K181431 · Cook Incorporated · GEX · Jul 30, 2018 · General, Plastic Surgery
Device Facts
Record ID
K181431
Device Name
Laser Ureteral Catheter
Applicant
Cook Incorporated
Product Code
GEX · General, Plastic Surgery
Decision Date
Jul 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Indications for Use
The Laser Ureteral Catheter is intended for protection and delivery of a laser fiber in the urinary tract.
Device Story
Single lumen catheter; 7.1 French diameter; 40 cm length; manufactured from polyethylene tubing with proximal/distal female and male Luer lock adapters and connector caps. Device protects and delivers laser ablation fibers within the urinary tract. Used by clinicians in a clinical setting. Provides a conduit for laser fibers; ensures fiber integrity during navigation of the urinary tract. Benefits patient by facilitating safe, controlled laser delivery for urological procedures.
Clinical Evidence
Bench testing only. Performance testing included compatibility and dimensional testing, tensile strength, radiopacity, and kink radius. Biocompatibility testing included cytotoxicity (ISO MEM Elution), sensitization (Guinea Pig Maximization), and irritation/intracutaneous reactivity. All predetermined acceptance criteria were met.
Technological Characteristics
Single lumen catheter; 7.1 Fr; 40 cm length. Material: Polyethylene radiopaque tubing. Components: Proximal/distal female and male Luer lock adapters and connector caps. Sterilization: Ethylene Oxide (EtO).
Indications for Use
Indicated for protection and delivery of a laser fiber in the urinary tract.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
July 30, 2018
Cook Incorporated Carly Powell Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404
Re: K181431
Trade/Device Name: Laser Ureteral Catheter Regulation Number: 21 CFR§ 878.4810 Regulation Name: Laser Surgical Instrument for use in General and Plastic Surgery and in Dermatology Regulatory Class: II Product Code: GEX Dated: May 31, 2018 Received: June 1, 2018
Dear Carly Powell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Glenn B. Bell -S
### for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181431
Device Name Laser Ureteral Catheter
Indications for Use (Describe)
The Laser Ureteral Catheter is intended for protection and delivery of a laser fiber in the urinary tract.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in large, white, sans-serif font at the top. Below "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font.
2.0 510(k) Summary
750 DANIELS WAY, I
K181431 Page 1 of 5
# Laser Ureteral Catheter 21 CFR 878.4810 Date Prepared: July 30, 2017
Cook Incorporated
Cook Incorporated
Carly Powell Karthik Pillai
# Submitted By:
Submission: Applicant: Contact:
Applicant Address:
Contact Phone Number: Contact Fax Number:
# Device Information:
750 Daniels Way Bloomington, IN 47404 (812) 335-3575 x104913 (812) 332-0281
Traditional 510(k) Premarket Notification
| Trade Name: | Laser Ureteral Catheter |
|------------------------------------|----------------------------------------------------------------------------------------|
| Common Name: | Powered Laser Surgical Instrument |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Classification Regulation: | 21 CFR 878.4810, Product Code GEX |
| Device Class/Classification Panel: | Class II, General & Plastic Surgery |
# Predicate Device:
The Laser Ureteral Catheter is substantially equivalent to the following device:
- K960768 Laser Fiber Delivery/Cleaning Catheter (Cook Urological Inc.). The Laser Fiber Delivery/Cleaning Catheter is used to protect the delivery of a laser ablation fiber for periodic, intraoperative cleaning of charred, thermally degraded tissue from the laser fiber and for irrigation of the surgical site.
# Reference Device:
- . K171662 Ureteral catheters are indicated for access and catheterization of the urinary tract, including the following applications:
- Delivery of contrast media
- I Drainage of fluids from the urinary tract
- Delivery of irrigation fluids to the urinary tract
- Navigation of a tortuous ureter
{4}------------------------------------------------
2
- . Access, advancement, or exchange of wire guides (open-ended catheters only)
- The Pediatric Ureteral Catheter is indicated for access and catheterization of the o urinary tract in pediatric patients, including the following applications:
- . Delivery of contrast media
- I Drainage of fluids from the urinary tract
- I Delivery of irrigation fluids to the urinary tract
- I Navigation of a tortuous ureter
## Device Description:
The Laser Ureteral Catheter is a single lumen catheter available in a diameter of 7.1 French and a length of 40 centimeters. The Laser Ureteral Catheter is manufactured from polyethylene tubing, with a proximal female Luer lock adapter and connector cap, and male Luer lock adapter and connector cap with a cap seal. The device is provided sterile and intended for one-time use.
## Indications for Use:
The Laser Ureteral Catheter is intended for protection and delivery of a laser fiber in the urinary tract.
# Comparison to Predicate:
The Laser Ureteral Catheter and the predicate device, Laser Fiber Delivery/Cleaning Catheter (K960768), are substantially equivalent in indications for use; these devices have similar design, technological characteristics, construction and function. Both are catheters intended to protect the delivery of a laser fiber. The Laser Ureteral Catheter and the reference device Ureteral Catheter (K171662) are similar in dimension, material, and these devices are both used in the ureter.
The intended uses for the Laser Fiber Delivery/Cleaning Catheter (K960768), and the subject device Laser Ureteral Catheter are fundamentally identical; "intended to be used to protect and deliver a laser fiber". The predicate has a broader indication for general surgery, whereas the subject device is specifically indicated for the urinary tract. The addition of this specific indication for the use in urinary tract does not raise any new questions of safety and effectiveness, since predicate is indicated for general surgery. Additionally, the access and catheterization of the urinary tract using a ureteral catheter is supported by the reference device Ureteral Catheter (K171662). The subject device and the reference device, Ureteral Catheters (K171662), are both indicated to access and catheterize the urinary tract.
