as an accessory to Depth Electrodes)
Applicant
Ad-Tech Medical Instrument Corporation
Product Code
GZL · Neurology
Decision Date
Aug 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1330
Device Class
Class 2
Indications for Use
The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode. Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes.
Device Story
Anchor Bolts are mechanical accessories for Depth Electrodes; used in intraoperative and neurological monitoring settings. Device is threaded into a pre-drilled hole in the skull; provides a secure access point for electrode insertion; stabilizes electrode position; includes an inner lumen silicone gasket to minimize CSF leakage. Operated by physicians/surgeons; requires a placement/removal wrench. Output is physical stabilization of the electrode; benefits patient by reducing infection risk and CSF leakage during intracranial monitoring. Device is disposable; single-patient use.
Clinical Evidence
Bench testing only. Performance verified via torque insertion tests, axial tension pull tests (anchor bolt from skull and electrode from anchor bolt), and CSF leakage resistance testing at 15 mmHg hydrostatic pressure. Biocompatibility testing included cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, pyrogenicity, genotoxicity, hemocompatibility, neurotoxicity, and implantation testing per ISO standards.
Technological Characteristics
Materials: Titanium, Silicone (inner lumen gasket), Parylene. Dimensions: 13-26 mm length; compatible with 0.86-1.3 mm diameter electrodes. Mechanical fixation via threaded screw into pre-drilled skull hole. Sterilization: Ethylene Oxide (EtO) per ANSI/AAMI/ISO 11135-1:2007. Single-patient use, disposable.
Indications for Use
Indicated for use as an accessory to Depth Electrodes to provide an access point through the skull and stabilize the electrode while minimizing CSF leakage and infection risk. Contraindicated in patients with skull softening, low skull bone density, or scalp infection.
Regulatory Classification
Identification
A depth electrode is an electrode used for temporary stimulation of, or recording electrical signals at, subsurface levels of the brain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services-USA, which features a stylized depiction of human profiles. To the right of this symbol, there is the FDA logo, which includes the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. The text is arranged in two lines, with 'U.S. FOOD & DRUG' on the top line and 'ADMINISTRATION' on the bottom line.
August 9, 2018
Ad-Tech Medical Instrument Corporation Gary Syring Principal Consultant Quality and Regulatory Associates, LLC 800 Levanger Lane Stoughton, Wisconsin 53589
Re: K181544
Trade/Device Name: Anchor bolt (as an accessory to Depth Electrodes) Regulation Number: 21 CFR 882.1330 Regulation Name: Depth Electrode Regulatory Class: Class II Product Code: GZL Dated: June 8, 2018 Received: June 12, 2018
Dear Mr. Syring:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jay R. Gupta -S
For Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181544
Device Name
Anchor Bolts as Accessories to Depth Electrodes
Indications for Use (Describe)
The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode.
Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:12pt"> <span style="font-family:Symbol"> ☑ </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:12pt"> <span style="font-family:Symbol"> ☐ </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 5.0 510(k) Summary
This summary is provided to support the 510(k) pre-market notification for the Anchor Bolts, as accessories to Depth Electrodes.
| Company Name: | Ad-Tech Medical Instrument Corporation<br>400 West Oakview Parkway<br>Oak Creek, WI 53154<br>Phone: (262) 634-1555 |
|------------------------|--------------------------------------------------------------------------------------------------------------------|
| Company Contact: | Lisa Theama, Chief Operating Officer |
| Date Summary Prepared: | June 8, 2018 |
| Trade Name: | Anchor Bolts |
| Common Name: | Anchor Bolt (Accessory to Depth Electrodes) |
| Classification Name: | Depth Electrode<br>21 CFR 882.1330<br>Product Code: GZL<br>Class II |
| Predicate Device: | K970418 Anchor Bolt, Accessory Depthalon® Depth Electrode<br>PMT Corporation |
#### Product Description 5.1
The device under review is a family of Anchor Bolts. Anchor Bolts are optional accessories to Depth Electrodes. These Anchor Bolts provide an optional access point through the skull and stabilization support for Depth Electrodes.
#### 5.2 Intended Use of the Device
The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired to minimize concerns about potential cerebrospinal fluid (CSF) leakage and infection of the subdural space while stabilizing the electrode.
Anchor Bolts are secured in the skull to provide an access point for Depth Electrodes.
