K181443 · Mesa Biotech, Inc. · OCC · Nov 23, 2018 · Microbiology
Device Facts
Record ID
K181443
Device Name
Accula RSV Test
Applicant
Mesa Biotech, Inc.
Product Code
OCC · Microbiology
Decision Date
Nov 23, 2018
Decision
SESE
Submission Type
Dual Track
Regulation
21 CFR 866.3980
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Accula RSV Test performed on the Accula Dock is a molecular in vitro diagnostic test utilizing polymerase chain reaction (PCR) and lateral flow technologies for the qualitative, visual detection of respiratory syncytial virus (RSV) viral RNA. The Accula RSV Test uses a nasal swab specimen collected from patients with signs and symptoms of respiratory infection. The Accula RSV assay is intended as an aid in the diagnosis of RSV infection in children and adults in conjunction with clinical and epidemiological risk factors. Negative results do not preclude RSV virus infection and should not be used as the sole basis for treatment or other patient management decisions.
Device Story
The Accula RSV Test is a semi-automated, point-of-care molecular diagnostic system. It processes unprocessed nasal swabs without prior nucleic acid extraction. The system comprises a reusable Accula Dock and a single-use disposable test cassette containing all necessary reagents. The device utilizes OscARTM, a proprietary RT-PCR nucleic acid amplification technology, followed by hybridization-based visual detection on a test strip. The Dock controls the amplification process; the user performs visual interpretation of colored lines on the cassette strip to determine the presence or absence of RSV viral RNA. The test is used in clinical labs and point-of-care settings by healthcare personnel to aid in the diagnosis of RSV infection. By providing rapid molecular results, the device assists clinicians in timely patient management and treatment decisions.
Clinical Evidence
Prospective multi-site clinical study (n=694 evaluable nasal swab specimens) conducted during 2017-2018 RSV season. Performance compared to an FDA-cleared molecular assay. Results: Sensitivity 90.2% (95% CI: 84.2-94.1%), Specificity 95.6% (95% CI: 93.6-97.1%). Invalid rate after repeat testing was 2.3%. Analytical studies included LoD determination, inclusivity (9 strains), specificity (41 microorganisms), and interference testing (19 substances), all showing no cross-reactivity or interference.
Technological Characteristics
Semi-automated, colorimetric, multiplex RT-PCR nucleic acid amplification (OscARTM technology). Uses single-use disposable test cassettes and a reusable electronic Dock. Detection via hybridization-based visual identification of amplification products on a test strip. No nucleic acid extraction required. Sample type: nasal swab.
Indications for Use
Indicated for qualitative, visual detection of RSV viral RNA in nasal swab specimens from children and adults with signs/symptoms of respiratory infection. Intended as an aid in diagnosis in conjunction with clinical/epidemiological risk factors. Negative results do not rule out infection.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K181443
B. Purpose for Submission:
This is a new 510(k) submission for the determination of Substantial Equivalence for the Mesa Biotech Accula RSV Test. Mesa Biotech, Inc. has submitted a combined 510(k) and CLIA waiver package for dual review.
C. Measurand:
RSV L Gene RNA
D. Type of Test:
RT-PCR amplification followed by hybridization and colorimetric visualization of amplified products on a test strip
E. Applicant:
Mesa Biotech, Inc.
F. Proprietary and Established Names:
Accula RSV Test
G. Regulatory Information:
1. Regulation section:
21 CRF 866.3980, Respiratory viral panel multiplex nucleic acid assay
2. Classification:
Class II
3. Product code:
OCC – Respiratory Virus Panel Nucleic Acid Assay System
{1}
4. Panel:
Microbiology (83)
H. Intended Use:
1. Intended use(s):
The Accula RSV Test performed on the Accula Dock is a molecular *in vitro* diagnostic test utilizing polymerase chain reaction (PCR) and lateral flow technologies for the qualitative, visual detection of respiratory syncytial virus (RSV) viral RNA. The Accula RSV Test uses a nasal swab specimen collected from patients with signs and symptoms of respiratory infection. The Accula RSV assay is intended as an aid in the diagnosis of RSV infection in children and adults in conjunction with clinical and epidemiological risk factors.
