K181396 · Passion For Life Healthcare (Uk) Limited · LRK · Nov 16, 2018 · Dental
Device Facts
Record ID
K181396
Device Name
Oral Device OA, Oral Device S
Applicant
Passion For Life Healthcare (Uk) Limited
Product Code
LRK · Dental
Decision Date
Nov 16, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Oral Device-OA is intended to reduce night-time snoring and / or mild to moderate obstructive sleep apnea (OSA) for adults 18 and older. The Oral Device-S is intended to reduce night time snoring for adults 18 and older.
Device Story
Oral Device-OA and Oral Device-S are mandibular repositioning devices (MRDs) designed to reduce snoring and/or mild-to-moderate OSA. Devices consist of pre-formed upper and lower trays; user-fitted via impression material to teeth. Oral Device-OA is fitted by a dentist; Oral Device-S is fitted by the user at home. Mechanism involves screw-based adjustment to advance the mandible, holding the lower jaw forward to maintain airway patency. Used in home (both) and clinical (OA only) settings. Output is physical repositioning of the mandible; healthcare providers use the device to manage airway obstruction during sleep. Benefits include reduced snoring and improved sleep apnea symptoms.
Clinical Evidence
Bench testing only. Testing included functional durability (180-day equivalent), flexural strength, fracture toughness, water absorption/solubility, mechanical/tensile testing, and drop testing. Biocompatibility testing (cytotoxicity, sensitization, irritation) performed per ISO 10993-1 and FDA guidance for intraoral devices.
Technological Characteristics
Mandibular repositioning device; pre-formed trays; screw-based advancement mechanism; allows lateral/vertical movement. Materials biocompatibility per ISO 10993-1. No energy source; mechanical operation. Single-patient, multi-use.
Indications for Use
Indicated for adults 18+ to reduce night-time snoring (Oral Device-S) or snoring and mild-to-moderate obstructive sleep apnea (Oral Device-OA). Contraindicated for central sleep apnea, severe respiratory disorders (asthma/emphysema), patients under 18, jaw joint disorders, severe jaw pain, loose teeth, periodontal disease, full dentures, fixed braces/retainers, or dental implants <1 year old.
Regulatory Classification
Identification
Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.
Special Controls
*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
Passion for Life Healthcare (UK) Limited Paul Dryden Consultant Pioneer House, Pioneer Business Park North Road Ellesmere Port, CH65 1AD GB
November 16, 2018
Re: K181396
Trade/Device Name: Oral Device OA. Oral Device S Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices For Snoring And Intraoral Devices For Snoring And Obstructive Sleep Apnea Regulatory Class: Class II Product Code: LRK Dated: October 16, 2018 Received: October 17, 2018
Dear Paul Dryden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Mary S. Mary S. Runner -S3 Runner -S3 Date: 2018.11.16
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K181396
Device Name Oral Device-OA
Indications for Use (Describe)
The Oral Device-OA is intended to reduce night-time snoring and / or mild to moderate obstructive sleep apnea (OSA) for adults 18 and older.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
## PDF Page 19 of 28
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K181396
Device Name Oral Device-S
Indications for Use (Describe)
The Oral Device-S is intended to reduce night time snoring for adults 18 and older.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| | 510(k) Summary<br>9-Nov-18 |
|-----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | Passion for Life Healthcare (UK) Limited<br>Pioneer House, Pioneer Business Park North Road<br>Ellesmere Port, CH65 1AD, United Kingdom |
| | Paul Earps - Quality Controller<br>Tel - +44 151 550 4708<br>E-mail - paul.earps@passionforlife.com |
| Submission Correspondent: | Paul Dryden<br>Tel - (239) 307-6061<br>E-mail - paul.dryden@promedic.cc |
| Proprietary or Trade Name: | Oral Device-OA and Oral Device-S |
| Common/Usual Name: | Intra-oral appliance |
| Product Code:<br>Product Classification Name: | LRK<br>Device, anti-snoring, Intraoral devices for snoring<br>and intraoral devices for snoring and obstructive sleep apnea |
| CFR Number:<br>Class: | 21 CFR 872.5570<br>2 |
| Predicate Devices: | Oral Device-OA - K113569 – ApneaRx<br>Oral Device-S - K170825 - SnoreRx |
#### Device Description
Passion for Life Healthcare UK Limited is offering two devices: 1) the Oral Device-S is an OTC device intended as an aid for the reduction of snoring; and 2) the Oral Device-OA is a Rx device intended as an aid for the reduction of mild to moderate obstructive sleep apnea, and/or snoring.
