Cryotop®US-flash and Cryotop®US-scoop are cryopreservation storage devices that are intended for use in vitrification procedures to contain and maintain human oocytes (MII) and embryos.
Device Story
Cryopreservation storage devices for human oocytes (MII) and embryos; used in assisted reproduction laboratories. Devices consist of a sample-loading tip (flat for Flash, U-shaped for Scoop) attached to a handle, which is inserted into a straw enclosure to create a hermetic seal. System isolates samples from liquid nitrogen during storage. Operated by embryologists/lab technicians. Provides a closed system for vitrification; maintains sample integrity during cooling (3,000°C/min) and warming (44,000°C/min). Benefits include secure containment and preservation of reproductive cells for future use.
Clinical Evidence
No new clinical trials conducted. Substantial equivalence supported by bench testing (cooling/warming rates, durability, tensile strength, bacterial endotoxins, Mouse Embryo Assay, package integrity, shelf-life) and leveraging clinical data from K171748 regarding vitrification procedures for oocytes and 2PN embryos.
Technological Characteristics
Materials: Polypropylene, stainless steel, acrylonitrile butadiene styrene (Flash handle), polyethylene terephthalate (Flash tip), polystyrene (Scoop handle/tip). Cooling rate: 3,000°C/min. Warming rate: 44,000°C/min. Design: Handle with sample-loading tip and straw enclosure. Sterilization: ISO 11137-1:2006 and ISO 11137-2:2013. Standards: USP <85>, ANSI/AAMI ST72:2002/(R)2010, ASTM D4169-09, ASTM F1929-15, ASTM F88/F88M-15.
Indications for Use
Indicated for cryopreservation storage of human oocytes (MII) and embryos during vitrification procedures.
Regulatory Classification
Identification
Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Kitazato Corporation Vitrification Kit and Thawing Kit (K171748)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym followed by the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a blue square and the agency's name in a sans-serif font.
November 9, 2018
Kitazato Corporation % Diane Sudduth Senior Consultant, RA Emergo Global Consulting. LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, TX 78746
Re: K181469
Trade/Device Name: Cryotop®US-flash and Cryotop®US-scoop Regulation Number: 21 CFR& 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: II Product Code: MQK Dated: October 10, 2018 Received: October 11, 2018
Dear Diane Sudduth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Michael T. Bailey -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181469
Device Name
Cryotop®US-flash and Cryotop®US-scoop
#### Indications for Use (Describe)
Cryotop@US-flash and Cryotop®US-scoop are cryopreservation storage devices that are intended for use in vitrification procedures to contain and maintain human oocytes (MII) and embryos.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <b> </b> </span> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary — K181469
#### 1. Submission Sponsor
Kitazato Corporation 81 Nakajima, Fuji-city Shizuoka 416-0907 JAPAN Phone number: +(81) 546-66-2202 Contact: Futoshi Inoue Title: President
#### 2. Submission Correspondent
Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327.9997 Contact: Diane Sudduth Title: Senior Consultant, RA
#### 3. Date Prepared
November 8, 2018
#### 4. Device Identification
| Trade Name: | Cryotop®US-flash and Cryotop®US-scoop |
|--------------------|---------------------------------------|
| Common Name: | Cryopreservation Storage Device |
| Regulation Name: | Assisted Reproduction Labware |
| Regulation Number: | 21 CFR 884.6160 |
| Product Code: | MQK (Labware, Assisted Reproduction) |
| Regulatory Class: | II |
#### 5. Predicate Device
Cryotop®US (K153027) manufactured by Kitazato Corporation (formally KITAZATO BIOPHARMA CO., LTD.) This predicate device has not been subject to any design related recalls.
#### 6. Device Description
This 510(k) covers two subject devices, the Cryotop®US-flash and Cryotop®US-scoop that are modified versions of the predicate device (Cryotop®US, K153027). Both devices are cryopreservation storage devices intended for embryo/oocyte vitrification. Cryotop®US-flash and Cryotop®US-scoop are provided sterile and are for single-use only.
{4}------------------------------------------------
The Cryotop®US-flash is composed of a polypropylene straw and an acrylonitrile butadiene styrene handle with a polyethylene terephthalate fine tip where oocytes or embryos are loaded. The fine device is flat with a triangular-shaped area for oocyte/embryo loading. The Cryotop®US-scoop is composed of a polypropylene straw and a polystyrene handle and fine tip where oocytes or embryos are loaded. The fine tip of this device is U-shaped. After oocyte or embryo loading, the handle of the subject devices is inserted into the straw enclosure. A hermetic seal is created via a tapered handle with a stop location integrated into the handle. As the handle is placed into the straw enclosure it creates a closed system keeping the fine tip isolated from the liquid nitrogen. The straw component of both devices also includes a weight at its distal end to aid in maintaining correct device position during storage in liquid nitrogen.