Negative results do not preclude RSV virus infection and should not be used as the sole basis for treatment or other patient management decisions.
2. Indication(s) for use:
Same as Intended Use.
3. Special conditions for use statement(s):
For Prescription Use
4. Special instrument requirements:
To be used only with the Accula Dock Instrument
I. Device Description:
The Accula RSV Test is a semi-automated, colorimetric, multiplex reverse-transcription polymerase chain reaction (RT-PCR) nucleic acid amplification test to qualitatively detect RSV A and RSV B viral RNA from unprocessed nasal swabs that have not undergone prior nucleic acid extraction. The system integrates nucleic acid extraction, reverse transcription, nucleic acid amplification using a novel Mesa Biotech technology, and hybridization-based visual detection into a completely self-contained and semi-automated system. The Accula RSV system consists of a small reusable Dock to drive the automated testing process and a single-use disposable test cassette that contains all the reagents.
Upon insertion of a Test Cassette, the Dock will detect and identify the cassette type. After the user transfers a clinical sample into the cassette and closes the dock lid, the embedded firmware will control fluid flow of the sample into the various chambers of the cassette.
{2}
Amplicon detection requires the hybridization of two internal probes to generate a signal on the Accula RSV detection strip. Dyed polystyrene microspheres are conjugated to oligonucleotide probes to form an amplicon-microsphere complex by hybridization to an internal region of the amplicon. The complex migrates through the pores of the detection strip membrane and across capture zones which contain oligonucleotides complementary to an amplicon region distinct from the detection probe binding site. Hybridization of the amplicon-microsphere complex to a capture zone probe retards the flow of the specific amplicon. This allows for the generation of a visible signal in the form of a colored line at the capture zone.
## Interpretation of results:
Results are interpreted visually by the operator after the test has completed. A colored line of any intensity at the "RSV" location indicates a positive result if the test is valid. A Negative Control line at the end of the test strip controls for non-specific binding or amplification and must be absent for a valid test. A positive control line at the beginning of the strip tests for amplification effectiveness and is necessary to interpret a test as "negative" for RSV.
## J. Substantial Equivalence Information:
1. Predicate device name(s):
Alere i RSV
2. Predicate 510(k) number(s):
K161375
3. Comparison with predicate:
Table 1: Similarities Between Accula RSV and Predicate Device
| Similarities | | |
| --- | --- | --- |
| Item | Mesa Biotech Accula RSV Test | Alere i RSV Test |
| Assay Targets | RSV virus | Same |
| Sample Type | Nasal Swab | Same |
| Assay Results | Qualitative | Same |
| Intended Users and Locations | Clinical Lab and Point of Care | Same |
| Nucleic Acid Purification | No | Same |
| Internal Control | Yes | Same |
| Positive and Negative Control Swabs | Yes | Same |
{3}
Table 2: Differences Between Accula RSV and Predicate Device
| Differences | | |
| --- | --- | --- |
| Item | Mesa Biotech Accula RSV Test | Alere i RSV Test |
| Intended Use | The Accula RSV Test performed on the Accula Dock is a molecular in vitro diagnostic test utilizing polymerase chain reaction (PCR) and lateral flow technologies for the qualitative, visual detection of respiratory syncytial virus (RSV) viral RNA. The Accula RSV Test uses a nasal swab specimen collected from patients with signs and symptoms of respiratory infection. The Accula RSV assay is intended as an aid in the diagnosis of RSV infection in children and adults in conjunction with clinical and epidemiological risk factors.
