← Product Code [FZP](/productcode/FZP) · K181488

# LIGACLIP 12mm L Endoscopic Rotating Multiple Clip Applier (K181488)

_Ethicon Endo-Surgery, LLC · FZP · Oct 30, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181488

## Device Facts

- **Applicant:** Ethicon Endo-Surgery, LLC
- **Product Code:** [FZP](/productcode/FZP.md)
- **Decision Date:** Oct 30, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4300
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Ligaclip® Endoscopic Rotating Multiple Clip Applier is intended for use on tubular structures or vessels wherever a metal ligating clip is indicated. The tissue being ligated should be consistent with the size of the clip.

## Device Story

Sterile, single-patient use endoscopic device; delivers 20 large titanium clips for ligation of tubular structures or vessels. Operated by surgeons via Endopath surgical trocars; features 360° shaft rotation and one-handed trigger operation. Device includes a low-glare shaft and a yellow indicator bar signaling when 3 or fewer clips remain. Mechanical design ensures clip advancement after each firing; trigger resistance increases when empty. Used in endoscopic procedures to provide secure vessel/tissue ligation, reducing bleeding risk and facilitating surgical workflow.

## Clinical Evidence

Bench testing only. Performance assessed via ex-vivo testing including trocar compatibility, clip retention strength, ratchet mechanism strength, ergonomic torque, tactile lockout feedback, and clip occlusion/security. Biocompatibility and product stability data leveraged from primary predicate. Pre-clinical studies included rabbit pyrogenicity and limulus amoebocyte lysate (LAL) testing. No clinical studies were performed.

## Technological Characteristics

Pistol handle, rotation knob, and 34.1 cm shaft (11.84 mm OD). Materials identical to predicate. Mechanical actuation; manual trigger-fired. Features 360° rotation, low-glare shaft coating, and mechanical yellow clip-count indicator. Sterile, single-use.

## Regulatory Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

## Predicate Devices

- Ligaclip® 10mm M/L Endoscopic Rotating Multiple Clip Applier ([K150840](/device/K150840.md))
- Ligaclip® 10mm M/L Endoscopic Rotating Multiple Clip Applier ([K864102](/device/K864102.md))

## Reference Devices

- Ligamax™ 5 ([K110699](/device/K110699.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 30, 2018

Ethicon Endo-Surgery, LLC % Ms. Melissa Iwu Senior Regulatory Affairs Specialist Ethicon Endo-Surgery, Inc 4545 Creek Rd Cincinnati, Ohio 45245

Re: K181488

Trade/Device Name: LIGACLIP 12mm L Endoscopic Rotating Multiple Clip Applier Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: FZP Dated: August 30, 2018 Received: August 31, 2018

Dear Melissa Iwu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Image /page/1/Picture/5 description: The image shows the name "Joseph Nielsen -S" in a large, bold font. To the right of the name is a digital signature from Joseph Nielsen -S. The signature includes the date 2018.10.30 08:03:39-04'00' and other information such as DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, cn=Joseph Nielsen -S, 0.9.2342.19200300.100.1.1=2000367505.

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

#### K181488

Device Name

LIGACLIP 12mm L Endoscopic Rotating Multiple Clip Applier

Indications for Use (Describe)

The Ligaclip® Endoscopic Rotating Multiple Clip Applier is intended for use on tubular structures or vessels wherever a metal ligating clip is indicated. The tissue being ligated should be consistent with the size of the clip.

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D) | <div> <span>                     ☒                 </span> </div> |
|----------------------------------------------|-------------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C)  | <div> <span>                     ☐                 </span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the Ethicon logo in red, with the words "PART OF THE Johnson & Johnson FAMILY OF COMPANIES" written in smaller letters below. The Ethicon logo is in a bold, sans-serif font. The Johnson & Johnson text is in a script font. The overall design is simple and clean.

