K181400 · Techlab, Inc. · LYR · Aug 21, 2018 · Microbiology
Device Facts
Record ID
K181400
Device Name
H. PYLORI CHEK™
Applicant
Techlab, Inc.
Product Code
LYR · Microbiology
Decision Date
Aug 21, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3110
Device Class
Class 1
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K181400 · Aug 21, 2018
H. PYLORI CHEK™
Techlab, Inc.
Sample repositories (frozen stool samples)
A retrospective study was conducted using 196 frozen stool samples from repositories to evaluate the performance of the H. PYLORI CHEK test. The results were compared against an FDA-cleared ELISA to determine positive and negative percent agreement.
Patients with H. pylori status determined by FDA-cleared ELISA; Sample Size: 196; Number of Sites: 1 (TECHLAB)
FDA-cleared ELISA
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA)
Indications for Use
The TECHLAB® H. PYLORI CHEK™ test is an enzyme immunoassay for the qualitative detection of Helicobacter pylori specific antigen. It is intended for use with human fecal specimens to aid in the diagnosis of H. pylori infection and to demonstrate loss of H. pylori antigen following treatment. The test can be used with unpreserved fecal specimens and fecal specimens preserved in transport media from patients suspected of H. pylori infection. Testing of patients to demonstrate loss of H. pylori antigen following treatment should be performed no sooner than 4 weeks after completion of the treatment regimen. Test results should be taken into consideration by the physician in conjunction with the patient history and symptoms.
Device Story
H. PYLORI CHEK™ is an enzyme immunoassay (EIA) for qualitative detection of H. pylori antigen in human fecal specimens. The device utilizes a microassay plate coated with capture antibodies and a conjugate of antibodies coupled to horseradish peroxidase. Fecal specimen aliquots are added to wells; if antigen is present, it binds to the conjugate and capture antibody. After washing, substrate (tetramethylbenzidine and peroxide) is added, producing a color change proportional to the amount of antigen-antibody complex. The test is performed in clinical laboratories; results are interpreted visually or via dual-wavelength spectrophotometric analysis. Results assist physicians in diagnosing H. pylori infection and verifying eradication post-treatment, providing a non-invasive alternative to endoscopy for patients without alarm symptoms.
Clinical Evidence
Prospective clinical study (n=109) compared device to composite reference method (CRM) for initial diagnosis, showing 100% sensitivity and 96.1% specificity. Post-therapy evaluation (n=9) showed 77.8% sensitivity. Retrospective study (n=196) compared device to FDA-cleared ELISA, showing 100% PPA and 100% NPA. Bench testing included precision, LoD, cross-reactivity, and interference studies.
Technological Characteristics
Qualitative enzyme-linked immunosorbent assay (ELISA). Uses polyclonal antibodies. Form factor: microassay plate. Energy source: spectrophotometric plate reader (450 nm or 450/620 nm). Storage: 2°C-8°C. Standards: CLSI EP07-A2, EP15-A3, EP17-A2, EP25-A.
Indications for Use
Indicated for qualitative detection of H. pylori antigen in human fecal specimens from patients (≥18 years) suspected of H. pylori infection, including diagnosis and post-treatment monitoring (≥4 weeks post-therapy).
Regulatory Classification
Identification
Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identify Campylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases. Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
K181400
B. Purpose for Submission:
To obtain a substantial equivalence determination for the detection of *Helicobacter pylori* antigens in human stool.
C. Measurand:
*Helicobacter pylori* antigen
D. Type of Test:
Qualitative enzyme immunoassay (EIA)
E. Applicant:
TECHLAB, Inc.
F. Proprietary and Established Names:
H. PYLORI CHEK
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3110; *Campylobacter fetus* serological reagents
2. Classification:
Class I
3. Product code:
LYR – *Campylobacter pylori*
4. Panel:
83-Microbiology
{1}
H. Intended Use:
1. Intended use(s):
The TECHLAB H. PYLORI CHEK test is an enzyme immunoassay for the qualitative detection of *Helicobacter pylori* specific antigen. It is intended for use with human fecal specimens to aid in the diagnosis of *H. pylori* infection and to demonstrate loss of *H. pylori* antigen following treatment. The test can be used with unpreserved fecal specimens and fecal specimens preserved in transport media from patients suspected of *H. pylori* infection. Testing of patients to demonstrate loss of *H. pylori* antigen following treatment should be performed no sooner than 4 weeks after completion of the treatment regimen. Test results should be taken into consideration by the physician in conjunction with the patient history and symptoms.