Negative results do not preclude RSV virus infection and should not be used as the sole basis for treatment or other patient management decisions. | The Alere i RSV assay performed on the Alere i Instrument is a rapid, molecular, in vitro diagnostic test utilizing an isothermal nucleic acid amplification technology for the qualitative detection of respiratory syncytial virus (RSV) viral RNA in direct nasopharyngeal swabs and nasopharyngeal swabs eluted in viral transport media from patients with signs and symptoms of respiratory infection. It is intended for use as an aid in the diagnosis of RSV in children <18 years and adults >60 years in conjunction with clinical and epidemiological risk factors. |
| RSV Target | L viral polymerase gene | NS2 gene and nucleocapsid gene N |
| Assay Technology | PCR amplification and visual identification of amplified products by hybridization to a test strip | Isothermal nucleic acid amplification and detection of amplified products using molecular beacon probes |
| Detection | Dyed microparticle conjugates specifically detect amplified products | Fluorescently-labeled molecular beacons identify amplified RNA targets |
| Instrument | Amplification controlled by the Accula Dock | Amplification performed on the Alere i Instrument |
| Results Interpretation | Visual interpretation of colored lines on a test strip | Optical detection of fluorescence by the Alere i instrument |
4
{4}
K. Standard/Guidance Document Referenced (if applicable):
Not applicable
L. Test Principle:
Nucleic acid amplification plus hybridization to a membrane and chromatographic visual detection
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
The reproducibility of the Accula RSV Test was tested in a study using contrived nasal swabs at three CLIA-waived sites (sites 1 through 3) and one moderately complex site (site 4) based in the United States. The objective of this study was to test panels of contrived nasal swab samples with the Accula RSV Test to demonstrate reproducibility of the assay in the hands of multiple users at multiple sites over multiple non-consecutive days.
Samples were provided to testing operators in panels of 3 samples (RSV Low Positive, Moderate Positive, and Negative). The targeted concentrations for the Moderate Positive samples were approximately 3X the respective limit of detection (LoD), the targeted concentrations for the Low Positive samples were approximately 1X the respective LoD, and the Negative samples contained no RSV. Samples were blinded and randomized. Each operator tested one panel per day, testing a maximum of three samples at a time. Each sample was tested in triplicate (from separate swabs) (2 operators x 1 run x 3 swabs x 5 non-consecutive days = 30 observations for each site per sample type). Results are reported for counts and percent agreement with expected results. Reproducibility was evaluated by site, by operator and by day.
Table 3: Site-to-Site Reproducibility: Percent Agreement and Total Counts
| | Agreement By Site | | | | | | | | Overall Percent Agreement and 95% CI | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | Site 1 | | Site 2 | | Site 3 | | Site 4 | | | |
| | % | Count | % | Count | % | Count | % | Count | | |
| LP^{1} RSV | 100% | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% | 30/30 | 100% (120/120) | (96.9%, 100%) |
| MP^{1} RSV | 100% | 30/30 | 100% | 30/30 | 100% | 31/31* | 100% | 30/30 | 100% (121/121) | (96.9%, 100%) |
| TN^{1,2} | 100% | 30/30 | 100% | 30/30 | 100% | 29/29* | 100% | 30/30 | 100% (119/119) | (96.9%, 100%) |
1 LP = Low Positive, MP = Moderate Positive; TN = True Negative
2 Percent agreement is for negative results
* One negative swab was mistakenly spiked with MP RSV
{5}
Agreement of actual results with expected results was 100%. There were no differences observed within run (replicates tested by one operator), between runs (five different days), between sites (four sites), or between operators (eight operators).
b. Linearity/assay reportable range:
Not Applicable; this is a qualitative assay.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Quality Control:
Each test cassette contains two internal process controls: an internal positive control and a negative control. The positive control is a non-infectious RNA molecule of the MS2 bacteriophage. The negative control is a non-RSV nucleic acid capture probe intended to monitor for non-specific binding. Each test kit also contains separate control swabs with inactivated RSV virus as well as a negative control swab to verify reagent integrity. The manufacturer recommends positive controls be run for each new lot or shipment of kits received, and for each new operator performing the test. Additional positive control swabs are available for purchase from the manufacturer.