# 510(k) Summary

## Company

Ethicon Endo-Surgery, LLC 475 Calle C Guaynabo, PR 00969

## Contact

Melissa Iwu, Senior Specialist, Regulatory Affairs Ethicon Endo-Surgery, Inc. Telephone: 513.337.7623 Email: miwu2@its.jnj.com

| Date Prepared:              | October 19, 2018                                                                                                         |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Trade Name:                 | Ligaclip® 12mm L Endoscopic Rotating Multiple Clip Applier                                                               |
| Common Name:                | Clip Applier                                                                                                             |
| Classification Name:        | Implantable Clip                                                                                                         |
| Device Class:               | Class II                                                                                                                 |
| Classification Regulation:  | 878.4300                                                                                                                 |
| Panel:                      | 79, General and Plastic Surgery                                                                                          |
| Classification Code:        | FZP                                                                                                                      |
| Primary Predicate Device:   | Ligaclip® 10mm M/L Endoscopic Rotating Multiple Clip Applier, K150840                                                    |
| Secondary Predicate Device: | Ligaclip® 10mm M/L Endoscopic Rotating Multiple Clip Applier submitted as Ligaclip® 20/20 Multiple Clip Applier, K864102 |
| Reference Device:           | Ligamax™ 5, K110699                                                                                                      |

# Device Description

The Ligaclip® 12mm L Endoscopic Rotating Multiple Clip Applier is a sterile, single-patient use device designed to provide a means of ligation through Endopath surgical trocars. The device delivers 20 large titanium clips that individually advance after each firing. The shaft is made of a low glare material that minimizes reflective distortion. It is designed to rotate 360° in either direction. The rotating knob is located to allow for a one-handed technique.

#### Indications for Use

The Ligaclip Endoscopic Rotating Multiple Clip Applier is intended for use on tubular structures or vessels wherever a metal ligating clip is indicated. The tissue being ligated should be consistent with the size of the clip.

{4}------------------------------------------------

PART OF THE Johnson Johnson FAMILY OF COMPANIES

#### Technological Characteristics

The Subject Device, Ligaclip® 12mm L Endoscopic Rotating Multiple Clip Applier's configuration consists of a pistol handle, a rotation knob, and a shaft with an outer diameter of approximately 11.84 mm and length of 34.1 cm. The shaft is made of a low glare material that minimizes reflective distortion. At the distal end of the shaft are the jaws, which form ligating clips. The force to squeeze the trigger increases when no clips remain in the device. The shaft contains a yellow clip counter indicator bar, which appears yellow when only 3 clips or fewer remain in the device.

The Subject Device is comprised of the same technological characteristics as the Primary Predicate Device with respect to materials, design, energy source, and other features. The key differences between the two devices are the working shaft length (28.9 cm for Primary Predicate, and 34.1 cm for Subject), trocar sizes (10/11 mm for Primary Predicate and 12 mm for Subject), and closed clip length (8.8 mm for Primary Predicate and 11.0 mm for Subject). The clip continues to ligate tubular structures and vessels and the clip formation remains unchanged.

## Performance Data

Ex-vivo tests were performed to verify that the performance of the Subject Device, Ligaclip® 12mm L Endoscopic Rotating Multiple Clip Applier, meets the definition of substantial equivalence to the Primary Predicate Device, Ligaclip® 10mm M/L Endoscopic Rotating Multiple Clip Applier. Device performance was assessed against the Subject Device's design requirement and comparative testing against the Primary and Secondary Predicate Devices. Bench testing includes trocar compatibility, retention strength of unformed clip in jaws, ratchet mechanism strength, ergonomic torque to fire device, tactile indication torque for "no-clip lockout" function, low clip indicator bar performance, clip occlusion of the formed clip, and clip security of the formed clip. Product Stability and Biocompatibility data from the Primary Predicate Device was leveraged in this submission as the Subject and Primary Predicate Device materials remain unchanged. Data from Pre-Clinical studies, specifically rabbit pyrogenicity and limulus amoebocyte lysate (LAL) testing, was included in the submission. This submission does not contain data from Clinical studies.

The conclusions of the testing criteria demonstrate that the Ligaclip® 12mm L Endoscopic Rotating Multiple Clip Applier device is substantially equivalent to the legally marketed Primary Predicate Device, K150840, and Secondary Predicate Device, K864102.

---

**Source:** [https://fda-staging.innolitics.com/device/K181488](https://fda-staging.innolitics.com/device/K181488)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).