K181430 · Lensar, Inc. · OOE · Aug 9, 2018 · Ophthalmic
Device Facts
Record ID
K181430
Device Name
LENSAR Laser System - fs 3D (LLS-fs 3D)
Applicant
Lensar, Inc.
Product Code
OOE · Ophthalmic
Decision Date
Aug 9, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4390
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LENSAR Laser System - fs 3D (LLS-fs 3D) with Streamline™ is an ophthalmic surgical laser indicated for use: · in the creation of an anterior capsulotomy; - · in patients undergoing surgery requiring laser-assisted fragmentation of the cataractous lens; - · in the creation of full and partial thickness single-plane and multi-plane arc cuts/incisions in the cornea; - in patients undergoing ophthalmic surgery or other treatments requiring pocket cuts/incisions in the cornea; and - · in the creation of a corneal flap in patients undergoing treatment requiring initial lamellar resection of the cornea.
Device Story
Ophthalmic surgical laser system; utilizes pulsed laser for precision tissue cutting. Inputs: pre-op analysis data; patient eye imagery. Processing: automated Iris Registration with cyclorotation adjustment; IntelliAxis-C/L markers for Toric IOL alignment; treatment planning tools. Outputs: precision anterior capsulotomy; cataractous lens fragmentation; corneal incisions (arc, pocket, flap). Used in clinical settings by surgeons. Provides automated control of incision size, depth, and location; enhances surgical precision; facilitates IOL alignment; improves patient outcomes in cataract and corneal surgeries.
Clinical Evidence
No clinical data. Bench testing only. Porcine eye studies compared elliptical PID to predicate PID regarding IOP rise, eye stability, and corneal fold potential. Iris Registration and Limbus detection functions verified via image analysis. Biocompatibility testing (cytotoxicity, irritation, sensitization) performed on molded PMMA contact lens.
Technological Characteristics
Ophthalmic femtosecond laser. Materials: molded PMMA contact lens. Features: elliptical Patient Interface Device (PID), automated Iris Registration, IntelliAxis-C/L markers. Connectivity: integration with pre-op analysis devices. Sterilization: autoclaving for multi-use PID Ring Arm. Software: major level of concern.
Indications for Use
Indicated for patients undergoing ophthalmic surgery requiring anterior capsulotomy, laser-assisted cataract fragmentation, corneal arc cuts/incisions (full/partial thickness), corneal pocket cuts/incisions, or corneal flap creation via lamellar resection.
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 9, 2018
LENSAR, Inc. Keith Peck Director, Ouality Assurance 2800 Discovery Drive, Suite 100 Orlando, FL 32826
Re: K181430
Trade/Device Name: LENSAR Laser System - fs 3D (LLS-fs 3D) Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: OOE Dated: May 31, 2018 Received: June 6, 2018
Dear Keith Peck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
# Bradley S. Cunningham -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K181430
Device Name
LENSAR Laser System - fs 3D (LLS-fs 3D)
Indications for Use (Describe)
The LENSAR Laser System - fs 3D (LLS-fs 3D) with Streamline™ is an ophthalmic surgical laser indicated for use:
· in the creation of an anterior capsulotomy;
- · in patients undergoing surgery requiring laser-assisted fragmentation of the cataractous lens;
- · in the creation of full and partial thickness single-plane and multi-plane arc cuts/incisions in the cornea;
- in patients undergoing ophthalmic surgery or other treatments requiring pocket cuts/incisions in the cornea; and
- · in the creation of a corneal flap in patients undergoing treatment requiring initial lamellar resection of the cornea.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
This 510(k) Summary has been prepared in accordance with the requirements of 21 CFR 807.92.