Specimen Stability
Contrived samples of RSV A were tested in triplicate on the Accula RSV assay in an experiment to determine specimen stability. Samples were prepared near the empirical LoD in pooled negative matrix containing the Accula RSV nasal swab buffer solution and stored under various conditions. The contrived specimens were prepared by spiking cultured viruses onto rayon-tipped swabs, which were then dipped into pooled negative nasal swab matrix (PNNS); the material was pooled and aliquoted into appropriately sized tubes for storage. Test conditions and the results of the testing are summarized in the table below:
Table 4: Specimen Stability After Elution in Sample Buffer
| Test Sample | Storage Condition | Duration | Pos. RSV / Expected | Pass/Fail |
| --- | --- | --- | --- | --- |
| 1 | None | 0 | 3/3 | Pass |
| 2 | 15-30°C | 1 Hour | 3/3 | Pass |
| 3 | 15-30°C | 4 Hours | 3/3 | Pass |
| 4 | 15-30°C | 24 Hours | 3/3 | Pass |
| 5 | 2-8°C | 24 Hours | 3/3 | Pass |
| 6 | 2-8°C | 72 Hours | 3/3 | Pass |
| 7 | -20°C | 72 Hours | 3/3 | Pass |
| 8 | -20°C | 1 Week | 3/3 | Pass |
| 9 | -80°C | 72 Hours | 3/3 | Pass |
| 10 | -80°C | 2 Weeks | 3/3 | Pass |
| 11 | -20°C | 1 Freeze/thaw | 3/3 | Pass |
{6}
7
| 12 | -20°C | 2 Freeze/thaw | 3/3 | Pass |
| --- | --- | --- | --- | --- |
| 13 | -20°C | 3 Freeze/thaw | 3/3 | Pass |
Mesa Biotech has indicated that specimens should be transferred to the nasal swab buffer sample tube immediately after collection and stored at 2-8°C for no longer than 24 hours. Swabs that have not been eluted into nasal swab buffer can be stored in the original swab packaging for up to two hours at room temperature. Tests supporting the storage of the un-eluted swab sample in the original packaging are described in the decision summary for CW180006.
## Shelf Life
Accula RSV test kits are being tested for shelf life by storage at 30°C for different lengths of time. At each time interval (day 0, 2 weeks, 4 weeks, and at 3 month intervals beyond) stored test kits are used to test control swabs and simulated RSV samples prepared in clinical matrix. Contrived specimens are prepared in clinical matrix near the empirical LoD. Negative samples and control swabs are tested as one replicate. Acceptance criteria for shelf life studies are as follows: all positive samples and controls must yield a 100% positive rate and all negative samples and controls must yield a 100% negative rate. Tests were considered stable as long as they continued to produce positive results for positive samples and negative results for negative samples. Data generated so far demonstrates that the three lots tested are stable for the first three time points tested (up to and including 4 weeks). The firm will continue stability testing and the kit storage conditions for the Accula RSV test will be labeled as stable at 15-30°C for the latest time period for which acceptance criteria have been met in the ongoing stability study. Expiration dating will be extended as the study continues provided that correct results are obtained for all three test lots during the testing period.
## d. Detection limit:
The objective of the Analytical Sensitivity Study was to identify the limit of detection (LoD) of the Accula RSV assay using characterized strains of RSV A and RSV B. Two RSV A strains (RSV A2 and RSV A Long) and two RSV B strains (RSV B1 and RSV B 18537) were tested in replicates of twenty (20) for each concentration. Clinical matrix was used as dilution material for all LoD determination/confirmation experiments.