#### 5.3 Summary of Technological Characteristics
The following table provides a side-by-side comparison of the Anchor Bolts to the predicate device applied to support this pre-market notification.
{4}------------------------------------------------
| Table 5.3-1: Substantial Equivalence Technical Characteristics | | | |
|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | Anchor Bolts<br>(Under Review) | Anchor Bolt, Accessory<br>Depthalon® Depth Electrode<br>(Predicate K970418) | Comment |
| Indications<br>for Use | The Ad-Tech Anchor Bolts are<br>optional accessories for use with<br>Depth Electrodes. The Anchor Bolts<br>may be applied when it is desired to<br>minimize concerns about potential<br>cerebrospinal fluid (CSF) leakage and<br>infection of the subdural space while<br>stabilizing the electrode.<br>Anchor Bolts are secured in the skull<br>to provide an access point for Depth<br>Electrodes. | The PMT depth electrode anchor<br>bolt is an accessory used in cases<br>where it is necessary to minimize<br>concerns about potential<br>cerebrospinal fluid leakage and<br>infection of the subdural space will<br>stabilizing the electrode during<br>intraoperative electro clinical<br>characterization. The PMT depth<br>electrode anchor bolts rest on the<br>skull with the insertion of the<br>anchor bolt tapered screw threads<br>into a predrilled burr hole through<br>a skin incision. | Same.<br>Both Anchor Bolts<br>are applied to a<br>pre-drilled hole in<br>the skull. The<br>Anchor Bolts<br>support Depth<br>Electrode<br>placement. |
| Clinical<br>Application | Threaded into a pre-drilled hole in the<br>skull. | Threaded into a pre-drilled hole in the<br>skull. | Same |
| Duration of<br>use | < 30 days | < 30 days | Same |
| Contra-<br>indications | Anchor Bolts should not be used on<br>any patient whom the physician/<br>surgeon considers at risk for infection<br>or on whom the use cannot be<br>performed safely.<br><br>The Anchor Bolt should not be used<br>with patients that have softening of<br>the skull or low skull bone density. | Depthalon® Depth Electrodes<br>should not be used in the presence<br>of any infection of the scalp. Other<br>infections may be considered a<br>contraindication if determined by<br>the attending physician.<br><br>The Anchor bolt accessory should<br>not be used if the patient has a<br>softening of the skull or low skull<br>bone density. | Equivalent<br>Anchor Bolts are<br>selected and<br>applied by a<br>physician<br>knowledgeable in<br>their use. |
| Anchor Bolt<br>Single patient<br>use,<br>Disposable | Yes | Yes | Same |
| Provided<br>Sterile | Yes<br>(Anchor Bolts provided sterile,<br>optional for Placement/Removal<br>Wrench ) | Yes<br>(Anchor bolts can be provided<br>sterile or non-sterile) | Equivalent |
| User<br>Sterilizable | Yes<br>(Placement/Removal Wrench only) | Yes<br>(Anchor Bolts supplied non-sterile) | As a convenience<br>to the user, the<br>Placement /<br>Removal Wrench<br>can be re-<br>sterilized. |
| Environment<br>of Use | Intraoperative and Neurological<br>monitoring locations | Intraoperative and Neurological<br>monitoring locations | Same |
| Duration of<br>Use | < 30 days | < 30 days | Same |
| Patient<br>contact<br>material | Titanium<br>Silicone (inner lumen gasket)<br>Parylene | Titanium Alloy<br>Silicone (inner lumen gasket) | Equivalent |
| Length | 13 mm to 26 mm | 25 mm to 40 mm | Equivalent |
| Table 5.3-1: Substantial Equivalence Technical Characteristics | | | |
| Feature | Anchor Bolts<br>(Under Review) | Anchor Bolt, Accessory<br>Depthalon® Depth Electrode<br>(Predicate K970418) | Comment |
| Compatible<br>Depth<br>Electrode<br>Body<br>Diameter | 0.86 mm to 1.3 mm | 1.27 mm | Equivalent |
| Depth<br>Electrode<br>Retention<br>Force | > 100 grams | > 100 grams | Same |
| Placement /<br>Removal<br>Wrench | Yes | Unknown | Given the unique<br>design of Anchor<br>Bolts it is very<br>likely that the<br>predicate device<br>requires a<br>placement and<br>removal tool. Hand<br>tightening an<br>Anchor Bolt into a<br>pre-drilled skull<br>hole is not<br>probable. |