# 7. Indication for Use
Cryotop®US-flash and Cryotop®US-scoop are cryopreservation storage devices that are intended for use in vitrification procedures to contain and maintain human oocytes (MII) and embryos.
| Devices | K181469 (subject device) | K153027 (predicate device) |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | Cryotop®US-flash and Cryotop®US-scoop are<br>cryopreservation storage devices that are<br>intended for use in vitrification procedures to<br>contain and maintain human oocytes (MII) and<br>embryos. | The Cryotop®US is a cryopreservation<br>storage device that is intended for use in<br>vitrification procedures to contain and<br>maintain human 4-8 cell and blastocyst<br>stage embryos. |
| Fundamental<br>design | Same as the predicate | A handle incorporating a sample-loading<br>tip, and a cap (straw) to close the device. |
| Film tip shape | Cryotop®US-flash - Flat<br>Cryotop®US-scoop - U-shape | Flat |
| Dimensions | Cryotop®US-flash:<br>• Total length - 116 mm<br>• Film tip length - 18 mm<br>• Handle length - 98 mm<br>• Straw length - 90 mm<br>• Straw diameter - 3 mm<br>Cryotop®US-scoop:<br>• Total length - 113 mm<br>• Film tip length - 15 mm<br>• Handle length - 98 mm<br>• Straw length - 90 mm<br>• Straw diameter - 3 mm | • Total length - 105 mm<br>• Film tip length - 20 mm<br>• Handle length - 85 mm<br>• Straw length - 100 mm<br>• Straw diameter - 2.9 mm |
| Mode of action | Same as the predicate | Vitrification method - The loaded, closed<br>device is placed in liquid nitrogen |
| Cooling rate | Same as the predicate | 3000 °C/min |
| Warming rate | Same as the predicate | 44000 °C/min |
| Materials | Cryotop®US-flash:<br>• Handle – acrylonitrile butadiene styrene<br>• Fine tip – polyethylene terephthalate<br>• Straw – polypropylene/stainless steel<br>Cryotop®US-scoop:<br>• Handle and fine tip – polystyrene<br>• Straw - polypropylene/stainless steel | • Handle – acrylonitrile butadiene styrene<br>Film sheet – polyethylene terephthalate<br>• Straw – polypropylene/stainless steel |
# 8. Substantial Equivalence Discussion
{5}------------------------------------------------
The subject devices are indicated to contain, vitrify and maintain MII oocytes and all stages of embryo development (2 PN-blastocyst stage), whereas the predicate device is only indicated for cleavage and blastocyst stage embryos. Inclusion of MII oocytes and earlier stage embryos (2 PN) in the subject devices indications does not represent a new intended use because processing of the embryos or oocytes is applicable to same patient population with same clinical needs – infertility preservation. Therefore, the subject and predicate devices have the same intended use.
Cryotop®US-flash and the predicate device have technological characteristics including design, mode of action, materials and cooling/warming rates; however, they have differences in dimensions. Cryotop®USscoop and the predicate device have the same fundamental design, mode of action and cooling/warming rates; however, they have differences in dimensions and materials. The differences noted do not raise different questions of safety and effectiveness.
# 9. Summary of Non-Clinical Performance Data
Non-clinical performance testing was conducted to support substantial equivalence to the predicate device. Cryotop®US-flash and Cryotop®US-scoop passed all the tests shown below in accordance with internal requirements and applicable standards.
- . Cooling Rate Testing - 3,000°C/min
- Warming Rate Testing 44,000°C/min .
- Dimensional Testing - Outer diameter and length per specifications
- . Durability Testing: No burst or liquid nitrogen ingress into straw after 30 second immersion
- Mechanical Tensile Testing – Tensile strength to withstand 5N
- Bacterial Endotoxins – <0.5 EU/device per USP <85> and ANSI/AAMI ST72:2002/(R)2010
- . Mouse Embryo Assay (MEA) – ≥80% of 1-cell embryos developed to blastocysts at 96 hours
- Transportation simulation testing per ASTM D4169-09
- Shelf-life Testing: ●
- Package integrity testing:
- Dye penetration test per ASTM F1929-15 O
- O Seal strength test per ASTM F88/F88M-15
- Dimensional analysis, mechanical, durability, endotoxin, and MEA testing in accordance with the methods and acceptance criteria mentioned above
In addition, information regarding sterilization validation per ISO 11137-1:2006 and ISO 11137-2:2013 provided in K153027 was leveraged to support substantial equivalence.
#### 10. Clinical Performance Data
Clinical information provided in K171748 (Kitazato Corporation Vitrification Kit and Thawing Kit) that incorporates the predicate Cryotop®US device was leveraged in supporting the additional uses of the Cryotop®US-flash and Cryotop®US-scoop for oocyte an 2 PN embryo vitrification procedures.
{6}------------------------------------------------
# 11. Conclusion
The subject and predicate devices have intended use and comparable technological characteristics. The differences in technological characteristics between the subject and predicate devices do not raise different questions of safety and effectiveness. The performance data demonstrate that the subject devices are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.