To determine the limit of detection, a range-finding study was first conducted to establish the lower limit of analytical sensitivity using virus dilutions tested in replicates of n=5. Following identification of the lower end of the detectable concentrations in the range finding study, virus concentrations which produced at least 3/5 detectable results were tested in replicates of 20. If the dilution tested did not result in at least 19/20 "detected" results, a two-fold higher concentration was tested again with 20 replicates. The final concentrations whereby at least 95% of results (≥19/20) were positive for the target virus are shown in the table below:
{7}
Table 5: Accula RSV Limits of Detection
| RSV Strain Tested | Concentration | Detected (Observed / Expected) |
| --- | --- | --- |
| RSV A2 | 300 TCID50/mL | 20/20 |
| RSV A Long | 300 TCID50/mL | 20/20 |
| RSV B1 | 10 TCID50/mL | 20/20 |
| RSV B 18537 | 0.5 PFU/mL | 19/20 |
# e. Analytical reactivity:
Inclusivity verification was evaluated for the Accula RSV test using nine RSV strains. The chosen strains represented clinical RSV A and RSV B isolates that were obtainable for testing. Virus was diluted into a pooled clinical matrix to create contrived samples at an estimated 2X LoD concentration based on the empirical LoD studies. Each strain was tested in triplicate. The RSV strains, test concentrations, and results are shown below:
Table 6: Accula RSV Reactivity
| Virus Tested | Subtype | Concentration Tested | RSV Positive Results / Expected |
| --- | --- | --- | --- |
| RSV 2012-10 | RSV B | 1 PFU/mL | 3/3 |
| RSV 2012-11 | RSV B | 1 PFU/mL | 3/3 |
| RSV A 12/2014 | RSV A | 600 TCID50/mL | 3/3 |
| RSV A 3/2015 | RSV A | 600 TCID50/mL | 3/3 |
| RSV B 12/2014 | RSV B | 20 TCID50/mL | 3/3 |
| RSV 2006 Isolate | RSV A | 600 TCID50/mL | 3/3 |
| RSV CH93(18)-18 | RSV B | 20 TCID50/mL | 3/3 |
| RSV 9320 | RSV B | 20 TCID50/mL | 3/3 |
| RSV B 3/2015 | RSV B | 20 TCID50/mL | 3/3 |
# f. Analytical specificity:
To determine the analytical specificity of the Accula RSV assay, 41 commensal and common pathogenic microorganisms (21 viruses, 19 bacteria, 1 fungus) including some that may be present in the nasal cavity or nasopharynx were tested. No cross-reactivity was observed for any of the microorganisms tested. The organisms tested and the results are shown in the table below:
{8}
Table 7: Accula RSV Specificity/Cross-Reactivity
| Organism Tested | 1,2Concentration Tested | RSV Positive Results / Tested |
| --- | --- | --- |
| Adenovirus Type 1 | 5.10E+05 | 0/3 |
| Adenovirus Type 7A | 3.31E+04 | 0/3 |
| Human Coronavirus 229E | 1.10E+04 | 0/3 |
| Human Coronavirus OC43 | 2.95E+05 | 0/3 |
| Human CMV AD-169 | 1.10E+04 | 0/3 |
| Human CMV Towne | 1.90E+04 | 0/3 |
| Human CMV Merlin | 2.52E+04 | 0/3 |
| Echovirus Type 11 | 2.95E+05 | 0/3 |
| Human Enterovirus 71 | 1.04E+04 | 0/3 |
| Human Metapneumovirus | 1.01E+05 | 0/3 |
| Measles virus | 2.95E+05 | 0/3 |
| Mumps virus | 9.75E+04 | 0/3 |
| *Parainfluenza 1 | 2.52E+04 | 0/3 |
| Parainfluenza 2 | 1.10E+04 | 0/3 |