2. Indication(s) for use:
Same as the Intended Use.
3. Special conditions for use statement(s):
For prescription use only.
4. Special instrument requirements:
A spectrophotometric plate reader capable of reading the following wavelengths: 450 nm or 450/620 nm.
I. Device Description:
The *H. PYLORI CHEK* test uses antibodies specific to *H. pylori* infection. The “Microassay Plate” in the kit contains immobilized capture antibodies against *H. pylori* antigen. The “Conjugate” consists of antibodies specific to *H. pylori* antigen conjugated to horseradish peroxidase (HRP). In the assay, an aliquot of diluted fecal specimen is transferred to a microassay well containing the Conjugate. If the antigen is present in the specimen, it will bind to the Conjugate and to the immobilized capture antibody during the incubation phase. Any unbound material is removed during the washing steps. Following the addition of the “Substrate”, a color is detected due to the enzyme-antibody-antigen complexes that formed in the presence of antigen. The color if positive is a yellow color that may be read visually or by using a spectrophotometer at the following wavelengths: 450 nm or 450/620 nm.
J. Substantial Equivalence Information:
1. Predicate device name(s):
Premier Platinum HpSA PLUS
{2}
2. Predicate 510(k) number(s):
K053335
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device: H. PYLORI CHEK (K181400) | Predicate: Premier Platinum HpSA PLUS (K053335) |
| Product Code | LYR | LYR |
| Intended Use | The TECHLAB H. PYLORI CHEK test is an enzyme immunoassay for the qualitative detection of Helicobacter pylori specific antigen. It is intended for use with human fecal specimens to aid in the diagnosis of H. pylori infection and to demonstrate loss of H. pylori antigen following treatment. The test can be used with unpreserved fecal specimens and fecal specimens preserved in transport media from patients suspected of H. pylori infection. Testing of patients to demonstrate loss of H. pylori antigen following treatment should be performed no sooner than 4 weeks after completion of the treatment regimen. Test results should be taken into consideration by the physician in conjunction with the patient history and symptoms. | Premier Platinum HpSA PLUS enzyme immunoassay (EIA) is an in vitro qualitative procedure for the detection of Helicobacter pylori antigens in human stool. Test results are intended to aid in the diagnosis of H. pylori infection and to monitor response during and post-therapy in patients. Accepted medical practice recommends that testing by any current method, to confirm eradication, be done at least four weeks following completion of therapy. |
| Measured analyte | Detection of H. pylori antigen | Same |
| Type of Test | Qualitative | Same |
| Controls | Positive and negative control included in the kit | Same |
| Target Population | Persons suspected of having H. pylori infection | Same |
| Storage | Refrigerated (2°C-8°C) | Same |
| Reading Method | Spectrophotometric or Visual | Same |
{3}
| Differences | | |
| --- | --- | --- |
| Item | H. PYLORI CHEK
(K181400) | Predicate: Premier Platinum
HpSA PLUS (K053335) |
| Specimen Type | Fresh or frozen formed, semi-solid, or liquid fecal specimens. Fecal specimens in Cary-Blair and C&S Transport Media | Fresh or frozen formed, semi-solid, or liquid fecal fecal specimens |
| Specimen Storage | Specimens may be held up to 96 hours at 2°C-8°C or at 20°C-25°C prior to testing | Specimens may be held up to 72 hours at 2°C-8°C prior to testing |
| Assay Incubation Temperature | 37°C ± 2°C | 19°C-27°C |
| Time to Result | Approximately 1 hour | Approximately 1 hour |
| Antibody Format | Polyclonal/Polyclonal | Monoclonal/Monoclonal |
## K. Standard/Guidance Document Referenced (if applicable):
CLSI EP07-A2. Interference Testing in Clinical Chemistry; Approved Guideline – Second Edition, May 5th, 2007.
CLSI EP15-A3. User Verification of Precision and Estimation of Bias; Approved Guideline – Third Edition. August 14th, 2015.
CLSI EP17-A2. Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition. January 15th, 2013.
CLSI EP25-A. Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline. January 15th, 2013.