{5}------------------------------------------------
Cook Incorporated - Traditional 510(k) Laser Ureteral Catheter July 17, 2018
The subject and predicate device have similar intended uses, and therefore, no new questions of safety or effectiveness are raised. A comparison of the intended uses and technological characteristics of the subject, reference and predicate devices is provided in Table 2.0-1.
| | PREDICATE DEVICE | REFERENCE DEVICE | SUBJECT DEVICE |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| | Laser Fiber Delivery/Cleaning<br>Catheter | Ureteral Catheters | Laser Ureteral Catheter |
| Manufacturer | Cook Urological, Inc | Cook Incorporated | Cook Incorporated |
| 510(k) | K960768 | K171662 | Subject of this submission |
| Regulation | 21 CFR §878.4810<br>Laser surgical instrument for use<br>in general and plastic surgery<br>and in dermatology. | 21 CFR §876.5130<br>Catheter, Ureteral, General & Plastic<br>Surgery | 21 CFR §878.4810<br>Laser surgical instrument for use in<br>general and plastic surgery and in<br>dermatology. |
| Product Code | GEX | KOD | GEX |
| Classification | II | II | Identical |
| Indications for<br>Use | The Laser Fiber<br>Delivery/Cleaning Catheter is<br>intended to be used to protect the<br>delivery of a laser ablation fiber<br>for periodic, intraoperative<br>cleaning of charred, thermally<br>degraded tissue from the laser<br>fiber and for irrigation of the<br>surgical site. The 5.7 Fr Laser<br>Fiber Delivery/Cleaning<br>Catheter will accept laser fibers<br>up to 1250 microns (0.050 inch).<br>The 8.0 Fr Laser Fiber<br>Delivery/Cleaning Catheter will<br>accept fibers up to 1500 microns<br>(0.060 inch). | Ureteral catheters are indicated for<br>access and catheterization of the<br>urinary tract, including the following<br>applications:<br>• Delivery of contrast media<br>· Drainage of fluids from the urinary<br>tract<br>• Delivery of irrigation fluids to the<br>urinary tract<br>• Navigation of a tortuous ureter<br>• Access, advancement, or exchange of<br>wire guides (open-ended catheters<br>only)<br>The Pediatric Ureteral Catheter is<br>indicated for access and catheterization<br>of the urinary tract in pediatric patients,<br>including the following applications:<br>• Delivery of contrast media<br>• Drainage of fluids from the urinary<br>tract<br>• Delivery of irrigation fluids to the<br>urinary tract<br>· Navigation of a tortuous ureter | The Laser Ureteral Catheter is<br>intended for protection and<br>delivery of a laser fiber in the<br>urinary tract. |
Table 2.0-1 Comparison Table
{6}------------------------------------------------
| | PREDICATE DEVICE | REFERENCE DEVICE | SUBJECT DEVICE |
|----------------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|--------------------------------|
| | Laser Fiber<br>Delivery/Cleaning Catheter | Ureteral Catheters | Laser Ureteral Catheter |
| Catheter<br>Diameter (Fr) | 5.7-accept fibers up to 1250<br>microns (0.05 inch),<br>8.0-accept fibers up to 1500<br>microns (0.06 inch) | 3-9 | 7.1 |
| Catheter<br>Length (cm) | 25-40 | 10, 15, 70, 85, 120 | 40 |
| Catheter<br>Material | Polyethylene | Radiopaque Polyvinyl chloride or<br>Polyurethane Radiopaque tubing or<br>Polytetrafluoroethylene | Polyethylene Radiopaque Tubing |
| Maximum<br>shelf-life<br>(years) | Unknown | 3 | 3 |
| Sterilization<br>Method | EtO | EtO | Identical |
| Packaging | Tyvek-Poly pouch | Tyvek polyethylene peel-open pouch | Tyvek-PET/ LDPE |
### Performance Data:
The subject device underwent the applicable testing listed below to ensure reliable design and performance under the testing parameters. Performance and biocompatibility testing were conducted in accordance with the following applicable FDA guidance documents to confirm the reliable performance of critical device characteristics.
- . Performance - Testing shows that the subject device conforms to the performance testing requirements based on intended use. All predetermined acceptance criteria were met.
| Test Performed | Guidance Document |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|
| Compatibility and Dimensional Testing – Testing ensures dimensional and<br>compatibility requirements were met. | N/A |
| Tensile Strength – Testing shows that there should be no fracture of catheter shaft and<br>hub-to-shaft. | |
| Radiopacity – Testing shows that the mean radiopacity met the acceptance criteria and<br>evaluated radiopacity by subjecting the ureteral catheters to a comparative<br>fluoroscopic evaluation. | |
| Kink Radius – Testing determined the kink radius of the ureteral catheter tubing. | |
- . Biocompatibility - Testing shows that the subject device conforms to the biocompatibility requirements based on its intended use. All predetermined acceptance criteria were met.
4
{7}------------------------------------------------
5
| Testing Performed | Guidance Document |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity – ISO MEM Elution | Use of International Standard<br>ISO 10993-1, "Biological<br>evaluation of medical devices -<br>Part 1: Evaluation and testing<br>within a risk management<br>process" |
| Sensitization – Guinea Pig Maximization | |
| Irritation/Intracutaneous Reactivity - Intracutaneous Study | |
All predetermined acceptance criteria were met.
### Conclusion:
The data included in this submission indicate that the subject device does not raise new questions of safety or effectiveness compared to the predicate device, Laser Fiber Delivery/Cleaning Catheter (K960768), which supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.