| Parainfluenza 3 | 1.18E+04 | 0/3 |
| Rhinovirus Type A1 | 3.31E+04 | 0/3 |
| Rhinovirus Type A16 | 3.31E+04 | 0/3 |
| Rhinovirus Type B14 | 3.02E+04 | 0/3 |
| 3Epstein-Barr virus | 3.98E+07 | 0/3 |
| Influenza A/CA/07/2009 | 2.15E+04 | 0/3 |
| Influenza B/Mass/2/2012 | 5.00E+05 | 0/3 |
| Candida albicans | 9.80E+05 | 0/3 |
| Bordetella pertussis | 4.22E+06 | 0/3 |
| Escherichia coli | 1.92E+07 | 0/3 |
| Haemophilus influenzae | 1.20E+06 | 0/3 |
| Klebsiella pneumoniae | 4.15E+07 | 0/3 |
| Lactobacillus sp. | 3.00E+06 | 0/3 |
| Legionella longbeachae | 9.65E+06 | 0/3 |
| Moraxella catarrhalis | 1.99E+05 | 0/3 |
| Mycobacterium tuberculosis | 3.62E+06 | 0/3 |
| Neisseria gonorrhoeae | 6.30E+06 | 0/3 |
| Neisseria meningitidis | 1.28E+06 | 0/3 |
| Neisseria subflava | 7.30E+06 | 0/3 |
| Proteus vulgaris | 2.07E+07 | 0/3 |
| Pseudomonas aeruginosa | 6.05E+05 | 0/3 |
| Staphylococcus aureus | 6.95E+07 | 0/3 |
| Staphylococcus epidermidis | 3.24E+07 | 0/3 |
| Streptococcus pneumonia | 2.09E+06 | 0/3 |
| Streptococcus pyogenes | 2.72 E+07 | 0/3 |
| Streptococcus salivarius | 2.32E+06 | 0/3 |
| 4Mycoplasma pneumoniae | 2.81E+05 | 0/3 |
* Replicate repeated due to an invalid result.
{9}
10
1 Virus concentrations in TCID₅₀/mL
2 Bacteria and C. albicans concentrations in CFU/mL
3 EBV concentration in cp/mL
4 Mycoplasma pneumoniae concentration in CCU/mL
g. Interfering Substances:
To assess substances with the potential to interfere with the performance of the Accula RSV test, two RSV strains or a negative control were tested in replicates of three (3) with each interfering substance at the "worst case" concentration. The RSV strains selected for testing were RSV A2 and RSV B1. Virus was serially diluted into a pooled clinical matrix to achieve a 1.5X LoD concentration.
Each RSV strain was tested with the "worst case" interferent concentration, representing the highest concentration likely to be found in a respiratory sample.
Table 8: Interfering Substances Tested
| Substance | Concentration |
| --- | --- |
| Mucin | 5mg/mL |
| Whole Blood | 1% (v/v) |
| Phenylephrine Nasal Spray | Neat |
| Oxymetazoline Nasal Spray | Neat |
| Ocean Saline Nasal Spray | Neat |
| Chloraseptic Max | Neat |
| Nasacort | Neat |
| Zicam Allergy Relief | Neat |
| Cepacol (Benzocaine) | 3 mg/mL |
| Beclomethasone | 1.6 mg/mL |
| Budesonide | 3.2 mg/mL |
| Dexamethasone | 30 mg/mL |
| Flunisolide | 1.6 mg/mL |
| Fluticasone Propionate | 0.25 mg/mL |
| Mometasone furoate | 1 mg/mL |
| Triamcinolone | 0.05 mg/mL |
| Zanamivir (Relenza) | 50 mg/mL |
| Tobramycin | 75 mg/mL |
| Mupircoin | 20 mg/mL |
No interference was observed for any of the substances tested above at the concentrations listed. No false positive results were observed for negative (control) samples.
h. Assay cut-off:
Not applicable; the detection signal is read visually.
{10}
11
2. Comparison studies:
a. Method comparison with predicate device:
Not applicable. Performance of the Accula RSV Test was evaluated against the comparator method in a prospective clinical study.
b. Matrix comparison:
Not applicable.
