EndoStream(TM)

K181457 · Partnership Medical Limited · OCX · Jul 11, 2019 · Gastroenterology, Urology

Device Facts

Record IDK181457
Device NameEndoStream(TM)
ApplicantPartnership Medical Limited
Product CodeOCX · Gastroenterology, Urology
Decision DateJul 11, 2019
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The EndoStream™ tubing is intended to provide irrigation via sterile water supply during gastrointestinal (GI) endoscopic procedures when used in conjunction with an irrigation pump.

Device Story

EndoStream™ is a disposable, single-lumen PVC irrigation tubing system; connects sterile water bottles to flexible endoscopes via peristaltic irrigation pumps. Device includes a cap for water bottle attachment and a distal connector with an integrated medical-grade plastic check valve. Operation involves conveying sterile water from an external source to the endoscope to flush the GI tract or clean the endoscope lens during procedures. Used in clinical settings by healthcare professionals. Integrated check valve ensures unidirectional flow, preventing backflow and reducing cross-contamination risks. Benefits include improved visualization and effective irrigation during GI endoscopy.

Clinical Evidence

Bench testing only. Evidence includes sterilization validation per ISO 11135, shelf-life/aging testing per ASTM 1980 (dye penetration, seal strength, visual seal), and biocompatibility testing per ISO 10993 (cytotoxicity, intracutaneous irritation, sensitization). Functionality and performance integrity were verified through bench testing.

Technological Characteristics

Single-lumen PVC tubing; medical-grade plastic check valve; peristaltic pump interface; ethylene oxide sterilization. Compatible with commercial sterile water bottles and various endoscope/pump systems (Olympus, Pentax, EndoStratus, Scope Assist, EGP-100E, ERBE).