Both the Oral Device-S and Oral Device-OA are an oral appliance design concept based upon the use of a standard set of upper and lower trays of one size.
The Oral Device-OA is fitted by a dentist to act as a Mandibular Repositioning Device (MRD).
Whereas the Oral Device-S, an OTC device, is fitted by the user at home and any adjustments can be made without clinician interaction.
### Indications for Use
#### Oral Device-OA
The Oral Device-OA is intended to reduce night-time snoring and / or mild to moderate obstructive sleep apnea (OSA) for adults 18 and older.
### Oral Device-S
The Oral Device-S is intended to reduce night time snoring for adults 18 and older.
### Environment of Use Oral Device-OA Home and clinical settings
{5}------------------------------------------------
#### Oral Device-S Home
Cautions, Warning, and Contraindications Oral Device-OA
## Warnings
Use of this device may cause:
- Tooth movement or changes in dental occlusion ●
- Dental sensitiveness after removing the Oral Device
- Gingival (gum) or dental soreness
- Pain or soreness of the jaw
- . Obstruction of oral breathing
- Excessive salivation ●
### Contraindications
The device is contraindicated for patients who:
- Suffer with central sleep apnea ●
- Have a severe respiratory disorder, such as asthma or emphysema
- Are under 18 years old
- . Have been diagnosed with a joint disorder related to the jaw
- Have severe jaw pain, loose teeth or periodontal disease
- Have full dentures that are removed at night
- Wear fixed braces or a retainer at night ●
- Have had an implant less than 1 year ago
### Cautions, Warning, and Contraindications Oral Device-S
### Warnings
Use of this device may cause:
- Tooth movement or changes in dental occlusion
- Dental sensitiveness after removing the Oral Device
- Gingival (gum) or dental soreness ●
- Pain or soreness of the jaw ●
- Obstruction of oral breathing ●
- Excessive salivation .
### Contraindications
The device is contraindicated for patients who:
- . You suffer with central sleep apnea
- You have a severe respiratory disorder, such as asthma or emphysema ●
- . You are under 18 years old
- You have been diagnosed with a joint disorder related to the jaw .
- You have severe jaw pain, loose teeth or periodontal disease ●
- You have full dentures that are removed at night
- You wear fixed braces or a retainer at night ●
- You had a dental implant less than 1 year ago
{6}------------------------------------------------
### 510(k) Summary 9-Nov-18
# Comparison of Oral Device-OA to Predicate
We present in Table 1 a comparison of the subject device, Oral Device-OA compared to the predicate ApneaRx - K113569.
| | Proposed Device<br>Oral Device-OA | Predicate Device<br>ApneaRx<br>K113569 |
|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attributes | | |
| Indications for Use | The Oral Device-OA is intended to reduce night-time snoring and / or mild to moderate obstructive sleep apnea (OSA) for adults 18 and older. | The Apnea Sciences Corporation "ApneaRx" is intended for use on adult patients 18 years of age or older as an aid for the reduction of mild to moderate obstructive sleep apnea (OSA), and/or snoring. |
| Environments of use | Home and clinical settings | Home and clinical settings |
| Patient Population | Adult patients 18 years and older | Adult patients 18 years and older |
| Contraindications | Suffer with central sleep apnea Have a severe respiratory disorder, such as asthma or emphysema Are under 18 years old Have been diagnosed with a joint disorder related to the jaw Have severe jaw pain, loose teeth or periodontal disease Have full dentures that are removed at night Wear fixed braces or a retainer at night Have had an implant less than 1 year ago | Suffer with central sleep apnea Have a severe respiratory disorder, such as asthma or emphysema Are under 18 years old Have been diagnosed with a joint disorder related to the jaw Have severe jaw pain, loose teeth or periodontal disease Have full dentures that are removed at night Wear fixed braces or a retainer at night Have had an implant less than 1 year ago |
| Prescription | Prescription use | Prescription use |
| Duration of Use | Single patient, multi-use | Single patient, multi-use |
| Principle of operation / means of<br>mandibular advancement | Adjustment of the relative position of the trays by the use of screw adjustment to obtain a fixed advancement. | Adjustment of the relative position of the trays by the use of a latch adjustment to obtain a fixed advancement. |
| Design | | |
| Tray Design | Pre-formed and fixed | Customized trays |
| Method to advance the mandible | Adjustable upper / lower tray for mandible repositioning | Adjustable upper / lower tray for mandible repositioning |