3. Clinical studies:
Clinical performance was evaluated in a multi-site study in the U.S. during the 2017-2018 RSV season. Accula RSV was used to evaluate fresh, prospectively collected nasal swab specimens from children and adults with suspected respiratory infection and meeting the inclusion/exclusion criteria. Samples were tested with the Accula RSV within one hour of sample collection. At all sites, one swab specimen was tested directly on the Accula RSV according to product instructions. The other swab was eluted in UTM and shipped to a reference laboratory for testing with a high performance, FDA-cleared molecular RSV assay. Elution and shipping of the reference swab sample was performed according to the kit instructions. Discrepant results were tested using alternative FDA-cleared molecular assays at the reference laboratory.
A total of 749 subjects were enrolled in the study with 55 specimens found to be unevaluable. Of those 55 specimens that were unevaluable, 39 samples were rejected due to protocol deviations, and 16 samples returned invalid results after repeat testing. A total of 694 nasal swab specimens were considered evaluable for the purpose of data analysis in the accuracy study. Patient age and gender distribution for the evaluable specimens is presented in the table below:
Table 9: Prospective Clinical Study Age and Gender Distribution
| Age Group (Years) | Female | Male | Total |
| --- | --- | --- | --- |
| <5 | 164 | 207 | 371 |
| 6-18 | 67 | 67 | 134 |
| 19-59 | 78 | 31 | 109 |
| ≥60 | 54 | 26 | 80 |
| Total | 363 | 331 | 694 |
During the prospective clinical study, the initial invalid rate was 4.2% (30/710) (95% CI: 3.0% to 6.0%). After repeat testing (according to the product instructions) the invalid rate was 2.3% (16/710) (95% CI: 1.4% to 3.6%). Clinical performance of the Accula RSV assay compared to an FDA-cleared molecular assay for nasal swab samples is presented below.
{11}
12
Table 10: Accula RSV Performance Compared to FDA-cleared Molecular Comparator.
| Accula RSV | Comparator | | |
| --- | --- | --- | --- |
| | Positive | Negative | Total |
| Positive | 129 | 24^{a} | 153 |
| Negative | 14^{b} | 527 | 541 |
| Total | 143 | 551 | 694 |
| Sensitivity: | 90.2% (95% CI: 84.2-94.1%) | | |
| Specificity: | 95.6% (95% CI: 93.6-97.1%) | | |
a RSV was detected in 22/24 false positive specimens and 7/24 false positive specimens using two alternative FDA-cleared molecular RSV assays.
b RSV was not detected in 12/14 false negative specimens and 13/14 false negative specimens using two alternative FDA-cleared molecular RSV assays.
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
In the Accula RSV clinical study (described in the "Clinical Studies" section above), a total of 694 nasal swab specimens were evaluable by the Accula RSV assay. The number and percentage of RSV positive cases per specified age group, as determined by the Accula RSV assay, are presented in the tables below:
Table 12: RSV Expected Values
| Age Group (Years) | Number of Specimens | Number of RSV Positives | RSV Positivity Rate |
| --- | --- | --- | --- |
| <5 | 371 | 134 | 36.1% |
| 6-18 | 134 | 9 | 6.7% |
| ≥19 | 189 | 10 | 5.3% |
| Total | 694 | 153 | 22.0% |
N. Instrument Name:
Accula Dock
O. System Descriptions:
1. Modes of Operation:
Does the applicant's device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ______ or No ☐ X
{12}
Does the applicant's device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ or No ☑
2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ☑ or No ☐
3. Specimen Identification:
Specimen ID is entered by hand directly onto the test cassette.
4. Specimen Sampling and Handling:
Not applicable. The specimens are manually inserted into the test cassette in the instrument.
5. Calibration:
The Accula Dock is factory calibrated and does not require any further calibration at the user site.
6. Quality Control:
Quality control is addressed for each specific FDA-cleared assay to be run on the Accula Dock instrument (separately cleared).
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above: N/A
Q. Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.