K181486 · El.En Electronic Engineering Spa · GEX · Aug 30, 2018 · General, Plastic Surgery
Device Facts
Record ID
K181486
Device Name
DEKA Motus AY
Applicant
El.En Electronic Engineering Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Aug 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Alexandrite 755nm laser source: Temporary hair reduction. Stable long-term or permanent hair reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6,9, and 12 months after the completion of a treatment regime, on all skin types (Fitzpatrick I - VI) including tanned skin. Treatment of benign pigmented lesions. Treatment of wrinkles. Photocoagulation of benign dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias). Nd:YAG 1064nm laser source: Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB . The lasers are indicated on all Skin Types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of benign pigmented and benign vascular lesions, such as but not limited to port wine stains, hemangioma, warts, teleangiectasia, rosacea, venus lake, leg veins and spider veins. Coagulation and hemostasis of soft tissue. Benign pigmented lesions such as, but not limited to, lentigos (sun spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, cloasma, verrucae, skin tags, keratosis and plaques. The laser is indicated for benign pigmented lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments. The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar. Treatment of wrinkles.
Device Story
Medical laser system utilizing dual sources: 755nm Alexandrite and 1064nm Nd:YAG. Laser output delivered via lens-coupled, user-replaceable optical fiber; interchangeable handpieces feature electronic spot recognition. Operated by physician/clinician in clinical setting via footswitch or fingerswitch. Integrated and external skin cooling system (15°C) protects epidermis during treatment. Device targets melanin in hair follicles or vascular/pigmented lesions to achieve photocoagulation, hemostasis, or hair reduction. Output parameters (fluence, pulse duration, repetition rate) adjusted by clinician based on lesion type and patient skin characteristics. Clinical benefit includes stable long-term hair reduction and resolution of dermatological lesions.
Clinical Evidence
No clinical data provided. Substantial equivalence established via non-clinical bench testing and standards conformance.
Technological Characteristics
Dual-source laser (755nm Alexandrite, 1064nm Nd:YAG). Pulse duration 0.2-50ms; repetition rate up to 10Hz. Integrated/external skin cooling (15°C). Handpieces feature electronic spot recognition. Conforms to AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1.
Indications for Use
Indicated for patients of all skin types (Fitzpatrick I-VI), including tanned skin, requiring hair reduction, treatment of benign pigmented/vascular lesions, wrinkle treatment, soft tissue coagulation/hemostasis, and scar pigmentation reduction.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 30, 2018
Elen Electronic Engineering Spa Paolo Peruzzi Regulatory Affairs Manager Via Baldanzese, 17 Calenzano, 50041 IT FI
Re: K181486
Trade/Device Name: DEKA MOTUS AY Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: May 31, 2018 Received: June 6, 2018
Dear Paolo Peruzzi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entiled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181486
Device Name DEKA MOTUS AY
| Indications for Use (Describe) | |
|---------------------------------|--|
| Alexandrite 755nm laser source: | |
Alexandrite 755nm laser source:
Temporary hair reduction.
Stable long-term or permanent hair reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6.9, and 12 months after the completion of a treatment regime, on all skin types (Fitzpatrick I - VI) including tanned skin. Treatment of benign pigmented lesions.
Treatment of wrinkles.
Photocoagulation of benign dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).
#### Nd:YAG 1064nm laser source:
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB . The lasers are indicated on all Skin Types Fitzpatrick I-VI including tanned skin.
Photocoagulation and hemostasis of benign pigmented and benign vascular lesions, such as but not limited to port wine stains, hemangioma, warts, teleangiectasia, rosacea, venus lake, leg veins and spider veins.
Coagulation and hemostasis of soft tissue.
Benign pigmented lesions such as, but not limited to, lentigos (sun spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, cloasma, verrucae, skin tags, keratosis and plaques.
The laser is indicated for benign pigmented lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
|---------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
## DEKA Motus AY
#### Submitter:
El.En. S.p.A. Via Baldanzese, 17 50041 Calenzano (FI), Italy
## Contact:
Paolo Peruzzi Regulatory Affairs Manager & Official Correspondent Phone: +39.055.8826807 E-mail: p.peruzzi@elen.it
## Date Summary Prepared:
August 28, 2018
Device Trade Name:
DEKA MOTUS AY
Common Name:
Medical Laser system
#### Classification Name:
Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology (GEX)
#### Classification Number:
21 CFR 878.4810
#### Predicate Devices:
Primary predicate: Synchro Repla: Y family of laser systems (K150516). Reference predicate: DEKA Motus AX (K162886)
#### Device Description:
The DEKA MOTUS A Y is a medical device equipped with two laser sources: 755nm and 1064mm. The laser sources deliver the laser output through a lens coupled user replaceable optical fiber with a wide range of interchangeable, quick release laser handpieces with electronic spot recognition. Handpiece activation is either by footswitch or fingerswitch.
#### Intended Use:
Alexandrite 755nm laser source:
Temporary hair reduction.
Stable long-term or permanent hair reduction through selective targeting of melanin in hair follicles. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs
{4}------------------------------------------------
regrowing when measured at 6,9, and 12 months after the completion of a treatment regime, on all skin types (Fitzpatrick I - VI) including tanned skin.