## L. Test Principle:
Enzyme-linked immunosorbent assay
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The precision of the H. PYLORI CHEK test was determined using an 8-member masked fecal specimen panel. The panel consisted of 2 negative, 1 high negative (2 times below the C₅), 1 high negative (C₅), 1 low positive (C₉₅), 1 low positive (2x
{4}
higher than the C₉₅), 1 moderate positive (3x higher than the C₉₅), and 1 moderate positive (4x higher than the C₉₅). Each fecal specimen was spiked with H. pylori antigen (ATCC strain 43526) to achieve the desired level. The specimens were tested twice per day over a 12-day period by multiple technicians using two different kit lots. By spectrophotometric analysis using dual OD (450/620 nm wavelength), positive specimens tested positive 98.3% of the time and negative specimens tested negative 97.8% of the time. By visual read, positive specimens tested positive 93.1% of the time and negative specimens tested negative 94.8% of the time.
The reproducibility of the H. PYLORI CHEK test was determined using the same 8-member fecal specimen panel as described for the precision study. Testing was performed at 2 independent laboratories and on-site at TECHLAB, Inc. The panel members were tested twice a day, in triplicate, over a five-day period by multiple technicians at each site using two different kit lots. Positive and negative controls were run with each sample panel of masked specimens. The results from each laboratory were submitted to TECHLAB, Inc. and compared with in-house results. The results were consistent among all three locations and produced the expected results (by dual OD) for the positive and negative samples 97.5% and 97.5% of time, respectively. By visual read, positive specimens tested positive 96.9% of the time and negative specimens tested negative 92.8% of the time
b. Linearity/assay reportable range:
Not Applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Sample stability study:
The effect of specimen storage on antigen stability was evaluated for both fresh stool samples and stool amples in transport media. The following transport media were used for the study: Thermo Scientific Protocol Cary Blair media and Thermo Scientific Protocol C&S media. For the analysis, a total of 32 fecal specimens were tested with the H. PYLORI CHEK test. The samples were prepared using a negative fecal matrix and spiked with H. pylori antigen (ATCC strain 43526). The panel consisted of 2 negative, 5 high negative (C₅), 10 low positive (1-2x C₉₅), and 15 positive specimens covering the range of the test (For fresh, unpreserved specimens, this range was 50 ng/mL – 1200 ng/mL).
Fresh samples were stored at refrigerated temperatures (between 2°C and 8°C) and room temperatures (between 20°C and 25°C) and were tested at 0, 24, 48, 72 and 96 hours. Positive and negative controls were also tested at each timepoint.
Fecal specimens added to Cary Blair and C&S media were transported as recommended in their respective package inserts; samples were stored at refrigerated temperatures (between 2°C and 8°C) and room temperatures (between 20°C and 25°C) and were tested at 24 hour intervals from 0 to 96 hours.
5
{5}
For fresh samples stored at refrigerated and room temperatures, the positive and negative samples gave the expected results 96.8% and 99.1% of the time, respectively. Storage in Cary Blair transport media did not affect the stability of the samples. Storage in C&S transport media did not affect the stability of samples stored at room temperature, but a slight decrease in sample stability was observed at refrigerated temperatures. Specifically, 93.8% of samples yielded the expected results when stored in C&S transport media at refrigerated temperatures. Based on these results, the recommended storage time for fresh, Cary Blair, and C&S stored samples is up to 96 hours when stored between 2°C and 8°C and between 20°C and 25°C.
## Frozen sample stability study
Stability of frozen fecal matrix samples compared to fresh samples was established using the 32 masked fecal specimen panel prepared as described for the storage stability study. Samples were not diluted into transport media for this study. Samples were stored at ≤ -10°C or ≤ -70°C for 14 days. Specimens were tested at 0, 5, 10, and 14 days. The results showed that all positive samples remained positive and negative samples remained negative throughout the study.
## Freeze/Thaw study
A study was conducted to determine stability after 3 freeze/thaw cycles using the 32-masked fecal specimen panel described for the storage stability study. Samples were not diluted into transport media for this study. The results showed that up to 3 freeze/thaw cycles neither enhanced nor impaired the performance of the H. PYLORI CHEK test. Therefore, the package insert will indicate that if samples are not tested fresh, frozen stool samples may be tested and may undergo up to no more than 2 freeze/thaw cycles.