#### SUBMITTER INFORMATION 1.0
The submitter of this 510(k) Summary is:
| Company: | Contact Person: |
|---------------------------------|-------------------------|
| LENSAR, Inc. | Keith Peck |
| 2800 Discovery Drive, Suite 100 | (888) 575-6412 (Office) |
| Orlando, FL 32826 | (407) 386-7228 (Fax) |
Summary Preparation Date: May 31, 2018
#### DEVICE INFORMATION 2.0
| | Trade/Proprietary Name: LENSAR Laser System - fs 3D (LLS-fs 3D) |
|-------------------------|-----------------------------------------------------------------|
| Common/Usual Name: | Ophthalmic Femtosecond Laser |
| Classification Name(s): | Ophthalmic Femtosecond Laser (21 CFR 886.4390) |
| Product Code(s): | OOE |
| Review Panel: | Ophthalmic |
| Regulatory Class: | II |
#### 3.0 PREDICATE DEVICE
The legally marketed (predicate) device to which LENSAR is claiming substantial equivalence to is:
| 510(k) Number | Device Name | Manufacturer |
|---------------|-----------------------------------------|--------------|
| K173346 | LENSAR Laser System – fs 3D (LLS-fs 3D) | LENSAR, Inc. |
#### DEVICE DESCRIPTION 4.0
The LENSAR Laser System - fs 3D (LLS-fs 3D) with Streamline™ is a medical device for use in ophthalmic surgery. The device utilizes a pulsed laser that can be used to cut a precision capsulotomy in the anterior lens capsule, to fragment the cataractous lens for removal during cataract surgery, and to create full and partial thickness single-plane and multi-plane arc cuts/ incisions in the cornea, each of which may be performed either individually or consecutively during the same procedure. The device is also intended for use in the creation of pocket cuts/ incisions in the cornea in patients undergoing ophthalmic surgery, and in the creation of a corneal flap in patients undergoing treatment requiring initial lamellar resection of the cornea, each of which may only be performed individually.
{4}------------------------------------------------
Use of the laser provides automated precision control of the size of the capsular opening; the type and parameters of laser fragmentation treatment within the lens; the size, architecture, and location of full thickness incisions within the cornea; the size, architecture, location, depth, and quantity of partial thickness incisions within the cornea; and the size, architecture, and depth of pocket and flap cuts.
The LENSAR Laser System – fs 3D (LLS-fs 3D) with Streamline™ includes the integration with pre-op analysis devices, automated Iris Registration with automatic cyclorotation adjustment, IntelliAxis-C™ (corneal) and IntelliAxis-L™ (lens) markers for simple alignment of Toric IOLs as well as treatment planning tools for precision-guided laser treatments.
#### INDICATIONS FOR USE 5.0
The LENSAR Laser System – fs 3D (LLS-fs 3D) with Streamline™ is an ophthalmic surgical laser indicated for use:
- in the creation of an anterior capsulotomy; ●
- in patients undergoing surgery requiring laser-assisted fragmentation of the cataractous ● lens:
- in the creation of full and partial thickness single-plane arc cuts/incisions in the cornea;
- in patients undergoing ophthalmic surgery or other treatments requiring pocket cuts/ incisions in the cornea; and
- in the creation of a corneal flap in patients undergoing treatment requiring initial lamellar resection of the cornea.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE 6.0 PREDICATE DEVICE
All elements of the proposed device remain unchanged from that cleared under K173346 except for the following technological difference:
- . introduction of a new elliptical-shaped Patient Interface Device (PID) Kit which can be used as an alternate to the currently cleared PID for use with all indications except for the indications related to corneal pockets and flaps;
- . a software change to ensure that Limbus detection is not impacted by the change in dimensions of the elliptical PID as mentioned above;
- . replacing the machined Poly methyl methacrylate (PMMA) contact lens of the Curved Contact PID (as cleared in K173346) with a molded PMMA contact lens to allow for a more cost-effective and easier manufacturing process.
A software change unrelated to the requested change for the elliptical-shaped PID is a speed-up of the Iris Registration function and a slight expansion of the cyclorotation range over which the Iris Registration can be used, both without compromising safety and effectiveness.
{5}------------------------------------------------
#### PERFORMANCE DATA 7.0
The following performance data were undertaken in support of the substantial equivalence determination.