Indications for Use

Indicated for patients undergoing gastrointestinal (GI) endoscopic procedures requiring irrigation via sterile water supply.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. July 11, 2019 Partnership Medical Limited Silbiano Gonzales Consultant QRC Consulting, LLC 10422 Huebner Road, Apt# 508 San Antonio, TX 78240 Re: K181457 Trade/Device Name: EndoStream™ Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: OCX Dated: June 3, 2019 Received: June 11, 2019 Dear Silbiano Gonzales: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shanil P. Haugen, Ph.D. Acting Assistant Division Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181457 Device Name EndoStream™ Indications for Use (Describe) The EndoStream™ tubing is intended to provide irrigation via sterile water supply during gastrointestinal (GI) endoscopic procedures when used in conjunction with an irrigation pump. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ K181457 510(k) Summary – Page 1 of 6 #### l. SUBMITTER | Applicant Name:<br>Address: | Partnership Medical Limited<br>Hartley House, Galveston Grove<br>Fenton, Stoke on Trent, ST4 3PE | |-----------------------------|--------------------------------------------------------------------------------------------------| | Contact Person: | Rob Hartley, Director | | US Contact / Correspondent: | QRC Consulting, LLC.<br>Silbiano Gonzales, Consultant | | Address: | 5623 Hamilton Wolfe Rd.,<br>Apt. #921 San Antonio, TX<br>78240 | | Phone:<br>Email: | (210) 287-9558<br>silbianog@sbcglobal.net | II. DEVICE > Trade Name: Catalog Number: EndoStream™ | CATALOGUE (PART) NUMBERS | | | |--------------------------|-------------------------------|---------------------------------------------------| | Catalogue<br>number | Description | Compatibility | | PFE130 | EndoStream™ Irrigation Tubing | Olympus®, Pentax®,<br>EndoStratus™, Scope Assist® | | PFE230 | EndoStream™ Irrigation Tubing | EGP-100E®, Olympus®,<br>ScopeAssist® and ERBE™ | | Common or Usual Name: | | |-----------------------|--| | Regulation Number: | | | Regulatory Name: | | | Regulatory Class: | | | Product Code: | | Endoscopic Irrigation/Suction System 21 CFR 876.1500 Endoscope and accessories Class II ОСХ The Gastroenterology/Urology devices panel has classified Endoscopic Irrigation/Suction System as Class II under 21 CFR §870.1500. OCX is the product code that has been assigned for these types of devices. PART 876 -- GASTROENTEROLOGY-UROLOGY-UROLOGY DEVICES Subpart B--Diagnostic Devices {4}------------------------------------------------ K181457 510(k) Summary – Page 2 of 6 Sec. 876.1500 Endoscope and accessories (a) Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations. (b) Classification. (1) Class II (performance standards). (2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube. rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b) (2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in 876.9. Product Code: OCX Device: Endoscopic irrigation/suction system Definition: To supply sterile water, other solutions and/or suction to endoscopes during endoscopic procedures. Premarket Review: Office of Product Evaluation and Quality OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices THT3A2: Gastroenterology and Endoscopy Devices Team {5}------------------------------------------------ K181457 510(k) Summary – Page 3 of 6 #### III. DEVICE DESCRIPTION The EndoStream™ is available in two (2) configurations: | CATALOGUE (PART) NUMBERS | | | |--------------------------|-------------------------------|---------------------------------------------------| | Catalogue number | Description | Compatibility | | PFE130 | EndoStream™ Irrigation Tubing | Olympus®, Pentax®,<br>EndoStratus™, Scope Assist® | | PFE230 | EndoStream™ Irrigation Tubing | EGP-100E®, Olympus®,<br>ScopeAssist® and ERBE™ | Partnership Medical Ltd introduced the EndoStream™ (product code PFE230) in 2014 which fits the EGP-100, Olympus and ERBE pumps and this product has been supplied to over 50 Hospitals in the UK since that time. The Scope Assist tubing version (product code PFE130) was introduced in 2016 to fit our own Scope Assist pump, but also fits the EndoStratus and Olympus OPF-2 pumps. The EndoStream™ is used to convey sterile water from an external water source to a flexible endoscope to aid its function. The EndoStream™ consists of the following: 1) PVC tubing is used to provide the method of transferring the sterile water from the external source to the flexible endoscope. This sterile water is used to clean the lens of the flexible endoscope during operation or to assist in improving the flexible endoscopes visual field. 2) An endoscope connector is located at the distal end of the device. The EndoStream™ uses a medical grade plastic check valve, which connects to the flexible endoscope via an adaptor. The EndoStream™ functions by using a single lumen in conjunction with a peristaltic pump to provide water to the flexible endoscope. EndoStream™ connects a sterile water supply into the proximal end of an endoscope. Once connected, EndoStream™ allows sterile water to flow into the endoscope which carries it into the Gastrointestinal (GI) tract to flush out the GI tract during endoscopic procedures. There are 2 configurations for use with different pumps. | CATALOGUE (PART) NUMBERS | | | |--------------------------|-------------------------------|---------------------------------------------------| | Catalogue<br>number | Description | Compatibility | | PFE130 | EndoStream™ Irrigation Tubing | Olympus®, Pentax®,<br>EndoStratus™, Scope Assist® | | PFE230 | EndoStream™ Irrigation Tubing | EGP-100E®, Olympus®,<br>ScopeAssist® and ERBE™ | {6}------------------------------------------------ K181457 510(k) Summary – Page 4 of 6 The EndoStream™ tubing is designed to be used in conjunction with a peristaltic pump which attaches to a flexible endoscope to supply sterile water for irrigation during endoscopic procedures. The EndoStream™ tubing is compatible with all commercial sterile water bottles. EndoStream™ is a tubing with a cap which directly screws onto a sterile water bottle. The tubing extremity connects to the endoscope via the pump head on a peristaltic irrigation pump. A back flow valve is integrated in the extremity of each EndoStream™ tubing ensuring the unidirectional flow from the sterile water bottle to the endoscope, thus quaranteeing the prevention of risks of cross contamination. #### IV. INDICATIONS FOR USE The EndoStream™ tubing is intended to provide irrigation via sterile water supply during gastrointestinal (GI) endoscopic procedures when used in conjunction with an irrigation pump. #### V. PREDICATE DEVICES The EndoStream™ is designed as a disposable water irrigation tubing for irrigation pumps (direct connection between the endoscope and sterile water bottle via an irrigation pump.) The EndoStream™ Iegally marketed predicates, to which Partnership Medical Limited is claiming equivalence are: | Predicate Device for the EndoStream™ | | | | | |--------------------------------------|--------------------------------------------------|-------------------|--|--| | Company | Predicate Device Name | FDA 510(k) Number | | | | Byrne Medical, Inc. | EndoGator | K092429 | | | | EndoChoice, Inc. | EndoChoice Water Bottle Cap<br>Irrigation System | K16482 | | | {7}------------------------------------------------ K181457 510(k) Summary – Page 5 of 6 ### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICES The EndoStream™ is substantially equivalent to the Byrne Medical, Inc. (EndoGator, K092429), and the EndoChoice, Inc. (EndoChoice Water Bottle Cap Irrigation System, K161482) | Characteristic<br>Compared | Our Device | Predicate 1 | Predicate 2 | |------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|-----------------------------------------------------------| | Manufacturer | Partnership<br>Medical, Ltd | Byrne Medical,<br>Inc. | EndoChoice,<br>Inc. | | Device | EndoStream™ | EndoGator | EndoChoice<br>Water Bottle<br>Cap<br>Irrigation<br>System | | 510(k) Number | - | K092429 | K161482 | | Indication | The<br>EndoStream<br>™ tubing is<br>intended to<br>provide<br>irrigation via<br>sterile water<br>supply during<br>gastrointestin<br>al (GI)<br>endoscopic<br>procedures<br>when used in<br>conjunction<br>with an<br>irrigation<br>pump. | Same | Same | | Product<br>Code/Class | OCX | OCX | OCX | | Characteristic<br>Compared | Our Device | Predicate 1 | Predicate 2 | | Principle of<br>Operation | Connects to a<br>peristaltic<br>pump to either<br>the<br>endoscopes<br>auxiliary<br>channel to<br>allow flushing<br>of water | Identical | Identical | | Difference | None | None | None | | Materials of<br>Construction | Per section 16 | Identical | Identical | | Sterilization | Ethylene<br>Oxide | Ethylene Oxide | Ethylene<br>Oxide | {8}------------------------------------------------ K181457 510(k) Summary - Page 6 of 6 #### VII. PERFORMANCE DATA The following performance testing was completed (per respective standards). - Sterilization Validation (ISO 11135) - Shelf-Life / Aging Testing (ASTM 1980) . - Dye Penetration o - Seal Strength Testing o - Visual Seal O Biocompatibility: The materials used were tested for biocompatibility per ISO 10993. The tests performed were for cytotoxicity, Intracutaneous Irritation Test and Sensitization. The EndoStream™ product was also test to demonstrate functionality and performance integrity. #### VIII. CONCLUSIONS Partnership Medical Limited considers the EndoStream™ to be substantially equivalent to the legally marketed predicate device listed above, and safe and effective for the intended use. The conclusion is based on similarities in indications for use, materials, performance testing, technological characteristics, principle of operation and design features. Any differences do not raise any new issues of safety or effectiveness.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...