## d. Detection limit:
The limit of detection (LoD) for the H. PYLORI CHEK test was determined by spiking purified flagellar antigen from H. pylori (ATCC strain 43526) into negative fecal matrix and Cary Blair and C&S transport media. The LoD is defined as the concentration of H. pylori antigen that yields a positive result 95% of the time and a negative result 5% of the time. The LoD was determined as the concentration that provided the correct result 95% of the time based on testing dilutions of antigen in a negative fecal sample matrix, in replicates of 24 for two sets. LoD testing was read at both single (450 nm) and dual (450/620 nm) wavelength reads. The LoD values were similar between single wavelength reads and dual wavelength reads, however single wavelength reads trended slightly higher for some, but not all, sample types. Table 1 lists the LoD values based on a dual wavelength reading for antigen spiked in negative fecal matrix, Cary Blair, and C&S transport media.
6
{6}
Table 1. LoD Values for H. PYLORI CHEK
| ng/mL | ng/test |
| --- | --- |
| Fecal Matrix | |
| 6.70 | 0.13 |
| Cary Blair | |
| 26.57 | 0.33 |
| C & S | |
| 18.19 | 0.23 |
# e. Analytical specificity:
# Cross-reactivity
The H. PYLORI CHEK test was evaluated for cross-reactivity with the bacteria, fungi, and viruses listed below. H. pylori purified flagellar antigen (ATCC strain 43526) was spiked in at $2 - 3\mathrm{x}$ LoD. H. pylori was spiked into clinical matrix that was negative for H. pylori. Bacteria and fungi were spiked at concentration of $>10^{8}$ CFU/mL and viruses at a range from $10^{3.3}$ to $10^{8.25}$ TCID50 units per $0.2\mathrm{mL}$ . Due to safety concerns for Listeria monocytogenes, freeze-dried biomaterial from ATCC was reconstituted at $>1\times 10^{4}$ CFU/mL in PBS and spiked with purified H. pylori flagellar antigen (ATCC strain 43526) at 2-3X LoD. No bacteria, fungi, or viruses tested showed interference with the performance of the H. PYLORI CHEK test.
# Bacteria/Fungi
Acinetobacter baumannii
Bacillus cereus
Borrelia burgdoferi
Campylobacter coli
Campylobacter fetus
Campylobacter helveticus
Campylobacter hyointestinalis
Campylobacter jejuni
Campylobacter lari
Campylobacter upsaliensis
Candida albicans
Clostridium bifermentans
Clostridium perfringens
Edwardsiella tarda
Enterobacter cloacae
Enterococcus faecalis
Escherichia coli
Escherichia coli EIEC
Escherichia coli EIEC
# Viruses
Adenovirus Type 2
Adenovirus Type 40
Human Coxsackievirus B6
Echovirus 9
{7}
Coronavirus, Human
Echovirus 22
Coxsackievirus B2
Echovirus 60
Coxsackievirus B3
Human Coxsackievirus B1
Human rotavirus*
*tested at 10¹·⁷⁵
## Interfering substances
The *H. PYLORI CHEK* test was evaluated for interfering substances with the substances and concentrations listed in Table 2. None of the substances were shown to interfere with the performance of the *H. PYLORI CHEK* test.
Table 2. Interfering Substances
| Barium sulfate (5% w/v) | Mylanta (4.2 mg/mL) |
| --- | --- |
| Benzalkonium Chloride (1% w/v) | Naproxen Sodium (5% w/v) |
| Ciprofloxacin (0.25% w/v) | Nonoxynol-9 (1% w/v) |
| Ethanol (1% w/v) | Nystatin (1% w/v) |
| Hog gastric mucin (3.5% w/v) | Palmitic acid (fecal fat) (40% w/v) |
| Human blood (40% w/v) | Pepto-Bismol (5% v/v) |
| Hydrocortisone (1% w/v) | Phenylephrine (1% w/v) |
| Imodium (5% v/v) | Prilosec OTC (5 ug/mL) |
| Kaopectate (5%v/v) | Sennosides (1% w/v) |
| Leukocytes (0.05% v/v) | Simethicone (10% w/v) |
| Maalox Advanced (5% v/v) | Steric acid (fecal fat) (40% w/v) |
| Mesalazine (10% w/v) | Tagamet (5 ug/mL) |
| Metronidazole (0.25% w/v) | TUMS (50 ug/mL) |
| MiraLAX (7% w/v) | Human Urine (5% v/v) |
| Mineral Oil (10% w/v) | Vancomycin (0.25% w/v) |
## Inclusivity study
The reactivity of six *H. pylori* strains spanning the 3 major clades (hpEastAsia, hpEurope, and hpAfrica1) was evaluated. Samples were prepared by spiking negative fecal matrix with purified flagellar antigen from each *H. pylori* strain strain at 2-3x LoD. All results were read at dual-wavelength and were positive, demonstrating that the *H. PYLORI CHEK* test can detect antigen from strains representing the major *H. pylori* clades.