### Performance Evaluation of the New Elliptical PID Modification:
Verification and validation testing were completed to demonstrate that the proposed device performance complies with specifications and requirements identified for the LENSAR Laser System - fs 3D. Each function and/or feature was verified through unit testing and system testing by means of the appropriate test case or test specification. The unit/system verification test reports provide the test cases, expected results for each test case, and the actual results obtained. All criteria for this testing were met and the results demonstrate that the LENSAR System with the new elliptical PID meets all performance specifications and requirements. The objectives defined in the validation plan were achieved according to the validation results. The software on the proposed LLS-fs 3D System has been updated in support of the changes outlined above.
Performance data supporting substantial equivalence is summarized as follows:
- . An analysis of IOP rise as a result of the new elliptical PID was completed. A study was performed using porcine eyes, comparing the new elliptical PID to LENSAR's existing PID (already pre-market cleared). The results show the new proposed elliptical PID was consistent with (as it relates to IOP pressure rise) LENSAR's existing PID.
- An analysis of eye stability during surgery with the new elliptical PID as compared to . LENSAR's current PID (already pre-market cleared) was performed. The results show that after the application of the suction, the minimum force necessary to detach the porcine eye from the suction ring was comparable to that of LENSAR's existing PID.
- . An analysis of the potential to have corneal folds as a result of the new elliptical PID as compared to LENSAR's existing PID (already pre-market cleared) was assessed. Using OCT images, no visible folds were noted.
- . An analysis of the impact of the new elliptical PID on the Iris Registration and Limbus detection functions was performed. The analysis incorporated images of eyes with the elliptical PID and showed that all specifications of the Iris Registration and the Limbus Detection functions were achieved.
- A review of the hazard analysis of all potential hazards to the patient, surgeon and other . system operators was performed to consider all changes to the proposed LLS-fs 3D device. The hazard analysis demonstrates that all potential hazards have acceptable levels of probability/severity characteristics.
In all cases, the test results showed that the new elliptical PID meets the performance specifications and requirements. The comparison shows that the proposed LENSAR Laser System - fs 3D with the new elliptical PID is substantially equivalent to the LENSAR Laser System - fs 3D (LLS-fs 3D) (K173346).
{6}------------------------------------------------
The minor differences between the additional LENSAR device feature and the predicate device do not raise any new questions of safety or effectiveness.
# Biocompatibility Testing:
Biocompatibility evaluations of materials coming in contact with the patient fluid path in the LENSAR Laser System - fs 3D (LLS-fs 3D) were conducted in accordance with national/ international recognized standards.
The only difference between the new elliptical PID and the currently approved PID is a slight dimensional change in the suction ring portion of the PID. No materials changes have been made as a result of this change. Because of this and the Biocompatibility risk evaluation for the new elliptical PID Kits no new biocompatibility testing was deemed necessary.
As mentioned earlier, the PMMA contact lens that is part of the Curved Contact PID (currently machined) is changing to a molded PMMA contact lens. The new part was evaluated to ensure materials used were biocompatible and the relevant biocompatibility testing completed. Based on this testing (cytotoxicity, irritation and sensitization) no issues were found and all testing passed.
The PID Ring Arm is multi-use and is sterilized by autoclaving. There were no changes to the material or assembly of the PID Ring Arm as part of this submission and thus remains unchanged from prior device files.
### Software Verification and Validation Testing:
A complete software verification and validation testing was conducted covering all cited changes and updates since the prior clearance (K173346), and documentation was provided as recommended by FDA's "Guidance for Industry and FDA Staff, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "major" level of concern, since a failure or latent flaw in the software could directly result in serious injury to the patient or operator.
### Summary of Clinical Studies:
The addition of the elliptical PID for use with the LENSAR Laser System did not require clinical performance data to demonstrate substantial equivalence to the predicate device.
#### 8.0 CONCLUSIONS
The activities used to evaluate the LENSAR Laser System – fs 3D (LLS-fs 3D) and the information provided in this 510(k) submission do not identify any new issues of safety or effectiveness.
Based on the above supportive information, the proposed LENSAR Laser System - fs 3D (LLS-fs 3D) with the new elliptical PID and molded PMMA curved contact PID are substantially equivalent with respect to safety and effectiveness and indication for use as cleared in the LENSAR Laser System - fs 3D (LLS-fs 3D) predicate device (K173346).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.