K181445 · Amsino International, Inc. · EZD · Oct 1, 2018 · Gastroenterology, Urology
Device Facts
Record ID
K181445
Device Name
AMSure Hydrophilic Intermittent Catheter
Applicant
Amsino International, Inc.
Product Code
EZD · Gastroenterology, Urology
Decision Date
Oct 1, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5130
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The AMSure® Hydrophilic Intermittent Catheter is inserted into the patient's bladder and indicated for intermittent use for the purpose of drainage of urine from the bladder. The catheter does not contain a balloon on its tip. An optional Water sachet is included to activate the hydrophilic-coated surface prior to use. The target populations for the subject device are Adults and Pediatrics.
Device Story
AMSure Hydrophilic Intermittent Catheter is a sterile, single-use urethral catheter for intermittent bladder drainage. Device consists of a flexible medical-grade PVC tube and funnel; surface features a hydrophilic polyvinyl pyrrolidone (PVP) coating to reduce friction upon activation with sterile water (provided in optional sachet). Used by clinicians or patients in clinical or home settings. Insertion into bladder allows urine drainage; absence of balloon prevents indwelling use. Hydrophilic coating enhances lubricity, potentially reducing urethral trauma and patient discomfort during insertion. Device is provided sterile.
Clinical Evidence
Bench testing only. Biocompatibility testing (cytotoxicity, sensitization, irritation) conducted per ISO 10993-5 and ISO 10993-10. Performance testing included leakage, kink resistance, strength, connector security, flow rate, friction value, chemical analysis (reducing matter, pH, metal ions, EO residue), sterility, water sachet integrity, and shelf-life stability. Human factor testing also performed.
Technological Characteristics
Materials: Medical-grade PVC (tubing/gripper), PVP hydrophilic coating, USP grade sterile water. Dimensions: Various French sizes (10Fr-16Fr). Connectivity: None. Sterilization: Sterile, single-use. Principle: Hydrophilic lubrication via water activation.
Indications for Use
Indicated for intermittent urinary drainage in adult and pediatric patients. Contraindicated for patients requiring indwelling catheterization (no balloon).
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 1, 2018
Amsino International Inc. Zoe Wu RA Specialist 708 Corporate Center Drive Pomona, CA 91768
Re: K181445
Trade/Device Name: AMSure Hydrophilic Intermittent Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological catheter and accessories Regulatory Class: II Product Code: EZD Dated: August 15, 2018 Received: August 21, 2018
Dear Zoe Wu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Glenn B. Bell -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known) K181445
Device Name
AMSure® Hydrophilic Intermittent Catheter
Type of Use (Select one or hoth, as annlicable)
Indications for Use (Describe)
INTENDED USE:
The AMSure® Hydrophilic Intermittent Catheter is inserted into the patient's bladder and indicated for intermittent use for the purpose of drainage of urine from the bladder. The catheter does not contain a balloon on its tip. An optional Water sachet is included to activate the hydrophilic-coated surface prior to use. The target populations for the subject device are Adults and Pediatrics.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the AMSINO logo. The logo is blue and white, with the word "AMSINO" in large, bold letters. Below the word "AMSINO" is the phrase "Advancing Healthcare Worldwide" in smaller letters. The logo is simple and professional, and it conveys the company's mission of improving healthcare around the world.
