← Product Code [GCJ](/productcode/GCJ) · K181395

# Universal Cannula Seal (5-12 mm) (K181395)

_Intuitive Surgical, Inc. · GCJ · Jul 27, 2018 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181395

## Device Facts

- **Applicant:** Intuitive Surgical, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jul 27, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

To provide a port of entry for endoscopic instruments.

## Device Story

The Universal Cannula Seal (5-12 mm) is a sterile, single-use accessory for da Vinci surgical systems. It attaches to da Vinci cannulas, obturators, and reducers to provide a port of entry for endoscopic instruments. The device maintains an air-tight seal and an insufflated body cavity during surgery. It features an insufflation port with a universal luer fitting and an insufflation lever. Used in clinical settings by surgeons and surgical staff, the device allows for the passage of instruments of varying diameters. By maintaining pneumoperitoneum, it facilitates endoscopic access, enabling the surgeon to perform procedures while ensuring the surgical site remains accessible and stable.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, animal model testing, and cadaver studies. Bench testing confirmed physical, mechanical, user interface, and packaging requirements, including leak testing, flow rates, and seal latch strength. Reliability/life testing confirmed performance throughout one procedure. Biocompatibility testing was performed per ISO 10993-1. Sterilization validation and bacterial endotoxin testing were completed. Packaging transit and shelf-life testing verified integrity.

## Technological Characteristics

Sterile, single-use, disposable seal for da Vinci cannulas. Features universal luer fitting for insufflation. Compatible with 5-12 mm diameter instruments. Materials tested for biocompatibility per ISO 10993-1. Mechanical design includes seal latches and stopcock for insufflation control. No software or electronic components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- 8MM Trocar Kit ([K133845](/device/K133845.md))
- EndoWrist® Stapler 45 and Stapler 45 Reloads ([K170508](/device/K170508.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 27, 2018

Intuitive Surgical, Inc. Ms. Cheryl Wu Regulatory Affairs Engineer 1266 Kifer Road Sunnyvale, California 94086

Re: K181395

Trade/Device Name: Universal Cannula Seal (5-12 mm) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: May 25, 2018 Received: May 29, 2018

Dear Ms. Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# David Krause -S

for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181395

Device Name Universal Cannula Seal (5-12 mm)

Indications for Use (Describe)

The da Vinci Trocar has application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Traditional 510(k)

# 510(k) Summary

| 510(k) Submitter:                     | Intuitive Surgical, Inc.                               |
|---------------------------------------|--------------------------------------------------------|
|                                       | 1266 Kifer Road                                        |
|                                       | Sunnyvale, CA 94086                                    |
| Contact:                              | Cheryl Wu                                              |
|                                       | Regulatory Affairs Engineer                            |
|                                       | Phone: 408-523-6401                                    |
|                                       | Fax: 408-523-1390                                      |
|                                       | Email: cheryl.wu@intusurg.com                          |
| Date Summary Prepared:                | July 26, 2018                                          |
| Trade Name:                           | Universal Cannula Seal (5-12 mm)                       |
| Common Name:                          | Endoscope and accessories                              |
| Classification:                       | Class II                                               |
|                                       | 21 CFR 876.1500, Endoscope and Accessories             |
| Product Codes:                        | GCJ                                                    |
| Classification Advisory<br>Committee: | General and Plastic Surgery                            |
| Predicate Device(s):                  | K133845 – 8MM Trocar Kit                               |
|                                       | K170508 – EndoWrist® Stapler 45 and Stapler 45 Reloads |

#### Intended Use:

To provide a port of entry for endoscopic instruments.

## Indications for Use:

The da Vinci Trocar has application in a variety of endoscopic procedures to provide a port of entry for endoscopic instruments.

# Device Description:

The Universal Cannula Seal (5-12 mm) is a sterile, single-use (disposable) seal used with da Vinci cannula, obturators, and reducers. The Universal Cannula Seal has an opening that is sized to allow specific diameter ranges of instruments and surgical accessories to pass

Image /page/3/Picture/11 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and in a light gray color. Below that, the word "SURGICAL" is in a smaller font and also in a light gray color, with the registered trademark symbol to the right of it.

