The ACCUFUSE Cervical System is used to maintain disc space distraction in skeletally mature adults requiring intervertebral body fusion of the cervical spine.
Device Story
ACCUFUSE Cervical System comprises intervertebral body fusion implants, cervical plates, and screws; used to maintain disc space distraction and facilitate fusion in the cervical spine. Implants feature hollow centers for autogenous bone graft/allograft placement; superior/inferior ridges grip endplates to prevent migration. Plates attach to vertebral bodies via self-tapping or self-drilling screws using an anterior approach. System used by surgeons in clinical settings; provides temporary stabilization during fusion development. Benefits include restoration of disc space height and structural support for spinal fusion. Device is non-sterile, single-use.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including static and dynamic axial compression, static and dynamic torsion, and subsidence testing.
Technological Characteristics
Materials: PEEK (ASTM F2026), Tantalum (ASTM F560), Titanium alloy Ti-6Al-4V (ASTM F136). Components: Intervertebral body fusion cages, cervical plates, and bone screws. Mechanical testing standards: ASTM F2077 (cages), ASTM F2267 (subsidence), ASTM F1717 (plates). Provided non-sterile for single use.
Indications for Use
Indicated for skeletally mature patients at one or two contiguous levels (C2-T1) for intervertebral body fusion in cases of degenerative disc disease (DDD) with discogenic pain. Also indicated for anterior screw fixation as an adjunct to fusion for temporary stabilization in patients with DDD, spondylolisthesis, trauma (fractures/dislocations), tumors, deformity (kyphosis, lordosis, scoliosis), pseudarthrosis, failed previous fusion, or spinal stenosis. Requires at least six weeks of non-operative treatment prior to surgery.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
August 29, 2018
Absolute Advantage Medical, LLC % Mr. J.D. Webb The OrthoMedix Group, Inc. 1001 Oakwood Boulevard Round Rock, Texas 78681
Re: K181471
Trade/Device Name: ACCUFUSE Cervical System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: ODP, KWO Dated: May 31, 2018 Received: June 4, 2018
Dear Mr. Webb:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Ronald P. Jean -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181471
Device Name ACCUFUSE Cervical System
#### Indications for Use (Describe)
The ACCUFUSE Cervical System (intervertebral body fusion device) is indicated for use as an intervertebral body fusion device in skeletally mature patients at one or two contiguous levels of the (C2-T1) to facilitate fusion in case of degenerative disc disease (DDD) defined as discogenic pain with degeneration of the disc confirmed by patient history and radiographic studies. The ACCUFUSE Cervical System implants are placed via an anterior approach using autogenous bone graft and/or allograft comprised of cancellous and/or corticocancellous bone graft. The ACCUFUSE Cervical System implants are to be used with supplemental fixation. Patients should have at least six weeks of nonoperative treatment prior to surgery.
The ACCUFUSE Cervical System (cervical plate/screws) is intended for anterior screw fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to fusion of the cervical spine (C2 to T1). The system is indicated for use in the temporary stabilization of the anterior spine during the development of cervical spimal fusion in patients with:
| •degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| •spondylolisthesis, |
| •trauma (i.e. fractures or dislocations), |
| •tumors, |
| •deformity (defined as kyphosis, lordosis, or scoliosis), |
| •pseudarthrosis, |
| •failed previous fusion, |
| •spinal stenosis |
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary: ACCUFUSE Cervical System
| Date Prepared | May 31, 2018 |
|-----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitted By | Absolute Advantage Medical, LLC<br>2731 Millbrook Road<br>Fayetteville, NC 28303 |
| Primary Contact | J.D. Webb<br>1001 Oakwood Blvd<br>Round Rock, TX 78681<br>512-388-0199 Tele<br>e-mail: jdwebb@orthomedix.net |
| Trade Name | ACCUFUSE Cervical System |
| Classification Name | Intervertebral Body Fusion Device Spinal Intervertebral body fixation orthosis |
| Common Name | Intervertebral Fusion Device with Bone Graft, Cervical Appliance, Fixation Spinal Intervertebral Body |
| Class | II |
| Product Code | ODP KWQ |
| CFR Section | 21 CFR section 888.3080 21 CFR section 888.3060 |
| Device Panel | Orthopedic |
| Primary Predicate | Crystal, Spinal Elements, Inc. (K133218) |
| Additional Predicate<br>Devices | UNIPLATE Anterior Cervical Plate System, DePuy Spine, Inc. (K042544 / K082273 / K100070) BAK/C Vista Interbody Fusion, Centerpulse Spine-Tech (P980048 S3) CLSP Plate, Synthes, (K000536) |