# Section 5: 510(k) Summary
| Submitter Information | 1 |
|-----------------------|--------------------------------------------------|
| Submitter: | Amsino International Inc. |
| | 708 Corporate Center Drive, Pomona CA 91768, USA |
| Contact Person: | Zoe Wu |
| | RA Specialist |
| Phone: | +86(21)-69117118 |
| Fax: | +86(21)-59148142 |
| E-mail: | Zoe_Wu@amsino.com |
Date Prepared: Sep 28, 2018
- വ Device Information
Trade Name: AMSure® Hydrophilic Intermittent Catheter Device Common Name: Urological Catheter and Accessories Regulation Name: Urological Catheter and Accessories Regulation Number: 21 CFR §876.5130 Product Code: EZD
Device Class: Class II
Review Panel: Gastroenterology/Urology
Table 5-1
| Product Code | Description |
|--------------|----------------------------------------------------------------------|
| HG971610 | Hydrophilic Intermittent Catheter, 10Fr |
| HG971612 | Hydrophilic Intermittent Catheter, 12Fr |
| HG971614 | Hydrophilic Intermittent Catheter, 14Fr |
| HG971616 | Hydrophilic Intermittent Catheter, 16Fr |
| HG971610SS | Hydrophilic Intermittent Catheter with water sachet and sleeve, 10Fr |
| HG971612SS | Hydrophilic Intermittent Catheter with water sachet and sleeve, 12Fr |
| HG971614SS | Hydrophilic Intermittent Catheter with water sachet and sleeve, 14Fr |
| HG971616SS | Hydrophilic Intermittent Catheter with water sachet and sleeve, 16Fr |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for AMSINO, a healthcare company. The logo is blue and white, with the word "AMSINO" in large, bold letters. Below the word "AMSINO" is the tagline "Advancing Healthcare Worldwide". The logo is simple and professional, and it conveys the company's mission of improving healthcare around the world.
- 3 Predicate Device Information
- Cure Catheter ™ Hydrophilic Coated (K132500) -
- AMSure® PVC Intermittent catheter (K091306) -
#### Device Description: ব
The AMSure® Hydrophilic Intermittent Catheter is a urethral catheter intended to be passed through the urethra during urinary catheterization and into the bladder to drain urine. It is hydrophilic and lubricous coated flexible tube owns a hydrophilic polymer in the surface which makes the catheter highly lubricious and good adhesion upon contact with water, saline solution, body fluids. The catheter and gripper is manufactured with medical grade PVC, the surface is coated with a hydrophilic lowfriction coating (polyvinyl pyrrolidone, or PVP), the water sachet is with USP grade sterile water. Each catheter is provided in sterile, single-use packages, intended to be used by adults and pediatrics.
- 5 Intended use
The AMSure® Hydrophilic Intermittent Catheter is inserted into the patient's bladder and indicated for intermittent use for the purpose of urine from the bladder. The catheter does not contain a balloon on its tip. An optional Water sachet is included to activate the hydrophilic-coated surface prior to use. The target populations for the subject device are Adults and Pediatrics.
#### 6 Product Comparison Summary
The proposed AMSure® Hydrophilic Intermittent Catheter and the predicate devices are intended for patients who drain the urine from bladder. These products are urethral catheters that have the same technological characteristics and material for tubing and funnel, same intended use, similar indications for use, the same function, and the same general characteristics.
The design and materials of the proposed devices are identical to the predicate Cure CatheterTM Hydrophilic Coated (K132500) and AMSure® PVC Intermittent catheter (K091306).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for AMSINO, a company that is advancing healthcare worldwide. The logo is blue and white, with the word "AMSINO" in large, bold letters. Below the company name is the slogan "Advancing Healthcare Worldwide" in a smaller font. The logo is simple and professional, and it conveys the company's mission to improve healthcare around the world.
- Nonclinical Testing 7
- � Biocompatibility Testing
Cytotoxicity, Sensitization and Irritation have been conducted per ISO 10993-5 and ISO 10993-10. The Biocompatibility testing demonstrated the biological safety of the proposed devices which may directly contact the patients.
- � Performance Testing
## Performance Testing of AMSure® Hydrophilic Intermittent Catheter was conducted as below:
Table 5-2
| SN | Test items |
|----|---------------------------------------|
| 1 | Packaging inspection |
| 2 | Appearance inspection |
| 3 | Leakage |
| 4 | Kink resistance |
| 5 | Strength test |
| 6 | Connector security |
| 7 | Flow test |
| 8 | Friction value |
| 9 | Chemical performance: Reducing matter |
| 10 | Chemical performance: pH value |
| 11 | Chemical performance: Metal ions |
| 12 | Chemical performance: EO residue |
| 13 | Biological performance: Sterility |
| 14 | Physical performance: Water sachet |
| 15 | Chemical performance: sterile water |
| 16 | Aging (shelf life/stability testing) |
| 17 | Human factor testing |
#### ထ Conclusions
The information provided within this pre-market notification demonstrates that the AMSure® Hydrophilic Intermittent Catheter is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.