K181500 · Coltene/Whaledent AG · EBF · Aug 15, 2018 · Dental
Device Facts
Record ID
K181500
Device Name
BRILLIANT Crios
Applicant
Coltene/Whaledent AG
Product Code
EBF · Dental
Decision Date
Aug 15, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
BRILLIANT Crios is indicated for: - Crowns, inlays, onlays and veneers - Implant-supported crowns
Device Story
BRILLIANT Crios is a radiopaque composite material consisting of crosslinked methacrylate-based polymerized monomers and dental glass fillers. Supplied as cured blocks in various sizes, shades, and translucencies; intended for fabrication of permanent tooth restorations (crowns, inlays, onlays, veneers, implant-supported crowns) using CAD/CAM milling equipment. Used in dental clinics by dental professionals. The material is milled into the final restoration shape; provides a durable, tooth-colored restorative solution for patients requiring dental reconstruction.
Clinical Evidence
Bench testing only. No clinical data. Testing included flexural strength, flexural modulus, compressive strength, and biocompatibility per ISO 4049, ISO 7405, and ISO 10993-1. Results confirmed the device meets performance requirements.
Technological Characteristics
Radiopaque composite material; crosslinked methacrylate-based monomers (Bis-GMA, Bis-EMA, TEGDMA, UDMA) and silica/barium glass fillers. Supplied as cured blocks for CAD/CAM milling. Non-sterile. Conforms to ISO 4049 (polymer-based restorative materials) and ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for permanent tooth restorations including crowns, inlays, onlays, veneers, and implant-supported crowns.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
August 15, 2018
Image /page/0/Picture/1 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is in blue and consists of the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" stacked on top of each other.
Coltene/Whaledent AG % Stuart Goldman Senior Consultant Emergo Global Consulting. LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746
### Re: K181500
Trade/Device Name: BRILLIANT Crios Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: Class II Product Code: EBF Dated: June 1, 2018 Received: June 7, 2018
### Dear Stuart Goldman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181500
Device Name BRILLIANT Crios
Indications for Use (Describe) BRILLIANT Crios is indicated for:
- Crowns, inlays, onlays and veneers
- Implant-supported crowns
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
K181500
# BRILLIANT Crios
#### 1. Submission Sponsor
Coltène/Whaledent AG Feldwiesenstrasse 20 Altstätten Sankt Gallen 9450 Switzerland
Contact: Dr. Daniel Meier Title: Manager Regulatory Affairs Phone: 41 71 757 5852
#### 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746
Contact: Stuart R. Goldman Title: Senior Consultant RA/QA Phone: (512) 327-9997
#### 3. Date Prepared
June 1, 2018
#### 4. Device Identification
| Trade Name: | BRILLIANT Crios |
|-----------------------|----------------------------|
| Common Name: | Restorative resin material |
| Classification Name: | Tooth shade resin material |
| Regulation Number: | 872.3690 |
| Product Code: | EBF |
| Class: | Class II |
| Classification Panel: | Dental |
#### 5. Legally Marketed Predicate Device(s)
- CERASMART™ (K133824); Manufacturer: GC America, Inc. ●
- . Coltène/Whaledent AG is also using their own device, SYNERGY® (K974465) as a ReferenceDevice.
{4}------------------------------------------------
#### 6. Indication for Use
BRILLIANT Crios is indicated for:
- Crowns, inlays, onlays and veneers
- Implant-supported crowns.
#### 7. Device Overview
BRILLIANT Crios is a radiopaque composite material composed of crosslinked methacrylate based polymerized monomers and dental glass fillers. The subject device is made available in block forms of different sizes, shades and translucencies for the fabrication of permanent tooth restorations such as inlays, onlays, crowns and veneers using CAD/CAM milling equipment.
#### 8. Substantial Equivalence Discussion
Table 5-1 compares BRILLIANT Crios blocks to the predicate device with respect to its physical state, structure, materials, mechanical properties, indications for use, packaging, biocompatibility and performance testing, and provides detailed information regarding the basis for the determination of substantial equivalence. The reference device is also included in this table for comparative purposes.