Treatment of benign pigmented lesions.
Treatment of wrinkles.
Photocoagulation of benign dermatological vascular lesions (such as port-wine stains, hemangiomas, telangiectasias).
#### Nd:YAG 1064nm laser source:
Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment of PFB . The lasers are indicated on all Skin Types Fitzpatrick I-VI including tanned skin. Photocoagulation and hemostasis of benign pigmented and benign vascular lesions, such as but not limited to port wine stains, hemangioma, warts, teleangiectasia, rosacea, venus lake, leg veins and spider veins.
Coagulation and hemostasis of soft tissue.
Benign pigmented lesions such as, but not limited to, lentigos, (age spots), solar lentigos (sun spots), cafe au lait macules, seborrheic keratosis, nevi, cloasma, verrucae, skin tags, keratosis and plaques.
The laser is indicated for benign pigmented lesions to reduce lesion size, for patients with lesions that would potentially benefit from aggressive treatment, and for patients with lesions that have not responded to other laser treatments.
The laser is also indicated for the reduction of red pigmentation in hypertrophic and keloid scars where vascularity is an integral part of the scar.
Treatment of wrinkles.
#### Substantial equivalence discussion:
The DEKA MOTUS AY is substantially equivalent to the DEKA Synchro Repla:Y family of laser systems (K150516).
755 nm laser:
| Device Trade Name | Proposed 510(k) Device<br>DEKA MOTUS AY | Predicate Device<br>K162886<br>DEKA Motus AX | Predicate Device<br>K150516<br>DEKA Synchro Repla: Y<br>family of laser systems |
|-----------------------------------------------|---------------------------------------------|----------------------------------------------|---------------------------------------------------------------------------------|
| Laser Type | Alexandrite | Alexandrite | Alexandrite |
| Wavelength (nm) | 755 nm | 755 nm | 755 nm |
| MAX Fluence (J/cm2) | 2-200 J/cm² | 2-600 J/cm² | 2-600 J/cm² |
| Handpiece Spot Sizes<br>(diameter millimeter) | 2.5, 5, 7, 10, 12, 14, 15, 16,<br>18, 20 mm | 2.5, 5, 7, 10, 12, 14, 15, 16,<br>18, 20 mm | 2.5, 5, 7, 10, 12, 14, 15, 16,<br>18, 20, 22, 24 mm |
| Pulse Duration<br>(milliseconds) | 2 to 50 ms | 2 to 50 ms | 0.2 to 50 ms |
| Pulse Repetition Rate<br>(Hz) | up to 10 Hz | up to 10 Hz | up to 10 Hz |
{5}------------------------------------------------
| Device Trade Name | Proposed 510(k) Device<br>DEKA MOTUS AY | Predicate Device<br>K162886<br>DEKA Motus AX | Predicate Device<br>K150516<br>DEKA Synchro Repla: Y<br>family of laser systems |
|-----------------------------|-----------------------------------------------|-----------------------------------------------|---------------------------------------------------------------------------------|
| Skin Cooling System | Yes<br>(handpiece integrated and<br>external) | Yes<br>(handpiece integrated and<br>external) | Yes<br>(handpiece integrated and<br>external) |
| Skin Cooling<br>temperature | 15°C | 15°C | 4-20°C |
1064 nm laser:
| Device Trade Name | Proposed 510(k) Device<br>DEKA MOTUS AY | Predicate Device<br>K150516<br>DEKA Synchro Repla:Y family of laser<br>systems |
|-----------------------------------------------|--------------------------------------------|--------------------------------------------------------------------------------|
| Laser Type | Nd: YAG | Nd:YAG |
| Wavelength (nm) | 1064 nm | 1064 nm |
| MAX Fluence (J/cm2) | 2-600 J/cm² | 2-600 J/cm² |
| Handpiece Spot Sizes<br>(diameter millimeter) | 2.5, 5, 7, 10, 12, 14, 15, 16, 18, 20 , mm | 2.5, 5, 7, 10, 12, 14, 15, 16, 18, 20, 22, 24<br>mm |
| Pulse Duration<br>(milliseconds) | 0.2 to 50 ms | 0.2 to 50 ms |
| Pulse Repetition Rate<br>(Hz) | up to 10 Hz | up to 10 Hz |
| Skin Cooling System | Yes<br>(handpiece integrated and external) | Yes<br>(handpiece integrated and external) |
| Skin Cooling<br>temperature | 15°C | 4-20°C |
The DEKA MOTUS AY has the same indications for use as the above mentioned predicate devices, with same principle of operation and essentially the same performances.
#### Clinical Performance Data:
None
#### Non-Clinical Performance Data:
The DEKA MOTUS AY was tested for standards conformance with the following standards:
AAMI/ANSI ES60601-1- Medical electrical equipment – Part 1: General requirements for basic safety and essential performance.
IEC 60601-1-2 - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
{6}------------------------------------------------
IEC 60601-2-22 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment.
IEC 60825-1- Safety of laser products – Part 1: Equipment classification and requirements.
# Conclusion:
Based on the outcome of non-clinical performance data performed, we can conclude that the DEKA Motus AY laser system is substantially equivalent to the predicate devices .
#### Additional Information:
None
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.