← Product Code [HWC](/productcode/HWC) · K181377

# PROSTEP MICA Screw System (K181377)

_Wrightmedicaltechnologyinc · HWC · Jan 11, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K181377

## Device Facts

- **Applicant:** Wrightmedicaltechnologyinc
- **Product Code:** [HWC](/productcode/HWC.md)
- **Decision Date:** Jan 11, 2019
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The MICA™ Screw is indicated for fixation of bone fractures or for bone reconstruction. Examples include: - Mono or Bi-Cortical osteotomies in the foot or hand - Distal or Proximal metatarsal or metacarpal osteotomies - Weil osteotomy - Fusion of the first metatarsophalangeal joint and interphalangeal joint - Fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc.) - Akin type osteotomy - Arthrodesis base first metatarsal cuneiform joint to reposition and stabilize metatarsus varus primus - Calcaneus/cuboid arthrodesis - Talar/navicular arthrodesis

## Device Story

PROSTEP MICA Screw System consists of fully threaded, cannulated titanium alloy screws; diameters 3-4 mm; various lengths. Used for bone reconstruction, osteotomy, arthrodesis, and fracture fixation in foot/hand. Device provides mechanical stabilization of bone segments. Operated by orthopedic surgeons in clinical/surgical settings. Benefits include stabilization of bone fractures and reconstruction sites to facilitate healing.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including packaging integrity (ISO 11607:2006), sterilization (ISO 11137-2:2013), and bacterial endotoxin testing (≤ 20EU/device).

## Technological Characteristics

Titanium alloy cannulated bone screws; 3-4 mm diameter; fully threaded. Sterilization per ISO 11137-2:2013. Packaging per ISO 11607:2006.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- MICA Screw System ([K162353](/device/K162353.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

January 11, 2019

Wright Medical Technology, Inc. Val Myles Senior Regulatory Affairs Specialist 1023 Cherry Road Memphis, Tennessee 38117

Re: K181377

Trade/Device Name: PROSTEP MICA Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: December 12, 2018 Received: December 14, 2018

Dear Val Myles:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Shumaya Ali-S

for

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2020

See PRA Statement below.

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

#### K181377

Device Name PROSTEPTM MICATM Screw System

Indications for Use (Describe)

The MICA™ Screw is indicated for fixation of bone fractures or for bone reconstruction. Examples include:

- · Mono or Bi-Cortical osteotomies in the foot or hand
- · Distal or Proximal metatarsal or metacarpal osteotomies
- · Weil osteotomy
- · Fusion of the first metatarsophalangeal joint and interphalangeal joint
- · Fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc.)
- · Akin type osteotomy
- · Arthrodesis base first metatarsal cuneiform joint to reposition and stabilize metatarsus varus primus
- · Calcaneus/cuboid arthrodesis
- · Talar/navicular arthrodesis

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Wright, a company that focuses on excellence. The logo consists of two overlapping trapezoids on the left, followed by the word "WRIGHT" in large, bold letters. Below the word "WRIGHT" is the phrase "FOCUSED EXCELLENCE" in smaller letters. The logo is simple and modern, and the use of gray gives it a professional look.

1023 Cherry Road Memphis, TN 38117 wright.com

## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the PROSTEP™ MICA™ Screw System.

| (a)(1). Submitted By:              | Wright Medical Technology, Inc.<br>1023 Cherry Road<br>Memphis, TN 38117         |
|------------------------------------|----------------------------------------------------------------------------------|
| Date:                              | May 11, 2018                                                                     |
| Contact Person:                    | Val Myles<br>Sr. Regulatory Affairs Specialist                                   |
| (a)(2). Proprietary Name:          | PROSTEP™ MICAT™ Screw System                                                     |
| Common Name:                       | Cannulated Bone Screw                                                            |
| Classification Name and Reference: | 21 CFR 888.3040- Class II- Smooth or threaded<br>metallic bone fixation fastener |
| Device Product Code, Device Panel: | HWC – Orthopedic                                                                 |
| (a)3. Predicate Device:            | K162353 - MICA Screw System                                                      |

#### (a)(4). Device Description

The PROSTEP™ MICA™ Screws are intended for use in bone reconstruction, osteotomy, arthrodesis, fracture repair, and fracture fixation of bones appropriate for the device. MICA™ Screws are fully threaded, cannulated screws manufactured from titanium alloy. The screws are available in diameters of 3-4 mm in various lengths.

#### (a)(5). Intended Use

The MICA" Screw is indicated for fixation of bone fractures or for bone reconstruction. Examples include:

- Mono or Bi-Cortical osteotomies in the foot or hand
- . Distal or Proximal metatarsal or metacarpal osteotomies
- Weil osteotomy
- . Fusion of the first metatarsophalangeal joint and interphalangeal joint
- Fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc.)
- . Akin type osteotomy

{4}------------------------------------------------

- Arthrodesis base first metatarsal cuneiform joint to reposition and stabilize metatarsus ● varus primus
- . Calcaneus/cuboid arthrodesis
- . Talar/navicular arthrodesis

# (a)(6). Technological Characteristics Comparison

The MICA™ Screws are substantially equivalent in indications, dimensions, and materials when compared to the legally marketed predicate device. The modified packaging configuration of the subject device meets the relevant packaging and sterility standards (listed below) and is therefore substantially equivalent to the predicate.

## (b)(1). Substantial Equivalence - Non-Clinical Evidence

| Validation and/or Verification<br>Method | Acceptance Value/Criteria          | Results |
|------------------------------------------|------------------------------------|---------|
| Packaging Integrity and Shelf-Life       | As established in ISO 11607:2006   | Pass    |
| Sterilization                            | As established in ISO 11137-2:2013 | Pass    |
| Bacterial Endotoxins Testing             | ≤ 20EU/device                      | Pass    |

# (b)(2). Substantial Equivalence- Clinical Evidence

N/A

# (b)(3). Substantial Equivalence- Conclusions

The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K181377](https://fda-staging.innolitics.com/device/K181377)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).