| Allows lateral and vertical movement | Allows for lateral and vertical movement | Allows for lateral and vertical movement |
| Maximum adjustment by the user | Up to 10 mm | Up to 10 mm |
| How it opens airway | Holds mandible / lower jaw forward | Holds mandible / lower jaw forward |
| Method of cleaning | Cleaned by simple rinsing with water and toothbrush | Cleaned by simple rinsing with water and toothbrush |
| | Proposed Device<br>Oral Device-OA | Predicate Device<br>ApneaRx<br>K113569 |
| Performance Testing | | |
| Biocompatibility and Patient contact | ISO 10993<br>Cytotoxicity<br>Sensitization<br>Irritation | ISO 10993<br>Cytotoxicity<br>Sensitization<br>Irritation |
| Durability testing | 180 days | Not available |
| Performance testing | • Functional testing for durability<br>after multiple cleanings equivalent<br>to 180 days use<br>• Flexural strength / Fracture<br>Toughness<br>• Water absorption / solubility<br>• Mechanical / Tensile testing<br>• Drop testing | Testing not available |
## Table 1 – Oral Device-OA vs. K113569 - ApneaRx
{7}------------------------------------------------
### 510(k) Summarv 9-Nov-18
## Discussion of Substantial Equivalence for Oral Device-OA
The Oral Device-OA is viewed as substantially equivalent to the predicate device because:
Indications - Similar to predicate - ApneaRx - K113569; indicated for treating snoring and mild to moderate obstructive sleep apnea (OSA).
Discussion - The indications for use between the subject device and predicate are similar and therefore they can be found as substantially equivalent.
Technology / Principle of Operation - Similar to predicate - ApneaRx - K113569. Both devices use upper and lower trays with a means to advance the mandible / lower jaw.
Discussion - Both devices use separate upper and lower trays with a means to advance the mandible / lower jaw are similar. The predicate uses customized trays while the proposed device uses preformed trays. Both utilize an impression material to mold to the individual's teeth and to keep the trays in place. There is a difference in the trays is only related to sizing and the mechanism of advancing the trays with the proposed device using screws and the predicate a latch and lock mechanism.
Environment of Use - Similar to predicate - ApneaRx - K113569. They are used in Home and clinical settings.
Discussion - Both devices have similar environments of use and therefore they can be found as substantially equivalent.
Patient Population - Similar to predicate - ApneaRx - K113569. 18 years and older Discussion - The patient population is similar and therefore they can be found as substantially equivalent.
### Non-clinical performance testing
The Oral Device underwent the following testing:
- Functional testing for durability after multiple cleanings equivalent to 180 days use ●
- . Flexural strength / Fracture Toughness
{8}------------------------------------------------
### 510(k) Summary 9-Nov-18
- Water absorption / solubility
- Mechanical / Tensile testing
- . Drop testing
## Biocompatibility of Materials
All the materials are considered per ISO 10993-1 per FDA Guidance Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA dated 11/12/2002
- Cytotoxicity ●
- Sensitization
- . Irritation
## Comparison of Oral Device-S to Predicate
### Table 2 – Oral Device-S vs. K170825 – SnoreRx OTC
| | Proposed Device<br>Oral Device-S | Predicate Device<br>SnoreRx OTC<br>K170825 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attributes | | |
| Indications for Use | The Oral Device-S is intended to<br>reduce night time snoring for adults<br>18 and older. | The SnoreRx is intended for use on<br>adult patients 18 years of age or older<br>as an aid for the reduction of snoring. |
| Environments of use | Home | Home |
| Patient Population | Adult patients 18 years and older | Adult patients 18 years and older |
| Contraindications | • You suffer with central sleep<br>apnea<br>• You have a severe respiratory<br>disorder, such as<br>asthma or emphysema<br>• You are under 18 years old<br>• You have been diagnosed with a<br>joint disorder<br>related to the jaw<br>• You have severe jaw pain, loose<br>teeth or periodontal disease<br>• You have full dentures that are<br>removed at night<br>• You wear fixed braces or a<br>retainer at night<br>• You had a dental implant less than<br>1 year ago | • You suffer with central sleep<br>apnea<br>• You have a severe respiratory<br>disorder, such as<br>asthma or emphysema<br>• You are under 18 years old<br>• You have been diagnosed with a<br>joint disorder<br>related to the jaw<br>• You have severe jaw pain, loose<br>teeth or periodontal disease<br>• You have full dentures that are<br>removed at night<br>• You wear fixed braces or a<br>retainer at night<br>• You had a dental implant less<br>than 1 year ago |