{4}------------------------------------------------

through to the surgical site while maintaining an air-tight system and an insufflated body cavity. Identical to the predicate seals, the subject seal has an insufflation port with a universal luer fitting and an insufflation lever.

## Technological Characteristics:

The Universal Cannula Seal is equivalent to the predicate devices in terms of its intended use, design, technology, and performance specifications. The Universal Cannula Seal combines the instrument and accessory size compatibility ranges from both predicate seals. Modifications also include a change in material formulation and a change in packaging configuration.

These modifications do not affect the intended use or the fundamental technology of the subject device relative to that of the predicate device.

## Performance Data:

In accordance with the Design Control process, risk analysis was conducted to evaluate the impact of the modifications on the predicate devices. Design verification and design validation testing were conducted on the subject devices to confirm that the design outputs meet design input requirements and that the devices are safe and effective for its intended use. The testing identified no new issues of safety or effectiveness and no new risks.

## Design Verification:

The subject device was subjected to a series of bench tests to evaluate performance and to demonstrate that the design outputs meet the design input requirements. Testing was performed with a compatible da Vinci surgical system. The design verification testing included confirmation that the subject device meets physical, mechanical, user interface, labeling, and packaging requirements. The verification test cases included:

- Number of seals per package ●
- Single use disposable product
- Diameter/instrument compatibility markings ●
- Position and use of insufflation valve ●
- Standard luer fitting attachment
- Insufflation valve open and close
- . Stopcock rotation
- No Natural rubber latex
- Stopcock axial load
- Depression of seal latches
- Seal latch strength
- Installation of seal onto cannula ●

Image /page/4/Picture/22 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and in a light gray color. Below that, the word "SURGICAL" is also in all caps and in a smaller font size, with the registered trademark symbol next to it.

{5}------------------------------------------------

- Rotation of seal on cannula
- Obturator compatibility ●
- Torque limits
- . Force limits
- Flow rate without device
- Leak test seal without device ●
- Leak test seal with compatible devices
- Seal removal force ●

## Reliability/Life Testing:

The subject device underwent reliability (life) testing to confirm that the subject device meets the design inputs throughout its intended life of one procedure.

#### Design Validation:

The design validation testing summarized in this submission validates general, functional, and interaction (compatibility) requirements for the subject device. Tests with an animal model and a cadaver were performed to confirm that the subject device functions in accordance with its intended use.

#### Human Factors:

In conformance with applicable FDA Guidance and international standards on human factors and usability engineering, a thorough analysis and internal risk assessment has been reviewed for the subject device. No new usability risk scenarios are associated with the differences between the subject device and the predicate devices. Therefore, additional human factors studies were not performed.

#### Sterilization:

Sterilization validation and Bacterial Endotoxins Testing have been performed on the subject device and demonstrates compliance with applicable FDA Guidance for sterile devices.

## Biocompatibility Testing:

Biocompatibility testing was successfully completed in accordance with applicable FDA Guidance and international standards related to ISO 10993-1.

#### Packaging Verification:

Package transit verification testing was performed to demonstrate that the packaging maintains package integrity and product functionality in simulated storage and shipping

Image /page/5/Picture/22 description: The image shows the logo for Intuitive Surgical. The word "INTUITIVE" is in all caps and is above the word "SURGICAL". There is a yellow dot above the word "INTUITIVE". There is a registered trademark symbol next to the word "SURGICAL".

{6}------------------------------------------------

conditions. Shelf life testing was performed to verify the Universal Cannula Seal and its packaging for the time specified.

# Summary:

Based on the intended use, indications for use, technological characteristics, and performance data, the subject device Universal Cannula Seal (5-12 mm) is substantially equivalent to currently marketed predicate devices in terms of safety, effectiveness, and performance.

Image /page/6/Picture/5 description: The image shows the logo for Intuitive Surgical. The logo consists of the word "INTUITIVE" in all caps, with a small yellow dot above the "I". Below "INTUITIVE" is the word "SURGICAL" in a smaller font size. The logo is simple and modern, and the use of the color yellow adds a touch of warmth.

---

**Source:** [https://fda-staging.innolitics.com/device/K181395](https://fda-staging.innolitics.com/device/K181395)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