| Attributes | Reference<br>Device | Predicate Device | Subject Device | Similarities /<br>Differences |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | SYNERGY® | CERASMART™ | BRILLIANT Crios | - |
| Manufacturer | Coltène/Whaledent | GC America | Coltène/Whaledent | - |
| 510(k) # | K974465 | K133824 | Pending | - |
| Product Code | EBF | EBF | EBF | Same |
| Regulation | §872.3690 | §872.3690 | §872.3690 | Same |
| Class | II | II | II | Same |
| Review Panel | Dental | Dental | Dental | Same |
| Device Image | Image: SYNERGY® | Image: CERASMART™ | Image: BRILLIANT Crios | - |
| Indications for<br>Use | SYNERGY is<br>indicated for:<br>- direct filling of<br>class I, II, III, IV<br>and V cavities,<br>- reconstruction | The product is<br>indicated for inlays,<br>onlays, veneers and<br>full crown<br>restorations,<br>including crowns on | BRILLIANT Crios is<br>indicated for:<br>- Crowns, inlays, onlays<br>and veneers<br>- Implant-supported<br>crowns | The subject and<br>predicate device have<br>the same intended<br>use and are used for<br>permanent tooth<br>restorations. |
| Attributes | Reference<br>Device | Predicate Device | Subject Device | K#####<br>Similarities /<br>Differences |
| | of natural<br>enamel and<br>dentine,<br>reconstruction<br>of fractured<br>anteriors, sealing of<br>extended<br>fissures in<br>molars and<br>premolars, stabilization of<br>mobile<br>anteriors, repair of<br>veneer facings, fixation of<br>splints, bonded<br>bridges, esthetic | implants | | |
| Physical State | Viscous paste | Cured blocks | Cured blocks | The subject and<br>predicate device are<br>supplied as cured<br>blocks and CAD/CAM<br>milled into the final<br>dental restoration. |
| Structure | Polymer resin<br>composite | Polymer resin /<br>ceramic hybrid<br>composite | Polymer resin<br>composite | The subject device<br>has an all polymer<br>resin matrix while the<br>predicate device has<br>a polymer resin /<br>ceramic hybrid<br>matrix. |
| Resin Matrix<br>Monomers | Bis-GMA, Bis-EMA,<br>TEGDMA | Bis-MEPP, UDMA,<br>DMA | Bis-GMA, Bis-EMA,<br>TEGDMA, UDMA | The subject and<br>predicate device both<br>contain UDMA in<br>their resin matrix. |
| Filler | Silica, barium glass | Silica, barium glass | Silica, barium glass | Same |
| Sizes | na | 12, 14 and 14L | 12, 14 | The subject device<br>size ranges fall within |
| Attributes | Reference Device | Predicate Device | Subject Device | Similarities / Differences |
| Flexural Strength | 125 (MPa) | 160 (MPa) | 198 (MPa) | those of the<br>predicate device. |
| Flexural Modulus of | 6.8 (GPa) | 8.3 (GPa) | 10.3 (GPa) | Similar |
| Compressive Strength | 390 (MPa) | 429 (MPa) | 426 (MPa) | Similar |
| Packaging | Plunger syringe type (2.3 g) | Mandrel mounted; five to a box | Mandrel mounted; five to a box | The subject and predicate device are packaged the same way based on their physical state. |
| Usage | Single Patient, multiple use | Single Patient, multiple use | Single Patient, multiple use | Same |
| Sterility | Non-Sterile | Non-Sterile | Non-Sterile | Same |
| Biocompatibility | Conforms with ISO 10993-1 | Conforms with ISO 10993-1 | Conforms with ISO 10993-1 | Same |
| Performance | Conforms with ISO 4049 | Conforms with ISO 4049 | Conforms with ISO 4049 | Same |
#### Table 5-1 - Comparison of Device Characteristics
{5}------------------------------------------------
{6}------------------------------------------------
#### 9. Non-Clinical Performance Data
As part of demonstrating substantial equivalence of BRILLIANT Crios blocks to the predicate device, Coltène/Whaledent submitted final finished devices for extensive testing in accordance with the applicable parts of the following voluntary standards, as well as to the company's own internal test protocols:
- . ISO 4049, Dentistry – Polymer-based Restorative Materials
- ISO 7405, Dentistry Evaluation of biocompatibility of medical devices used indentistry .
- . ISO 10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process
Testing evaluated flexural strength, flexural modulus, compressive strength, and biocompatibility of the subject device, as well as other related physical properties. The subject device passed all required testing.
#### 10. Statement of Substantial Equivalence
BRILLIANT Crios blocks have the same intended use, indications for use, physical attributes, and are fabricated into permanent tooth restorations using the same CAD/CAM manufacturing methods as CERASMART™ blocks. Any minor differences in the materials used to make the subject device when compared to the predicate device have been successfully evaluated by Coltène/Whaledent through extensive performance and biocompatibility testing on their device, such that the information submitted to the FDA demonstrates that the subject device is as safe and effective as the predicate device and
{7}------------------------------------------------
does not raise any new questions of safety and effectiveness. BRILLIANT Crios blocks, as designed and manufactured by Coltène/Whaledent, have been determined to be substantially equivalent to CERASMART™ blocks.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.