## *H. pylori* strains
ATCC 700392 (hpEurope)
ATCC 43526 (hpEurope)
ATCC 700824 (hpAfrica1)
ATCC 43504 (clade unknown)
ATCC 43579 (clade unknown)
JP26 (hpEastAsia)
{8}
f. Assay cut-off:
The following cut off-values and result interpretation criteria were established for the spectrophotometric single and dual wavelength reading methods:
Spectrophotometric Single Wavelength (450 nm)
Negative: <0.120
Positive: ≥0.120
Negative Control: <0.120
Positive Control: ≥0.500
Spectrophotometric Dual Wavelength (450/620 nm)
Negative: <0.080
Positive: ≥0.080
Negative Control: <0.080
Positive Control: ≥0.500
The following result interpretation criteria were established for the visual reading method:
Visual Reading
Negative = Colorless to very faint yellow color
Positive = Definitive yellow color
A positive result indicates that H. pylori antigens were detected. A negative result indicates that no H. pylori antigens were detected, or that the antigen levels are below what can be detected by the assay.
g. Prozone/Hook Effect:
The purpose of this study was to demonstrate that a high concentration of H. pylori antigen does not interfere with a positive reaction in the H. PYLORI CHEK test. High concentration samples were prepared by spiking negative fecal matrix with H. pylori antigen at 12 ug/mL (i.e. 10x the highest concentration observed in a positive clinical sample). A total of 5 different dilutions were prepared from the contrived sample. Samples were prepared by spiking the following concentrations of antigen into a negative fecal pool: 12 ug/mL, 6 ug/mL, 2.4 ug/mL, 1.2 ug/mL, 0.6 ug/mL and 0.12 ug/mL. Testing was performed in triplicate according to the Package Insert instructions. The results demonstrated that there was no overall hook effect, and that elevated levels of H. pylori antigen did not affect the test results.
2. Comparison studies:
a. Method comparison with predicate device:
Not Applicable.
{9}
b. Matrix comparison:
Not Applicable.
3. Clinical studies:
a. Clinical Sensitivity:
## Prospective study
### Initial Diagnosis:
The performance of the *H. PYLORI CHEK* test was evaluated by conducting a multi-site prospective clinical study. Six independent sites were included, five of which collected specimens from patients suspected of *H. pylori* infection. The sixth site, TECHLAB, did not conduct specimen collection. Of these five collection sites, three conducted testing using both the *H. PYLORI CHEK* test and the composite reference method (CRM). The remaining two sites performed CRM testing and sent samples to TECHLAB for testing using the *H. PYLORI CHEK* test. The CRM for diagnosis of *H. pylori* infection is based on endoscopally obtained gastric biopsy. The biopsied tissue was tested for the presence of *H. pylori* by histology or rapid urease test. The sensitivity and specificity for the *H. PYLORI CHEK* test was determined using the CRM.
Prospective testing consisted of 176 stool specimens collected from the patients with symptoms of dyspepsia, gastritis, or peptic ulcer who were scheduled to undergo endoscopy with gastric biopsy as part of routine care (Initial Diagnosis Group). Of these, 67 patients were excluded either because they were on a treatment regimen [i.e., proton-pump inhibitors (PPIs), or antibiotics] or had samples with CRM results that were rapid urease positive but histology negative. The remaining 109 patients, who were not taking PPIs or antibiotics at the time of specimen collection, were considered for final analysis. These specimens were tested at the following five sites: Carilion Clinic, International Centre for Diarrhoeal Disease Research Bangladesh, Mayo Clinic, University of Virginia, Kliniken Essen-Mitte, and TECHLAB (internal). Reading of the *H. PYLORI CHEK* was conducted by dual wavelength spectrophotometric analysis and by visual reading of the “Microassay Plate” for color change.
The ages of patients ranged from 19 to 82 years with 100% of the specimens coming from patients that were ≥18 years. Of the 109 patients tested, 66% were female and 34% were male. No difference in test performance was observed based on patient age or gender. The results are provided in Table 3, which shows the clinical performance of the *H. PYLORI CHEK* test for all 6 test sites combined. The results of the study show that the *H. PYLORI CHEK* test, by dual spectrophotometric analysis, exhibited a sensitivity of 100%, and a specificity of 96.1% compared to the CRM. Testing was also conducted by visual reading of “Microassay Plates”. Visual results were the same as dual wavelength spectrophotometric results 99% of the time.