| | | | The length of the<br>Anchor Bolt is<br>based upon need<br>and user<br>preferences. |
{5}------------------------------------------------
### 5.4 Non-clinical Performance Tests to Demonstrate Substantial Equivalency
To establish the technical equivalency of the Anchor Bolts, evaluations were conducted to confirm compliance with performance requirements, including:
| Test | Test Method Summary | Result |
|-------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Torque required to insert<br>to pre-drilled skull hole. | With a torque wrench, verify the torque required to insert<br>the Anchor Bolt into pre-drilled hole in 40 pcf polyurethane<br>foam (simulated skull bone) with the Placement/Removal<br>Wrench is ≤ 8 in-lbs. (N=10 samples). | Pass<br>All results ≤<br>1 in-lbs. |
| Axial tension to pull<br>Anchor Bolt out of pre-<br>drilled skull hole. | With a tensile tester, measure the axial tension to pull the<br>Anchor Bolt out of a pre-drilled hole in 40 pcf polyurethane<br>foam is ≥ 40 lbf. (N=10 samples) | Pass<br>All results<br>> 48 lbf. |
| Axial tension to pull<br>Depth Electrode from the<br>Anchor Bolt. | With a tensile tester, measure the axial tension to pull a<br>Depth Electrode from the Anchor Bolt, ≥ 100 grams. (N=10<br>samples). | Pass<br>All results<br>> 140<br>grams. |
| Anchor Bolt gasket resist<br>CSF leakage. | Insert Depth Electrode into the Anchor Bolt secured in 40<br>pcf polyurethane foam. Apply a metal tube with a diameter<br>larger than the Depth Electrode and Anchor Bolt body,<br>around both. Seal the metal tube to the polyurethane foam.<br>Fill the metal tube with water to a hydrostatic pressure of | Pass<br>No leaks<br>observed |
| Test | Test Method Summary | Result |
| | 15 mmHg. Verify no leakage at the Depth Electrode,<br>Anchor Bolt after 24 hours. (N=22 samples) | after 24<br>hours. |
| Sterility | Anchor Bolts and Placement/Removal Wrenches are<br>terminally ethylene oxide sterilized devices by a validated<br>EtO sterilization cycle, per ANSI/AAMI/ISO 11135-1:2007<br>half-cycle method. The sterility assurance level (SAL) is ≤<br>SAL of 10-6. | Pass |
| Sterile barrier integrity | The sterile barrier package is a double Tyvek pouch. The<br>methods applied to evaluate the Tyvek sterile barrier<br>package integrity included post terminal sterilization:<br>simulated distribution, seal peel and bubble emission<br>testing. | Pass |
| Anchor Bolt silicone<br>gasket shelf life | The ability of real time aged Anchor Bolts with Silicone<br>Gaskets to meet Depth Electrode retention requirements<br>and minimize cerebrospinal fluid (CSF) leakage were<br>evaluated to the methods above for Axial tension to pull<br>Depth Electrode from the Anchor Bolt and Anchor Bolt<br>gasket resist CSF leakage. | Pass<br>The same<br>acceptance<br>criteria were<br>met. |
| Biocompatibility | The following test methods were applied to final finished<br>form Anchor Bolts (patient contacting materials), with<br>passing results:<br>Cytotoxicity: L929 MEM Elution Test – ISO Sensitization: Kligman Maximization Test - ISO Intracutaneous Reactivity: Intracutaneous Injection Test - ISO Acute Systemic Toxicity: 28-Day Systemic Toxicity in Rabbits via Subcutaneous Implantation Material-Mediated Pyrogenicity: Rabbit Pyrogen Test (Material Mediated) - ISO Genotoxicity: Salmonella Typhimurium and Escherichia Coli Reverse Mutation Assay - ISO and Mouse Lymphoma Mutagenesis Assay - ISO Hemocompatibility: Rabbit Blood Hemolysis Test (Indirect Contact) – ASTM Neurotoxicity by: SK-N-MC MEM Elution Test. Subacute / Subchronic Toxicity: Systemic Injection Test - ISO Implantation: Neurological (Brain) Implantation Test - ISO. | Pass |
{6}------------------------------------------------
### 5.5 Conclusion
The Anchor Bolts meet performance requirements equivalent to the predicate device. The intended use and technology of the Anchor Bolts are the same as the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.