{9}------------------------------------------------
### 510(k) Summary 9-Nov-18
| | Proposed Device<br>Oral Device-S | Predicate Device<br>SnoreRx OTC<br>K170825 |
|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| OTC | OTC | OTC |
| Duration of Use | Single patient, multi-use | Single patient, multi-use |
| Principle of operation / means of<br>mandibular advancement | Adjustment of the relative position of<br>the trays by the use of screw<br>adjustment to obtain a fixed<br>advancement. | Adjustment of the relative position of<br>the trays by the use of latch<br>adjustment to obtain a fixed<br>advancement. |
| Design | | |
| Tray Design | Pre-formed and fixed | Pre-formed and fixed |
| Method to advance the mandible | Adjustable upper / lower tray for<br>mandible repositioning | Adjustable upper / lower tray for<br>mandible repositioning |
| Allows lateral and vertical movement | Allows for lateral and vertical<br>movement | Allows for lateral and vertical<br>movement |
| Maximum adjustment by the user | Up to 6 mm | Up to 6 mm |
| How it opens airway | Holds mandible / lower jaw forward | Holds mandible / lower jaw forward |
| Method of cleaning | Cleaned by simple rinsing with water<br>and toothbrush | Cleaned by simple rinsing with water<br>and toothbrush |
| | Performance Testing | |
| Biocompatibility and Patient contact | ISO 10993<br>Cytotoxicity<br>Sensitization<br>Irritation | ISO 10993<br>Cytotoxicity<br>Sensitization<br>Irritation |
| Durability testing | 180 days | Not available |
| Performance testing | • Functional testing for durability<br>after multiple cleanings equivalent<br>to 180 days use<br>• Flexural strength / Fracture<br>Toughness<br>• Water absorption / solubility<br>• Mechanical / Tensile testing<br>• Drop testing | Testing not available |
# Discussion of Substantial Equivalence for Oral Device-S
The Oral Device-S is viewed as substantially equivalent to the predicate device because:
Indications - Similar to predicate - SnoreRx OTC - K170825; indicated for reducing night-time snoring.
Discussion - The indications for use between the subject device and predicate are similar and therefore they can be found as substantially equivalent.
Technology / Principle of Operation - Similar to predicate - SnoreRx OTC - K170825. Both devices use upper and lower trays with a means to advance the mandible / lower jaw. Discussion - Both devices use separate pre-formed trays with a means to advance the mandible / lower jaw that are similar. Both utilize an impression material to the individual's teeth and to keep the trays in place. They use a screw or a latch and lock mechanism to advance and hold the trays in their advanced position.
{10}------------------------------------------------
#### 510(k) Summarv 9-Nov-18
Environment of Use - Similar to predicate - SnoreRx OTC - K170825. They are used in Home settings.
Discussion - Both devices have similar environments of use and therefore they can be found as substantially equivalent.
Patient Population - Similar to predicate - SnoreRx OTC - K170825. 18 years and older Discussion - The patient population is similar and therefore they can be found as substantially equivalent.
### Non-clinical performance testing
The Oral Device underwent the following testing:
- . Functional testing for durability after multiple cleanings
- Flexural strength / Fracture Toughness ●
- . Water absorption / solubility
- . Mechanical / Tensile testing
- . Drop testing
### Biocompatibility of Materials
All the materials are considered per ISO 10993-1 per FDA Guidance Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA dated 11/12/2002
- Cytotoxicity ●
- . Sensitization
- Irritation
### Discussion of Differences
The differences between the Oral Device - OA predicate - ApneaRx - K113569 and the Oral Device-S and the predicate SnoreRx - K170825 are:
- For the Oral Device-OA the predicate trays are customized but the principle of a boil and bite . tray with impression material is similar.
- . The mechanism for advancing the trays is a set of screws for the Oral Device while the predicate uses clips on the side to latch and hold the trays in the advanced position.
### Substantial Equivalence Conclusion
Based upon the performance testing and comparison to the legally marketed predicate device for indications for use, technology, and performance we believe we have demonstrated that the Oral Device-OA is substantially equivalent to the predicate ApneaRx - K113569 and the Oral Device-S is substantially equivalent to the predicate SnoreRx OTC - K170825.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.