| Device Description | The ACCUFUSE Cervical System consists of intervertebral body fusion devices along<br>with cervical plates and screws used to maintain disc space distraction in skeletally<br>mature adults requiring an intervertebral body fusion. It is designed to be used in<br>conjunction with supplemental spinal fixation instrumentation. The implant is available in<br>a range of footprints and heights to suit each individual's pathology and anatomical<br>conditions of the patient. The implant has a hollow center to allow placement of<br>autogenous bone graft to promote intervertebral body fusion. Ridges on the superior and<br>inferior surfaces of the device help to grip the endplates and prevent implant migration<br>and/or expulsion.<br>The ACCUFUSE Anterior Cervical Plate is intended for anterior screw fixation of the plate<br>to the cervical spine. The fixation construct consists of a cervical plate that is attached to<br>the vertebral body of the cervical spine with self-tapping or self-drilling bone screws using |
| | an anterior approach. Plates are available in a variety of lengths addressing single level<br>and two levels of fixation. |
| Materials | PEEK per ASTM F2026 Tantalum per ASTM F560 Titanium alloy (Ti-6Al-4V) per ASTM F136 |
| Intended Use | The ACCUFUSE Cervical System is used to maintain disc space distraction in<br>skeletally mature adults requiring intervertebral body fusion of the cervical spine. |
| Substantial<br>Equivalence Claimed to<br>Predicate Devices | The ACCUFUSE Cervical System is substantially equivalent to the predicate devices<br>in terms of intended use, design, materials used, mechanical safety and performances. |
| Indications for Use | The ACCUFUSE Cervical System (intervertebral body fusion device) is indicated for use<br>as an intervertebral body fusion device in skeletally mature patients at one or two<br>contiguous levels of the cervical spine (C2-T1) to facilitate fusion in case of degenerative<br>disc disease (DDD) defined as discogenic pain with degeneration of the disc confirmed<br>by patient history and radiographic studies. The ACCUFUSE Cervical System implants<br>are placed via an anterior approach using autogenous bone graft and/or allograft<br>comprised of cancellous and/or corticocancellous bone graft. The ACCUFUSE Cervical<br>System implants are to be used with supplemental fixation. Patients should have at least<br>six weeks of non- operative treatment prior to surgery.<br>The ACCUFUSE Cervical System (cervical plate/screws) is intended for anterior screw<br>fixation to the cervical spine. It is to be used in skeletally mature patients as an adjunct to<br>fusion of the cervical spine (C2 to T1). The system is indicated for use in the temporary<br>stabilization of the anterior spine during the development of cervical spinal fusion in<br>patients with:<br>degenerative disc disease (as defined by neck pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), spondylolisthesis, trauma (i.e. fractures or dislocations), tumors, deformity (defined as kyphosis, lordosis, or scoliosis), pseudarthrosis, failed previous fusion, spinal stenosis |
| Summary of the<br>technological<br>characteristics<br>compared to predicate | Intended Use<br>The ACCUFUSE Cervical System and all the predicates have similar intended uses.<br>Materials<br>The ACCUFUSE Cervical System is composed of the same material as the predicate<br>device<br>Design Features/Functions<br>The ACCUFUSE Cervical System and cited predicate devices share similar basic design<br>features and functions.<br>Dimensions<br>The ACCUFUSE Cervical System is dimensionally similar to cited predicate devices.<br>Sterilization<br>The ACCUFUSE Cervical System is provided non-sterile and cited predicate devices are<br>non-sterile for single use only.<br>Performance Specification<br>Mechanical testing confirmed the ACCUFUSE Cervical System demonstrated equivalent<br>performance to the cited predicate device under the same test conditions. |
| Non-clinical Test<br>Summary | The following analyses were conducted:<br>• Static Axial Compression: Cage per ASTM F2077<br>• Dynamic Compression: Cage per ASTM F2077<br>• Static Torsion: Cage per ASTM F2077<br>• Dynamic Torsion: Cage per ASTM F2077<br>• Subsidence: Cage per ASTM F2267<br>• Static Axial Compression: Plate per ASTM F1717<br>• Dynamic Axial Compression: Plate per ASTM F1717<br>• Static Torsion: Plate per ASTM F1717<br>The results of these evaluations indicate that the ACCUFUSE Cervical System is<br>equivalent to predicate devices. |
| Clinical Test Summary | No clinical studies were performed |
| Conclusions: Non-<br>clinical and Clinical | Absolute Advantage Medical considers the ACCUFUSE Cervical System to be<br>equivalent to the predicate devices listed above. This conclusion is based upon the<br>devices' similarities in principles of operation, technology, materials, performance, and<br>intended use |
In accordance with 21 CFR 807.92 of the Federal Code of Regulations
{4}------------------------------------------------
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.