10
{10}
Table 3. Clinical performance of the H. PYLORI CHEK by dual wavelength spectrophotometric analysis.
| Initial Diagnosis | CRM Positive | CRM Negative |
| --- | --- | --- |
| H. PYLORI CHEK Positive | 32 | 3* |
| H. PYLORI CHEK Negative | 0 | 74 |
| Performance | | 95% CI |
| Sensitivity | 100% | 89.3% - 98.9% |
| Specificity | 96.1% | 89.2% - 98.7% |
* All three specimens tested positive initially by the H. PYLORI CHEK test, but negative upon re-testing with the H. PYLORI CHEK test.
# Post-Therapy Diagnosis:
Eradication (post-therapy) evaluation was conducted on patients enrolled prospectively at 3 sites. A total of 9 specimens were collected at least 4 weeks after completion of the treatment regimen. Post-therapy evaluation was conducted using a two-step algorithm of patient analysis. First, patients were screened for the continued presence of $H$ pylori using an FDA-cleared ELISA. Positive patients were reflected to a follow-up endoscopy and biopsy analysis by rapid urease test and histology. The results are provided in Table 4, which shows the clinical performance of the $H$ PYLORI CHEK test for all 6 test sites combined. The results of the study show that the $H$ PYLORI CHEK test exhibited a sensitivity of $77.8\%$ compared to the CRM.
Table 4. Clinical performance of the H. PYLORI CHEK
| Post-Therapy Diagnosis | CRM Positive | CRM Negative |
| --- | --- | --- |
| H. PYLORI CHEK Positive | 7* | 0 |
| H. PYLORI CHEK Negative | 2** | 0 |
| Performance | | 95% CI |
| Sensitivity | 77.8% | 45.3% - 93.7% |
*One specimen tested positive by visual read but negative by spectrophotometric interpretation by the H. PYLORI CHEK test (OD450/620 0.034).
{11}
**One specimen tested negative initially by the H. PYLORI CHEK test, but positive upon retesting with the H. PYLORI CHEK test.
The ages of patients ranged from 33 years to 72 years. Of the 9 patients tested, 6 were female and 3 were male.
# Retrospective study
A retrospective study was conducted to evaluate performance and to augment the prospective clinical study. Testing of retrospective samples was conducted at TECHLAB.
To provide assurance that the retrospective samples are representative of a wide range of OD readings, CRM positive samples and retrospective positive samples were both tested by an FDA cleared ELISA. The distribution and mean OD values obtained from the retrospective study, $n = 75$ samples, were compared to those values obtained from CRM positive samples, $n = 42$ , to ensure that the use of retrospective sample results reflects the OD distribution of CRM positive samples. This analysis demonstrated that the use of retrospective and prospective clinical samples have similar distributions and no concern of bias was noted.
Thee performance of the H. PYLORI CHEK test was evaluated by testing retrospective samples at TECHLAB. Retrospective testing consisted of 196 frozen stool samples (75 positive and 121 negative by an FDA-cleared ELISA) obtained from sample repositories. Positive percent agreement (PPA) and negative percent agreement (NPA) for the H. PYLORI CHEK test was determined by comparing to an FDA Cleared ELISA that was tested concurrently. The results of the study show that the H. PYLORI CHEK test exhibited a PPA of $100\%$ and a NPA of $100\%$ with an FDA cleared ELISA. The results are shown in Table 5.
Table 5. Clinical performance of the H. PYLORI CHEK test on retrospective samples.
| N = 196 | FDA Cleared ELISA (Positive) | FDA Cleared ELISA (Negative) |
| --- | --- | --- |
| H. PYLORI CHEK Positive | 75 | 0 |
| H. PYLORI CHEK Negative | 0 | 121 |
| Performance | | 95% CI |
| Positive Percent | 100% | 95.1%-100% |
{12}
| Agreement | | |
| --- | --- | --- |
| Negative Percent Agreement | 100% | 96.9%-100% |
b. Clinical specificity:
See section M3a. above.
c. Other clinical supportive data (when a. and b. are not applicable):
Not Applicable
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
Not Applicable
N. Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
O. Conclusion:
The submitted information in this premarket notification is complete and supports substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.